K210173 · GE Healthcare · KPS · Mar 29, 2021 · Radiology
Device Facts
Record ID
K210173
Device Name
StarGuide
Applicant
GE Healthcare
Product Code
KPS · Radiology
Decision Date
Mar 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The GE StarGuide system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging of tumors, planning, guiding, and monitoring therapy, including the nuclear medicine part of theragnostic procedures. The GE StarGuide system, combining a CZTbased, high energy and spatial resolution, Nuclear Medicine (NM) system and a Computed Tomography (CT) system, is intended to produce: NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types and by imaging features designed to enhance image quality. CT System: Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body, cardiac and vascular X-Ray Computed Tomography applications. NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based corrections of SPECT images as well as functional and anatomical mapping imaging (localization, registration and fusion). The GE StarGuide system includes digital processing of data and images, signal analysis and display equipment, patient and equipment supports, components and accessories. The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation. The system may be used for patients of all ages. FAME: Functional-Anatomical Mutual Enhancement (FAME) technology is an image processing method intended for Computed Tomography (CT) based corrections of Nuclear Medicine (NM) bone scintigraphy images. FAME adjusts the radioisotope tracer distribution to correlate with the skeletal anatomical structures in the CT image. FAME may be used for patients of all ages.
Device Story
StarGuide is a hybrid SPECT-CT imaging system; utilizes twelve CZT detectors arranged around a gantry bore; detectors perform rotational, radial, and sweep motions to acquire SPECT projections. CT component (Optima CT540) provides anatomical cross-sectional images. System used in clinical settings by trained healthcare professionals (physicians, technologists). Input: radioisotope tracer uptake (NM) and X-ray transmission data (CT). Processing: automated patient contouring and motion planning; FAME algorithm performs CT-based correction of NM bone scintigraphy. Output: fused functional-anatomical images, tomographic scans, and quantitative uptake data. Physicians use outputs for diagnosis, staging, and therapy guidance. Benefits include high spatial/energy resolution and optimized workflow via automated scan planning.
Clinical Evidence
Clinical reader study conducted at two sites using 42 SPECT-CT exams. Five experienced NM physicians evaluated images using 5-point Likert scales for overall image quality and resolution. All physicians reported acceptable diagnostic results. Bench testing included SPECT resolution, sensitivity, energy resolution, count rate linearity, temporal resolution, body contouring, and FAME lesion detectability.
Technological Characteristics
Hybrid SPECT-CT system. NM: 12 CZT detectors, 80 cm bore, 40-270 keV energy range, fixed collimators. CT: Optima CT540 (16-channel). Connectivity: Xeleris Workstation and SmartConsole. Software: Moderate level of concern. Standards: IEC 60601-1 Ed 3.1. Motion: Rotational, radial, and sweep.
Indications for Use
Indicated for patients of all ages for detection, localization, and diagnosis of diseases, trauma, abnormalities, and disorders (e.g., cardiovascular, neurological, orthopedic, cancer). Used for tumor staging, therapy planning/monitoring, and theragnostic procedures. Contraindications not specified.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{1}------------------------------------------------
March 29, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The text is stacked, with "ADMINISTRATION" on the bottom line.
GE Healthcare % Mr. George Mashour Regulatory Affairs Manager 4 Hayozma Street Tirat Hacarmel, 30200 ISRAEL
Re: K210173
Trade/Device Name: StarGuide Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS, JAK Dated: January 21, 2021 Received: January 22, 2021
Dear Mr. Mashour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210173
Device Name StarGuide
#### Indications for Use (Describe)
The GE StarGuide system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging of tumors, planning, guiding, and monitoring therapy, including the nuclear medicine part of theragnostic procedures. The GE StarGuide system, combining a CZTbased, high energy and spatial resolution, Nuclear Medicine (NM) system and a Computed Tomography (CT) system, is intended to produce:
· NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types and by imaging features designed to enhance image quality.
· CT System: Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body, cardiac and vascular X-Ray Computed Tomography applications.
· NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based corrections of SPECT images as well as functional and anatomical mapping imaging (localization, registration and fusion).
The GE StarGuide system includes digital processing of data and images, signal analysis and display equipment, patient and equipment supports, components and accessories. The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation. The system may be used for patients of all ages.
FAME: Functional-Anatomical Mutual Enhancement (FAME) technology is an image processing method intended for Computed Tomography (CT) based corrections of Nuclear Medicine (NM) bone scintigraphy images. FAME adjusts the radioisotope tracer distribution to correlate with the skeletal anatomical structures in the CT image. FAME may be used for patients of all ages.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has a white border, giving it a clean and modern look. The logo is simple, recognizable, and represents the brand identity of General Electric.
# Section 5: 510(k) Summary
StarGuide
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a white border with decorative swirls or flourishes around it. The logo is simple, recognizable, and represents the General Electric brand.
## 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
K210173
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
| Date: | January 21, 2021 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Israel, Functional Imaging (GE Healthcare)<br>4 Hayozma Street<br>Tirat Hacarmel, 30200, Israel |
| Primary Contact: | George Mashour<br>Regulatory Affairs Manager<br>GE Medical Systems Israel, Functional Imaging<br>Tel: +972-4-8563666 ext. 639<br>Email: george.mashour@ge.com |
| Secondary Contact: | John Jaeckle<br>Chief Regulatory Affairs Engineer and Strategist<br>GE Healthcare<br>Tel: 262-424-9547<br>email: john.jaeckle@ge.com |
| Device Trade Name: | StarGuide |
| Device Classification: | Class II |
| Regulation Number: | 21CFR 892.1200 & 21CFR 892.1750 |
| Product Codes: | 90 KPS & 90 JAK |
| Predicate Device Information | |
|-------------------------------------|----------------------------------------------------|
| Device Name: | Discovery NM/CT 670 CZT |
| Manufacturer: | GE Medical Systems Israel, Functional Imaging |
| 510(k) Number: | K153402 |
| Regulation Number/<br>Product Code: | 21CFR 892.1200 & 21CFR 892.1750<br>90 KPS & 90 JAK |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are three water droplet shapes evenly spaced around the outside of the circle.
| Reference Device Information | |
|-------------------------------------|----------------------------------------------------|
| Device Name: | NM/CT 850 and NM/CT 860 |
| Manufacturer: | GE Medical Systems Israel, Functional Imaging |
| 510(k) Number: | K173816 |
| Regulation Number/<br>Product Code: | 21CFR 892.1200 & 21CFR 892.1750<br>90 KPS & 90 JAK |
| Reference Device Information | |
|-------------------------------------|----------------------------------------------------|
| Device Name: | Veriton CT |
| Manufacturer: | Spectrum Dynamics Medical Ltd |
| 510(k) Number: | K182484 |
| Regulation Number/<br>Product Code: | 21CFR 892.1200 & 21CFR 892.1750<br>90 KPS & 90 JAK |
## Marketed Devices
StarGuide is a modification of the predicate device, Discovery NM/CT 670 CZT. The primary changes are related to replacing the dual head CZT detector NM system on Discovery NM/CT 670 CZT with a new, SPECT acquisition-only NM system that uses twelve, slim CZT detectors that are uniformly distributed around the gantry bore. The CT portion of the StarGuide system remains untouched.
## Device Description
StarGuide is a SPECT-CT system that combines an all-purpose Nuclear Medicine imaging system and the commercially available GE Optima CT540 CT system. It is intended for general purpose Nuclear Medicine (NM) imaging procedures as well as head, whole body, cardiac and vascular CT applications and CT-based corrections and anatomical localization of SPECT images. The StarGuide system does not introduce any new Intended Use.
Each of StarGuide's twelve CZT detectors can independently rotate about their long axis and "sweep" the field of view (FOV). The detectors can also move rotationally around the gantry and radially in and out, similar to that of the reference device, Spectrum Dynamics' Veriton CT. The detectors on StarGuide's NM system are built up from the identical same CZT modules that are used in the current production version of the predicate device.
