Blood Administration Set

K210335 · Baxter Healthcare Corporation · BRZ · Jun 10, 2021 · General Hospital

Device Facts

Record IDK210335
Device NameBlood Administration Set
ApplicantBaxter Healthcare Corporation
Product CodeBRZ · General Hospital
Decision DateJun 10, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

For the administration of blood, blood components or solutions from a container into the patient's vascular system through a vascular access device.

Device Story

The Blood Administration Set is a single-use, sterile, non-pyrogenic disposable device used to deliver blood, blood components, or solutions to a patient's vascular system. It functions as an extension set or a primary set for syringe pump administration. The device consists of non-DEHP PVC tubing, a notch clamp, a female Luer lock, and a male Luer lock. It is operated by clinicians in a clinical setting. The device connects to a fluid container (e.g., syringe) and a vascular access device. By providing a fluid path, it enables controlled delivery of therapeutic fluids. The device benefits patients by facilitating safe, reliable vascular access for fluid and blood product administration. It is designed to be used alone or as an extension to existing primary administration sets to provide additional length.

Clinical Evidence

Bench testing only. Performance data included ISO 80369-7 Luer testing, tensile strength, leak testing, clamp activation/shut-off testing, particulate matter (USP <788>), flow rate, and blood component compatibility (ISO 1135-4). Biocompatibility was evaluated per ISO 10993-1 for prolonged contact, external communicating, indirect blood path devices, including cytotoxicity, sensitization, irritation, systemic toxicity, material-mediated pyrogenicity, and hemocompatibility. All tests met acceptance criteria.

Technological Characteristics

Materials: Non-DEHP PVC tubing, Polymethyl methacrylate (acrylic) female Luer lock, ABS male Luer lock. Dimensions: 150 cm (59") length. Connectivity: Luer lock interfaces. Sterilization: Gamma radiation (SAL 10^-6) per ANSI/AAMI/ISO 11137-1/2. Standards: ISO 80369-7 (Luer), ISO 1135-4 (Blood sets), USP <788> (Particulates), USP <85>/<161> (Endotoxins).

