Disposable Surgical Mask

K210372 · Dongguan Shin YI Healthcare Products Factory · FXX · Jun 7, 2021 · General, Plastic Surgery

Device Facts

Record IDK210372
Device NameDisposable Surgical Mask
ApplicantDongguan Shin YI Healthcare Products Factory
Product CodeFXX · General, Plastic Surgery
Decision DateJun 7, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use ,disposable device, provided non-sterile.

Device Story

Disposable, three-layer, non-sterile surgical mask; intended for single use by healthcare personnel and patients. Constructed with inner/outer spun-bond polypropylene layers and a middle melt-blown polypropylene filter layer. Features a malleable aluminum wire nose piece for fit and either elastic ear loops or tie-on straps. Used in clinical or general settings for infection control to reduce exposure to blood, body fluids, and particulates. Provides physical barrier protection; performance meets ASTM F2100 Level 3 standards. No active components or software.

Clinical Evidence

No clinical data. Bench testing only, including fluid resistance (ASTM F1862), particulate filtration efficiency (ASTM F2299), bacterial filtration efficiency (ASTM F2101), differential pressure (EN 14683), flammability (16 CFR 1610), and biocompatibility (ISO 10993-5, ISO 10993-10).

Technological Characteristics

Three-layer flat mask; inner/outer layers: spun-bond polypropylene; middle layer: melt-blown polypropylene filter. Nose piece: malleable aluminum wire. Straps: elastic fiber or spun-bond polypropylene. Dimensions: (18±1)cm x (9±1)cm or (14.5±1)cm x (9±1)cm. Non-sterile. ASTM F2100 Level 3 compliant.

