The Insight Agile DRF is intended for use by a qualified/trained medical professionals, who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device provides fluoroscopic imaging and is used for guidance and visualization during routine fluoroscopic and image-guided procedures. The device is to be used in professional healthcare facilities, in a variety of image-guided procedures such as feeding tube and catheter insertion on pediatric and neonates (birth to one month). This device is not indicated for large/obese patients or interventional use. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The system is not intendedfor mammography applications. (Rx Only)
Device Story
Insight Agile DRF is a fluoroscopic imaging chain for integration with fixed or mobile x-ray positioning systems. Input: X-ray signals captured by flat-panel digital detectors (Vieworks 1212G or Rayence 1212FCA). Operation: Software-based acquisition, processing, and display of fluoroscopic images; includes Automatic Brightness Stabilization (ABS) control via graphical user interface. Output: Real-time fluoroscopic images on a workstation monitor. Used in professional healthcare facilities by trained medical professionals. Benefits: Provides high-quality, high-resolution visualization for pediatric/neonatal image-guided procedures. Clinical decision-making: Assists clinicians in guiding catheters and feeding tubes. System utilizes virtual serial ports for communication with GE R&F systems and a dedicated interface for Source-Ray x-ray generators.
Clinical Evidence
No clinical data; bench testing only. Phantom images compared to predicate device demonstrated equivalent imaging performance. Radiologist review of clinical exam images confirmed high quality and comparability to OEC 9800/9900 mobile c-arms. System passed electrical, mechanical, and radiation safety testing (IEC 60601-1-3, IEC 60601-2-54, 21 CFR 1020.32).
Technological Characteristics
Fluoroscopic imaging chain; Vieworks 1212G or Rayence 1212FCA (IZGO-based) flat-panel detectors. Source-Ray x-ray generator. Connectivity: Virtual serial ports for GE R&F system integration. Software-controlled ABS. Compliance: IEC 60601-1:2005+A1, IEC 60601-1-2:2014, IEC 60601-1-3:2008+A1, IEC 60601-2-54:2009+AMD1/2. Fixed aperture collimator with 0.1 mm Cu filter.
Indications for Use
Indicated for pediatric patients and neonates (birth to one month) for guidance and visualization during routine fluoroscopic and image-guided procedures (e.g., feeding tube/catheter insertion). Not indicated for large/obese patients, interventional use, or mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MobileRay Pulse SE Digital Imaging System (K191503)
Reference Devices
Insight Enhanced DRF Digital Imaging System (K200396)
Solid State X-Ray Imager (Flat Panel/Digital Imager) (K202722)
OEC 9800 mobile c-arm
OEC 9900 mobile c-arm
Submission Summary (Full Text)
{0}------------------------------------------------
July 27, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Imaging Engineering, LLC % Mr. George Jachode President 1318 NW 7th Ave CAPE CORAL FL 33993
Re: K210469
Trade/Device Name: Insight Agile DRF Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA, OXO, RCC, QHY Dated: July 12, 2021 Received: July 19, 2021
Dear Mr. Jachode:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K210469
Device Name
### Insight Agile DRF
Indications for Use (Describe)
The Insight Agile DRF is intended for use by a qualified/trained medical professionals, who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device provides fluoroscopic imaging and is used for guidance and visualization during routine fluoroscopic and image-guided procedures. The device is to be used in professional healthcare facilities, in a variety of image-guided procedures such as feeding tube and catheter insertion on pediatric and neonates (birth to one month). This device is not indicated for large/obese patients or interventional use. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The system is not intendedfor mammography applications. (Rx Only)
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows three overlapping squares of different colors. The top square is orange, the bottom left square is dark blue, and the center square where the other two overlap is light blue. The squares are made up of smaller squares arranged in a grid pattern.