Indications for Use

Indicated for administration of blood, blood components, or solutions from a container into the vascular system of patients of all ages (neonatal, pediatric, adult) via a vascular access device.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 10, 2021 Baxter Healthcare Corporation James Vangeisen Principal Specialist, Regulatory Affairs 25212 West Illinois Route 120 Round Lake, Illinois 60073 Re: K210335 Trade/Device Name: Blood Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: BRZ Dated: May 4, 2021 Received: May 11, 2021 Dear James Vangeisen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210335 Device Name Blood Administration Set Indications for Use (Describe) For the administration of blood, blood components or solutions from a container into the patient's vascular system through a vascular access device. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The word is the logo for the company Baxter International, a global healthcare company. ## 510(k) Summary #### K210335 June 10, 2021 ## OWNER: Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015 ## CONTACT PERSON: James Vangeisen Principal Specialist, Regulatory Affairs 25212 West Illinois Route 120 Round Lake, IL 60073 Telephone: (224) 270 3308 Fax: (224) 270 4119 # IDENTIFICATION OF THE DEVICE: Trade/Device Name: Blood Administration Set Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: BRZ | Code # | 2N3383 | |--------------------|------------------------------------------------| | Device Description | Blood Extension Set, 59" (150 cm), Vol. 4.7 mL | | 1. | Non-Vented Cap for Female Luer Lock | | 2. | Female Luer Lock | | 3. | Notch Clamp | | 4. | Tube | | 5. | Male Luer Lock | | 6. | Filter Vented Cap for Male Luer Lock | #### Table 1. Proposed Set Configuration {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The word is the logo for the healthcare company Baxter International. #### PREDICATE DEVICE: | Device | Company | Predicate 510(k) | Clearance Date | |------------------------------|----------------------------------|------------------|-------------------| | Blood Administration<br>Sets | Baxter Healthcare<br>Corporation | K993120 | November 17, 1999 | #### Table 2. Predicate Device #### REASON FOR SUBMISSION: The basis for this premarket notification is the intent to market an Intravascular (IV) Administration Set (Blood Administration Set). The proposed device in this submission is a single-use, disposable device, intended for the administration of fluids from a container into the patient's vascular system through a vascular access device. ## DESCRIPTION OF THE DEVICE: The proposed device is an IV Administration Set (Blood Administration Set). It is a single use, non-pyrogenic, sterile disposable device intended for the administration of fluids from a container into the patient's vascular system. It can be used to administer solutions, blood, blood products to patients of all ages ranges - neonatal, pediatric, and adult. The proposed set consists of non-DEHP PVC (< 0.1% DEHP) tubing, a notch clamp, a female Luer lock, a non-vented cap for a female Luer lock, a male Luer lock, and a filter vented cap for a male Luer lock. It can be used to administer solutions, blood, and blood products to the patient. The most common use of this proposed set would be for syringe pump administration as a primary set and as an extension set. This proposed set can also be used as an extension set in combination with gravity and large volume infusion pump blood administration sets (primarily to add length). See Figure 1, Figure 2, and Figure 3 for the clinical use set ups for the proposed device. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is presented in a simple, clean design. ### Figure 1. Clinical Use Set Up of Proposed Device when used directly with a Syringe as a Primary Syringe Set Image /page/5/Figure/2 description: This image shows a medical setup with various components labeled. A "Male Luer Lock" is connected to a patient's vascular access device. A "Female Luer Lock" is connected to a syringe, which is labeled as a "Syringe (Container)". A "Shut Off Clamp" is positioned in the middle of the tube. Figure 2. Clinical Use Set Up of Proposed Device when used as an Extension to a Primary Syringe Set Image /page/5/Figure/4 description: This image shows a medical setup for administering fluids or medication to a patient. The setup includes a syringe connected to a primary syringe set, which is then connected to an extension set. A male luer lock connects to the patient's vascular access device, while a female luer lock connects to the syringe. A shut-off clamp is also included in the setup. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is likely a logo or brand name. Image /page/6/Figure/1 description: The image shows a diagram of an intravenous (IV) infusion setup. The setup includes a solution container at the top, connected to a primary administration set. Further down, a female Luer lock connects the primary administration set to an extension set, which is then connected to a male Luer lock that is inserted into the patient's arm. ## Figure 3. Clinical Use Set Up of Proposed Device when used as an Extension to a Primary Administration Set # INDICATIONS FOR USE: For the administration of blood, blood components or solutions from a container into the patient's vascular system through a vascular access device. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly slanted to the right, giving it a dynamic appearance. The word is the primary focus of the image, and the background is plain. ## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE: The proposed device is substantially equivalent to the predicate device, previously cleared under 510(k) premarket notification K993120 on November 17, 1999. The intended use and function of the proposed device is equivalent to the predicate device. Table 3 is a device comparison table outlining the differences between the predicate and proposed devices. | Features | Predicate Device<br>K993120 | Proposed Device<br>K210335 | Assessment of Differences | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For the administration of<br>fluids from a container into<br>the patient's vascular<br>system through a vascular<br>access device. | Same | N/A | | Indication for Use | For the administration of<br>blood, blood components or<br>solutions from a container<br>into the patient's vascular<br>system through a vascular<br>access device. | Same | N/A | | Regulation Number | 21 CFR 880.5440 | Same | N/A | | Product Code | BRZ | Same | N/A | | Sterile | Gamma radiation | Same | N/A | | Sterility Assurance<br>Level (SAL) | 10⁻⁶ | Same | N/A | | Non-Pyrogenic | Yes | Same | N/A | | Single Use | Yes | Same | N/A | | Length | 112" (2.8 m)<br>(2C8750) | 59" (150 cm)<br>(2N3383) | The predicate device is<br>comparatively longer than the<br>proposed device.