The O-arm O2 Imaging System is a mobile x-ray system designed for adults and pediatric patients weighing 601bs or greater and having an abdominal thickness greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomical structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System in compatible with certain image guided surgery systems.
Device Story
Mobile cone-beam X-ray system providing 2D and 3D imaging for intraoperative use. System comprises Image Acquisition System (IAS) and Mobile View Station (MVS). IAS features motorized gantry with X-ray source and flat-panel detector; performs 360-degree rotation for 3D reconstruction. New 4.2.0 software adds '2D Long Film' protocol, utilizing automated gantry motion and slot-filter collimation to stitch projection images into a single long-format 2D image. System displays gantry rotor/tilt angles on pendant. MVS provides visualization, window/leveling, and oblique view rendering. Integrates with image-guided surgery systems. Operated by physicians/technicians in clinical settings. Enhanced cybersecurity includes LDAP authentication, PHI encryption, and audit logging. Output assists surgical planning and navigation; benefits include improved visualization of bony anatomy and metal implants during procedures.
Clinical Evidence
Bench testing only. Included software V&V, hardware verification, dosimetry measurements, and usability testing per FDA human factors guidance. Image quality assessment compared 4.2.0 to 4.1.0. A cadaver image pair study evaluated clinical utility of images between versions.
Technological Characteristics
Mobile cone-beam X-ray system. 32 kW generator; 40x30 cm CsI flat-panel detector. Motorized gantry with isocentric motion. Connectivity: Ethernet/networked for hospital IT integration (LDAP). Power: 100/120/240 VAC. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-43, IEC 60601-1-6.
Indications for Use
Indicated for 2D and 3D imaging in adult and pediatric patients weighing ≥60 lbs with abdominal thickness >16 cm. Used where physician benefits from 2D/3D anatomical information and high-attenuation objects (bony anatomy, metal). Compatible with image-guided surgery systems.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
April 24, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medtronic, Inc. % Mr. Dean Honkonen Sr. Regulatory Affairs Manager 300 Foster Street LITTLETON MA 01460
Re: K200074
Trade/Device Name: Medtronic O-arm™ O2 Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: March 27, 2020 Received: March 30, 2020
Dear Mr. Honkonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200074
Device Name Medtronic O-arm™ O2 Imaging System
#### Indications for Use (Describe)
The O-arm O2 Imaging System is a mobile x-ray system designed for adults and pediatric patients weighing 601bs or greater and having an abdominal thickness greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomical structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
The O-arm O2 Imaging System in compatible with certain image guided surgery systems.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| Submitter: | Medtronic Navigation, Inc. (Littleton)<br>300 Foster Street<br>Littleton, MA 01460 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dean Honkonen (Primary)<br>Sr. Regulatory Affairs Manager<br>Phone: (978) 698-6065<br>Fax: (978) 698-6090<br>e-mail: dean.honkonen@medtronic.com<br>Tijana Prodanovic (Alternate)<br>Sr. Regulatory Affairs Specialist<br>Phone: (978) 339-4134<br>Fax: (978) 698-6090<br>e-mail: tijana.prodanovic@medtronic.com |
| Date Summary Prepared: | January 13, 2020 |
| Device Trade Name: | Medtronic O-arm™ O2 Imaging System |
| Common Name: | Interventional Fluoroscopic X-ray System |
| Device Classification: | Class II |
| Product Code: | Primary OWB<br>Secondary OXO, JAA |
| Classification Name: | 892.1650 - Image Intensified Fluoroscopic X-ray System, Mobile |
| Predicate Device: | K173664 – O-arm™ O2 Imaging System 4.1.0 software |
#### Device Description:
The O-arm™ O2 Imaging System is a mobile x-ray system that provides 3D and 2D imaging. O-arm O2™ Imaging System was originally cleared for market under the original 510(k) K151000 and subsequently via special 510(k) K173664. The device is classified under primary product code OWB (secondary product codes OXO, JAA) ref 21 CFR 892.1650.