# Exhibit 5: 510(k) Summary
#### 1. Trade/proprietary name: Insight Agile DRF
| 510(k) Clearance #: | K210469 |
|---------------------|---------------------------------------------|
| Regulation Name: | Image-Intensified Fluoroscopic X-ray System |
| Regulation Number: | 21 CFR 892.1650 |
| Regulatory Class: | Class II |
| Product Code: | JAA, OXO, RCC and QHY |
| Common/Usual Name: | Mobile Fluoroscopic System |
#### 2. Predicate Device: K191503
| Trade Name: | MobileRay Pulse SE Digital Imaging System |
|--------------------|-------------------------------------------|
| Manufacturer: | PortaVision Medical, LLC |
| Regulation Number: | 21 CFR 892.1650 |
| Regulatory Class: | Class II |
| Product Code: | OWB, JAA, QHY, OXO and RCC |
| Common/Usual Name: | Mobile Fluoroscopic System |
#### 3. Reference Devices: (Used with this device)
| Trade Name: | Insight Enhanced DRF Digital Imaging System |
|-----------------------|------------------------------------------------------|
| 510(k) Clearance #: | K200396 |
| Clearance date: | 03/06/2020 |
| Classification Name: | Image Intensified Fluoroscopic X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21CFR 892.1650 (Produce Code: JAA) |
| Device Class: | Class II |
| Trade Name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| 510(k) Clearance #: | K202722 |
| Clearance date: | 10/26/2020 |
| Classification Name: | Image Intensified Fluoroscopic X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21CFR 892.1680 (Produce Code: MQB) |
| Device Class: | Class II |
#### 4. Device Description
The Insight Agile DRF Digital Imaging System is a modification of Insight Enhanced DRF, a previously cleared (K200396) fluoroscopic imaging chain employing a flatpanel digital detector (Vieworks 1212G) to replace the image intensifier and video camera on GE R&F systems (P500, Legacy and Advantx). This system has now been modified in the following manner: 1) the digital interface board (DIB) that provides communications with the GE systems has been replaced with a software program that utilizes virtual serial ports.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows three overlapping squares of different colors. The square on the left is dark blue, the square on the top is orange, and the square in the middle where the other two overlap is light blue. Each square is made up of smaller squares arranged in a grid pattern.
2) Another software program has been added to allow the system to communicate with and control a Source-Ray x-ray generator. The generator interface provides for Automatic Brightness Stabilization (ABS) control and optional manual control through a graphical user interface on the Insight Aqile image acquisition screen. 3) A second dynamic flat-panel detector (Rayence 1212FCA, K202722) has been integrated to provide an alternate digital detector. 4) the software risk assessment has been changed from minor to moderate. 5) the Indications for Use has been changed to that of the predicate device.
With these modifications, Insight Agile DRF is a fluoroscopic imaging chain that can acquire, process, and display fluoroscopic images when installed on any suitable positioning system in a fixed room, on a mobile or portable stand. Another purpose of this application is to expand the indications for use.
#### 5. Indications for Use
The Insight Agile DRF is intended for use by a qualified/trained medical professionals, who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device provides fluoroscopic imaging and is used for quidance and visualization during routine fluoroscopic and image-guided procedures. The device is to be used in professional healthcare facilities, in a variety of image-guided procedures such as feeding tube and catheter insertion on pediatric patients and neonates (birth to one month). This device is not indicated for large/obese patients or interventional use. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The system is not intended for mammography applications. (Rx Only)
#### 6. Summary of Design Control Risk Management
The Insight Agile DRF fluoroscopic imaging chain has been developed to provide OEMs and System Integrators a fluoroscopic imaging chain system that can be integrated with a suitable positioning system. It provides optimized workflow when imaging patients while meeting critical functional requirements and international safety standards. The risks and the hazardous impact of the device design were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the device design and production phase were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design and production were successfully mitigated and accepted.
{5}------------------------------------------------
Image /page/5/Figure/0 description: The image shows a grid of squares with three different colors. The bottom left section of the grid is dark blue, the top right section is orange, and the overlapping section is light blue. The squares are arranged in a rectangular pattern.
