K210792 · Auxein Medical Private Limited · HWC · Aug 25, 2021 · Orthopedic
Device Facts
Record ID
K210792
Device Name
Auxein Nailing System
Applicant
Auxein Medical Private Limited
Product Code
HWC · Orthopedic
Decision Date
Aug 25, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
AUXEIN Brand of Expert Tibia Nailing System- consists of variety of Bone nails and bone screws. Expert Tibia Nail System is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions. AUXEIN Brand of Expert Femoral Nailing System -is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions. AUXEIN Brand of Gamma Nailing System -The Gamma Nail with Anti -Rotation Screw is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections. AUXEIN Brand of AJAX Advance Nailing System, end cap and screws is intended to treat stable and unstable proximal fracture including Pertrochanteric fractures, Intertrochanteric fractures and High sub trochanteric fractures and combination of these fractures. The AUXEIN MEDICAL's Elastic Intramedullary Nail System is indicated for fixation of diaphyseal fractures where the canal is narrow or flexibility of the implant is important. This includes upper extremity fractures in all patients and lower extremity fractures in children above the age of 6 years and any patient below 65 kg. This system is also intended to treat metaphyseal and epiphyseal fractures, such as radial neck fractures. AUXEIN Brand of Humeral Nailing System is indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures, proximal humeral fractures with diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures. AUXEIN Brand of Retrograde Nailing System This nails are indicated for simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated.
Device Story
Auxein Nailing System comprises titanium alloy (Ti-6Al-4V ELI) intramedullary nails, locking bolts, screws, and end caps; used for orthopedic fracture fixation. System includes Expert Tibia, Expert Femoral, Gamma, AJAX Advance, Elastic, Humeral, and Retrograde nailing modules. Devices are implanted by surgeons in clinical settings to stabilize bone fractures; facilitate healing via rigid or flexible internal fixation. Implants are supplied sterile via gamma irradiation. System includes associated stainless steel instruments (screwdrivers, holders, containers). Output is mechanical stabilization of bone segments; aids in patient recovery and bone union.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via mechanical testing including fatigue testing, screw cut-out tests (ASTM F1264), and compliance with ASTM F543 and F384. Sterilization validated per ISO 11737-2 and accelerated aging per ASTM F1980.
Technological Characteristics
Materials: Titanium alloy Ti-6Al-4V ELI (ASTM F136) for implants; Stainless Steel (ASTM F899) for instruments. Principle: Intramedullary fixation rod. Form factor: Various nail diameters (1.5mm-13mm) and lengths. Sterilization: Gamma irradiation. Connectivity: None (mechanical device).
Indications for Use
Indicated for stabilization of long bone fractures (tibia, femur, humerus) including shaft, proximal, distal, comminuted, spiral, oblique, segmental, and pathologic fractures; non-unions; malunions; polytrauma; tumor resections; and bone lengthening/shortening. Elastic nails indicated for pediatric patients (>6 years, <65 kg) and narrow canals. Specific contraindications not listed.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Auxein Medical Private Limited Rahul Luthra Director Plot No.168,169,170 Phase-IV, Sector 57, Kundli Industrial Area Sonipat, Haryana 131028 India
## Re: K210792
Trade/Device Name: Auxein Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, JDS Dated: March 15, 2021 Received: March 15, 2021
#### Dear Rahul Luthra:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
August 25, 2021
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210792
Device Name Auxein Nailing System (Sterile)
#### Indications for Use (Describe)
Device Name: Auxein Nailing System are categorized to the following system:
1. Expert Tibia Nailing System 2. Expert Femoral Nailing System 3. Gamma Nailing System 4. AJAX Advance nailing system 5. Elastic Nailing System 6. Humeral Nailing System 7. Retrograde Nailing System
INDICATION FOR USE:
1. AUXEIN Brand of Expert Tibia Nailing System- consists of variety of Bone nails and bone screws. Expert Tibia Nail System is intended to stabilize fractures of the proximal and distal tibial shaft; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions.
2. AUXEIN Brand of Expert Femoral Nailing System -is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions.
3. AUXEIN Brand of Gamma Nailing System -The Gamma Nail with Anti -Rotation Screw is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections.