<br>Design control activities have been<br>conducted and confirmed, through<br>clinical studies, that the length of the<br>proposed device ensures a safe<br>delivery without excessive<br>manipulation of the set and patient<br>discomfort. | | Features | Predicate Device<br>K993120 | Proposed Device<br>K210335 | Assessment of Differences | | | Fluid Path Components/Materials | | The proposed device can also be<br>used as an extension set that allows<br>the clinician to extend the length of<br>the primary set, when and if needed.<br>A recommended clinical practice,<br>when selecting the overall length of<br>the set, is for the clinician to try and<br>avoid using excessive lengths to<br>minimize residual volumes and<br>retain flow accuracy. From this<br>assessment it can be concluded that<br>the proposed shorter set does not<br>raise different questions of safety<br>and effectiveness when compared to<br>the predicate device. | | Spike | Acrylonitrile Butadiene<br>Styrene<br>(2C8750) | N/A | The proposed device is an extension<br>set and does not contain a spike. | | Blood<br>Chamber | Styrene-Butadiene Blended<br>Copolymer (Chamber and<br>Filter Housing)<br>Polyester -Filter membrane<br>(Mesh)<br>(2C8750) | N/A | The proposed device is an extension<br>set and does not contain a blood<br>chamber. | | Tubing | Polyvinyl Chloride<br>(2C8750) | Same | N/A | | Injection Site | Clearlink<br>Polycarbonate (Inlet/Outlet<br>Housing)<br>Silicone (Gland)<br>Polycarbonate (Center Post)<br>(2C8750) | N/A | The proposed device is an extension<br>set and does not contain an injection<br>site. | | Female Luer<br>Lock | N/A | Polymethyl<br>methacrylate<br>(Acrylic) | The predicate device does not have a<br>female Luer Lock. Design control<br>activities have been conducted and<br>have confirmed that the different | | Features | Predicate Device<br>K993120 | Proposed Device<br>K210335 | Assessment of Differences | | | technological characteristics of the proposed devices do not raise different questions of safety and effectiveness. | | | | Male Luer<br>Lock | Acrylonitrile Butadiene<br>Styrene<br>(2C8750) | Same | N/A | #### Table 3. Device Comparison {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a corporate logo. The word is horizontally oriented and takes up most of the image space. # Table 3. Device Comparison {9}------------------------------------------------ Image /page/9/Picture/0 description: The image features the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized, giving the word a dynamic and modern look. The color of the text is a vibrant blue, which stands out against the white background. # Table 3. Device Comparison {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right. ## DISCUSSION OF NONCLINICAL TESTS: Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use. ## Performance Data: The following bench tests (Table 4) were conducted to evaluate the functional performance of the proposed devices: | Test | Acceptance Criteria | |------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ISO 80369-7 Luer Tests on male Luer Lock Connector | ISO 80369-7:2016, Clause 6.1.2 or 6.1.3,<br>ISO 80369-7:2016, Clause 6.2,<br>ISO 80369-7:2016, Clause 6.3,<br>ISO 80369-7:2016, Clause 6.4,<br>ISO 80369-7:2016, Clause 6.5,<br>ISO 80369-7:2016, Clause 6.6,<br>ISO 80369-7:2016, Clause 5 | | ISO 80369-7 Luer Tests on female Luer Lock Connector | ISO 80369-7:2016, Clause 6.1.2 or 6.1.3,<br>ISO 80369-7:2016, Clause 6.2,<br>ISO 80369-7:2016, Clause 6.3,<br>ISO 80369-7:2016, Clause 6.4,<br>ISO 80369-7:2016, Clause 6.5,<br>ISO 80369-7:2016, Clause 6.6,<br>ISO 80369-7:2016, Clause 5 | | Tensile Strength Test | ISO 1135-4:2015, Clause 5.3 | | Leak Test (Pressure Test) | ISO 1135-4:2015, Annex A.2 | | Notch Clamp Activation Force Test | Activation force ≤50N | | Notch Clamp Shut-Off Test | No liquid or air leakage when subjected to<br>50kPa for 15 sec | | Non-DEHP Claim Verification | <0.1% DEHP | | Particulate Matter Test | USP <788> | | Flow Rate Testing | ISO 1135-4:2015, Section 5.9 | | ISO 1135-4 Blood Component Compatibility Test | ISO 1135-4, Clause 7.6 and 7.7 | # Table 4. Performance Data All tests met the acceptance criteria. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is blue and appears to be a logo or brand name. The font is a slightly italicized and the letters are closely spaced together. ### Biocompatibility: Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged duration, external communicating device, indirect blood path and FDA-2013-D-0350 Guidance for Industry and FDA Staff, "Use of International Standard ISO-10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,"" as recommended in the Intravascular Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]." Biocompatibility assessments were conducted on a worst case/representative final, finished device for all fluid path materials of the proposed device. The following tests were conducted as part of the biocompatibility assessment for the proposed device: - . Cytotoxicity ISO 10993-5 - Sensitization ISO 10993-10 - Intracutaneous (Irritation) Reactivity ISO 10993-10 - . Acute Systemic Toxicity ISO 10993-11 - . 30 Day Systemic Repeat Dose Toxicity Study ISO 10993-11 - Material Mediated Pyrogen ISO 10993-11 - Hemocompatibility ISO 10993-4 Based upon the results, the data supports an ISO 10993-1 categorization of external communicating device, indirect blood path, prolonged contact duration. The proposed device is biocompatible and appropriate for its intended use. ### Sterility: The proposed device is sterilized with gamma radiation. The product is in the bioburden (sub) category "General Sets Labeled Sterile". The Minimum Sterilizing Dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2, "Sterilization of health care products – Radiation-Part 2: Establishing the sterilization dose." The dose setting method used includes, but is not limited to, Method 1 or VDmax. Generally, the MSDs are between 14.2 - 25.0 kGy. The continued validity of the MSD for this (sub) category is confirmed via periodic dose audit studies. Bacterial endotoxins tests were conducted in conformance to USP <85>. The endotoxin limit is 20 EU/device per USP <161>. In addition, routine periodic pre-sterilization bioburden testing is performed for each (sub) category. The sterilization process for the proposed {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is the logo for the healthcare company Baxter International. device was established in accordance with ANSI/AAMI/ISO 11137-1, "Sterilization of health care products-Radiation-Part 1; Requirements for development, validation and routine control of a sterilization process for medical devices." ## Shelf-Life: Baxter has performed aging testing to support a shelf-life claim of 3 (three) years. # CONCLUSION: The non-clinical data demonstrate that the subject device is substantially equivalent and performed comparably to the predicate device that is legally marketed for the same intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...