StarGuide's table is the same as the one used on the NM/CT 860 reference systems with only slight modifications. StarGuide's "SmartConsole" is the same as that on the NM/CT 850 and NM/CT 860 with modifications made primarily in support of StarGuide image processing. StarGuide's image processing (i.e. reconstruction and post reconstruction processing) uses known algorithms and methods that have been cleared for emission computed tomography (i.e. SPECT, PET). However, StarGuide introduces a new post reconstruction image processing algorithm, FAME, for CT-based correction of NM bone scintigraphy images for better correlation with the skeletal anatomical structures in the CT image.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. There are decorative flourishes around the letters, giving the logo a classic and recognizable appearance.
## Intended Use
The GE StarGuide system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using tomographic scanning mode supported by various acquisition types, and a Computed Tomography system which is intended for enabling CT-based corrections and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.
FAME: The Functional-Anatomical Mutual Enhancement (FAME) technology is an image processing method for Nuclear Medicine (NM) bone scintigraphy images.
#### Indications for Use
The GE StarGuide system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders, orthopedic disorders, and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy, including the nuclear medicine part of theragnostic procedures. The GE StarGuide system, combining a CZT-based, high energy and spatial resolution, Nuclear Medicine (NM) system and a Computed Tomography (CT) system, is intended to produce:
- NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types and by imaging features designed to enhance image quality.
- CT System: Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body, cardiac and vascular X-Ray Computed Tomography applications.
- NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based corrections of SPECT images as well as functional and anatomical mapping imaging (localization, registration and fusion).
The GE StarGuide system includes digital processing of data and images, signal analysis and display equipment, patient and equipment supports, components and accessories. The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation. The system may be used for patients of all ages.
FAME: Functional-Anatomical Mutual Enhancement (FAME) technology is an image processing method intended for Computed Tomography (CT) based corrections of Nuclear Medicine (NM) bone scintigraphy images. FAME adjusts the radioisotope tracer distribution to correlate with the skeletal anatomical structures in the CT image. FAME may be used for patients of all ages.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle and the letters are in a light blue color. The logo is simple and recognizable, representing the well-known multinational corporation.
## Technological Characteristics
StarGuide and its predicate use the same principles of SPECT acquisitions. The two systems rotate the detectors around the patient to acquire projections from multiple angular positions and reconstruct them into a SPECT image. The difference is that the detectors on StarGuide acquire the projections using a sweeping motion. Because of the multiple projections acquired with sweep motion at each rotational angular position, StarGuide SPECT scans require fewer angular positions than dual detector NM systems. The sweep motion allows SPECT scans to be acquired either uniformly over the sweep's full range or "focused" on a user-defined region of interest (ROI).
StarGuide automates the setup of SPECT scans for enhancements in workflow and scan motion efficiency. Based on a user-defined linear range (in the z direction), the system determines the contour of the patient's body and cradle and combines it with protocol-specific settings to plan the systems and detectors' motion and position during SPECT scans, without additional user intervention.
The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
{10}------------------------------------------------
# GE Healthcare 510(k) Premarket Notification Submission – StarGuide
Image /page/9/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has a decorative border that resembles water droplets or stylized leaves.
