The Serena Sleep Appliance is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Device Story
Serena Sleep BMA and EMA are custom-fitted, removable intraoral devices for snoring and mild-to-moderate OSA. Input: digital dental scans provided by a treating dental physician. Transformation: CAD/CAM design based on physician-prescribed mandibular advancement; 3D printing using Nylon 12 (PA 2200). BMA uses integrated positioning blocks; EMA uses exterior pins and rubber bands for jaw repositioning. Output: custom trays that reposition the mandible forward, increasing pharyngeal space to improve airflow. Used in home setting by patient; prescribed and fitted by dental professionals. Benefits: reduced snoring and improved airway patency during sleep. Clinical decision-making: clinician selects advancement setting based on patient needs; device provides range of motion and lateral movement.
Clinical Evidence
No clinical testing was required. Evidence consists of bench testing comparing device performance to literature-based maximum jaw strengths, including compression, shear, and bruxism force testing, and one-year durability simulation.
Technological Characteristics
Custom-fitted intraoral trays; material: Nylon 12 (PA 2200); manufacturing: 3D printing (additive manufacturing); EMA straps: thermoplastic polyurethane/cured elastomer; non-sterile; biocompatibility: ISO 10993-5 compliant; adjustment: clinician-adjustable blocks (BMA) or clinician/patient-adjustable connectors (EMA).
Indications for Use
Indicated for adults with snoring and mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
March 10, 2021
Meris Investment Group % Sarah Moss Regulatory Affairs Consultant Medavice, Inc 11218 Zest Ct NE Blaine, Minnesota 55449
Re: K203606
Trade/Device Name: Serena Sleep Block Mandibular Advancement (BMA), Serena Sleep Elastic Mandibular Advancement (EMA) Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: LRK Dated: December 10, 2020 Received: December 10, 2020
Dear Sarah Moss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203606
#### Device Name
Serena Sleep Block Mandibular Advancement (BMA) Serena Sleep Elastic Mandibular Advancement (EMA)
Indications for Use (Describe)
The Serena Sleep Appliance is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 5. 510(k) Summary - K203606
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Serena Sleep | |
|------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person: | Phone:<br>Email: | Gary Maas, President<br>(651) 402-5577<br>Gary.Maas@serenasleep.com |
| Submission<br>Correspondent: | Address:<br>Phone:<br>Email: | Sarah Moss, Regulatory Affairs Consultant<br>11218 Zest Ct. NE, Blaine, MN 55449<br>(281) 795-1812<br>smoss@medavice.com |
| Date Prepared: | December 2020 | |
| Proprietary Name: | Serena Sleep Block Mandibular Advancement and Elastic Mandibular<br>Advancement Devices | |
| Common Name: | Device, Anti-snoring | |
| Product Code: | LRK | |
| Device Classification: | Class 2 | |
| Predicate Devices: | Block Mandibular Advancement: Prosomnus MicrO2 (Primary<br>Predicate - K133683) | |
| Device Description: | Elastic Mandibular Advancement: O2Vent Optima (Reference Device<br>(K190236)) | |
The Serena Sleep Block Mandibular Advancement is an intraoral device that treats snoring and sleep apnea through mandibular repositioning the lower jaw forward from its normal location, the patient's pharyngeal space is increased and their ability to exchange air during sleep is improved. This device consists of custom fitted trays which fit over the upper and lower teeth. Based on the physician's prescription, positioning blocks are built into the upper and lower trays The mandibular advancement is achieved through the physician's selection of a specific upper and lower tray that when fitted, sets the optimal mandibular advancement. The separate two-piece construction is desirable for patients that want greater range of motion and lateral movement.
{5}------------------------------------------------
The Elastic Mandibular Advancement model is similar to the Block Mandibular Advancement but has exterior pins that allow for the use of rubber bands rather than the blocks. These bands provide additional treatment options and forces for repositioning the bottom jaw.
