K203514 · Philips Healthcare (Suzhou) Co., Ltd. · JAK · Jun 17, 2021 · Radiology
Device Facts
Record ID
K203514
Device Name
Precise Position
Applicant
Philips Healthcare (Suzhou) Co., Ltd.
Product Code
JAK · Radiology
Decision Date
Jun 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Patient positioning workflow
Convolutional Neural Network
—
Up to 23% time reduction in patient positioning; up to 50% increase in vertical position accuracy; up to 70% increase in vertical and horizontal position consistency.
The Precise Position is intended for use with Philips Incisive CT systems. The following guided workflow. - Patient orientation identification - · Surview range recommendation - · Automatic centering the patient anatomy - · Provide visual images of patient on the table Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies. Patient population limitation: Patient younger than 16 years are not supported.
Device Story
Precise Position is an optional feature for Philips Incisive CT systems to assist patient positioning. A ceiling-mounted camera with color and depth sensing captures images of the patient on the table. Data is transmitted via high-speed fiber/copper hybrid USB cable to the system. An AI algorithm (Convolutional Neural Network) processes input to detect patient joints, orientation, and anatomy center, and recommends surview start/end positions. The system displays visual images of the patient to the operator. Used in hospitals and clinics by radiology staff, the device automates positioning, reducing setup time and improving vertical/horizontal centering accuracy and consistency. It benefits patients by streamlining the workflow and enhancing positioning precision before CT scans.
Clinical Evidence
Clinical evaluation involved 80 scan positions (40 with Precise Position, 40 without) assessed by 5 experts. Results showed up to 23% reduction in patient positioning time, up to 50% increase in vertical (iso)center accuracy, and up to 70% increase in vertical/horizontal positioning consistency. No radiation was delivered to volunteers. Study concluded the device performs as intended with a favorable benefit-risk ratio.
Technological Characteristics
Ceiling-mounted color/depth camera; high-speed fiber/copper hybrid USB connectivity; powered by CT gantry. Software utilizes a Convolutional Neural Network (CNN) for joint detection and anatomy localization. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 14971, and IEC 62304. Operates as an integrated feature of the Philips Incisive CT system.
Indications for Use
Indicated for CT imaging of head, chest, abdomen, pelvis, and combinations thereof. Patient population: adults and children 16 years and older. Contraindications: patients younger than 16 years; decubitus orientations.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Auto Positioning (GE Hangwei Medical System Co., Ltd.) (K192956)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Healthcare (Suzhou) Co., Ltd. % Shiguang An Advanced Regulatory Engineer No. 258, Zhong Yuan Road, Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA
Re: K203514
Trade/Device Name: Precise Position Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: May 19, 2021 Received: May 19, 2021
Dear Shiguang An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
June 17, 2021
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K203514
Device Name
Precise Position
Indications for Use (Describe)
The Precise Position is intended for use with Philips Incisive CT systems. The following guided workflow.
- Patient orientation identification
- · Surview range recommendation
- · Automatic centering the patient anatomy
- · Provide visual images of patient on the table
Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies.
