K210912 · Globus Medical, Inc. · OWB · Aug 12, 2021 · Radiology
Device Facts
Record ID
K210912
Device Name
Excelsius3D
Applicant
Globus Medical, Inc.
Product Code
OWB · Radiology
Decision Date
Aug 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Excelsius3D™ is a mobile X-ray system designed for 2D fluoroscopy, 2D digital radiography, and 3D imaging of adult and pediatric patients. The system is indicated for use where a physician benefits from 2D and 3D information on anatomic structures and high contrast objects with high x-ray attenuation such as bony anatomy and metallic objects. Excelsius3D™ images are compatible with image guided systems such as ExcelsiusGPS®.
Device Story
Mobile X-ray system; performs 2D fluoroscopy, 2D digital radiography, and 3D imaging. Inputs: X-ray radiation; outputs: 2D/3D images for visualization and DICOM storage. Used in clinical/surgical settings by physicians to localize pathology for preoperative, intraoperative, and postoperative analysis. Assists in complex procedures like spinal deformity correction or deep brain stimulation (DBS) placement. Integrates with ExcelsiusGPS® image-guided systems. Provides real-time image capture and post-capture processing to aid clinical decision-making and surgical navigation.
Clinical Evidence
No clinical trials performed. Evidence consists of non-clinical system and software verification/validation, electrical safety/EMC testing per IEC standards (60601 series, 62304, 62366), image quality assessment compared to predicates, and qualitative human cadaveric validation in clinically relevant scenarios.
Technological Characteristics
Mobile X-ray system; isocentric gantry with 360° rotation; rotating anode X-ray tube; cone-beam collimation with x-y shutters. Connectivity via industry-standard protocols for DICOM storage. Software lifecycle per IEC 62304. Safety compliance: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-43, 60601-2-54, IEC 60825-1.
Indications for Use
Indicated for adult and pediatric patients requiring 2D fluoroscopy, 2D digital radiography, or 3D imaging of anatomic structures and high-contrast objects (bony anatomy, metallic objects).
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
August 12, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Globus Medical Inc. % Kelly Baker, Ph.D. Senior Vice President, Regulatory and Clinical Affairs 2560 General Armistead Ave. AUDUBON PA 19403
### Re: K210912
Trade/Device Name: Excelsius3D™ Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: July 20, 2021 Received: July 21, 2021
Dear Dr. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
. for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210912
Device Name Excelsius3D™
#### Indications for Use (Describe)
Excelsius3D™ is a mobile X-ray system designed for 2D fluoroscopy, 2D digital radiography, and 3D imaging of adult and pediatric patients. The system is indicated for use where a physician benefits from 2D and 3D information on anatomic structures and high contrast objects with high x-ray attenuation such as bony anatomy and metallic objects. Excelsius3D™ images are compatible with image guided systems such as ExcelsiusGPS®.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary: Excelsius3D™
| Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Kelly Baker, Ph.D.<br>Senior Vice President, Regulatory and Clinical Affairs |
| Date Prepared: | July 20, 2021 |
| Device Name: | Excelsius3D™ |
| Common Name: | Computer-assisted mobile image-intensified fluoroscopic and<br>X-ray system |
| Classification: | Per 21 CFR as follows:<br>§892.1650 Image-intensified Fluoroscopic X-ray System,<br>Mobile<br>Product Code(s): OWB, OXO, JAA<br>Regulatory Class: II |
Primary Predicate: Medtronic O-Arm Imaging System (K092564) Other Predicates: Medtronic O-Arm O2 Imaging System (K200074) Reference Device: GE Optima XR220amx/XR200amx (K142383)
#### Purpose:
The purpose of this submission is to request clearance of the Excelsius3D™
### Device Description:
Excelsius3D™ is a mobile X-ray imaging system that combines 2D fluoroscopy, 2D digital radiography, and 3D imaging into one unit for acquisition and visualization of 2D or 3D images of patient anatomy. Excelsius3D™ provides real-time image capture, post-capture processing and visualization, and DICOM storage of images.