## Indications for Use:
The Serena Sleep appliance is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
{6}------------------------------------------------
| Specification | Serena Sleep<br>Block Mandibular Advancement | Prosomnus MicrO2<br>(Predicate Device) | Comparison<br>Result |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| 510(k) Number | | K133683 | |
| Device Photo | Image: Serena Sleep Device | Image: Prosomnus MicrO2 Device | |
| Indication for<br>Use | The Serena Sleep Appliance is<br>a removable medical device<br>that is fitted in the patient's<br>mouth and is intended to<br>reduce or alleviate snoring and<br>mild to moderate obstructive<br>sleep apnea (OSA) in adults. | The MICRODENTAL, Inc. MicrO2<br>device is intended to reduce night<br>time snoring and mild to moderate<br>obstructive sleep apnea (OSA) in<br>adults. | Equivalent |
| Product Code | LRK | LRK | Equivalent |
| Regulations | 21CFR 872.5570 | 21CFR 872.5570 | Equivalent |
| Class | 2 | 2 | Equivalent |
| Use of<br>Device | Removable intraoral device. Single<br>patient multiple use. Prescription<br>use only. | Removable intraoral device. Single<br>patient multiple use. Prescription<br>use only. | Equivalent |
| Target<br>Population | People over 18 of age to<br>reduce or alleviate snoring<br>and mild to moderate<br>obstructive sleep apnea (OSA)<br>in adults. | Intended to reduce night time<br>snoring and mild to moderate<br>obstructive sleep apnea (OSA) in<br>adults | Equivalent |
| Device | Repositions the lower jaw<br>forward. | Repositions the lower jaw<br>forward. | Equivalent |
| Functionality | Acts by increasing the | Acts by increasing the pharyngeal | Equivalent |
| Specification | Serena Sleep<br>Block Mandibular Advancement | Prosomnus MicrO2<br>(Predicate Device) | Comparison<br>Result |
| | pharyngeal space to improve<br>the patient's ability to<br>exchange air. | space to improve the patient's<br>ability to exchange air. | |
| | Retains the top and bottom<br>teeth using rigid trays (PA<br>2200). | Retains the top and bottom teeth<br>using rigid trays (PMMA). | Similar – Different<br>Material |
| | Upper and lower trays are<br>separate | Upper and lower trays are<br>separate | Equivalent |
| Device Design<br>and Principle or<br>Operation | Mandibular advancement is<br>achieved through positioning<br>blocks that are built into the<br>upper and lower trays. | Mandibular advancement<br>achieved through twin-mated<br>positioning posts built into the<br>upper and lower trays | Similar – The BMA<br>uses blocks and the<br>MicrO2 uses posts |
| Means of<br>advancing the<br>mandible | Mandibular advancement<br>achieved via the use of<br>interlocking blocks placed on<br>the trays that are in line with<br>the teeth. The amount of<br>mandibular advancement is<br>based on the location of the<br>"step" relative to the<br>centerline of the block. | Mandibular advancement<br>achieved through twin-mated<br>positioning posts built into the<br>upper and lower trays | Equivalent |
| Adjustment | Can be adjusted by the<br>clinician. | Can be adjusted by the clinician. | Equivalent |
| Design | Serena Sleep devices will be<br>designed from digital scans of<br>a patient's dentition. This<br>baseline model of the<br>dentition is provided by the<br>treating dental physician.<br>Using the 3D image of the<br>patient's dentition and the<br>treating physician's indicated<br>mandibular advancement, the<br>patient's custom device is<br>designed by adding blocks.<br>The device is then 3D printed<br>using the material Nylon 12<br>(PA 2200). | CAD/CAM generated specifically<br>for each prescription and made<br>with a hard PMMA material | Similar – different<br>material |
| Specification | Serena Sleep<br>Block Mandibular Advancement | Prosomnus MicrO2<br>(Predicate Device) | Comparison<br>Result |
| Supplied<br>Sterile/Non-<br>Sterile | Non-sterile | Non-sterile | Equivalent |
| Materials:<br>Upper and<br>Lower Trays | Nylon 12 (PA 2200) | Hard PMMA | Similar — different<br>material |
| Cleaning<br>Instructions | Clean the device after use in<br>water with a soft<br>toothbrush. Rinse, dry, and<br>store in the case provided. | Cleaned daily. | Similar |
| Biocompatibility<br>: ISO 10993-5 | Passed | Not performed as the materials<br>are identical as in the<br>Somnomed MAS RXA (K050592). | Equivalent |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| Comparison to Predicate Devices: Elastic Mandibular Advancement Device | | | |
|------------------------------------------------------------------------|--|--|--|
|------------------------------------------------------------------------|--|--|--|
| Specification | Serena Sleep<br>Elastic Mandibular Advancement | O2Vent Optima<br>(Predicate Device) | Comparison<br>Result |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| 510(k) Number | | K190236 | |
| Device Photo | Image: Serena Sleep Device | Image: O2Vent Optima Device | |
| Indication for<br>Use | The Serena Sleep Appliance is<br>a removable medical device<br>that is fitted in the patient's<br>mouth and is intended to<br>reduce or alleviate snoring and<br>mild to moderate obstructive<br>sleep apnea (OSA) in adults. | The O2Vent Optima is a<br>removable medical device that is<br>fitted in the patient's mouth and<br>is intended to reduce or alleviate<br>snoring and mild to moderate<br>obstructive sleep apnea (OSA).<br>The device is indicated for use<br>during sleep to aid in the<br>treatment of these conditions. | Equivalent |
| Product Code | LRK | LRK | Equivalent |
| Regulations | 21CFR 872.5570 | 21CFR 872.5570 | Equivalent |
| Class | 2 | 2 | Equivalent |
| Use of<br>Device | Removable intraoral device. Single<br>patient multiple use. Prescription<br>use only. | Removable intraoral device. Single<br>patient multiple use. Prescription<br>use only. | Equivalent |