# 7. Substantial Equivalence
The Insight Agile™ DRF fluoroscopic imaging chain conforms to the FDA recognized standards as like the predicate device. Based on the recognized standard conformity evidence related to the electro-, mechanical, software-, clinical-and risk management, it is the sponsor's opinion that the subject device is a safe and effective device.
| Characteristic | MobileRay Pulse SE Digital<br>Imaging System<br>(K191503) | Insight Agile DRF<br>Digital Imaging<br>System |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Power source | 120 VAC 50/60 HZ<br>2.5 amps | 120 VAC 50/60 HZ<br>3.5 amps |
| Generator Type | High frequency inverter type | SAME |
| Output Power | 4 kW | 80 W |
| Fluoroscopy | 0.5 - 10 mA (pulsed)<br>40 - 25kV | 0.1 - 5 mA (Cont.)<br>35 - 80 kV |
| Image acquisition | Up to up to 7.5 fps (fluoro) | Up to 30 fps (fluoro) |
| X-ray Tube | Stationary Anode | SAME |
| Indicators | Display on workstation monitor | SAME |
| Collimator | Multi-leaf adjustable motorized,<br>Collimare Touch LED | Fixed Aperture with 0.1 mm<br>Cu fixed filter |
| Digital Panel Specification | PerkinElmer XRpad2 3025 4346 | Vieworks 1212G<br>Rayence 1212FCA |
| Pixel Pitch | 100 μm | Vw 1212G 145 μm<br>Ray.1212FCA 194 μm |
| Matrix | 2508 x 3004 | Vx 2000 x 2000<br>Ray 1536 x 1536 |
| A/D conversion | 16 bits | SAME |
| Image Acquisition | Amorphous Si direct deposition<br>Csl:TI | Vw 1212G SAME<br>Ray IZGO CsI:TI |
| Performance Standard | 21 CFR 1202.30 | SAME |
| Electrical Safety | IEC60601-1:2005 + A1 (2012)<br>IEC60601-1-2:2007<br>IEC60601-1-3:2008<br>IEC60601-2-28:2010<br>IEC60601-2-43:2010<br>IEC60601-2-54:2009 | IEC60601-1-<br>2:2014+ANSI+AAMI<br>IEC60601-1-<br>3:2008+A1:2013<br>IEC60601-2-54:2009<br>+AMD1:2015+AMD2:2018 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows two overlapping squares, each made up of smaller squares. The square on the left is primarily blue, while the square on the right is orange. Where the squares overlap, the smaller squares are light blue, indicating a combination of the two original colors. The arrangement suggests a concept of intersection or overlap between two distinct entities.
The following table compares The MobileRay Pulse SE software (predicate device) to the subject device software.
| Feature | Predicate Device<br>MobileRay Pulse SE<br>K191503 | Subject Device |
|----------------------------------|---------------------------------------------------|----------------|
| Acquiring image from<br>detector | Yes | Yes |
| Viewing image | Yes | Yes |
| Change window/level | Yes | Yes |
| Invert | Yes | Yes |
| Lookup Table | Yes | Yes |
| Zoom | Yes | Yes |
| Pan | Yes | Yes |
| Noise Reduction | Yes | Yes |
| Patient Information | Yes | Yes |
| Annotation | Yes | Yes |
| Image rotation | Yes | Yes |
| X-Ray generator control | Yes | Yes |
| DICOM worklist and Send | Yes | Yes |
## 8. Difference Discussion
The predicate device is a portable system that employs a more powerful generator than the subject device. The predicate device has an indication for use statement that includes radiography and interventional use. The subject device has a lower power generator but does have similar technological characteristics as it relates to fluoroscopic imaging as the predicate device. The indication for use statement of the subject device therefore does not include either radiography or interventional use and further restricts the patient population it is indicated for. Accordingly, the subject device is only substantially equivalent to the fluoroscopic imaging characteristics of the predicate device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a square grid of small squares. The grid is composed of three different colors: orange, blue, and light blue. The orange squares are located in the upper right corner, the blue squares are located in the lower left corner, and the light blue squares are located in the center where the orange and blue squares overlap.
