Tigertriever and Tigertriever 17 Revascularization Devce
K203592 · Rapid Medical Ltd. · NRY · Mar 23, 2021 · Cardiovascular
Device Facts
Record ID
K203592
Device Name
Tigertriever and Tigertriever 17 Revascularization Devce
Applicant
Rapid Medical Ltd.
Product Code
NRY · Cardiovascular
Decision Date
Mar 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.
Device Story
Stentriever device for mechanical thrombectomy in ischemic stroke; used in neurovasculature. Comprised of adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire, and handle. Physician delivers device via microcatheter under fluoroscopy; pulls slider on handle to manually expand mesh to conform to vessel diameter; mesh encapsulates clot for retrieval. Provides physician control over mesh diameter during procedure. Used by neurointerventionalists in clinical settings. Benefits include effective clot removal and restoration of blood flow in large intracranial vessels.
Clinical Evidence
TIGER prospective clinical trial (N=148 mITT cohort) evaluated safety and efficacy. Primary effectiveness: successful revascularization (mTICI ≥2b) in ≤3 passes achieved in 73% (95% CI: 66.3%). Primary safety: composite of all-cause mortality at 90 days and/or sICH at 24 hours was 17.7% (upper bound 95% CI: 24.8%). Secondary outcomes included 54.7% good clinical outcome (mRS ≤2) at 90 days. Data supports substantial equivalence to predicate.
Indicated for patients 18-85 years old experiencing ischemic stroke with large-vessel occlusion (ICA, MCA, M1, M2, vertebral, or basilar arteries) within 8 hours of symptom onset, who are ineligible for or failed IV t-PA therapy. Contraindicated in patients with excessive arterial tortuosity, stenosis, or proximal vessel occlusion preventing access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
March 23, 2021
Rapid Medical Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, 23rd Floor Philadelphia, Pennsylvania 19103
Re: K203592
Trade/Device Name: Tigertriever and Tigertriever 17 Revascularization Device Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: February 19, 2021 Received: February 19, 2021
Dear Janice Hogan:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventiona1 and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203592
#### Device Name
Tigertriever and Tigertriever 17 Revascularization Device
#### Indications for Use (Describe)
The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> | <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(K) SUMMARY Tigertriever Revascularization Device K203592
Submission Sponsor
Rapid Medical Ltd. Carmel Building, P.O. Box 337 Yokneam, 2069205 Israel Company Phone No .: +972-72-250-3331
Contacts:
Dr. Orit Yaniv, VP of QA/RA Email: orit@rapid-medical.com
Ronen Eckhouse, CEO Email: ronen@rapid-medical.com
### Date Prepared
March 23, 2021
### Device Identification
Trade/Proprietary Name: Tigertriever and Tigertriever 17 Revascularization Device Common/Usual Name: Catheter, Thrombus Retriever Classification Name: Percutaneous catheter Regulation Number: 21 CFR 870.1250 Product Code: NRY Device Class: II Classification Panel: Neurology
### Legally Marketed Predicate Device(s)
Solitaire 2 Revascularization Device (K141491)
### Indication for Use Statement
The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.
{5}------------------------------------------------
## Device Description
The Tigertriever device is a stentriever that is comprised of an adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire and a handle. The shaft connects the mesh and the handle by the core wire that runs inside the shaft from the distal end of the mesh to the slider activation element in the handle. The mesh is expanded when the physician pulls the slider, since the wires of the mesh are completely radiopaque, the physician sees the mesh under fluoroscopy and controls it until it conforms to the vessel diameter. The design of the wire mesh is optimized to penetrate the clot and encapsulate it during retrieval. Two versions of the device are available. The standard version Tigertriever (TRPP7155) has a net length of 32mm (unexpanded form) and it is delivered through a microcatheter with an internal diameter of 0.021 inches. The shorter version Tigertriever 17 (TRPP7166) has a net length of 23 mm (unexpanded form) and it is delivered through a microcatheter with an internal diameter of 0.017 inches. The Tigertriever is provided with a 3.5 Fr peelable loading sheath.