{5}------------------------------------------------
## Modified Device:
This submission for the O-arm™ O2 Imaging System 4.2.0 software release introduces following features:
- . "2D Long Film" Imaging Protocol (Intraoperative Radiographic Scan)
- . Gantry Rotor Angle and Tilt Angle Display
- User Access Enhancements ●
These features are described in more detail in the Substantial Equivalence section.
The O-arm™ O2 Imaging System consists of two main assemblies that are used together:
- The Image Acquisition System (IAS) ●
- The Mobile View Station (MVS) .
The two units are interconnected by a single cable that provides power and signal data. The IAS has an internal battery pack that provides power for motorized transportation and gantry positioning. In addition, the battery pack is used to power the X-ray tank. The MVS has an internal UPS to support its function when mains power is disconnected.
The O-arm™ O2 Imaging System operates off standard line voltage within the following voltages:
- VAC 100, 120 or 240 ●
- Frequency 60Hz or 50Hz ●
- Power Requirements 1440 VA
### Indications for Use:
The O-arm™ O2 Imaging System is a mobile x-ray system designed for 2D and 3D imaging for adult and pediatric patients weighing 601bs or greater and having an abdominal thickness greater than 16cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
The O-arm™ O2 Imaging System is compatible with certain image guided surgery systems.
### Substantial Equivalence:
O-arm™ O2 Imaging System with 4.2.0 software is substantially equivalent to the following device:
K173664 - O-arm™ O2 Imaging System 4.1.0 software The unmodified device (predicate - K173664) and the modified device are compared in table below:
{6}------------------------------------------------
# Device Comparison Table
| | Predicate Device | Modified Device | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | O-arm O2 Imaging System<br>K173664 | With revision 4.2.0 software | Discussion |
| Classification | Class II | Class II | Identical |
| Product Code | OWB; 892.1650 | OWB; 892.1650 | Identical |
| Indications for Use | The O-arm O2 Imaging System is a<br>mobile x-ray system designed for 2D<br>fluoroscopic and 3D imaging for adult<br>and pediatric patients weighing 60lbs or<br>greater and having an abdominal<br>thickness greater than 16cm and is<br>intended to be used where a physician<br>benefits from 2D and 3D information of<br>anatomic structures and objects with high<br>x-ray attenuation such as bony anatomy<br>and metallic objects.<br>The O-arm O2 Imaging System is<br>compatible with certain image guided<br>surgery systems. | The O-arm O2 Imaging System is a<br>mobile x-ray system designed for 2D and<br>3D imaging for adult and pediatric<br>patients weighing 60lbs or greater and<br>having an abdominal thickness greater<br>than 16cm and is intended to be used<br>where a physician benefits from 2D and<br>3D information of anatomic structures and<br>objects with high x-ray attenuation such<br>as bony anatomy and metallic objects.<br>The O-arm O2 Imaging System is<br>compatible with certain image guided<br>surgery systems. | Changed to remove the word<br>"fluoroscopic".<br>O-arm O2 Imaging System 4.2.0<br>software uses more than one 2D<br>mode. Modes of operation are<br>described in the Instructions for Use. |
| Cone Beam CT | The O-arm O2 Imaging System is a<br>mobile cone-beam x-ray system with<br>isocentric motion options. It allows 3D<br>image reconstruction using a 360 degree<br>rotation of the x-ray source and detector<br>within closed gantry. | The O-arm O2 Imaging System is a<br>mobile cone-beam x-ray system with<br>isocentric motion options. It allows 3D<br>image reconstruction using a 360 degree<br>rotation of the x-ray source and detector<br>within closed gantry. | Identical |
| | Predicate Device | Modified Device | |
| | O-arm O2 Imaging System K173664 | With revision 4.2.0 software | Discussion |
| Detector Technology | 40 x 30 cm (RoHS compliant, Flat-Panel Detector using a CsI scintillation) | 40 x 30 cm (RoHS compliant, Flat-Panel Detector using a CsI scintillation) | Identical |
| Generator Technology | 32 kW, RoHS compliant generator | 32 kW, RoHS compliant generator | Identical |
| 2D Imaging | 2D Fluoroscopic | 2D Fluoroscopic | Identical |