4. AUXEIN Brand of AJAX Advance Nailing System, end cap and screws is intended to treat stable and unstable proximal femoral fracture including Pertrochanteric fractures and High sub trochanteric fractures and combination of these fractures.
5. The AUXEIN MEDICAL's Elastic Intramedullary Nail System is indicated for fixation of diaphyseal fractures where the canal is narrow or flexibility of the important. This includes upper extremity fractures in all patients and lower extremity fractures in children above the age of 6 years and any patient below 65 kg. This system is also intended to treat metaphyseal and epiphyseal fractures, such as radial neck fractures.
6. AUXEIN Brand of Humeral Nailing System is indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures, proximal humeral fractures with diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures.
7. AUXEIN Brand of Retrograde Nailing System This nails are indicated for simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated."
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Auxein. The logo consists of a stylized letter "A" inside a circle on the left, followed by the word "Auxein" in a bold, sans-serif font. The color of the logo is blue.
CORPORATE OFFICE:
7-A, Nirankari Colony,
Delhi - 110 009, India
MANUFACTURING UNIT
Plot No. 169-170, Phase-IV,
Sector-57, Kundli Industrial Area,
HSIIDC, Sonepat-131028, Haryana, India
+91 99106 43658
+91 86077 70197 (FAX)
www.auxein.com
# 510K Summary
Premarket Notification 510(k) Summary as required by Section 807.92
General Company Information as required by 807:92(a).
(A.1) The submitter's name, address, telephone number, a contact person, and the date the summary was prepared.
| Submitter's Name: | Auxein Medical Private Limited |
|----------------------|------------------------------------------------------------------|
| Address: | Auxein Medical Pvt. Ltd. |
| | Plot No. 168-169-170, Phase- IV, Kundli Industrial Area, HSIIDC, |
| | Sector-57, Sonepat-131028, Haryana. |
| Contact Person Name: | Mr. Rahul Luthra |
| Title: | Director |
| Phone Number: | +91-9811720999 |
| Dated: | 10/03/2021 |
Throughout the submission Auxein Nailing System is covered under 510(k) Submission.
# (A.2) The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
#### Proprietary Name:
Auxein Nailing System
#### Common or Usual Name:
Rod, Fixation, Intramedullary and Accessories
Nail, Fixation, Bone
#### Classification Name:
- Intramedullary Fixation Rod.
- Single/multiple component metallic bone fixation appliances and accessories.
#### Product Code:
HSB, JDS
Device Class: II
Review Panel: Orthopaedic
Regulation Number: 21 CFR 888.3020, 21 CFR 888.3030
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Auxein. The logo consists of a blue circle with a white "A" inside of it, followed by the word "Auxein" in blue. The font is sans-serif and the logo is simple and modern.
CORPORATE OFFICE:
7-A, Nirankari Colony,
Delhi - 110 009, India
MANUFACTURING UNIT
Plot No. 169-170, Phase-IV,
Sector-57, Kundli Industrial Area,
HSIIDC, Sonepat-131020, Haryana, India
+91 99106 43658
+91 86077 70197 (FAX)
www.auxein.com
### Variants/Types:
An intramedullary rod, also known as an intramedullary nail (IM nail) or inter-locking nail is a metal rod forced into the medullary cavity of a bone. IM nails have long been used to treat fractures of long bones of the body.
The device is made from a Titanium alloy, Ti-6Al-4V ELI (ASTM F136).
# (A.3) Identification of the Predicate Device:
Following are the predicate device 510(k) with which we are declaring substantial equivalence:
The following is the range of variants covered with their corresponding predicate devices.
| S.No. | Item Description | Predicate Device |
|-------|-----------------------|------------------------------------------------------------------------------------------------------------------|
| 1. | Auxein Nailing System | K192003, Auxein Nailing System, (Primary Predicate)<br>K011857, Synthes USA PFNA-II Blade, (Reference Predicate) |
Note: K011857, Synthes USA PFNA-II Blade is a Reference predicate device for (Ajax Blade)
## (A.4). A description of the device that is the subject of the premarket notification submission, such as might be found in the labelling or promotional material for the device.