Patient population limitation: Patient younger than 16 years are not supported.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span>☒</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div style="display:flex; align-items:center;"> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
collection of information is estimated to average 79 hours p time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
## 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | November 18, 2020 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, ZhongYuan Road, Suzhou Industrial Park, 215024<br>Suzhou, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA<br>Establishment Registration Number: 3009529630 |
| Primary Contact<br>Person: | Shiguang An<br>Advanced Regulatory Engineer<br>Phone: +86-13940106467<br>E-mail: shiguang.an@philips.com |
| Secondary Contact<br>Person: | Erhong Wang<br>Senior Regulatory Affairs Manager<br>Phone: +86-13021019589<br>E-mail: erhong.wang@philips.com |
| Device Name: | Precise position |
| Classification: | Classification Name<br>Computed tomography x-ray system<br>Classification Regulation:<br>21CFR §892.1750<br>Classification Panel:<br>Radiology<br>Device Class:<br>Class II<br>Primary product code:<br>JAK |
| Predicate Device: | Trade Name:<br>Philips Incisive CT<br>Manufacturer:<br>Philips Healthcare (Suzhou) Co., Ltd.<br>510(k) Clearance:<br>K180015-March 20, 2018<br>Classification Regulation:<br>21 CFR, Part 892.1750<br>Classification Name:<br>Computed tomography x-ray system<br>Classification Panel:<br>Radiology<br>Device Class:<br>Class II<br>Product Code<br>JAK |
| Reference Device: | Manufacturer:<br>Auto Positioning |
| | GE Hangwei Medical System<br>Co., Ltd. |
| | |
| 510(k) Clearance: | K192956 (January 16, 2020) |
| Classification Regulation: | 21 CFR, Part 892.1750 |
| Classification Name: | Computed tomography x-ray<br>system |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Product Code | JAK |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
| Device<br>description: | Precise Position is an optional feature to assist user for position the<br>patient before the body examination such as CT scan. The purpose<br>of this feature is to reduce the patient position time via the camera<br>detection and calculation result. It includes automatic detect patient<br>orientation, patient anatomy scan range and center of patient<br>anatomy. |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Precise Position including a camera with both color and depth<br>function is installed in the ceiling of the scan room, in such a way<br>to cover the entire patient on the patient table. The camera control<br>and image data transmit via the high speed fiber and copper hybrid<br>USB cable. The power supply of the camera is from the gantry.<br>Precise position adopts the AI algorithm (Convolution Neural<br>Network) to detect the joints of the patient body, and then identify<br>surview start/end position and patient orientation. The algorithm<br>can also support detect center of patient anatomy. |
| | Limitation for Precise Position<br>There is no limitation for Precise Position except below items:<br>• Patients below the age of 16 are not supported.<br>• Decubitus orientations are not supported. |
| | The Precise Position display results may get affected by the<br>following conditions:<br>• When the patient is covered by sheet, blanket etc.,<br>• When the patient is not completely covered by the ceiling camera<br>view, e.g. blocked by the gantry or out of camera's FOV etc.<br>• When the patient is wearing clothes that reflects light, e.g. plastic-<br>like clothes.<br>• When the patient is wearing black clothes.<br>• When the patient is wearing thick clothes.<br>• When there are other people around the patient. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the words "Traditional 510(k)". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Traditional" appearing first, followed by "510(k)".
Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
| Indications for Use: | The Precise Position is intended for use with Philips Incisive CT systems. The device provides the following guided workflow.<br>Patient orientation identification Surview range recommendation Automatic centering the patient anatomy Provide visual images of patient on the table Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies.<br>Patient population limitation: Patient younger than 16 years are not supported. |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the primary focus of the image and dominates the visual space.
| Fundamental<br>scientific<br>technology: | Based on the information provided above, the precise position is<br>considered substantially equivalent to the primary currently<br>marketed and predicate device Philips Incisive CT (K180015,<br>20/March/2018) in terms of fundamental scientific technology. | | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Attribute | Predicate Device<br>Philips Incisive CT<br>(K180015) | Proposed Device Precise<br>Position |
| | Patient<br>positioning<br>workflow | Manually position<br>the patient via the<br>couch motion button<br>and laser. Normally,<br>need several round<br>adjustments by user. | Precise position provides<br>the auto workflow to set<br>the surview start/end<br>position, center of patient<br>anatomy, and patient<br>orientation. |
| | Detection<br>algorithm | The manual<br>workflow does not<br>need special<br>algorithm. | Precise position adopt the<br>AI algorithm (Convolution<br>Neural Network) to detect<br>the joints of the patient<br>body automatically, and<br>then automatic detect<br>surview start/end position,<br>center of patient anatomy<br>and patient orientation. |