Visualization of patient anatomy assists in localizing regions of pathology for preoperative examination, in vivo surgical procedures, and post-surgical analysis. The three imaging modalities can assist physicians with performing complex surqical procedures, such as spinal deformity correction or deep brain stimulation (DBS) placement.
{5}------------------------------------------------
### Indications for Use:
Excelsius3D™ is a mobile X-ray system designed for 2D fluoroscopy, 2D digital radiography, and 3D imaging of adult and pediatric patients. The system is indicated for use where a physician benefits from 2D and 3D information on anatomic structures and high contrast objects with high x-ray attenuation such as bony anatomy and metallic objects. Excelsius3D™ images are compatible with image quided systems such as ExcelsiusGPS®.
### Technological Characteristics:
The Excelsius3D™ has similar technological characteristics to the predicate devices including the main system components, workflow, user interface, software features, and design. The Excelsius3D™ is comparable to the predicates in terms of intended use, fundamental scientific technology, technological characteristics and principle of operation.
| | Subject<br>Excelsius3D™ | Predicate<br>O-Arm<br>(K092564, K200074) | Reference<br>Optima XR220amx<br>(K142383) |
|------------------------------------|--------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------|
| Intended<br>Applications | Orthopaedic, cranial | Orthopaedic, cranial | Orthopaedic, cranial<br>(& gastrointestinal,<br>thoracic, pulmonary) |
| Imaging<br>Modalities | 2D fluoroscopy<br>3D imaging<br>2D digital radiography | 2D fluoroscopy<br>3D imaging | 2D digital radiography |
| Gantry<br>Motion | Isocentric | Isocentric | N/A |
| Gantry<br>Rotation | 360° | 360° | N/A |
| Gantry Stored<br>Positions | Yes | Yes | N/A |
| X-ray Tube<br>Type | Rotating anode | Rotating anode | Rotating anode |
| X-ray Beam/<br>Collimation<br>Type | Cone beam with x-y<br>shutters | Cone beam with x-y<br>shutters | Cone beam with x-y<br>shutters |
| Image<br>Acquisition<br>Presets | Yes | Yes | Yes |
| Method of<br>Image<br>Acquisition | Hand switch<br>or foot pedal | Hand switch<br>or foot pedal | Hand switch |
| Image<br>Transfer | Yes | Yes | Yes |
| Cybersecurity | Industry standard<br>protocols | Industry standard<br>protocols | Industry standard<br>protocols |
#### Comparison of Principles of Operation and Technological Characteristics
{6}------------------------------------------------
# Performance Testing:
Verification and validation testinq were conducted on Excelsius3D™ to confirm that the device meets performance requirements under the indications for use and to ensure the system performs as intended.
- Non-clinical system and software verification and validation. ●
- Software validation and verification testing was performed in accordance o with the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). The software for this device is considered a "MAJOR" level of concern.
- Electrical Safety and Electromagnetic Compatibility ●
- o Testing was performed to assure compliance with recognized safety standards for electrical safety and electromagnetic compatibility
- IEC 60601-1:2012 Medical electrical equipment Part 1: General . requirements for basic safety and essential performance
- . IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
- IEC 60601-1-3:2013 Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- l IEC 60601-1-6:2013 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- .. IEC 60601-2-43:2010 Medical electrical equipment - Part 2-43: Particular requirements for the safety and essential performance of Xray equipment for interventional procedures
- IEC 60601-2-54:2015 Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- I IEC 60825-1:2007 Safety of laser products - Part 1: Equipment classification, and requirements
- . IEC 62304:2015 Medical device software - Software lifecycle processes
- . IEC 62366:2015 Medical devices - Part 1: Application of usability engineering to medical devices
- Image quality assessment comparison and paired image analysis to ● predicates
- Human cadaveric qualitative validation under clinically relevant scenarios ●
# Basis of Substantial Equivalence:
Excelsius3D™ has been found to be substantially equivalent to the predicate devices with respect to technological characteristics, imaging performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.