#### Device Description:
#### Nail Fixation
Auxein Brand of Expert Tibia Nail system consists of various shapes and sizes of sterile nails featuring proximal & distal holes, end caps & locking bolts of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones. The diameter of the nails varies from 8mm to 12mm; the length varies from 255mm. The screw implants are in corresponding diameter of 4.4mm & 4.8mm diameters with lengths varying as per the requirements and minimum length: 24mm to maximum length 90mm.
Auxein Brand of Expert Femoral Nail system consists of various shapes and sizes of sterile nails featuring proximal & distal holes, end caps, locking screw & Anti-rotation screw of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones.The diameter of the nails varies from 9mm to 13mm; the length varies from 320mm to 440mm.The screw implants are in corresponding diameter of 4.8mm, 6.4mm & 10.5mm diameters with lengths varying as per the requirements and minimum length: 25mm to maximum length 130mm.
Auxein Brand of Ga-mma Nailing system consists of various shapes and sizes of sterile nails featuring proximal & distal holes, end caps, Inner screw, locking bolts, Anti-rotation
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Auxein. The logo consists of a stylized letter "A" inside of a circle on the left, followed by the word "Auxein" in blue text on the right. The font is sans-serif and the color is a medium blue.
Image /page/6/Picture/1 description: This image shows contact information for a company. The corporate office is located at 7-A, Nirankari Colony, Delhi-110009, India. The manufacturing unit is located at Plot No. 169-170, Phase-IV, Sector-57, Kundia Industrial Area, 15HOC, Sonepat-131028, Haryana, India. The phone number is +91 9910643658, and the fax number is +91 86077 70197. The website is www.auxein.com.
screw & Cephalic screw of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones. The diameter of the nails varies from 9mm to 13mm; the length varies from 180mm. The screw implants are in corresponding diameter of 4.8mm, 6.4mm & 10.5mm diameters with lengths varying as per the requirements and minimum length: 26mm to maximum length 120mm.
Auxein Brand of Humeral Nailing system consists of various shapes and sizes of sterile nails featuring proximal & distal holes, end caps, locking bolt, compression screw, and proximal screw of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones. The diameter of the nails varies from 6mm to 9mm; the length varies from 180mm to 320mm. The screw implants are in corresponding diameter of 3.5mm & 4.5mm diameters with lengths varying as per the requirements and minimum length: 20mm to maximum length 70mm.
Auxein Brand of AJAX Advance Nailing system consists of various shapes and sizes of sterile nails featuring proximal & distal holes, end caps, locking bolts, Cephalic screw, Ajax Blade, and inner screw of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones. The diameter of the nails varies from 9mm to 12mm: the length varies from 170mm to 420mm.The screw implants are in corresponding diameter of 4.9mm & 10.5mm diameters with lengths varying as per the requirements and minimum length: 26mm to maximum length 120mm.
Auxein Brand of Elastic Nailing system consists of various shapes and sizes of sterile nails featuring end caps of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones. The diameter of the nails varies from 1.5mm to 4.0mm; the total length of 440mm. The end caps are in corresponding diameter ranges from 1.5mm to 4.0mm diameters with length of 14mm & 25mm.
Auxein Brand of Retrograde Nailing system consists of various shapes and sizes of sterile nails featuring proximal & distal holes, end caps & locking bolts of different sizes. The nails and screws are fabricated from titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of bones. The diameter of the nails varies from 9.5 to 12mm; the length varies from 180mm to 360mm. The screw implants are in corresponding diameter of 5.0mm diameters with lengths varying as per the requirements and minimum length: 32mm to maximum length 80mm.
Auxein Nailing System covered same products as of covered in K192003. In this addition Auxein Nailing System is offering sterile version of the nailing devices.
The device is made from a Titanium alloy, Ti-6Al-4V ELI (ASTM F136).
However, we are adding a new device i.e. Ajax Blade in this submission. The associated Instruments with the newly added device are tabulated below.
| S.No. | Item Description | Material |
|-------|-------------------------------|---------------------------------|
| 1. | Inner Screwdriver, Hex 4.0mm | Stainless Steel as Per ASTM 899 |
| 2. | Inner Screw Holder, Hex 4.0mm | Stainless Steel as Per ASTM 899 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Auxein. The logo consists of a blue circular emblem with a stylized "A" inside, followed by the word "Auxein" in a blue, sans-serif font. The overall design is clean and professional, suggesting a corporate or business identity.