| | Hardware<br>need to<br>support the<br>patient<br>positioning | Manual patient<br>positioning does not<br>need any unique<br>hardware except<br>traditional exiting<br>hardware on the CT<br>or other system.<br>Exiting hardware<br>includes table<br>motion button<br>(in/out, up/down) on<br>gantry panel or<br>CTBOX, lasers and<br>etc. | Precision position feature<br>need color and depth<br>camera, high speed USB<br>transmission cable, and<br>power cable, and relative<br>the enhanced workflow.<br>The tradition exiting<br>patient positioning<br>hardware still be available<br>for user all the time, it is<br>convenient to switch<br>between the camera<br>automatically detecting<br>mode and manual mode. |
| | Patient<br>population | All ages | More than 16 years old. |
| | Environment<br>of use | Hospitals, outpatient<br>clinics, research<br>institutions, and<br>other clinical<br>facilities. | Hospitals, outpatient<br>clinics, research<br>institutions, and other<br>clinical facilities. |
| Summary of Non-<br>Clinical | The Precise Position complies with the following international and<br>FDA-recognized consensus standards: | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
| Performance data: | • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012<br>(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements. For Basic Safety and Essential Performance (IEC 60601-1:2012, MOD).<br>FDA/CDRH recognition number 19-4<br>• IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance -- Collateral standard: Electromagnetic disturbances - Requirements and tests.<br>FDA/CDRH recognition number 19-8<br>• ISO 14971 Medical devices – Application of risk management to medical devices. FDA/CDRH recognition number 5-40.<br>• IEC 62304:2015, Medical device software -- Software life cycle processes<br>FDA/CDRH recognition number 13-79<br>There are no risks identified in risk management documentation that require clinical data for the purpose of clinical evaluation; Risk Management Plan as Appendix 001, Risk Management Report as Appendix 002, and Risk Management Matrix as Appendix 003 of Precise Position.<br>There are no clinical risks identified by the evaluated clinical data.<br>Sufficient evidence is available to demonstrate the ability of Precise Position achieve the intended performances during normal condition of use.<br>Full consistency exists between the state-of-the-art, the evaluated data, the risk management documentation and the information materials supplied.<br>Therefore, the Precise Position is substantially equivalent to the primary currently marketed and predicate device (K180015, 20/March/2018) in terms of safety and effectiveness. |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Clinical Data: | Precise Position is evaluation covered total 80 clinical scan positions in which 40 cases used with Precise Position and another 40 cases without the usage of Precise Position in order to meet the sample size calculation of 40 cases performed average on a CT scanner per day.<br>The thorough clinical evaluation of this feature is done by 5 Clinical experts.<br>The testing did not deliver radiation to volunteers as Precise Position is only for supporting the positioning of the patient for the localization radiograph and therefore “radiation" was not required for those volunteers. |
| | |
| | The clinical evaluation was done with 3 major objectives as follows, |
| | To calculate the time saved per surview planning with and without |
| | Precise Position. |
| | To calculate the accuracy of vertical (iso)center positioning and |
| | Surview (horizontal) start /end position with and without Precise<br>Position |
| | Intra operator consistency in positioning the patient and surview scan<br>range |
| | The summary of clinical evaluation testing was clearly demonstrating<br>the objectives in terms of user benefits as shown below<br>Average Time taken to position without Precise Position and with<br>Precise Position is recorded. Time at user select the patient from<br>Gantry Panel, to time when user pressed Go button from Gantry Panel<br>is measured and concluded that up to 23% time reduction in patient<br>positioning achieved with “Precise Positioning workflow”. |
| | The average offset in mm for vertical (iso) center position among 5<br>operators without Precise Position and with Precise Position is<br>recorded and results shown that with "Precise Position" the vertical<br>position accuracy is increased up to 50%. |
| | Standard deviation in mm for vertical (iso)center positioning &<br>Surview (horizontal) start position among 5 operators without Precise<br>Position and with Precise Position is recorded and results shown up to<br>70% increase in Vertical and horizontal position consistency with<br>Precise Position. |
| | Overall, the Precise Position Clinical Review Report. (Appendix_007)<br>concluded that Philips Incisive CT systems with Precise Position,<br>under normal condition of use, perform as intended, are safe for its<br>intended use and have a favorable benefit-risk ratio. Further clinical<br>investigations are not necessary, as sufficient evidence exists to<br>support these conclusions. |
| Substantial | The Precise Position is substantially equivalent to the primary |
| Equivalence | currently marketed and predicate device (K180015, 20/March/2018) in |
| Conclusion: | terms of design features, fundamental scientific technology, |
| | indications for use, and safety and effectiveness. Additionally, |
| | substantial equivalence was demonstrated with non-clinical |
| | performance tests, which complied with the requirements specified in |
| | the international and FDA-recognized consensus standards, IEC 62304 |
| | and ISO 14971. The results of these tests demonstrate that Precise |
| | Position met the acceptance criteria and is adequate for this intended |
| | use. |
{8}------------------------------------------------
## Traditional 510(k)
Image /page/8/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.