The Fingertip Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in clinical institution and home environments.
Device Story
Fingertip pulse oximeter for non-invasive spot-checking of SpO2 and pulse rate; intended for adult patients in clinical or home environments. Device uses dual-wavelength light source (660 nm red, 905 nm infrared) and photodetector to measure light absorption in pulsating arteriolar vascular bed. Photodetector converts light intensity to electronic signals; system calculates SpO2 based on ratio of light absorption during systole and diastole. Powered by 2 AAA batteries; features PCB, display (OLED/LED), and plastic shell. Operated by patient or clinician; provides numerical output on screen. No alarms; not for life-support. Benefits include portable, rapid assessment of oxygenation status.
Clinical Evidence
Clinical hypoxia testing performed on adult human volunteers per ISO 80601-2-61. Accuracy validated by comparing device SpO2 readings against arterial oxygen saturation (SaO2) measured via co-oximetry. Results support accuracy claims within the specified saturation range.
Technological Characteristics
Surface-contacting device; materials evaluated per ISO 10993-1 (cytotoxicity, sensitization, irritation). Dual-wavelength sensing (660 nm/905 nm). Powered by 2 AAA alkaline batteries. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home environment), and ISO 80601-2-61 (pulse oximeter performance).
Indications for Use
Indicated for adult patients in clinical and home settings for non-invasive spot-checking of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate (PR).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
K072825 — FINGERTIP PULSE OXIMETER, MODEL MD300I · Beijing Choice Electronic Technololgy Co., Ltd. · Jan 14, 2008
K092652 — FINGERTIP PULSE OXIMETER, MODEL MD300CA ND300CA · Beijing Choice Electronic Technololgy Co., Ltd. · Jan 8, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 30, 2021
Zhuhai Linte Medical Instrument Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 508, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong 518067 China
Re: K210274
Trade/Device Name: Fingertip Pulse Oximeter, Model LT-F20 and LT-F21 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: June 29, 2021 Received: July 2, 2021
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K210274
### Device Name
Fingertip Pulse Oximeter, model: LT-F20 and LT-F21
Indications for Use (Describe)
The Fingertip Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in clinical institution and home environments.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# 510(K) Summary
### Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: 2021/07/25
#### 1. Submission sponsor
Name: Zhuhai Linte Medical Instrument Co., Ltd.
Address: 4th Floor, Building 1, No.66, Yongda Road, Hongqi Town, Jinwan District, Zhuhai, Guangdong 519090 P.R. China
Contact person: Kezheng Ma
Title: Management Representative
E-mail: mark.ma(@)lintemed.com
Tel: +86 135 30230360
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: Room 508, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P. R. China 518067
Contact person: Kevin Wang
E-mail: kevin(@chonconn.com
Tel: +86-755 33941160
| 3. Subject Device Information | |
|-------------------------------|---------------------------------|
| Trade/Device Name | Fingertip Pulse Oximeter |
| Model | LT-F20 and LT-F21 |
| Common Name | Fingertip Pulse Oximeter |
| Regulatory Class | Class II |
| Classification | 21CFR 870.2700 / Oximeter / DQA |
| Submission type | Traditional 510(K) |
#### 3. Subject Device Information
#### 4. Predicate Device
Manufacturer: Shenzhen Aeon Technology Co., Ltd. Device name: Pulse Oximeter 510(K) Number: K190869
### ട്. Device Description
The Fingertip Pulse Oximeter is intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).
The Fingertip Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 905 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a
{4}------------------------------------------------
constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The Pulse Oximeter is powered by 2 AAA alkaline batteries.
The device mainly composed of PCB board, On/Off button, OLED or LED or LED screen, battery compartment, and plastic shell.
The device is a spot-check pulse oximeter and does not include alarms. The device is not intended for lifesupporting or life-sustaining.