Image /page/7/Picture/1 description: The image shows contact information for a company. The corporate office is located at 7-A, Nirankari Colony, Delhi - 110 009, India. The manufacturing unit is located at Plot No. 169-170, Phase-IV, Sector-57, Kundli Industrial Area, SIIDC, Sonepat-131020, Haryana, India. The phone number is +91 99106 43658, the fax number is +91 86077 70197, and the website is www.auxein.com.
| 3. | Container for AJAX Cephalic Screw<br>Instrument Set | Stainless Steel as Per ASTM 8 |
|----|-----------------------------------------------------|-------------------------------|
|----|-----------------------------------------------------|-------------------------------|
These implants will be supplied in Sterile State (Gamma Irradiation) Only.
# (A.5) A statement of the intended use of the device.
#### Indications for Use:
#### AUXEIN Brand of Expert Tibia Nailing System
AUXEIN Brand of Expert Tibia Nailing System- consists of variety of Bone nails and bone screws. Expert Tibia Nail System is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions.
#### AUXEIN Brand of Expert Femoral Nailing System
AUXEIN Brand of Expert Femoral Nailing System -is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions.
#### AUXEIN Brand of Gamma Nailing System
AUXEIN Brand of Gamma Nailing System -The Gamma Nail with Anti -Rotation Screw is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections.
#### AUXEIN Brand of AJAX Advance Nailing System,
AUXEIN Brand of AJAX Advance Nailing System, end cap and screws is intended to treat stable and unstable proximal fracture including Pertrochanteric fractures, Intertrochanteric fractures and High sub trochanteric fractures and combination of these fractures.
#### The AUXEIN MEDICAL's Elastic Intramedullary Nail System
The AUXEIN MEDICAL's Elastic Intramedullary Nail System is indicated for fixation of diaphyseal fractures where the canal is narrow or flexibility of the implant is important. This includes upper extremity fractures in all patients and lower extremity fractures in children above the age of 6 years and any patient below 65 kg. This system is also intended to treat metaphyseal and epiphyseal fractures, such as radial neck fractures.
#### AUXEIN Brand of Humeral Nailing System-
AUXEIN Brand of Humeral Nailing System is indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures, proximal
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Auxein. The logo consists of a stylized letter "A" inside of a circle, followed by the word "Auxein" in a sans-serif font. The logo is blue in color and the wordmark is to the right of the symbol.
Image /page/8/Picture/1 description: The image shows contact information for a company. The corporate office is located at 7-A, Nirankari Colony, Delhi - 110009, India. The manufacturing unit is located at Plot No. 169-170, Phase-IV, Sector-57, Kundli Industrial Area, 15HDC, Sonepat-131020, Haryana, India. The phone number is +91 9910643638, and the fax number is +91 86077 70197. The website is www.auxein.com.
humeral fractures with diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures.
### AUXEIN Brand of Retrograde Nailing System
AUXEIN Brand of Retrograde Nailing System This nails are indicated for simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated."
#### (A.6). Summary of Technological Characteristics as compared to the predicate devices:
#### Substantial equivalence including comparison with predicate devices.
A comparison between the Auxein Nailing System and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria.
| S.No. | Characteristics | Predicate Device Versus New Device<br>(Auxein Brand) | Remarks |
|-------|--------------------------|----------------------------------------------------------------------------------------------------------|------------|
| 1. | Indications for use | Identical Indications for use. | Equivalent |
| 2. | Material | Same material used in New Device and<br>Predicate device. | Equivalent |
| 3. | Performance Standards | Same performance standards used in<br>both New Device as well as predicate<br>device. | Equivalent |
| 4. | Sterilization (Gamma) | Earlier the predicate device was supplied<br>Non-sterile but the now device will be<br>supplied Sterile. | Equivalent |
| 5. | Dimensional Verification | Same dimensions are found in both New<br>Devices as well as Predicate device. | Equivalent |
Following is the summary of parameters in which the comparison has been verified:
#### (C.1). Discussion on the non-clinical testing performed
Following are the applicable product standards considered for non-clinical standards:
A: Material Standards
#### B: Performance Standards
#### A: Material Standards:
The material standards are the essential part to be complied with first, as it is the basis of
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Auxein. The logo consists of a stylized letter "A" inside of a circle on the left, followed by the word "Auxein" in blue text on the right. The font is sans-serif and the overall design is simple and modern.