## Comparison of Technological Characteristics with the Predicate Device
The Tigertriever and the Solitaire 2 predicate have the same indications for use, similar technological characteristics and principle of operations. Both stentrievers share similar design and similar construction materials. Both devices are used to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke. Both stentrievers are delivered to the target vessel by means of a microcatheter and expanded once deployed from the microcatheter. The expansion of the subject device is controlled by the physician, while the predicate mesh is self-expanded to a fixed diameter upon deployment. However, this difference does not raise new questions of safety or effectiveness. A detailed comparison between the Tigertriever and the predicate device is provided in the table below.
| | Tigertriever Revascularization<br>Device | Solitaire 2 Revascularization Device<br>(Predicate Device) |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K203592 | K141491 |
| Regulation | 21 CFR 870.1250 | 21 CFR 870.1250 |
| Product Code | NRY | NRY |
| | Tigertriever Revascularization<br>Device | Solitaire 2 Revascularization Device<br>(Predicate Device) |
| Indications for<br>Use | The Tigertriever Revascularization<br>Device is intended to restore blood<br>flow by removing thrombus from a<br>large intracranial vessel in patients<br>experiencing ischemic stroke<br>within 8 hours of symptom onset.<br>Patients who are ineligible for<br>intravenous tissue plasminogen<br>activator (IV t-PA), or who fail IV<br>t-PA therapy, are candidates for<br>treatment. | The Solitaire 2 Revascularization<br>Device is intended to restore blood<br>flow by removing thrombus from a<br>large intracranial vessel in patients<br>experiencing ischemic stroke within 8<br>hours of symptom onset. Patients who<br>are ineligible for intravenous tissue<br>plasminogen activator (IV t-PA) or<br>who fail IV t-PA therapy are<br>candidates for treatment. |
| Anatomical<br>Location | Neurovasculature | Neurovasculature |
| Sterilization<br>Method | Ethylene oxide | Ethylene oxide |
| Single Use | Yes | Yes |
| Packaging | Placed into a Dispenser hoop,<br>blister, Tyvek pouch, and Carton<br>box | Placed into a Dispenser hoop, Tyvek<br>pouch, and Carton box |
| Distal OD<br>(expanded<br>configuration) | 3mm Tigertriever 17<br>6mm Tigertriever | 4mm<br>6mm |
| Stent length<br>(unexpanded<br>configuration) | 23mm Tigertriever 17<br>32mm Tigertriever | 31mm |
| Overall length | 200 cm | 184 cm |
| Design of distal<br>portion | Close end braided nitinol mesh,<br>manually expandable | Open end laser cut stent, self-<br>expanded |
| Mode of<br>operation | Manual expansion of the braided<br>distal portion into the clot using the<br>handle component | Self-expansion of the distal stent<br>portion into the clot following<br>retraction of the delivery catheter |
| Materials | | |
| Stent | Nitinol | Nitinol |
| Markers | 90% Platinum/ 10% Iridium | 90% Platinum/ 10% Iridium |
| | Tigertriever Revascularization Device | Solitaire 2 Revascularization Device (Predicate Device) |
| Core wire (shaft) | Nitinol core wire and stainless steel shaft | Nitinol wire |
| Push wire shrink tubing | PTFE | PTFE |
| Introducer Sheath | PTFE/Grilamid | PTFE |
{6}------------------------------------------------
{7}------------------------------------------------
# Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the device and in showing substantial equivalence to the predicate device, Rapid Medical Ltd. completed a number of non-clinical performance tests. The device meets all the requirements of overall design, sterilization, and biocompatibility. Testing results confirm that the design output meets the design specification for the device.
# Biocompatibility
Biocompatibility of the Tigertriever was based on the biocompatibility testing data for the Comaneci Embolization Assist Device (DEN170064). The two devices share the same manufacturing process and same manufacturing environment. In addition, the two devices are intended to be used in the same anatomical locations, and are identical in terms of frequency and duration of exposure. Biocompatibility testing was completed for Comaneci device and consisted of the following tests: Cytotoxicity, Irritation (Intracutaneous Reactivity), Sensitization, Hemocompatibility, Pyrogenicity, Acute Systemic and Toxicity Testing. In addition, in vivo thrombogenicity was performed for the Tigertriever. Results of the biocompatibility testing indicate that Tigertriever revascularization device is biocompatible and is substantially equivalent for its intended use.