| 2D Imaging | Manually stitched fluoroscopy | Added Feature:<br>Automatically stitched 2D Radiographic<br>(Long Film) | Added Feature:<br>The O-arm O2 2D Long Film feature<br>leverages the motion capability of<br>the O-arm to provide an imaging<br>mode, whereby 2D narrowly<br>collimated projection images would<br>be automatically acquired while the<br>gantry is moving from two<br>preprogramed gantry locations.<br>These projections would then be<br>automatically stitched together using<br>knowledge of the motion profile to<br>form a single 2D Long Film. |
| 3D Imaging<br>(20 cm FOV) | Full Fan (20cm FOV) scan acquisition | Full Fan (20cm FOV) scan acquisition | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
| | Predicate Device | | Modified Device | | | Discussion |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|------------------------------------------------------------------------------------|
| | O-arm O2 Imaging System<br>K173664 | | With revision 4.2.0 software | | | |
| 3D Imaging Protocols<br>(20 cm FOV) | Available presets:<br>Standard 3D<br>HD3D (High Definition)<br>Enhanced Cranial<br>Low Dose 3D | | Available presets:<br>Standard 3D<br>HD3D (High Definition)<br>Enhanced Cranial<br>Low Dose 3D | | | Identical |
| 3D Imaging<br>(40 cm FOV) | Half-fan single scan acquisition | | Half-fan single scan acquisition | | | Identical |
| 3D Imaging Protocols<br>(40 cm FOV) | Available presets:<br>HD3D (high definition) equivalent to<br>750 projections<br>Stereotaxy protocols | | Available presets:<br>HD3D (high definition) equivalent to<br>750 projections<br>Stereotaxy protocols | | | Identical |
| Annotation | Allows for adding arrows lines and text to<br>2D images<br>Additional annotation capability to<br>perform angular measurements onto a 2D<br>images. These measurements include<br>closed, open and Cobb angles. It also<br>provides the ability to place a right angle<br>on the image. | | Allows for adding arrows lines and text to<br>2D images<br>Additional annotation capability to<br>perform angular measurements onto a 2D<br>images. These measurements include<br>closed, open and Cobb angles. It also<br>provides the ability to place a right angle<br>on the image. | | | Changed to accommodate the larger<br>image size of 2D Radiographic<br>(Long Film). |
{9}------------------------------------------------
| | Predicate Device | Modified Device | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | O-arm O2 Imaging System<br>K173664 | With revision 4.2.0 software | Discussion |
| Image Transfer | Automatically transfers auto-registered<br>navigation scans.<br><br>Easy Image Transfer:<br>Depending upon the clinical application<br>and workflow within the procedure, this<br>will automatically transfer non-auto-<br>registered (non-navigated) images to the<br>navigation system | Automatically transfers auto-registered<br>navigation scans.<br><br>Easy Image Transfer:<br>Depending upon the clinical application<br>and workflow within the procedure, this<br>will automatically transfer non-auto-<br>registered (non-navigated) images to the<br>navigation system | Identical |
| 3D Visualization<br>(Enhanced Dynamic<br>Range) | 3D visualization of CBCT image on the<br>MVS. It allows the user to window level<br>the images as well as render oblique<br>views<br>Improved visualization of images that<br>contain objects of high-x-ray attenuation<br>such as metal implants on the Mobile<br>View Station. | 3D visualization of CBCT image on the<br>MVS. It allows the user to window level<br>the images as well as render oblique<br>views<br>Improved visualization of images that<br>contain objects of high-x-ray attenuation<br>such as metal implants on the Mobile<br>View Station. | Identical |
{10}------------------------------------------------
| Cybersecurity | | Added Feature: | Added Feature |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Industry standard protocols with error<br>detection for data transmission and<br>storage.<br><br>Authentication that includes user<br>names and passcodes<br><br>Software integrity check | Enhanced user access | • The ability to have a user log in to<br>the O-arm system through the<br>Lightweight Directory Access<br>Protocol (LDAP) by using hospital<br>domain credentials set up by hospital<br>IT.<br>• The ability for the hospital IT team<br>to configure group policies on their<br>hospital IT systems and assign the<br>proper user group to individuals with<br>O-arm user and/or O-arm hospital<br>administrator privileges.<br>• The ability to maintain legacy O-<br>arm O2 4.1.0 local authentication<br>methods if customers prefer this<br>method over LDAP.<br>• The ability for Medtronic Service<br>or Manufacturing to configure the<br>authentication feature based on<br>customer requests.<br>• The ability for Authorized<br>personnel to audit changes made<br>within the system related to<br>Protected Health Information (PHI),<br>User Management and System<br>Configuration.<br>• Encryption of PHI information<br>stored on MVS computer for easier<br>integration of the improved User<br>Access Enhancements into various<br>Hospital IT systems. |