Image /page/9/Picture/1 description: The image shows contact information for a company. The corporate office is located at 7-A, Nirankari Colony, Delhi-110009, India. The manufacturing unit is located at Plot No. 169-170, Phase-IV, Sector-57, Kundli Industrial Area, Sonepat-131020, Haryana, India. The phone number is +91 99106 43638, and the fax number is +91 86077 70197. The website is www.auxein.com.
manufacturing metallic surgical implants.
We have complied with the following material standards.
1. ASTM F 136: Standard specification for wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra low interstitial) Alloy for surgical implant applications.
2. ASTM F 899-12: Standard Specification for Wrought Stainless Steels for surgical instruments.
We have verified the purchased material compliance to these standards and copies of the relevant test results are attached herewith.
## B: Performance Standards:
The device's performance of Auxein Nailing System has been demonstrated against following applicable standards.
- ASTM F 543-17 ●
- ASTM F1264
- ASTM F384 ●
- Custom Fatigue Testing
#### (B.1). Discussion on the clinical evaluation referenced and relied upon:
Auxein Nailing System is of similar design and pattern as well as similar intended use.
#### Conclusion:
#### General, Safety and Performance conclusion:
| S.No. | Parameter of<br>Conclusion | Proposed Device | Predicate Device |
|-------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| 1. | Product Code | HSB, JDS | Same |
| 2. | Regulation Number | 21 CFR 888.3020, 21 CFR 888.3030 | Same |
| 3. | Regulatory Class | Class II | Same |
| 4. | Intended Use | AUXEIN Brand of Expert Tibia Nailing<br>System<br>AUXEIN Brand of Expert Tibia Nailing<br>System- consists of variety of Bone nails<br>and bone screws. Expert Tibia Nail<br>System is intended to stabilize fractures<br>of the proximal and distal tibia and the<br>tibial shaft; open and closed tibial shaft<br>fractures; certain pre- and post-isthmic<br>fractures; and tibial malunions and non-<br>unions.<br>AUXEIN Brand of Expert Femoral Nailing | Same |
| | System | | |
| | AUXEIN Brand of Expert Femoral Nailing<br>System -is intended to stabilize femoral<br>shaft fractures, subtrochanteric<br>fractures, ipsilateral neck/shaft fractures,<br>impending pathologic fractures,<br>non-unions and malunions. | | |
| | AUXEIN Brand of Ga-mma Nailing<br>System | | |
| | AUXEIN Brand of Gamma Nailing System<br>-The Gamma Nail with Anti -Rotation<br>Screw is intended to treat stable and<br>unstable proximal fractures of the femur<br>including pertrochanteric fractures,<br>intertrochanteric fractures, high<br>subtrochanteric fractures and<br>combinations of these fractures,<br>including non-union, malunion and tumor<br>resections. The Long Nail system is<br>additionally indicated to treat<br>pertrochanteric fractures associated<br>with shaft fractures, pathologic fracture<br>in osteoporotic bone (including<br>prophylactic use) of the trochanteric and<br>diaphyseal fractures,<br>areas, impending<br>pathological fractures, long<br>subtrochanteric fractures, ipsilateral<br>femoral fractures, proximal or distal<br>non-unions, malunions, revision<br>procedures and tumor resections. | | |
| | AUXEIN Brand of AJAX Advance Nailing | | |
| | System,<br>AUXEIN Brand of AJAX Advance Nailing<br>System, end cap and screws is intended<br>to treat stable and unstable proximal<br>femoral fracture including<br>Pertrochanteric fractures,<br>Intertrochanteric fractures and High sub<br>trochanteric fractures and combination<br>of these fractures. | | |
| | The AUXEIN MEDICAL'S Elastic<br>Intramedullary Nail System | | |
| | The AUXEIN MEDICAL'S Elastic | | |
| | | Intramedullary Nail System is indicated<br>for fixation of diaphyseal fractures where<br>the canal is narrow or flexibility of the<br>implant is important. This includes upper<br>extremity fractures in all patients and<br>lower extremity fractures in children<br>above the age of 6 years and any patient<br>below 65 kg. This system is also<br>intended to treat metaphyseal and<br>epiphyseal fractures, such as radial neck<br>fractures. | |