| Biological Endpoint | Test Results | Conclusion |
|--------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------|
| Cytotoxicity - ISO<br>Elution Method | Grade 0 reactivity observed 48 hours<br>post exposure to test article extract. | Non-cytotoxic |
| Irritation -<br>Intracutaneous Reactivity<br>in Rabbit | Difference of overall mean score<br>between test article and control was 0. | Non-irritant |
| Sensitization - Guinea<br>Pig Maximization Test | Grade 0, no evidence of causing<br>delayed dermal contact sensitization. | Does not elicit<br>sensitization<br>response |
{8}------------------------------------------------
| Biological Endpoint | Test Results | Conclusion |
|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Hemocompatibility -<br>Complement activation<br>Assay | SC5b-9 concentration of the test article<br>sample was statistically lower than the<br>activated NHS control (p<0.05), and<br>was not statistically higher than the<br>negative control. | Pass |
| Hemocompatibility – In<br>Vitro Hemolysis | Test article = 0.0% hemolysis. | Non-Hemolytic |
| Pyrogenicity - Material<br>Mediated Pyrogenicity in<br>Rabbit | The total rise of rabbit temperatures<br>during the 3-hour observation<br>period was within acceptable USP<br>limits. | Non-pyrogenic |
| Systemic Toxicity -<br>Systemic Toxicity Study<br>in Mice | No mortality or evidence of systemic<br>toxicity from the extracts injected<br>into mice. | Non-toxic |
| Thrombogenicity - Acute<br>Pre-Clinical Evaluation<br>of the Safety of<br>Tigertriever device in a<br>Swine Model<br>Thrombogenicity | Test device did not show higher<br>thrombogenicity rate compared to the<br>predicate device. | Pass |
## Sterilization and Shelf Life
The device is sterilized by 100% Ethylene Oxide.
The shelf life testing for Tigertriever revascularization device has been conducted (T = 2.5 years) with test results confirmed that all acceptance criteria were met.
## Bench Tests
The device passed all performance bench testing in accordance with internal requirements, national standards and international standards as shown in the table below to support substantial equivalence of the device.
| Performance Bench Testing | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Test Method Summary | Conclusions |
| Simulated use test | Simulated use testing of the<br>Tigertriever Revascularization<br>Device was performed in an | The device was tested for<br>handling and clot retrieval in<br>an in vitro tortuous path |
| Performance Bench Testing | | |
| Test | Test Method Summary | Conclusions |
| anatomical model which simulated<br>the tortuosity of the<br>neurovasculature. Devices were<br>delivered through the tortuous<br>anatomical model to evaluate the<br>effectiveness of the device at<br>retrieval of firm and soft clots. | anatomical model, which has<br>been used in the evaluation of<br>other similar devices. The<br>subject device effectively<br>retrieved clot and restored flow<br>in the test model. | |
| Radial force | The radial force of the subject device<br>was measured within a range of<br>lumen diameters applicable to the<br>intended vasculature and compared<br>with the radial forces measured for<br>the predicate devices. | The radial force of the subject<br>device when tested in<br>applicable lumen sizes is<br>comparable to the predicate<br>device. |
| Durability | Damage was evaluated after delivery<br>and withdrawal of the device beyond<br>the recommended number of passes<br>and resheathings recommended in<br>the instructions for use. | Devices tested demonstrated<br>no damage after delivery and<br>withdrawal testing. Durability<br>established acceptable<br>performance for 3 passes,<br>which is at least equivalent to<br>the number of passes specified<br>in the predicate labeling (2<br>passes per device). |
| Delivery,<br>deployment and<br>retrieval | The delivery, deployment and<br>retrieval forces were measured<br>during simulated use of the subject<br>device. | The device was tested for<br>delivery, deployment, and<br>retrieval in an in vitro tortuous<br>path anatomical model, which<br>has been used in the evaluation<br>of other similar devices. The<br>subject device demonstrated<br>acceptable performance with<br>respect to delivery,<br>deployment and retrieval. |
| Torque strength | Devices were tracked through a<br>microcatheter in a tortuous path<br>anatomical model and evaluated for<br>damage following a number of<br>rotations with the distal end<br>constrained | The device demonstrated the<br>ability to withstand 5 rotations<br>without damage. Like the<br>predicate device, rotational<br>maneuvers are not expected<br>under the intended conditions |
| Performance Bench Testing | | |
| Test | Test Method Summary | Conclusions |
| | | of clinical use. Therefore, the |
| | | results demonstrate acceptable |
