Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts.
Device Story
Insight Enhanced™ DRF Digital Imaging System is an add-on imaging acquisition device for existing R&F fluoroscopic systems (GE Advantx, GE Legacy, GE P500, Shimadzu YSF300). It integrates dynamic flat-panel detectors (dFPD) (Vivix-D 1212G or 1717G) alongside or replacing the existing CMOS camera. The system captures high-definition fluoroscopic and spot images (up to 30 fps); processes images via an image processor PC; and displays them on monitors for clinical review. Operators use the system to enhance images to visualize diagnostic details. Images are stored locally or transmitted over a network for long-term storage/hardcopy printing. The device assists clinicians in diagnostic imaging by providing high-resolution visualization of anatomical structures, facilitating clinical decision-making through improved image clarity.
Clinical Evidence
No clinical data. Bench testing only. Phantom images were acquired using the subject device and compared to the predicate device on the same GE Legacy R&F system to assess image quality and radiation dose.
Technological Characteristics
System includes dynamic flat-panel detectors (Vivix-D 1212G/1717G), fluoroscopic video camera, monitors, and image processor PC. Resolution: 1k x 1k 12-bit. Connectivity: DICOM. Power: 120 VAC 50/60 Hz. Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3. Software is identical to the predicate, providing GUI, image processing, and workflow functions.
Indications for Use
Indicated for use by qualified doctors or technicians on adult and pediatric patients to obtain fluoroscopic radiographic images of the skull, spine, chest, abdomen, extremities, and other body parts.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Insight Essentials™ DRF Digital Imaging System (K191310)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
March 6, 2020
Imaging Engineering, LLC % Mr. George Jachode President 1318 NW 7th Avenue CAPE CORAL FL 33993
Re: K200396
Trade/Device Name: Insight Enhanced™ DRF Digital Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA Dated: February 12, 2020 Received: February 18, 2020
Dear Mr. Jachode:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K200396
Device Name
Insight Enhanced™ DRF Digital Imaging System
Indications for Use (Describe)
Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image shows three overlapping squares. The square on the left is dark blue, the square on the top right is orange, and the square in the middle where the other two squares overlap is light blue. The squares are made up of smaller squares.
#### 1. Special 510(k) SUMMARY
#### Assigned Submission Number: K200396
This summary of Special 510(k) is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.
Date Special 510K summary prepared: February 10, 2020
| Submitter's Name: | Imaging Engineering, LLC |
|------------------------|--------------------------------------------|
| Submitter's Address: | 1318 NW 7th Ave, Cape Coral, Florida 33993 |
| Submitter's Telephone: | (239) 223-1371 |
| Contact person: | George Jachode/President |
| Official Correspondent: | George Jachode (g.jachode@imagingengineeringllc.com) |
|-------------------------|------------------------------------------------------|
| Address: | 1318 NW 7th Ave, Cape Coral, Florida 33993 |
| Telephone: | (239) 223-1371 |
#### Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| 510K Number: | |
|-------------------------|------------------------------------------------|
| Trade/proprietary name: | Insight Enhanced™ DRF Digital Imaging System |
| Model Number: | EN-1000-00 |
| Regulation Name: | Image-Intensified Fluoroscopic X-ray System |
| Regulation Number: | 21 CFR 892. 1650 |
| Regulatory Class: | Class II |
| Product Code: | JAA and LLZ |
| Predicate Device | |
| Trade Name: | Insight Essentials™ DRF Digital Imaging System |
| 510(k) Clearance #: | K191310 |
| Clearance date: | 06/10/2019 |
| Classification Name: | Image Intensified Fluoroscopic X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21CFR 892.1650 (Produce Code; JAA and LLZ) |
| Device Class: | Class II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows two overlapping squares. The square on the left is blue, and the square on the right is orange. The overlapping area is light blue. Each square is made up of smaller squares.
Insight Enhanced™ DRF Digital Imaqing System
# 2. Device Description
The Insight Enhanced™ DRF Digital Imaging System is designed to provide an additional imaging acquisition device to the existing image intensifier, camera and user interface with a dynamic flat-panel detector (dFPD) on the following FDA approved R&F fluoroscopic imaging systems employing an image intensifier: GE Advantx, GE Legacy, GE P500 and Shimadzu YSF300. The subject fluoro system Insight Enhanced is based on the predicate Insight Essentials. The following dynamic digital detectors are added to the predicate system: Vivix-D 1212G and Vivix-D 1717G. The x-ray generator, beam-limiting device and patient positioner, necessary for a full fluoroscopy system are not part of the subject device software stays unchanged from the predicate Insight Essentials DRF.
The Insight Enhanced™ DRF Digital Imaging System allows the operator to view and enhance high-definition fluoroscopy images. High resolution digital spot images may also be acquired at single or rapid acquisition rates, up to 30 fps. Images may be viewed and enhanced enabling the operator to bring out diagnostic details difficult or impossible to see using conventional imaging techniques. Images can be stored locally for short term storage. The Insight Enhanced™ DRF Digital Imaging System enables the operator to produce hardcopy images with a laser printer or send images over a network for longer-term storage. The major system components include: a fluoroscopic video camera, monitors, and an image processor PC.
