Harmony Referral System (Harmony RS) is a comprehensive software platform intended for use in importing, processing, viewing, measurement and storage of clinical images and videos as well as in management and communication of patient data, diagnostic and clinical information and reports from ophthalmic diagnostic instruments through either direct connection with the instruments or through computerized networks. The system neither performs any interpretations nor provides treatment recommendations.
Device Story
Harmony Referral System (Harmony RS) is a web-based software platform for ophthalmic image and data management. It imports clinical images, videos, and reports from ophthalmic diagnostic instruments via direct connection or computerized networks. The system stores, processes, and displays this data at a PC workstation for review by trained professionals. It provides standard viewing tools (zoom, rotate, flip), image enhancement (contrast adjustment), and manual measurement tools (line, area, cup-disc ratio) based on user-marked pixels. It does not perform automated image analysis, clinical interpretation, or provide treatment recommendations. The system facilitates communication with other PACS and EMR systems. Security is maintained via SSL encryption, firewalls, VPN, and IP-based access restrictions. By centralizing diagnostic information, it assists clinicians in reviewing patient data to support clinical decision-making.
Clinical Evidence
No clinical data provided. Bench testing and software verification/validation demonstrate the device performs as intended and meets specifications.
Technological Characteristics
Web-based software platform; operates on Debian Linux (archive/application/integration servers) and Windows Server (integration server). Client-side requires browser support (Edge Chromium, Firefox). Supports DICOM communication, SSL encryption, VPN, and IP-based access control. Image formats: JPEG, JPEG 2000, RAW, encapsulated in DICOM. Manual measurement tools based on pixel spacing.
Indications for Use
Indicated for use by trained professionals for the management, viewing, and measurement of ophthalmic clinical images, videos, and patient data imported from diagnostic instruments.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
September 13, 2021
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Topcon Healthcare Solutions EMEA Oy % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180
Re: K210396
Trade/Device Name: Harmony Referral System Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management and Processing System Regulatory Class: Class II Product Code: NFJ Dated: August 18, 2021 Received: August 19, 2021
Dear Maureen O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K210396
Device Name Harmony Referral System (Harmony RS)
#### Indications for Use (Describe)
Harmony Referral System (Harmony RS) is a comprehensive software platform intended for use in importing, processing, viewing, measurement and storage of clinical images and videos as well as in management and communication of patient data, diagnostic and clinical information and reports from ophthalmic diagnostic instruments through either direct connection with the instruments or through computerized networks. The system neither performs any interpretations nor provides treatment recommendations.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span style="font-size: 100%;"> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <div> <span style="font-size: 100%;"> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY Topcon Healthcare Solutions Harmony Referral System (Harmony RS)
### 510(k) Owner
Topcon Healthcare Solutions EMEA Oy Makelininkatu 43 90100 Oulu, Finland Phone: (201) 599-5208 Contact Person: Ramya Sundaram
### Submission Correspondent
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham. MA 02180 Phone: (978) 207-1245 Email: Maureen@oconnellregulatory.com
Date Prepared: August 16, 2021
### Trade Name of Device
Harmony Referral System (Harmony RS)
#### Common or Usual Name
System, image management, ophthalmic
#### Classification Name
21 C.F.R. 892.2050 Medical Image Management and Processing System
#### Predicate Device
Topcon Harmony cleared in K182376
#### Indications for Use
Harmony Referral System (Harmony RS) is a comprehensive software platform intended for use in importing, processing, viewing, measurement and storage of clinical images and videos as well as in management and communication of patient data, diagnostic and clinical information and reports from ophthalmic diagnostic instruments through either direct connection with the instruments or through computerized networks. The system neither performs any interpretations nor provides treatment recommendations.
#### Device Description
Harmony Referral System is an internet-browser-based software platform that allows users to access examination data of a patient from different sources. Harmony Referral System may be used together with a number of computerized digital imaging devices and third party software. In
{4}------------------------------------------------
addition. Harmony Referral System software collects and manages patient demographics, image data, and clinical reports from a range of approved medical devices. Harmony Referral System enables a real-time review of diagnostic patient information at a PC workstation. The software uses SSL encryption in network communication and secure network infrastructure with firewalls and additionally also VPN and IP-based access restrictions to ensure secure networking environment. The Harmony Referral System does not perform automated image analysis but provides measurements based on pixels of an image, which were marked by the user manually on the screen including cup-disk ratio and line and area measurements.
#### Performance Data
No performance data was required or provided. Software validation and verification demonstrate that Harmony RS performs as intended and meets its' specifications.
### Substantial Equivalence
The Harmony RS is substantially equivalent to Topcon Healthcare Solutions' Harmony cleared in K182376. The Harmony RS has the same intended use and similar indications for use, technological characteristics, and principles of operation as the previously cleared predicate. Both devices are software only image management systems which manage, display and analyze ophthalmic images.
Regarding acquisition, Harmony RS and Harmony (K182376) do not offer capture components. Both devices import digital images, patient data, diagnostic data and clinical information from other software capture systems and directly from ophthalmic devices. The Harmony RS has an absolute measurement tool which is derived from pixel spacing information provided by the source device and does not perform any automated image analysis. It provides line and area measurements as well as cup-disc ratio based on the user manually marking the images on the screen which are the same measurements performed by the Harmony. The Harmony also performs MPS which is not performed by Harmony RS.