| | | torque strength. |
| Dimensions test | Dimensional inspection was<br>tested per engineering drawings. | The subject device dimensions<br>are within the range of existing<br>predicate dimensions for this<br>device type. The minor<br>differences in dimensions do<br>not affect performance, safety<br>or effectiveness. |
| Tip flexibility | Tip Flexibility was performed to<br>measure the force required to deflect<br>Tigertriever tips to 90 degrees at<br>7mm test lengths. | The subject device met<br>acceptance criteria based on<br>comparable device used in the<br>same anatomy and<br>demonstrated similar tip<br>flexibility. |
| Kink resistance | Tigertriever with ancillary<br>microcatheter was looped around<br>post of calibrated kink measurement<br>jig (0.25" diameter) about 50mm<br>from the distal end of the<br>microcatheter. | Kink resistance was evaluated<br>under conditions simulating<br>anatomic tortuosity,<br>comparable to predicates, and<br>demonstrated acceptable<br>performance. |
| Tensile test | The minimum force to break the<br>Tigertriever was tested for all joints. | The tensile strength of the<br>device met acceptance criteria<br>based on recognized standards<br>(ISO 10555-1). |
| Particulates | Particulate test was performed<br>according to the light obscuration<br>test method. Simulated use testing of<br>the subject and predicate devices was<br>performed in an anatomical model.<br>Devices were flushed and the fluid<br>was evaluated for particle sizes of<br>≥10, ≥25 and ≥50 $ μ $ m. | The particulate generated by<br>the subject device was similar<br>to the particulate generated by<br>the predicate device. |
| Austenite Finish (Af)<br>Temperature | The Active Af temperature was<br>determined from a graph of<br>displacement as a function of | The Af temperature of the<br>device met acceptance criteria. |
| Performance Bench Testing | | |
| Test | Test Method Summary | Conclusions |
| Coating integrity<br>assessment | temperature (Bend and Free<br>Recovery per ASTM F2082). | Results demonstrated no<br>damage to the coating<br>following simulated use. |
| | The test was performed with the | |
| | Comaneci device and not with the | |
| | subject device (the design of the two | |
| | device is the same in terms of the | |
| | core wire mechanism). Damage to | |
| | the PTFE core wire coating was | |
| | evaluated following simulated use. | |
| | Corrosion | |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
# Pre-Clinical Animal Testing Data
The safety of the Tigertriever was demonstrated in a controlled animal study in domestic swine comparing the Tigertriever device and the predicate device. The results of the study showed substantial equivalence between the subject device and the predicate device.
The study included four animals; two animals were used for the acute stage (3 days period) and two animals were used for the chronic stage (30 days period). The study included two procedures: the first procedure was performed on day zero (3 or 30 days before the animal was sacrificed), and the second procedure was performed on the sacrifice day (day 30).
The first procedure included simulation (both for test device and predicate device) at two sites in the femoral artery, each simulation included three passes, first pass with clot removal, and two additional passes without clot (total three device delivery and retrieval simulations at each site of the femoral artery). No damage was observed by angiography during procedure in the subject device and the predicate device treated sites, no safety related abnormalities were observed by the veterinarian and no abnormal gross findings were recorded for external surface, orifices, cavities, or muscles and tissues downstream of the femoral arteries. In addition, the histology evaluations of the arterial tested sites revealed a comparable range of observations for the subject and predicate devices.
{12}------------------------------------------------
The second procedure was performed on the study termination day before the animal was sacrificed. To evaluate the thrombogenicity of the subject and predicate device, the tested unit was inserted to the renal artery in one side of the kidney and deployed for 10 minutes, then an additional two passes were performed, resulting in a total of 30 minutes for each device. Thrombogenicity scores revealed comparable values for the subject and predicate devices; kidney microscopic and histology evaluations revealed comparable range of observations for the subject and predicate devices.
In conclusion, the animal study showed that thrombectomy using the Tigertriever was safe without evidence of vessel injuries or abnormal thrombogenicity. Clinical, angiographic, pathologic, and histologic data supported a similar safety profile between the Tigertriever and the Solitaire predicate.
