K210928 · Pentax of America, Inc. · EOQ · Oct 21, 2021 · Ear, Nose, Throat
Device Facts
Record ID
K210928
Device Name
PENTAX Medical Video Bronchoscope EB11-J10
Applicant
Pentax of America, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Oct 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PENTAX Medical Video Bronchoscope EB11-J10 has been designed to be used with a PENTAX Video Processor (including Light source), documentation equipment, video monitor, endo-therapy accessories (such as biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.
Device Story
The PENTAX Medical Video Bronchoscope EB11-J10 is a flexible endoscope used for visualization and therapeutic access to the airways and tracheobronchial tree. It features a flexible insertion tube, control body for angulation, and a working channel for endo-therapy accessories like biopsy forceps. The device uses light-carrying bundles for illumination and a CCD sensor to capture endoscopic image data. It connects to a PENTAX Video Processor, which processes signals for display on a monitor. Operated by clinicians in clinical settings, the device assists in diagnosing and treating airway conditions. The system provides real-time visualization, allowing physicians to navigate the tracheobronchial tree, perform biopsies, or conduct other endoscopic procedures. The device is reprocessed between uses via high-level disinfection or sterilization (STERRAD NX/100NX).
Clinical Evidence
No human clinical trials were conducted. Evidence includes bench testing (optical performance, electrical safety per IEC 60601-1/2-18, EMC per IEC 60601-1-2), reprocessing validation (simulated use, soil analysis, disinfection/sterilization), and biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity). An animal image capture study demonstrated that the subject device visualizes vascularity and mucosal surfaces as well as or better than the predicate device.
Technological Characteristics
Flexible video bronchoscope with insertion tube, control body, and PVE connector. Features light-carrying bundles for illumination and a CCD sensor for imaging. Compatible with PENTAX video processors. Sterilization via STERRAD NX/100NX. Biocompatible materials assessed for cytotoxicity and sensitization. Electrical safety and EMC compliance per IEC 60601 series.
Indications for Use
Indicated for use in endoscopic surgery and visualization within the airways and tracheobronchial tree. Intended for use by trained clinicians in a clinical setting.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
PENTAX Video Bronchoscopes EB-J10 Series (K200678)
PENTAX Medical Video Processors EPK-i5010 (K143727)
PENTAX Medical Video Processors EPK-i7010 (K173679)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
October 21, 2021
PENTAX of America, Inc. William Goeller Vice President, Ouality Assurance and Regulatory Affairs 3 Paragon Drive Montvale, New Jersey 07645-1782
Re: K210928
Trade/Device Name: PENTAX Medical Video Bronchoscope EB11-J10 Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: September 20, 2021 Received: September 22, 2021
Dear William Goeller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210928
Device Name PENTAX Medical Video Bronchoscope EB11-J10
Indications for Use (Describe)
The PENTAX Medical Video Bronchoscopes EB-J10 Series have been designed to be used with a PENTAX Video Processor (including Light source), documentation equipment, video monitor, endo-therapy accessories (such as biopsy forceps) and other ancillary equipment for endoscopic surgery within the airways and tracheobronchial tree.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### PENTAX Medical Video Bronchoscope EB11-J10 Traditional 510(k) Submission
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 221 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.
Applicant: PENTAX Medical HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782
Contact: William Goeller Vice President, Quality/Regulatory Affairs PENTAX of America. Inc. 3 Paragon Drive Montvale, New Jersey 07645-1782 Email: william.goeller@pentaxmedical.com Phoner: 800-431-5880 Ext 2318 Fax: 201-571-2340
Date Prepared: 3/22/2021
Common Name: Bronchoscope Name of the System: PENTAX Medical Video Bronchoscope EB11-J10
Regulation Number: 21 CFR Part 874.4680 Regulation Names: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOO Predicate Device: PENTAX Video Bronchoscope EB-1170K (K131028)
PENTAX Medical is seeking clearance of EB11-J10 Video Bronchoscope (Hereinafter "the subject device" which is a new line of Video Bronchoscopes EB-J10 Series (K200678) with the compatible PENTAX Medical Video Processors EPK-3000, EPK-i5010 (K143727), and EPKi7010 with EB Family (K173679).
This premarket notification also captures some minor design changes that have occurred during the evolution of the product line. Although the changes are believed to be minor, the 510(k) is being submitted to account for technological advances in associated compatible devices and to ensure that FDA has the most current information concerning the subject device.
The subject device has virtually the same indications for use, viewing directions, and image size as the PENTAX Medical Video Bronchoscope EB-1170K (Hereinafter "the predicate device").
#### The main differences between the subject device and predicate device are as follows:
- Introducing the subject device as a new video bronchoscope to the Video Bronchoscopes ● EB-J10 series.
- The subject device has a new control body design with a sterile single-use suction control ● valve, OF-B205.
{5}------------------------------------------------
#### PENTAX Medical Video Bronchoscope EB11-J10 Traditional 510(k) Submission
- . Due to technological advances, the optical design of the subject device has been modified. Specifically, its sensor module assembly and glass rod are different with respect to the predicate device. The optical performance of the subject and predicate devices has been confirmed to be substantially equivalent through bench testing and animal imaging study results.
- Device firmware and a charge coupled device (CCD) has been updated to reflect technological modernization with higher resolution CCD.