| Subsystem /<br>Specification | Predicate Device<br>Discovery NM/CT 670 CZT<br>(K153402) | Proposed Device<br>StarGuide |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| NM Gantry | Dual CZT detectors for Planar and<br>SPECT Imaging<br>70 cm bore | Twelve CZT detectors for SPECT<br>Imaging<br>80 cm bore |
| NM Detectors | Built using CZT modules<br>Rotational and radial motion. | Built using the identical CZT<br>modules<br>Rotational, radial, and sweep<br>motion. |
| NM Collimators | Wide Energy High Resolution<br>(WEHR)<br>Medium Energy High Resolution<br>and Sensitivity (MEHRS) | Fixed Collimator |
| Energy Range | 40 – 250 keV | 40 – 270 keV |
| Energy Resolution | ≤ 6.3%<br>for Tc-99m @ 20 kcps | ≤ 5.9%<br>for Tc-99m @ 20 kcps |
| SPECT Reconstructed<br>Spatial Resolution with<br>Scatter | Central: ≤ 6.0 mm<br>Radial: ≤ 5.5 mm<br>Tangential: ≤ 4.1 mm<br>*with WEHR collimator | Central: ≤ 4.5 mm<br>Radial: ≤ 4.1 mm<br>Tangential: ≤ 3.2 mm |
| Automated Patient<br>Contouring | Yes | Yes |
| CT System | Optima CT540, 16 channel CT<br>system | Identical |
| Standards Conformance | IEC 60601-1 and applicable<br>Collateral and Particular Standards. | Identical |
| Image Processing | Xeleris Workstation | Xeleris Workstation and Smart<br>Console |
| Software Level of Risk | Moderate | Moderate |
StarGuide's technological characteristics do not create new questions of safety or effectiveness, and did not introduce any new risks/hazards, warnings, or limitations.
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. There are also three water droplet shapes surrounding the circle. The logo is simple and recognizable, representing the company's brand identity.
## Determination of Substantial Equivalence
## Summary of Non-Clinical, Design Control Testing
StarGuide has successfully completed the design control testing per our quality system. No additional hazards were identified, and no unexpected test results were observed. StarGuide was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GE believes that the extensive bench testing and the physician evaluations performed are sufficient for FDA's substantial equivalence determination.
StarGuide has been independently tested and conforms with IEC 60601-1 Ed. 3.1 and its applicable Collateral and Particular Standards.
The following quality assurance measures have been applied to the development of the system:
- · Requirement Definition
- Risk Analysis
- Technical Design Reviews
- Formal Design Reviews
- · Software Development Lifecycle
- Testing on unit level (Module verification)
- · Integration testing (System verification)
- · System Testing:
- Safety Testing (Verification)
- System and Image Performance Testing (Verification)
- o Simulating Use Testing (Validation)
The testing and results did not raise new or different questions of safety and effectiveness than associated with predicate device. We consider the proposed device is substantially equivalent to the predicate device.
The substantial equivalence is also based on the software documentation for a "Moderate" level of concern. GE believes that StarGuide is of comparable type and substantially equivalent to the predicate device.
## Additional Non-Clinical Testing
Engineering bench testing was performed to support substantial equivalence, demonstrate performance, and substantiate the product claims. This included testing for:
- . SPECT resolution
- . Planar & volume sensitivity
- I Energy resolution and simultaneous multi-isotope acquisition
- . Count rate linearity
- 트 Temporal resolution for dynamic SPECT
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue.
- 트 Body contour & adaptive motion planning
- . Focused imaging acquisitions
- . Contrast to noise ratio
- Lutetium-177 imaging performance
- . Lesion detectability when using FAME
- . Patient contact safety
- . Generation of derived planar images
## Clinical Testing
A clinical reader study using 42 SPECT CT exams acquired in StarGuide was conducted at two clinical sites. The exams constituted a clinically representative sample for evaluation of StarGuide's performance. The results of the study support the determination of substantial equivalence. In total five experienced NM physicians scored the images using 5 point Likert scales for both overall image quality and image resolution. All of the physicians attested that their assessments demonstrated acceptable diagnostic results.
#### Substantial Equivalence Conclusion
The changes associated with StarGuide do not create a new Intended Use and represent equivalent technological characteristics, with no impact on the control mechanisms, operating principle, and energy type. GE's quality system's design verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
Based on development under GE Healthcare's quality system, the successful system and software verification and validation testing, conformance to standards, the additional engineering bench testing, and the clinical reader study demonstrates that StarGuide is substantially equivalent to, and hence as safe and as effective for its Intended Use, as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.