## Clinical Study
The TIGER (Treatment with Intent to Generate Endovascular Reperfusion) clinical trial assessed the efficacy and safety of the Tigertriever Revascularization Device against a performance goal derived from the TREVO 2, SWIFT, MR CLEAN, ESCAPE, REVASCAT and SWIFT PRIME clinical trials. Key inclusion criteria were: patients with a large-vessel occlusion who could be treated within 8 hours of stroke symptom onset; age 18-85; 8_NIHSS_29; angiographic confirmation of an occlusion of an ICA, MCA, M1 or M2, vertebral or basilar arteries, and IV t-PA, if used, was initiated within 3 hrs of stroke onset. Key exclusion criteria were: angiographically evident excessive arterial tortuosity, stenosis, or any occlusion, in a proximal vessel that required treatment or would prevent access to the site of occlusion.
One hundred and sixty (160) patients signed informed consent, treated with the Tigertriever device and included in the study. Of these. 148 patients met all inclusion and exclusion criteria as requested by FDA for the modified Intent-to-Treat (mITT) Cohort. Among the 12 patients excluded to meet the FDA defined mITT criteria, 1 had excessive arterial tortuosity, 2 had prior recent stroke in the past 3 months, 1 had a 100% occluded vessel which required stenting, 2 did not meet the criteria around laboratory ranges, 1 was given IV t-PA initiated >3 hours from symptom onset, 1 was treated with mRS 3, 2 missing pregnancy test, 1 was treated with the device over 8 hours from symptom onset, and 1 was treated with an aspiration device prior to the Tigertriever. Per FDA's request, within this mITT cohort, use of rescue therapy at any point in the procedure was imputed as a failure for the revascularization and clinical outcome endpoints. The table below summarizes the effectiveness and safety outcomes of this cohort.
The Primary Effectiveness Endpoint was successful revascularization defined as an mTICI score of at least 2b in the target vessel, following three or less passes of the Tigertriever device, using Core Laboratory adjudicated data. The Primary Safety Endpoint defined as the composite
{13}------------------------------------------------
of all-cause mortality at 90 days and/or symptomatic intracranial hemorrhage (sICH) within 24 (18-36) hours of the study procedure. The Primary Safety Endpoint was adjudicated by the Clinical Events Committee (CEC).
| TIGER Study Results<br>Effectiveness and Safety Results for use of only Tigertriever | |
|-------------------------------------------------------------------------------------------------------------|----------------------------|
| Endpoint | mITT (N=148) |
| Successful revascularization rate after Tigertriever<br>treatment1 -<br>Patients with mTICI ≥2b, n (%) | 108/148 (73%) |
| Lower Bound of 95% CI | 66.3% |
| Secondary Effectiveness Endpoints | |
| Successful revascularization rate after first pass with<br>Tigertriever - patients with mTICI ≥ 2b<br>n (%) | mITT (N=148)<br>81 (54.7%) |
| Good Clinical Outcome:<br>mRS ≤2 at 90 days2<br>n, (%) | N= 148<br>81/148 (54.7%) |
| Patient Reported Outcomes: EQ5D at 90 days3 | N= 107 |
| Mean (SD) | 73.5 (21.8) |
| Median | 80 |
| 25th to 75th | 62.5,90 |
| Patient Reported Outcomes: ALDS at 90 days3 | N= 109 |
| Mean (SD) | 11.1 (4.5) |
| Median | 13.0 |
| 25th to 75th | 8,15 |
| Primary Composite Safety Endpoint | |
| Endpoint | mITT (N=147)4 |
| Mortality at 90 Days and/or sICH at 24 Hours post<br>procedure4 | 26 (17.7%) |
| Upper Bound of 95% CI | 24.8% |
| Mortality at 90 Days4 | 26 (17.7%) |
| sICH at 24 Hours post procedure5 | 3 (2.0%) |
| Secondary Safety Endpoints | |
| Endpoint | mITT (N=147)6 |
| Total Asymptomatic ICH at 24 Hours (%), n (%) | 47 (31.9%) |
| HI-1 | 18 (12.2%) |
| HI-2 | 12 (8.8%) |
| PH-1 | 0 |
| PH-2 | 2 (1.4%) |
| SAH | 11 (7.5%) |
| SAH and HI-2 | 2 (1.4%) |
| SAH and PH-2 | 1 (0.7%) |
| Neurological Deterioration at 24 Hours, n (%) | 13 (8.8%) |
| Embolization to New Territory at End of Procedure, n<br>(%) | 4 (2.7%) |
{14}------------------------------------------------