| Target<br>Population | People over 18 of age to<br>reduce or alleviate snoring<br>and mild to moderate<br>obstructive sleep apnea (OSA)<br>in adults. | People over 18 years of age who<br>snore and/or have sleep apnea. | Equivalent |
| Device<br>Functionality | Repositions the lower jaw<br>forward. | Repositions the lower jaw forward. | Equivalent |
| | Acts by increasing the<br>pharyngeal space to improve<br>the patient's ability to<br>exchange air. | Acts by increasing the pharyngeal<br>space to improve the patient's<br>ability to exchange air. | Equivalent |
| Specification | Serena Sleep | O2Vent Optima | Comparison<br>Result |
| | Elastic Mandibular Advancement | (Predicate Device) | |
| | Retains the top and bottom<br>teeth using rigid trays (PA<br>2200). | Retains the top and bottom teeth<br>using rigid trays (PA 2200). | Equivalent |
| | Upper and lower trays are<br>separate | Upper and lower trays are separate | Equivalent |
| Device Design<br>and Principle or<br>Operation | Mandibular advancement<br>using exterior pins that allow<br>for the use of rubber bands<br>rather than the blocks. These<br>bands provide additional<br>treatment options and forces<br>for repositioning the bottom<br>jaw. | The trays have protrusions (lugs) at<br>the side of the upper and lower<br>parts with a connector between the<br>upper and lower parts to stabilize<br>and/or advance the lower jaw. | Equivalent |
| Means of<br>advancing the<br>mandible | Lower jaw adjusted by<br>attaching connectors of<br>varying lengths | Lower jaw adjusted by attaching<br>connectors of varying lengths | Equivalent |
| Adjustment | Can be adjusted by the<br>clinician and patient | Can be adjusted by the clinician and<br>patient | Equivalent |
| Design | Serena Sleep devices will be<br>designed from digital scans of<br>a patient's dentition. This<br>baseline model of the<br>dentition is provided by the<br>treating dental physician.<br>Using the 3D image of the<br>patient's dentition and the<br>treating physician's indicated<br>mandibular advancement, the<br>patient's custom device is<br>designed by adding pins. The<br>device is then 3D printed<br>using the material Nylon 12 | Customized for each patient in a<br>dental laboratory located at the<br>manufacturing site based on the<br>dentist prescription Use of computer<br>aided design (CAD) and computer<br>aided manufacturing (CAM) and is<br>made through additive<br>manufacturing using laser sintering<br>of Nylon 12 (PA2200) material. The<br>use of these technologies provides<br>for customization according to the<br>unique characteristics of the patient<br>oral anatomy based on the<br>prescription provided by the clinician | Equivalent |
| Adjustment<br>Accessory | Replacement/Re-supply<br>connectors (13-21 mm) | Replacement/re-supply connectors<br>(13-21 mm) | Equivalent |
| Supplied<br>Sterile/Non-<br>Sterile | Non-sterile | Non-sterile | Equivalent |
| Materials: Upper<br>and Lower Trays | Nylon 12 (PA2200) | Nylon 12 (PA2200) | Equivalent |
| Specification | Serena Sleep<br>Elastic Mandibular Advancement | 02Vent Optima<br>(Predicate Device) | Comparison<br>Result |
| Materials:<br>Straps | 100% thermoplastic<br>Polyurethane/cured<br>elastomer | 100% thermoplastic<br>Polyurethane/cured elastomer | Equivalent |
| Cleaning<br>Instructions | Clean the device after use in<br>water with a soft<br>toothbrush. Rinse, dry, and<br>store in the case provided. | Clean the device daily in lukewarm<br>water with a soft toothbrush. Rinse,<br>dry, and store in the case provided. | Equivalent |
| Biocompatibility<br>: ISO 10993-5 | Passed | Passed | Equivalent |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
## Performance Testing (Bench)
Testing was conducted to compare the Serena Sleep Block Mandibular Advancement device and Elastic Mandibular Advancement device to mechanical forces found through a literature search of maximum jaw strengths. Compression, Shear, and Bruxism forces were tested, as well as a product durability test for one year of use. The bench testing proved that they devices were able to withstand forces greater than expected jaw forces during nighttime jaw clenching and bruxing.
A key design advantage of the Serena Sleep products are its ability to customize the amount of mandibular advancement for each patient. This feature allows the doctor the ability to find a setting that is right for each patient.
Therefore, the performance of the Serena Sleep device were shown to be at least equivalent to the predicate devices.
## Clinical Testing
No Clinical Testing was required for this product.
### Statement of Equivalence
### Block Mandibular Advancement
As summarized above, the main differences between the subject (Serena Sleep BMA) and predicate (Prosomnus MicrO2) devices are:
- Predicate device does not use PA2200 in the additive manufacturing process
- Predicate device uses positioning pins instead of positioning blocks
Based on comparison of indications for use, user population, performance testing, mechanical and technological features, the Block Mandibular Advancement device has been shown to be substantially equivalent to the legally marketed predicate device. This device does not raise any new safety or effectiveness questions as compared to the predicate device.
## Elastic Mandibular Advancement
As summarized above, there are no functional differences between the subject (Elastic Mandibular Advancement Device) and predicate (O2Vent Optima) devices.
Based on comparison of indications for use, user population, performance testing, mechanical and technological features, the Elastic Mandibular Advancement device has been shown to be substantially equivalent to the legally marketed predicate device. This device does not raise any new safety or effectiveness questions as compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.