- The subject device is compatible with the video processor EPK-3000 in addition to compatibility with EPK-i5010, EPK-i7010.
- . The subject device can be sterilized using the STERRAD 100NX® system in addition to STERRAD NX® system.
## Device Description:
The subject device is used to provide visualization of, and therapeutic access to, the airways and tracheobronchial tree. There are three models of EB-J10 series: the subject device, and the previously cleared models: EB15-J10 and EB19-J10 (K200678). These models are identical in all parameters and only differ in French size: 11, 15 and 19, respectively.
The subject device is used with cleared PENTAX Video Processors (a software- controlled device). The subject device has a flexible insertion tube, a control body, and PVE connector. The PVE connector will be attached to the Video Processor and has connections for illumination, video signals, air/ water and suction.
The control body includes controls for up/ down angulation, air/ water delivery, and an accessory inlet port. The subject device contains light carrying bundles (LCB) to illuminate the body cavity, and a CCD to collect endoscopic image data. The instrument contains a working channel through which biopsy devices, or other devices, may be introduced.
The video processor contains a lamp that provides white light and is focused at the PVE connector light guide prong. The endoscope light carrying bundles present the light to the body cavity and the CCD collects endoscopic image data. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display.
The subject device is immersible (with the use of supplied cleaning accessories) as described in the endoscope reprocessing instructions.
## Intended Use /Indications for Use (EB11-J10)
The PENTAX Medical Video Bronchoscope EB11-J10 has been designed to be used with a PENTAX Video Processor (including Light source), documentation equipment, video monitor, endo-therapy accessories (such as biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.
## Compatible Video Processors
In addition to compatibility with EPK-i5010, EPK-i7010 previously cleared with other PENTAX bronchoscopes (K143727; K173679), EB11-J10 is compatible with a new processor EPK-3000 that has been previously cleared to be used with PENTAX Medical Video Bronchoscopes EB-
{6}------------------------------------------------
#### PENTAX Medical Video Bronchoscope EB11-J10 Traditional 510(k) Submission
J10 series (K200678). EPK-3000 labeling has been revised to reflect compatibility of the processor with the EB11-J10.
### Summary of Technology Characteristics:
The subject device is functionally equivalent to the predicate device, the PENTAX Medical Video Bronchoscope EB-1170K (EB Family) cleared by FDA in 2014 (K131028). The only difference between the two devices are minor technological changes.
The changes in the subject device have been evaluated through performance testing and do not raise different issues of safety and effectiveness of the device as these differences have no effect on the performance, function or general intended use of the device.
### Non-Clinical Performance Data
The subject device has been successfully tested for their functions, performance and safety as per FDA recognized consensus standards. The following performance data are provided in support of the substantial equivalence determination.
### Reprocessing Validation
Simulated use testing, soil accumulation analysis, cleaning, high-level disinfection, and rinsing validation studies of the subject device and accessories were conducted and confirmed the effectiveness of reprocessing procedures.
#### Sterilization and Shelf Life
PENTAX Medical coordinated with Advanced Sterilization Products, Inc. to validate the use of STERRAD NX®/100NX® for the sterilization of the subject device and accessories. The subject device is not provided sterile.
#### Biocompatibility
Biocompatibility of the EB11-J10 scope on direct and indirect contact materials was confirmed by assessing the cytotoxicity, sensitization, and intracutaneous reactivity. The risk levels of local toxicity were determined as "Acceptable" as a result of applying the risk level of local toxicity to the risk evaluation criteria.
#### Software
Software verification and validation including cybersecurity assessments were conducted according to IEC 62304: 2006 and FDA Guidances for Industory and Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.", "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" and "Postmarket Management of Cybersecurity in Medical Devices."
## EMC and Electrical Safety
The acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES) for the PENTAX Medical Video Bronchoscope EB11-J10 was confirmed by the following standards: IEC 60601-1-2:2014; IEC 60601-1:2005+CORR 1:2006+CORR 2:2007+A1:2012; and IEC 60601-2-18:2009, and IEC 60601-1-2:2014.
{7}------------------------------------------------
### Optical Testing
As a part of Design Verification and Validation, optical properties of imaging and illumination performances were measured for subject device in conjunction with the EPK-3000, EPK-i7010, and EPK-i5010 Video Processors. All results show that the optical characteristics of the subject device is equivalent to those of the predicate device.
#### Animal Image Capture Study
An animal image capture study was performed as a part of optical and color performance testing. The results indicate that the subject device is able to visualize vascularity and mucosal surface for each anatomical area as well or better than the predicate device.
### Substantial Equivalence Discussion:
After analyzing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials), labeling, and sterilization method, PENTAX Medical concludes that the subject device PENTAX Medical Video Bronchoscopes EB11-J10 is as safe and effective as the predicate device. There are no differences in indications for use and intended use between the subject and predicate devices and are therefore, substantially equivalent.
The technological differences in terms of design features, performance characteristics and constituent materials are not substantive.
#### Conclusion:
Accordingly, PENTAX Medical believes the PENTAX Medical Video Bronchoscopes EB11-110 is substantially equivalent to the identified the predicate device, the PENTAX Video Bronchoscope EB-1170K, cleared by FDA in 2014 (K131028).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.