Indications for Use

Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material; intended for infection control to reduce exposure to blood and body fluids. Suitable for over-the-counter use.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 7, 2021 Dongguan Shin Yi Healthcare Products Factory % Joyce Yang Consultant Shenzhen Jovantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 51800 China Re: K210372 Trade/Device Name: Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX, Dated: April 25, 2021 Received: May 3, 2021 Dear Joyce Yang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Clarence W. Murray III -S Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K210372 Device Name Disposable Surgical Mask #### Indications for Use (Describe) The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use ,disposable device, provided non-sterile. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary 510(k) Number: K210372 Date Prepared: June 5, 2021 #### 1.Submission Sponsor | Manufacturer information | Company: Dongguan Shin Yi Healthcare<br>Products Factory | |--------------------------|----------------------------------------------------------------------------------------------------------| | | Company address: No. 17 Ban Hu Road, Huang<br>Jiang Town, Dong Guan City, Guang Dong<br>Province, China. | | | Contact person: Zhao Shuge | | | Phone: +86-769-83361381 | | | Fax: +86-769-83631102 | | | E-mail: zhaosg.dg@master-frank.cn | # 2. Submission correspondent ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Name Address | Shenzhen Joyantech Consulting Co., Ltd<br>1713A, 17th Floor, Block A, Zhongguan Times Square,<br>Nanshan District, Shenzhen | |---------------------|-----------------------------------------------------------------------------------------------------------------------------| | Post Code Phone No. | 518000<br>86-755-86069197 | | Contact Person | Joyce Yang | | Email | joyce@cefda.com | # 3. Device Identification | Type of 510(k) submission: | Traditional | |----------------------------|----------------------------------------| | Trade Name: | Disposable Surgical Mask | | Model: | MFA05004, MFA05005, MFA05007, MFA05008 | | Classification name: | Mask, Surgical | | Review Panel: | Surgical Apparel | | Product Code: | FXX | | Device Class: | II | | Regulation Number: | 21 CFR 878.4040 | {4}------------------------------------------------ #### 4. Legally Marketed Predicate Device | Trade Name | Single-use Surgical Mask | |-------------------|-------------------------------------| | Regulation number | 21 CFR 878.4040 | | Regulation class | II | | Regulation name | Surgical Apparel | | 510(k) Number | K200923 | | Product Code | FXX | | Manufacturer | BYD Precision Manufacturer Co.,Ltd. | #### 5. Device Description The proposed devices are single use, three-layer, flat masks with straps and nose piece. The Surgical Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The proposed device is held in place over the user's mouth and nose by two elastic ear loops or four ties welded to the face mask. The elastic ear loops are not made with natural rubber latex. The nose piece contained in the proposed device is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Malleable aluminum wire. The proposed devices are sold non-sterile and are intended to be single use, disposable devices. #### 6. Intended Use/ Indications for Use The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile. #### 7. Technological characteristics comparison | Comparison<br>item | Subject Device: Disposable<br>Surgical Mask(K210372) | Predicate Device: Single-<br>use Surgical Mask<br>(K200923) | Comments | |----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------| | Product Code | FXX | FXX | Same | | Regulation Number | 21 CFR § 878.4040 | 21 CFR § 878.4040 | Same | | Classification n | Class II | Class II | Same | | OTC use | Yes | Yes | Same | | Intended use & | The Surgical Masks are | The Single-use Surgical | Same | | Comparison<br>item | Subject Device: Disposable<br>Surgical Mask(K210372) | Predicate Device: Single-<br>use Surgical Mask<br>(K200923) | Comments | | Indication<br>s for Use | intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These masks are<br>intended for use in infection<br>control practices to reduce<br>the potential exposure to<br>blood and body fluids. This is<br>a single use ,disposable<br>device, provided non-sterile. | Face Masks are intended<br>to be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single<br>use, disposable device,<br>provided non-sterile. | | | Design<br>feature | Ear-loop, Tie-on | Ear-loop | Similar<br>(Issue 1) | | Usage | Single use | Single use | Same | | Color | Blue | Blue | Same | | Size | $(18±1)cm×(9±1)cm$<br>$(14.5±1)cm×(9±1)cm$ | $(17.5±1)cm×(9.5±1)cm$ | Similar<br>(Issue 2) | | Sterile | Non-sterile | Non-sterile | Same | | Material | Outer layer: Spun-bond<br>polypropylene | Outer layer: Spun-bond<br>polypropylene | Same | | | Middle layer: Melt blown<br>polypropylene filter | Middle layer: Melt blown<br>polypropylene filter | Same | | | Inner layer: Spun-bond<br>polypropylene | Inner layer:Spun-bond<br>polypropylene | Same | | | Nose piece:Malleable<br>aluminum wire | Nose piece:Metal Core<br>Plastic | Different<br>(Issue 3) | | Comparison<br>item | Subject Device: Disposable<br>Surgical Mask(K210372) | Predicate Device: Single-<br>use Surgical Mask<br>(K200923) | Comments | | | Ear-loops: Elastic fiber<br>Tie-on:Spun-bond<br>polypropylene | Ear-loops: Polyester | Similar<br>(Issue 3) | | ASTM F<br>2100 Level | Level 3 | Level 3 | Same | | Fluid<br>Resistance<br>Performance<br>ASTM F<br>1862-13 | Pass at 160 mmHg | Pass at 160 mmHg | Same | | Particulate<br>Filtration<br>Efficiency<br>ASTM F<br>2299 | > 98% | > 98% | Same | | Bacterial<br>Filtration<br>Efficiency<br>ASTM F<br>2101 | > 98% | > 98% | Same | | Differential<br>Pressure<br>(Delta P) EN<br>14683:2019+<br>AC : 2019 | < 6.0mmH2O/cm² | < 6.0mmH2O/cm² | Same | | Flammability<br>16CFR 1610 | Class 1 | Class 1 | Same | | Cytotoxicity | Comply with ISO 10993-5<br>Non cytotoxic | Comply with ISO 10993-5<br>Non cytotoxic | Same | | Irritation | Comply with ISO 10993-10<br>Non irritating | Comply with ISO 10993-10<br>Non irritating | Same | | Sensitization | Comply with ISO 10993-10<br>Non sensitizing | Comply with ISO 10993-10<br>Non sensitizing | Same | {5}------------------------------------------------ {6}------------------------------------------------ Issue 1: Compared to the predicate device, the subject device has an additional design feature "Tie-on". The difference in design feature do not affect the performance of the face mask. Issue 2: The sizes of proposed device are a little smaller than the size of predicate device. The small size is to apply to the population with thin face, and there is no difference in performance. {7}------------------------------------------------ lssue 3: The nose piece is used to allow the user to fit the face mask around their nose. The difference of materials of nose piece between subject device and predicate device could not affect the performance of the face mask. The Ear-loops of the proposed device are made by elastic fiber, which of the predicate device is made by polyester. The major chemical composition of the elastic fiber is segmented polyurethane-urea, which is similar to polyester. In addition, the proposed devices have been tested for biocompatibility, and the test results confirm that the device is biocompatible. #### 8. Summary of Non-clinical Testing Non-clinical tests were conducted and conformed to the following standards and the requirements state in the Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification [510(k)] Submission issued on March 05, 2004. Standards: - ASTM F2100-19 Standard Specification For Performance of Materials used in Medical Face Masks. - ASTM F1862-13 Standard Test Method For Resistance of Medical Face Masks to Penetration by Synthetic Blood. - ASTM F2299-03 Stand Test Method For Determining The Initial Efficiency Of Materials Used In Medical Face Masks To Penetration By Particulates Using Latex Spheres. - 16 CFR 1610 Standard For The Flammability Of Clothing Textiles. - ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. - ANSI/AAMI/ISO 11135:2014 Sterilization of health care products Ethylene ● oxide - Requirements for Development, validation, and routine control of a sterilization process for Medical devices - ISO 10993-5: 2009 Biological Evaluation Of Medical Devices Part 5: Tests ● For In Vitro Cytotoxicity. - ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. | Test Methodology | Purpose | Acceptance Criteria | Results | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------|---------------------|---------------------| | Fluid Resistance<br>Performance<br>(ASTM F1862) | For evaluating the ability of<br>the mask's material to resist<br>the penetration of blood and<br>body fluids. | Pass at 160 mmHg | Pass at 160<br>mmHg | #### Table 2: Performance Characteristic {8}------------------------------------------------ | Test Methodology | Purpose | Acceptance<br>Criteria | Results | |--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------|----------------| | Particulate Filtration<br>Efficiency<br>( ASTM F2299 ) | For evaluating a submicron<br>efficiency performance | > 98% | > 98% | | Bacterial Filtration<br>Efficiency<br>(ASTM F2101) | For evaluating the ability of<br>the mask's material to<br>prevent the passage of<br>aerosolized bacteria | > 98% | > 98% | | Differential Pressure<br>(EN 14683:2019) | For evaluating the<br>breathability and comfort of<br>the surgical mask | < 6.0mmH2O/cm² | < 6.0mmH2O/cm² | | Flammability (16<br>CFR 1610) | For evaluating the<br>flammability of the material | Class 1 | Class 1 | #### Table 3: Biocompatibility | Test<br>Methodology | Purpose | Acceptance<br>Criteria | Results | |---------------------|--------------------------------------------------------|------------------------|-------------------------------------------------------------------| | Cytotoxicity | For evaluating the cytotoxicity<br>of the materials | Non-cytotoxicity | Under the conditions<br>of the study, non-<br>cytotoxicity effect | | Irritation | For evaluating the skin irritation<br>of the materials | Non-irritation | Under the conditions of<br>the study, non-irritation | | Sensitization | For evaluating the sensitization<br>of the materials | Non-sensitization | Under the conditions of<br>the study, non-<br>sensitization | #### 9. Brief discussion of clinical tests No clinical tests were performed. #### 10. Conclusions The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(k) submission K210372, the Disposable Surgical Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K200923.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...