#### 3. Indications for Use
Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts,
# 4. Summary of Design Control Risk Management
The Insight Enhanced™ DRF Digital Imaging System has been developed to provide medical professionals optimized workflow when imaging patients while meeting critical functional requirements and international safety standards. The risks and the hazardous impact of the device design were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the device design and production phase were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design and production were successfully mitigated and accepted.
{5}------------------------------------------------
Image /page/5/Figure/0 description: The image shows three overlapping squares of different colors. The square on the left is dark blue, the square on the top right is orange, and the square in the middle where the other two overlap is light blue. The squares are made up of smaller squares arranged in a grid pattern.
# 5. Substantial Equivalence
The Insight Enhanced™ DRF Digital Imaging System conforms to the FDA recognized standards as like the predicate device. Based on the recognized standard conformity evidences related to the electro-, mechanical, software-, clinical-and risk management, it's the sponsor's opinion that the subject device is a safe and effective device.
| Characteristic | Insight Essentials<br>DRF Digital Imaging<br>System<br>(K191310) | Insight Enhanced<br>DRF Digital<br>Imaging System |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Intended Use: | Intended for use by a<br>qualified/trained doctor or<br>technician on both adult<br>and pediatric subjects for<br>obtaining fluoroscopic<br>radiographic images of the<br>skull, spinal column, chest,<br>abdomen, extremities, and<br>other body parts. | SAME |
| Power source | 120 VAC 50/60 HZ<br>2.5 amps | 120 VAC 50/60 HZ<br>2.5 amps |
| Image acquisition | Up to 15 FPS (spot), up to<br>30 fps (fluoro) | SAME |
| File compatibility | DICOM | SAME |
| Digital Video Source | Basler ACE1920-40gm<br>CMOS camera | VIVIX-D 1212 and<br>VIVIX-D 1717 dFPD |
| Digital Resolution | 1k x 1k 12 bit | SAME |
| Electrical safety | IEC 60601-1:<br>IEC 60601-1-2<br>IEC 60601-1-3 | SAME |
{6}------------------------------------------------
Image /page/6/Figure/0 description: The image shows a square grid pattern with three different colors. The top right section of the grid is orange, the bottom left section is dark blue, and the overlapping center section is light blue. The grid appears to be a square shape, with each color occupying a distinct area within the square.
# 6. Difference Discussion
The dynamic flat-panel detectors Vivix-D1212G and Vivix-D 1717G have been added to the predicate system Insight Essentials in order to provide image acquisition support in addition to the existing CMOS camera.
# 7. Summary of the technological characteristics of the device compared to the predicate device
Both the subject device and the predicate device have the identical software program that provides a Graphical User Interface (GUI), image processing and patient workflow functions.
Two dFPD detectors have been added to the existing device Insight Essentials, to acquire fluoroscopic and spot images. The dFPDs are both manufactured by the same company Vieworks, use the same manufacturing technology but are of different sizes (12x12 inch and 17x17 inch). The subject device can now be configured with either a CMOS camera or one of the 2 dFPDs.
The VIVIX-D1212 dFPD has already been cleared by the FDA (K180473) for use on a mobile c arm. The Vivix-D 1717 uses the same architecture as the Vivix-D 1212 (K180473) and the only major difference is its larger size. A complete SSXI data report for the VIVIX-D 1717 is provided along with bench test data to establish substantial equivalency.
# 8. Description of non-clinical tests.
Bench testing was performed on the Vivix digital detectors used with the subject system Insight Enhanced, in order to assess substantial equivalence to the predicate device Insight Essentials. Phantom images were acquired with the subject device and compared to the predicate device, also installed on the same GE Legacy R&F system. These images and the doses used to acquire them were analyzed and compared. In conclusion, the tests obtained demonstrated substantial equivalence to the predicate device.
#### 9. Description of clinical tests.
No clinical data is necessary to evaluate safety or effectiveness for purposes of determining substantial equivalence of the proposed modification. Bench testing was performed to assess the device safety and effectiveness.
{7}------------------------------------------------
Image /page/7/Figure/0 description: The image shows three overlapping squares. The square on the left is dark blue, the square on the top right is orange, and the square in the middle where the other two overlap is light blue. The squares are made up of smaller squares arranged in a grid pattern.
# 10. Conclusion as to Substantial Equivalence.
The Insight Enhanced™ DRF Digital Imaging System, the subject device is substantially equivalent to the predicate (K191310). The intended use, the design principle, and the applicable standards for the subject device are identical to those of the predicate device. The performance test and non-clinical consideration result demonstrate that these differences do not raise any new questions of safety and effectiveness. Therefore, it is the sponsor's opinion that the subject device appears to be as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.