K210776 · 5r Med Instruments (Chengdu) Co., Ltd. · LZA · Jul 12, 2021 · General Hospital
Device Facts
Record ID
K210776
Device Name
Nitrile Examination Gloves
Applicant
5r Med Instruments (Chengdu) Co., Ltd.
Product Code
LZA · General Hospital
Decision Date
Jul 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
The Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
Device Story
Nitrile examination gloves; disposable; non-sterile; ambidextrous; blue color. Worn by healthcare examiners to provide barrier protection against contamination between patient and examiner. Manufactured to meet ASTM D6319, ASTM D5151, and ASTM D6124 standards. Device provides physical barrier; no active components or software. Used in general clinical settings.
Clinical Evidence
Bench testing only. No clinical data. Compliance verified via ASTM D6124 (residual powder), ASTM D5151 (freedom from holes), ASTM D6319 (physical properties), and ISO 10993-5/10 (biocompatibility/cytotoxicity/irritation/sensitization).
Technological Characteristics
Material: Nitrile. Form factor: Ambidextrous, non-sterile, blue. Standards: ASTM D6319-19 (specifications), ASTM D5151-19 (holes), ASTM D6124-06 (residual powder), ASTM D7160-16 (expiration). Biocompatibility: ISO 10993-5, ISO 10993-10. No energy source or software.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hands to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
K212311 — Medical Examination Gloves · Guangzhou Junda Gloves Co., Ltd. · Oct 21, 2021
K213176 — Disposable Nitrile Powder-Free Examination Gloves · Anhui Zhong Lian Latex Gloves Manufacturing Co., Ltd. · Dec 13, 2021
K212467 — Disposable Nitrile Examination Glove · Inner Mongolia Boming Medical Supplies Co., Ltd. · Nov 4, 2021
K210057 — Blue Nitrile Exam Glove, Powder Free · Real Star Medical Technology Co., Ltd. · Mar 31, 2021
K143477 — Nitrile Patient Examination gloves, Powderfree, Blue color · Primus Gloves Pvt Limited · Apr 14, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
July 12, 2021
5R MED Instruments (Chengdu) Co., Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K210776
Trade/Device Name: Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: June 9, 2021 Received: June 16, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210776
Device Name Nitrile Examination Gloves
Indications for Use (Describe)
The Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K210776
This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92.
# 1.0 submitter's information
Name: 5R MED Instruments (CHENGDU) Co., Ltd. Address: Building 9, Section 1, No.618 of West Kelin Road, Chengdu Cross-Straits Technology Industry Development Zone, Wenjiang District, Chengdu, Sichuan Province, 611130, China Phone Number: +86-13928606563 Contact: Yuzhanq Pan Date of Preparation: 2021.03.02
#### Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device information
Trade name: Nitrile Examination Gloves
Common name: Patient Examination Gloves
Classification name: Non-powdered patient examination glove
Model(s): XS, S, M, L, XL 3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I General Hospital Panel:
#### 4.0 Predicate device information
Ever Global (Vietnam) Enterprise Corp Manufacturer:
{4}------------------------------------------------
Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422
# 5.0 Indications for use
The Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
# 6.0 Device description
The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.
# 7.0 Summary comparing technological characteristics with predicate device
| Item | Proposed device | Predicated device | Remark |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 510(k) number | K210776 | K171422 | Same |
| Product Code | LZA | LZA | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Indications for use | The Nitrile Examination<br>Gloves is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | The Disposable Powder<br>Free Nitrile Examination<br>Glove, White/ Blue/ Black/<br>Pink Color is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | Same |
| Powdered or Powered free | Powdered free | Powdered free | Same |
| Design Feature | ambidextrous | ambidextrous | Same |
| Labeling Information | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Nitrile<br>Examination Gloves,<br>Non-Sterile | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Disposable<br>Powder Free Nitrile<br>Examination Glove, | Same |
#### Table1-General Comparison
{5}------------------------------------------------
| | | | Non-Sterile | |
|--|--|--|-------------|--|
|--|--|--|-------------|--|
| Predicate<br>Device(K171422) | Designation | Size | | | | | Tolerance |
|------------------------------|----------------|------------|-----|-----|-----|-----|-----------|
| | Length, mm | XS | S | M | L | XL | |
| | | 230 | 230 | 230 | 230 | 230 | min |
| | Width, mm | 75 | 85 | 95 | 105 | 115 | ±5 |
| | Thickness, mm: | | | | | | |
| | Finger | 0.05 | | | | | min |
| | Palm | 0.05 | | | | | min |
| Proposed Device | Designation | Size | | | | | Tolerance |
| | | XS | S | M | L | XL | |
| | Length, mm | 220 | 220 | 230 | 230 | 230 | min |
| | Width, mm | 70 | 80 | 95 | 110 | 120 | ±10 |
| | Thickness, mm: | | | | | | |
| | Finger | 0.05 | | | | | min |
| | Palm | 0.05 | | | | | min |
| Remark | | Analysis 1 | | | | | |
#### Table2 Device Dimensions Comparison
Analysis1: The sizes and tolerances of proposed device are different with those of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.
| Item | | | Proposed device | Predicated device | Remark |
|------------------------|------------------------|------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|-----------|
| Colorant | | | blue | White/ Blue/ Black/ Pink | Analysis2 |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | SAME |
| | | Ultimate<br>Elongation | 500%min | 500%min | SAME |
| | After<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | SAME |
| | | Ultimate<br>Elongation | 400%min | 400%min | SAME |
| | Comply with ASTM D6319 | | Comply with ASTM D6319 | | SAME |
| Freedom from Holes | | | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | Be free from holes when tested in accordance with ASTMD5151 AQL=2.5 | SAME |
| Powder Content | | | 0.1 | Meet the requirements of ASTM D6124 | SIMILAR |
# Table3 Performance Comparison
Analysis2: The proposed device has different color to the predicate device, but all proposed devices are conducted the biocompatibility test, the test results shown that the color difference do not effect the safety of proposed device.
{6}------------------------------------------------
| Item | | Proposed device | Predicated device | Remark |
|--------------------|---------------|--------------------------------------------------------------------------------|-------------------------|---------|
| Material | | Nitrile | Nitrile | SAME |
| Biocompatibility | Irritation | Under the conditions of the study, not an irritant | Comply with ISO10993-10 | SAME |
| | Sensitization | Under conditions of the study, not a sensitizer. | | |
| | Cytotoxicity | Under conditions of the study, did not show potential toxicity to L-929 cells. | Comply with ISO10993-5 | SIMILAR |
| Label and Labeling | | Meet FDA's Requirement | Meet FDA's Requirement | SAME |
# Table4 Safety Comparison
# 8.0 Discussion of non-clinical and clinical test performed
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19,Standard Specification For Nitrile Examination Gloves For Medical Application.
ASTM D7160-16,Standard Practice for Determination of Expiration Dating for Medical Gloves
Clinical testing is not needed for this device.
# 9.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.