Insight Agile™ DRF
Digital Imaging System
The predicate device is a portable fluoroscopic and radiographic system. Insight Agile DRF is a fluoroscopic imaging chain that can be installed on any suitable positioning system including the predicate device. Insight Enhanced DRF and the Vieworks flat panel detector have already been cleared as a fluoroscopic imaging chain component (K200396) to upgrade the image intensifier and video camera on GE R&F systems. The Rayence 1212FCA detector has been previously cleared for MQB (K202722).
### 9. Summary of the technological characteristics of the device compared to the predicate device.
The predicate device employs a different fluoroscopic imaging chain consisting of a Sedecal x-rav generator. Collimator. Perkin Elmer digital x-ray detector and console software provided by RADinfo Systems, Inc. Insight Agile DRF employs a Source-Ray x-ray generator, a manual collimator, a Vieworks or Rayence digital x-ray detector and the Insight Agile DRF console software.
Two dFPD detectors are being applied for clearance in this application. One of the dFPDs is manufactured by Vieworks and was cleared previously with Insight Enhanced (K200396). The second dFPD is manufactured by Rayence (1212FCA), this detector is based on 3rd generation flat panel television technology (Indium Zinc Gallium Oxide or IZGO). This technology provides a lower noise floor and much longer detector lifetime compared with a-Si based dFPDs due to its ten-fold increase in radiation damage resistance.
The Rayence 1212FCA was recently cleared for MQB (K202722). A complete SSXI data report for the Rayence 1212FCA detector is provided along with bench test data to establish substantial equivalency.
## 10. Description of non-clinical tests.
Bench testing was performed on the subject device with the Rayence 1212FCA on the system to assess substantial equivalence of the device and with the Vieworks detector in terms of image quality. Both detectors were integrated with the Source-Ray generator. Phantom images were acquired with the subject device and compared to the predicate device. These images and the doses used to acquire them were analyzed and compared. In conclusion, the tests obtained demonstrated substantial equivalence to the predicate device in terms of imaging performance.
Additional bench testing was performed by Imaging Engineering on the Insight Agile DRF system to determine if the system complies with IEC 60601-1-3, IEC60601-2-54 and 21 CFR 1020.32. The test results are provided in Exhibit 18. The system passed all electrical, mechanical and radiation safety testing. EMC and Electrical Safety performance for the Rayence digital receptor panel had previously been submitted to FDA in K202722.
{8}------------------------------------------------
Image /page/8/Figure/0 description: The image shows a square grid of smaller squares, with three distinct colors. The bottom left portion of the grid is dark blue, the top right portion is orange, and the overlapping center portion is light blue. The arrangement of colors creates a visually interesting pattern.
Insight Agile™ DRF
Digital Imaging System
A radiologist was provided a series of fluoroscopic patient exam images to evaluate the image quality produced by the subject device compared to 2 OEC (9800 and 9900) mobile c arms. A report of their findings is included in Exhibit 16. This report includes the following statement "I found the Imaging Engineering images to be of high quality, high resolution, and clinically acceptable for the intended use and fully comparable to the OEC images."
Software has been written and validated according to the FDA Software Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Cybersecurity concerns have been addressed in accordance with: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff (October 2, 2014).
## 11. Description of clinical tests.
No clinical data is necessary to evaluate safety or effectiveness for purposes of determining substantial equivalence of the proposed modification. Bench testing was performed to assess the device safety and effectiveness.
## 12. Conclusion as to Substantial Equivalence.
The Insight Agile DRF Digital Imaging System, the subject device is substantially equivalent to the fluoroscopic imaging properties of the predicate (K191503). The design principle and the applicable standards for the subject device are similar to those of the predicate device. The statement for use of the predicate device has been modified to remove radiographic and interventional use and to restrict the population demographics to reflect the limitations of the subject device. The performance test and non-clinical consideration result demonstrate that these differences do not raise any new questions of safety and effectiveness. Therefore, it is the sponsor's opinion that the subject device appears to be as safe and effective as the predicate device for fluoroscopic use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.