PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes

K173679 · Pentax of America, Inc. · PEA · Jul 3, 2018 · Gastroenterology, Urology

Device Facts

Record IDK173679
Device NamePENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes
ApplicantPentax of America, Inc.
Product CodePEA · Gastroenterology, Urology
Decision DateJul 3, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for bronchoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK -7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX video bronchoscopes.

Device Story

Video processor for bronchoscopic imaging; connects to PENTAX endoscopes, camera heads, light sources, and monitors. Transforms optical signals from endoscope into digital video for display. Features i-Scan (digital post-processing filter) and OE (optical enhancement using band-limited light) to improve visualization of mucosal microvasculature and surface structures. Used in clinical settings by physicians; provides real-time visual feedback to assist in diagnosis and treatment. Output viewed on external monitors; aids clinical decision-making by highlighting tissue characteristics. Does not replace histopathological sampling. Accessories include keyboard, foot switch, white balance adjuster, and condenser earth cable.

Clinical Evidence

No human clinical data. Evidence includes bench testing (Signal to Noise, Color, Resolution, Modulation Transfer Function, Distortion, Light Distribution, Spectral Distribution, Total Luminous Flux, Photobiological Safety) and a porcine animal study comparing visual experience of white light and i-Scan images between subject and predicate devices.

Technological Characteristics

Video processor with digital and optical imaging enhancement. Connectivity: keyboard, foot switch, white balance adjuster, condenser earth cable. Standards: CISPR 11:2009, IEC 60601-1:2005, IEC 60601-2-18:2009, ANSI/AAMI ES60601-1:2005, IEC 62304:2006. Software: moderate level of concern.