Harmony RS and Harmony both allow standard viewing operations such as zoom in/out, flip/rotate, etc. and standard image enhancements such as contrast adjustment, and drawing tools. Harmony RS provides DICOM communication with other PACS as does Topcon Healthcare Solutions' Harmony (K182376). Harmony RS and Harmony both provide similar customizable print templates and archive and backup functionality.
The minor differences including the operating system, hardware requirements, system access, and measurement and analysis. The predicate device is operated on a MS Windows Server and supports the Chrome browser while the Harmony RS uses additional servers and browsers. The predicate software is used on a Windows XP/2000 PC while the Harmony RS operates on a Desktop PC operating Windows. The predicate device is available both on a desktop and webbased while the Harmony RS is only web-based.
The different technological characteristics of the devices do not raise new questions of safety and effectiveness. The differences in hardware requirements and system access are all system features that can be evaluated during software validation and verification and were primarily revised to allow the system to operate with newer hardware, browsers and operating systems.
{5}------------------------------------------------
The Harmony RS shares similar technological characteristics with the predicate device, both in terms of the manner in which images are captured, and stored, as well as the operation of the device by the intended user. The software validation and verification shows that the Harmony RS performs as intended and supports substantial equivalence.
The following table compares the Harmony RS to the Harmony cleared in K182376.
| | Proposed Device | Predicate Device |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Harmony Referral System | Harmony |
| 510(k) number | K210396 | K182376 |
| Manufacturer | Topcon Healthcare Solutions<br>EMEA Oy | Topcon Healthcare Solutions, Inc. |
| Device Type | System, Image Management,<br>Ophthalmic | System, Image Management,<br>Ophthalmic |
| Regulation<br>Description | Picture archiving and<br>communications system | Picture archiving and communications<br>system |
| Regulation<br>Medical Specialty | Radiology | Radiology |
| Review Panel | Ophthalmic | Ophthalmic |
| Product Code | NFJ | NFJ |
| Regulation<br>Number | 892.2050 | 892.2050 |
| Device Class | 2 | 2 |
| Indications for Use | Harmony Referral System<br>(Harmony RS) is a comprehensive<br>software platform intended for use<br>in importing, processing, viewing,<br>measurement and storage of<br>clinical images and videos as well<br>as in management and<br>communication of patient data,<br>diagnostic and clinical information<br>and reports from ophthalmic<br>diagnostic instruments through<br>either direct connection with the<br>instruments or through<br>computerized networks. The | Harmony is a<br>comprehensive software<br>platform intended for use<br>in acquisition or<br>importing, processing,<br>measurement, analysis<br>and storage of clinical<br>images and videos of the<br>eye as well as in<br>management of patient<br>data, diagnostic data,<br>clinical information,<br>reports from ophthalmic<br>diagnostic instruments<br>through either a direct<br>connection with the |
| | | |
| | system neither performs any<br>interpretations nor provides<br>treatment recommendations. | instruments or through<br>computerized networks. |
| Device Type | Software only | Software only |
| User Population | Trained professionals | Trained professionals |
| Platform /<br>Operating System | Archive & application server:<br>Debian Linux<br><br>Integration server: Debian Linux,<br>Windows Server<br><br>Client: operating system<br>supporting the following browsers:<br><br>Microsoft Edge Chromium,<br>Mozilla Firefox (two latest<br>verssons) | Server: MS Windows Server 2012 R2<br><br>Client: operating system supporting the<br>following browsers:<br><br>Chrome |
| Hardware<br>requirements | Processor performance level: i3 7th<br>gen dual core, i5 4th gen quad core<br><br>Ram: 4GB<br>Full HD 1920x1080 resolution<br>monitor<br><br>Any operating system capable of<br>running supported browser<br><br>Network connection: 100 Mbps<br>(recommended), 5 Mbps<br>(minimum upload/download) | Window XP/2000 PC<br>Monitor<br>Keyboard<br>Mouse |
| Web based access | Harmony Referral System is web-<br>based system | Desktop and web-based access |
| Image Data<br>compression | JPEG (lossy), JPEG 2000 (lossy,<br>lossless), RAW (uncompressed)<br>All files encapsulated inside<br>DICOM format. | TIFF - uncompressed,<br>PNG - lossless compress,<br>JPEG - lossy compress,<br>DICOM - lossy & lossless |
| Virtualization | Yes | Yes |
| User<br>administration | centralized | centralized |
| HIS/EMR<br>integration | Yes | Yes |
| DICOM | Yes | Yes |
| communication<br>with<br>other PACS | Image processing | |
| Image processing | Standard image<br>viewing operation<br>such as zoom in/out,<br>flip/rotate, etc.<br>Standard image<br>enhancement<br>adjustment such as<br>contrast adjustment,<br>etc.<br>Text annotation<br>Line/circle/freehand<br>drawing | Standard image<br>viewing operation<br>such as zoom in/out,<br>flip/rotate, etc.<br>Standard image<br>enhancement<br>adjustment such as<br>contrast adjustment,<br>etc.<br>Text annotation<br>Line/circle/freehand<br>drawing |
| Measurement and<br>analysis | Scale to display pixel<br>dimensions of the image Line and area measurements Cup-disc ratio | Perform line/area<br>measurement with<br>retinal images, Cup-Disc ratio, MPS |
| Reporting | available | available |
TABLE 1 HARMONY REFERRAL SYSTEM SUBSTANTIAL EQUIVALENCE CHART
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.