| | | AUXEIN Brand of Humeral Nailing<br>System | |
| | | AUXEIN Brand of Humeral Nailing<br>System is indicated for fractures of the<br>proximal humerus, including 2-part<br>surgical neck fractures, 3-part fractures,<br>and 4-part fractures, proximal humeral<br>fractures with diaphyseal extension,<br>diaphyseal fractures of the humeral<br>shaft, and impending pathologic humeral<br>fractures. | |
| | | AUXEIN Brand of Retrograde Nailing<br>System | |
| | | AUXEIN Brand of Retrograde Nailing<br>System This nails are indicated for<br>simple long bone fractures; severely<br>comminuted, spiral, large oblique and<br>segmental fractures; nonunions and<br>malunions; polytrauma and multiple<br>fractures; prophylactic nailing of<br>impending pathologic fractures;<br>reconstruction, following tumor<br>resection and grafting; supracondylar<br>fractures; bone lengthening and<br>shortening. Interlocking intramedullary<br>nails are indicated for fixation of<br>fractures that occur in and between the<br>proximal and distal third of long bones<br>being treated." | |
| 5. | Sterilization<br>(Gamma) | Earlier the predicate device was supplied<br>Non-Sterile but the now scope device will<br>be supplied Sterile (Gamma Irradiation). | Same |
| 6. | Mechanical Test<br>Performance | • ASTM F543<br>• ASTM F384 | Same |
| | Screw-Cut Out Test ASTM F1264 | | |
| 7. | Material Standards ASTM F136 & ASTM F 899-12b. | Same | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Auxein. The logo consists of a blue stylized letter "A" inside of a circle, followed by the word "Auxein" in blue, sans-serif font. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.
CORPORATE OFFICE:
7-A, Nirankari Colony,
Delhi - 110 009, India
MANUFACTURING UNIT
Plot No. 169-170, Phase-IV,
Sector-57, Kundli Industrial Area,
HSIIDC, Sonepat-131028, Haryana, India
+91 99106 43638
+91 86077 70197 (FAX)
www.auxein.com
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Auxein. The logo consists of a stylized letter "A" with a circular design around it, followed by the text "Auxein" in a sans-serif font. The color of the logo and text is blue.
# 510K Number: K210792
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Auxein. The logo consists of a blue circle with a white triangle inside of it, and the word "Auxein" is written in blue letters to the right of the circle. The font is sans-serif and the letters are bolded.
Image /page/12/Picture/1 description: This image shows contact information for a company. The corporate office is located at 7-A, Nirankari Colony, Delhi-110 009, India. The manufacturing unit is located at Plot No. 169-170, Phase-IV, Sector-57, Kundli Industrial Area, HSIIDC, Sonepat-131020, Haryana, India. The phone number is +91 9910643638, and the fax number is +91 86077 70197. The website is www.auxein.com.
From the data available we can justify that the Auxein Nailing System is as safe, and as effective and performs as same indications for use as that of already marketed predicate devices identified in A.3. of 510(k) summary. Hence our devices can be considered safe and effective for their intended use.
#### Non Clinical Testing
Gap assessment was performed between the previously cleared 510k and the new submission. Since this submission includes sterile version of products as new additions and no other technical changes are being included. No additional risk for mechanical testing and biocompatibility testing was identified.
We have performed the Mechanical Testing as per ASTM F384 and Screw Cut-Out test for the Ajax Blade.
#### The sterilization method was validated using the below standards:
- ASTM F1980-Standard Guide for Accelerated Aging of Sterile Barrier Systems for ● Medical Devices.
- ISO 11737-2:2019-Sterilization of health care products-Microbiological methods-Part . 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process.
#### Conclusion
Bases upon the non-clinical testing, the Auxein Nailing System has been demonstrated to be substantially equivalent to the predicate device and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.