Indications for Use

Indicated for bronchoscopic diagnosis, treatment, and video observation in patients requiring endoscopic examination of the airway. Used as an adjunct to white light endoscopy for enhanced visualization of mucosal surface texture and blood vessels.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. July 3, 2018 PENTAX of America, Inc. William Goeller Vice President. Ouality/Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782 Re: K173679 Trade/Device Name: PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOO Dated: June 1. 2018 Received: June 4, 2018 Dear William Goeller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely yours, # Michael J. Ryan -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173679 Device Name PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes Indications for Use (Describe) The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for bronchoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK -7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX video bronchoscopes. Type of Use (Select one or both, as applicable) | <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |--------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in large, red, bold letters. Below that, there is a gray line, and below that, the word "MEDICAL" is in gray, bold letters. ## 510(k) Summary #### I. SUBMITTER PENTAX of America, Inc., HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782 Phone: 201-571-2318 Fax: 201-2340 Contact: William Goeller Date Prepared: November 8, 2017 #### II. DEVICE | Device Name | PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes | |---------------------|-------------------------------------------------------------------| | Common Name | Endoscopic Video Processor and Light Source | | Classification Name | Endoscopic video imaging system/component | | Regulation No. | 21 CFR Part 874.4680 | | Regulation Name | Bronchoscopes (Flexible or Rigid) and accessories | | Device Class | 2 | #### III. PREDICATE DEVICES The PENTAX Medical EPK-i5010 Video Processor with EB Family (K143727) is the predicate device for this 510(k) Premarket Notification as the processors have the same intended use and the minor technology differences do not raise new issues of safety and effectiveness. The PENTAX Medical EPK-i7010 Video Processor with GI Family (K150618) is utilized as a Reference Device for this 510(k) Premarket Notification as the processor is identical to the subject device. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold letters, and it is placed above a gray line. Below the gray line, the word "MEDICAL" is written in gray, bold letters. #### IV. DEVICE DESCRIPTION The purpose of this 510(k) is to expand the indications for the PENTAX Medical EPK-i7010 Video Processor with GI Family (K150618) to include the PENTAX Video Bronchoscopes, EB Family (k131028): EB-1970TK, EB-1170K, EB-1570K, EB-1970K, EB-1575K, EB-1975K, EB -1990i). The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for bronchoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK-i7010 Video Processor contains PENTAX i-Scan and Optical Enhancement (OE) technologies. The PENTAX i-Scan technology is a digital filter-based image enhancement technique with three modes, i-Scan 1, 2, and 3. PENTAX i-Scan™ is intended to give the user an enhanced view of the texture of the mucosal surface and blood vessels. In addition to i-Scan, OE is intended to provide alternative methods to improve blood vessel visibility by using band limited light illumination source. The OE technology emphasizes mucosal microvasculature and fine mucosal structures on the bronchial surface. The following accessories are provided for use with the PENTAX Medical EPK-i7010 Video Processor and EB Family of Scopes: - . Keyboard - input device for the video processor. - . Foot Switch - used to remotely control processor functions. - . White Balance Adjuster - used as the object of white balance feature. - Condenser Earth Cable - used to reduce high-frequency noise generated during highfrequency electro cautery device use with Pentax endoscopes. #### V. INDICATIONS FOR USE The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for bronchoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK-i7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX video bronchoscopes. #### COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE There are no differences between the subject and predicate devices with regard to intended use. Both subject and predicate devices are intended for illuminating the inside of the human body. {6}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Pentax Medical. The word "PENTAX" is written in red, bold letters on the top line. Below a gray line, the word "MEDICAL" is written in gray, bold letters. The subject and predicate video processors have virtually identical indications for use and are substantially equivalent with regard to design, technological characteristics, energy used, performance, safety, and effectiveness. The subject and reference processors are identical with regard to design, technological characteristics, energy used, performance, safety, and effectiveness. The sole difference is the addition of a bronchoscopic indication. There are no significant differences in technical characteristics that raise new questions of safety and effectiveness. The data provided in this 510(k) Premarket Notification support the equivalence of the subject and predicate devices. #### VI. PERFORMANCE DATA The following performance data are provided to support the equivalence of the subject and predicate devices. ## a. Electrical Safety and Electromagnetic Compatibility (EMC) The PENTAX Medical EPK-i7010 Video Processor and PENTAX Medical EB Family of Scopes were tested and comply with the following standards for electrical safety and EMC. - . CISPR 11:2009, A1:2010 Group 1 Class B - . IEC 61000-4-2:2008; IEC 61000-4-3:2006; A1:2007, A2:2010, IEC 61000-4-4:2012; IEC 61000-4-5:2005; IEC 61000-4-6:2013; IEC 61000-4-8:2009; IEC 61000-4-11:2004; IEC 61000-3-2:2014; IEC 61000-3-3:2013 - IEC 60601-1-2:2007; IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012; IEC 60601-2-18:2009 - ANSI/AAMI ES60601-1:2005 + A2:2010 + A1:2012 ● - IEC 62304:2006 ● - CAN/CSA-C22.2 NO. 60601-2-18:11 # b. Software Verification and Validation Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered a "moderate" level of concern based on the FDA guidance. #### c. Performance Testing #### Bench Testing Optical testing was performed with the PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes, as well as PENTAX Medical EKP-i5010 with EB Family. Testing consisted of Signal to Noise, Color (IEE), Resolution, Modulation Transfer Function, Distortion, Light Distribution, Spectral Distribution, Total Luminous Flux, and Photobiological Safety testing. It was concluded that the performance of the subject and predicate devices are equivalent. {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in large, red, bold letters. Below that is a gray line, and below that is the word "MEDICAL" in gray, bold letters. #### Animal Testing PENTAX Medical performed a porcine animal study to show the similarity of visual experience of the PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes (subject) in comparison to PENTAX Medical EKP-i5010 Video Processor with EB Family (predicate). A library of images was captured to show visual experience for both processors for white light images (WLE) and i-Scan images. Images of the pig larynx demonstrate similar visual experience between scopes. #### VIII. CONCLUSION The data submitted in support of the safety of the device and the hardware and software verification and validation demonstrate that the PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes performs as intended in the specified use conditions. The optical data analysis demonstrates the equivalence of the PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes to the predicate device. Furthermore, a porcine animal study shows the similarity of visual experience with PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes in comparison with the predicate device. The subject and predicate devices have the same intended use, and there are no significant differences in technical characteristics that raise different questions of safety and effectiveness. The data provided in this 510(k) Premarket Notification support the equivalence of the subject and predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...