#### 6. Intended use & Indication for use
The Fingertip Pulse Oximeter is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in clinical institution and home environments.
| | 7. Comparison to the Predicate Device | | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Features | Subject Device | Predicate Device K190869<br>AEON Pulse Oximeter | Comparison |
| | Fingertip Pulse Oximeter | | n |
| Applicant | Zhuhai Linte Medical<br>Instrument Co., Ltd. | Shenzhen Aeon Technology<br>Co., Ltd. | / |
| Classification<br>Regulation | 21CRF 870.2700 | 21CRF 870.2700 | Same |
| Classification and<br>Code | Class II, DQA | Class II, DQA | Same |
| Common name | Fingertip Pulse Oximeter | Fingertip Pulse Oximeter | Same |
| Intended use | The Fingertip Pulse Oximeter<br>is a non-invasive device<br>intended for spot checking of<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2)<br>and pulse rate (PR). This<br>portable device is indicated for<br>use in adult patients in clinical<br>institution and home<br>environments. | The Pulse Oximeter is a non-<br>invasive device intended for<br>spot checking of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse<br>rate (PR). This portable device<br>is indicated for use in adult<br>patients in clinical institution<br>and home environments. | Same |
| Patient populations | Adults | Adults | Same |
| Principle | The device displays numerical<br>values for functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and pulse | The device displays numerical<br>values for functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and pulse | Same |
| Features | Subject Device<br>Fingertip Pulse Oximeter | Predicate Device K190869<br>AEON Pulse Oximeter | Comparison |
| | rate by measuring the<br>absorption of red and infrared<br>(IR) light passing through<br>perfused tissue. Changes in the<br>absorption caused by the<br>pulsation of blood in the<br>vascular bed are used to<br>determine oxygen saturation<br>and pulse rate. | rate by measuring the<br>absorption of red and infrared<br>(IR) light passing through<br>perfused tissue. Changes in the<br>absorption caused by the<br>pulsation of blood in the<br>vascular bed are used to<br>determine oxygen saturation<br>and pulse rate. | |
| Light Emitting | Red: 660 nm<br>Infrared: 905 nm | Red: 660 nm<br>Infrared: 905nm | Same |
| Power source | 2 AAA alkaline batteries | 2 AAA alkaline batteries | Same |
| Display data | SpO2%, PR | SpO2%, PR | Same |
| SpO2 Measuring<br>Range | 35%-100% | 35%-100% | Same |
| SpO2 Resolution | 1% | 1% | Same |
| SpO2 Accuracy | 70% ~ 100%, Arms ± 3%.<br><70%, unspecified. | 70~ 100%, ±3%.<br><70%, unspecified; | Same |
| PR Range | 30 bmp – 250 bmp | 30 bmp – 250 bmp | Same |
| PR Resolution | 1 bpm | 1 bpm | Same |
| PR Accuracy | ±1% or ±3bpm, whichever is<br>greater | ± 2 | Different (1) |
| Sterile | No | No | Same |
| Application site | Finger | Finger | Same |
| Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Same |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Same |
| Performance | Complied with ISO 80601-2-<br>61 | Complied with ISO 80601-2-<br>61 | Same |
| Biocompatibility | Cytotoxicity | Complied with ISO 10993-5 | Same |
| | Skin Irritation | Complied with ISO 10993-10 | Same |
| | Sensitization | Complied with ISO 10993-10 | Same |
#### 7 Comparison to the Predicate Device
{5}------------------------------------------------
Justifications for differences between proposed device and the predicate device are shown as below: Different (1): The accuracy of PR is different. The PR measurement has been verified according to declared range and accuracy. Thus, this difference does not raise different questions of safety and effectiveness.
{6}------------------------------------------------
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the Fingertip Pulse Oximeter was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:
- . Cytotoxicity
- . Sensitization
- Irritation .
The subject devices are considered surface contacting for a duration of not exceed 24 hours.
### Non-clinical data
The Fingertip Pulse Oximeter has been tested according to the following standards:
- IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment-● Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
- IEC 60601-1-11: 2015. Medical electrical equipment- Part 1-11: General requirements for basic . safety and essential performance- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for . Basic Safety and Essential Performance of Pulse Oximeter Equipment.
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical . Devices.
- FDA Guidance for Pulse Oximeters Premarket Notification Submissions [510(k)s] .
The test was selected to show substantial equivalence between the subject device and the predicate.
## Clinical data
Clinical studies were conducted to verify the accuracy of proposed device. The clinical studies were conducted per following standards:
- . ISO 80601-2-61: 2017 Medical Electrical Equipment - Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.
- Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and . Drug Administration Staff
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical
{7}------------------------------------------------
hypoxia test results were obtained in human adult volunteers to validate the accuracy of Pulse Oximeter versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.
### 9. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.