| <sup>1</sup> Use of rescue therapy at any point in the procedure was imputed as a failure to achieve the endpoint. |
|--------------------------------------------------------------------------------------------------------------------|
| There were 12 cases imputed as failure. |
| <sup>2</sup> Five patients had missing 90 days mRS and were imputed as a failure to achieve the endpoint. |
| <sup>3</sup> Sample sizes vary due to missing data at 90 days. |
| <sup>4</sup> Reduced sample size due to one patient who withdrew consent prior to the 90 d follow up visit. |
| <sup>5</sup> All 3 subjects with sICH at 24 hours post-procedure died. |
| <sup>6</sup> Reduced sample size due to missing 24 hour CT or final angiogram. |
Definitions:
| mRS: | modified Rankin Score. |
|-------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EQ5D: | EuroQol Five Dimensions. |
| ALDS: | Academic Medical Center Linear Disability Score. |
| SICH: | symptomatic Intracranial Hemorrhage, any parenchymal hematoma type 2, remote intracerebral hemorrhage, subarachnoid hemorrhage, or intraventricular hemorrhage that is the predominant cause of ≥4 points NIHSS deterioration at 24 hours. |
| HI: | Hemorrhagic Infarction. |
| PH: | Parenchymal Hematoma. |
| SAH: | Subarachnoid Hemorrhage. |
TIGER Summary of Serious Adverse Events (SAEs)
One hundred and sixty-five (165) SAE occurred within all the study population, all patients enrolled and treated with Tigertriever. The table below summarizes the frequency of the SAEs classified by System Organ.
| System Organ Class (SOC) | All<br>(165 Events) |
|-------------------------------------------------------------------------|---------------------|
| Blood and lymphatic system disorders | 3 (1.8%) |
| Cardiac disorders | 22 (13.3%) |
| Gastrointestinal disorders | 6 (3.6%) |
| General disorders and administration site conditions | 2 (1.2%) |
| Infections and infestations | 11 (6.7%) |
| Injury, poisoning and procedural complications | 12 (7.3%) |
| Injury, poisoning and procedural complications Vascular disorders | 1 (0.6%) |
| Investigations | 3 (1.8%) |
| Metabolism and nutrition disorders | 3 (1.8%) |
| Musculoskeletal and connective tissue disorders Psychiatric disorders | 1 (0.6%) |
| Nervous system disorders | 40 (24.2%) |
| Nervous system disorders Injury, poisoning and procedural complications | 1 (0.6%) |
| Nervous system disorders Nervous system disorders | 1 (0.6%) |
| Nervous system disorders Surgical and medical procedures | 1 (0.6%) |
| Psychiatric disorders | 1 (0.6%) |
| Renal and urinary disorders | 4 (2.4%) |
| Reproductive system and breast disorders | 1 (0.6%) |
{15}------------------------------------------------
| System Organ Class (SOC) | All<br>(165 Events) |
|-------------------------------------------------|---------------------|
| Respiratory, thoracic and mediastinal disorders | 25 (15.2%) |
| Skin and subcutaneous tissue disorders | 1 (0.6%) |
| Surgical and medical procedures | 2 (1.2%) |
| Vascular disorders | 24 (14.5%) |
To conclude, the TIGER study was successfully met all pre-defined success criteria. Based on the results of this clinical study, the Tigertriever device, when used for the revascularization in ischemic stroke due to LVOs, has reperfusion rates and a safety profile similar to alternative devices. Therefore, the clinical data support the substantial equivalence.
## Statement of Substantial Equivalence
The Tigertriever device has the same intended use and indications for use, and similar technological characteristics compared to the Solitaire 2 predicate device. The differences in technological characteristics between the Tigertriever and the predicate device were evaluated in bench, animal and clinical testing as discussed above and do not raise different questions regarding the safety and effectiveness and demonstrate similar performance and safety characteristics between the Tigertriever and predicate device. Therefore, the results from these tests support the conclusion that the Tigertriever device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.