{11}------------------------------------------------
| | Predicate Device | Modified Device | |
|-------------------------------------------------|--------------------------------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | O-arm O2 Imaging System<br>K173664 | With revision 4.2.0 software | Discussion |
| Hardware – Collimator<br>Assembly | No slot filter present in Collimator<br>Assembly | Changed hardware<br>Collimator Assembly includes a slot filter<br>for 2D Long Film Imaging | Changed hardware<br>Collimator Assembly includes a slot<br>filter for 2D Long Film Imaging<br>capability. New filter ladder<br>assembly designed for 2D Long Film<br>allows for higher repeatability of<br>filter positioning than the current<br>design. |
| Software - Gantry Tilt<br>& Rotor Angle Display | Not present | Added Feature<br>Display of the Gantry Rotor Angle and<br>Gantry Tilt Angle on the Pendant | Added feature:<br>Gantry Tilt & Rotor Angle Display<br>will provide following capability:<br>• Will enable users to view the tilt<br>angles of the gantry from the O-<br>arm pendant.<br>• Will enable users to view the<br>relative angle/location of X-ray<br>source & detector. |
{12}------------------------------------------------
## Performance Testing:
Testing conducted demonstrates the product will perform as intended according to the outlined design requirements. The following testing was conducted on the O-arm™ O2 Imaging System device to establish substantial equivalence with the predicate device and verify that device will perform as intended in meeting all the design inputs:
- AAMI/ANSI ES 60601-1 2005+A1:2012 Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance
- · IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General requirements for safety; Electromagnetic Compatibility - Requirements and Tests
- IEC 60601-1-3:2008 + A1:2013 Medical Electrical Equipment Part 1-3: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Radiation Protection in Diagnostic X-ray Equipment
- IEC 60601-2-28:2010 Medical electrical equipment part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
- IEC 60601-2-43:2010 + A1:2017 Medical electrical equipment Part 2-43: Particular requirements for the basic safety and essential
- IEC60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- · Software Verification and Validation testing verifying the software requirements perform as intended.
- · Hardware verification ensuring the hardware requirements identified for the system perform as intended.
- · The Dosimetry Report documents the dosimetry measurements for the various modes of the O-arm™ O2 Imaging System with 4.2.0 software.
- · Usability Testing was conducted according to the FDA guidance Applying Human Factors and Usability to Optimize Medical Device Design. Users conducted a series of imaging functions under simulated use conditions.
- · The Image Quality Assessment of the O-arm™ O2 Imaging System provides a quantitative image quality assessment of the O-arm O2 Imaging System with 4.2.0 in comparison to the predicate O-arm™ O2 Imaging System with 4.1.0.
- · The O-arm™ Cadayer Image Pair Study evaluated the clinical utility of the images obtained using the O-arm™ O2 Imaging System with 4.2.0 compared to the images obtained using the predicate O-arm™ O2 Imaging System with 4.1.0.
## Conclusion:
The O-arm™ O2 Imaging System with 4.2.0 software is similar in technological characteristics, imaging performance and indications for use as the predicate device listed. These aspects, along with the functional testing conducted to the FDA recognized standards, demonstrate that Oarm™ O2 Imaging System with 4.2.0 software does not raise new risks of safety and effectiveness when compared to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.