RS85 Diagnostic Ultrasound System, RS80 EVO Diagnostic Ultrasound System
Applicant
Samsung Medison Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Jun 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal(Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Device Story
General-purpose, mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via probes; displays images in 2D, M-mode, Color/Power/Spectral Doppler, Tissue Doppler, 3D/4D, and ElastoScan modes. Used in hospitals, clinics, and private practices by trained healthcare professionals. Provides anatomical measurements and analysis packages to aid clinical diagnosis. Features real-time acoustic output display (mechanical/thermal indices). System aids clinicians in visualizing internal structures and fluid flow; supports diagnostic decision-making through image review and reporting.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including acoustic output measurement, biocompatibility, cleaning/disinfection validation, and thermal/electromagnetic/mechanical safety testing in accordance with recognized standards (ISO 14971, IEC 60601 series, AAMI/ANSI/ISO 10993-1).
Technological Characteristics
Mobile ultrasound system; supports multiple imaging modes (2D, Doppler, 3D/4D, Elastoscan). Materials evaluated per ISO 10993-1. Acoustic output complies with NEMA UD 2-2004 and IEC 60601-2-37. Electrical safety per ANSI/AAMI ES60601-1; EMC per IEC 60601-1-2. Software-driven functionality; includes measurement and analysis packages.
Indications for Use
Indicated for fetal/obstetrics, abdominal, gynecological, intraoperative, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal, urological, cardiac (adult/pediatric/trans-esophageal), and peripheral vessel imaging in patients requiring diagnostic ultrasound.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
June 24, 3021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Samsung Medison Co., LTD. % Scully Kim Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA
# Re: K210959
Trade/Device Name: RS85 Diagnostic Ultrasound System, RS80 EVO Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: May 25, 2021 Received: May 27, 2021
# Dear Scully Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210959
## Device Name RS85 Diagnostic Ultrasound System RS80 EVO Diagnostic Ultrasound System
## Indications for Use (Describe)
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal(Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, Power Doppler (PD) mode, ElastoScan™ Mode, Multi-Image mode(Dual, Quad), Combined modes, 3D/4D mode
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">X</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
SAMSUNG MEDISON CO., LTD.
# K210959
RS85 / RS80 EVO Diagnostic Ultrasound System
## 510(K) Summary:
In accordance with 21 CFR 807.92 the following summary of information is provided:
- Date Prepared March 29, 2021 1.
- 2. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
- 3. Primary Contact Person Scully Kim Regulatory Affairs Specialist Phone: +82.2.2.2194.1312 Fax: +82. 2.2194.0273 Email: scully.kim@samsungmedison.com
- 4. Secondary Contact Person Ninad Gujar Vice President of Regulatory & Quality Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com
- 5. Proposed Device
| Common Name: | Diagnostic Ultrasound System and Accessories |
|-----------------------------|----------------------------------------------------------------------------|
| Trade/Device Name: | RS85 Diagnostic Ultrasound System<br>RS80 EVO Diagnostic Ultrasound System |
| Additional Marketing Name : | RS85 Prestige Diagnostic Ultrasound System |
| Regulation Name: | Ultrasound pulsed Doppler imaging system |
| Panel/ Regulatory Class: | Radiology / II |
| Product Code: | IYN, IYO, ITX |
| Regulation No. | 892.1550, 892.1560, 892.1570 |
- 6. Predicate Device
- RS85 Diagnostic Ultrasound System (K192903)
- 7. Device Description
The RS85 / RS80 EVO are a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode, MV-Flow Mode or as a combination of these modes.
The RS85 / RS80 EVO also give the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals.
The RS85 / RS80 EVO have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
{4}------------------------------------------------
#### 8. Indications for Use
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal(Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, Power Doppler (PD) mode, ElastoScan™ Mode, Multi-Image mode(Dual, Quad), Combined modes, 3D/4D mode
### 9. Technologv
The RS85 / RS80 EVO employ the same fundamental scientific technology as its predicate device.
- 10. Determination of Substantial Equivalence
Comparison to Predicate: The RS85 / RS80 EVO are substantially equivalent to the predicate device with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis. ●
- . The proposed RS85 / RS80 EVO and predicate RS85(K192903) have the same clinical intended use.
- . The proposed RS85 / RS80 EVO and predicate RS85(K192903) have the same imaging modes and modes of operation.
- . The proposed RS85 / RS80 EVO added two new transducers (LA4-18A, LA3-22AI).
- . The proposed RS85 / RS80 EVO and predicate RS85(K192903) have same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- . The proposed RS85 / RS80 EVO and predicate RS85(K192903) have been designed in compliance with approved electrical and physical safety standards
- . The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- . The systems have acoustic power levels which are below the applicable FDA limits.
RS85 and RS80 EVO are the same systems that provide only different option configurations. There is an addition of new transducers (LA4-18A, LA3-22AI) compare to previously cleared RS85 (K192903) system.
The material and design changes of MMPT3-7 transducer will be applied to all SAMSUNG
{5}------------------------------------------------
MEDISON products after 510(k) clearance.
The material change of Endocavity transducers (EV2-10A, EA2-11AR/EA2-11AV) will be applied to all SAMSUNG MEDISON products after 510(k) clearance.
## 11. Summary of Non-Clinical Test
The proposed devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The RS85 / RS80 EVO and its applications comply with voluntary standards.
| Test | Standards and FDA Guidance |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Risk Management | ISO 14971 Second edition 2007 Medical devices - Application of risk<br>management to medical devices |
| Electrical Safety | ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012<br>Medical Electrical Equipment - Part 1: General Requirements for basic safety<br>and essential performance. |
| Electromagnetic<br>Compatibility | IEC60601-1-2: 2014(4th Edition) Medical Electrical Equipment -- Part 1-2:<br>General Requirements For Basic Safety And Essential Performance --<br>Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests |
| Biocompatibility | AAMI / ANSI / ISO 10993-1:2009/(R)2013, Biological evaluation of medical<br>devices - Part 1: Evaluation and testing within a risk management process |
| Reprocessing Medical<br>Devices | FDA Guidance: Reprocessing Medical Devices in Health Care Settings:<br>Validation Methods and Labeling issued on March 17, 2015, revised June 9,<br>2017. |
| Software/Firmware-<br>driven Functionality | All migrated software functionality was evaluated using the same test criteria as<br>the predicate for all applicable imaging modes to ensure that migration into a<br>new system design did not compromise image quality with respect to the<br>intended use of each feature.<br>Guidance for the Content of Premarket Submissions for Software Contained in<br>Medical Devices issued on May 11, 2005 |
| Ultrasound Safety | Marketing Clearance of Diagnostic Ultrasound Systems and Transducers issued<br>June 27, 2019<br>IEC60601-2-37:2007 + A1:2015, Particular requirements for the safety of<br>ultrasonic medical diagnostic and monitoring equipment<br>NEMA UD 2-2004 (R2009)<br>Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment<br>Revision 3 |
## 12. Summary of Clinical Tests
The subject of this premarket submission, RS85 / RS80 EVO, did not require clinical studies to support substantial equivalence.
## 13. Conclusion
SAMSUNG MEDISON considers the RS85 and RS80 EVO to be as safe, as effective, and performance is substantially equivalent to the predicate device.
{6}------------------------------------------------
- END of 510(K) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.