ProLift Wedge Expandable Spacer System
K203361 · Life Spine, Inc. · MAX · Mar 17, 2021 · Orthopedic
Device Facts
| Record ID | K203361 |
| Device Name | ProLift Wedge Expandable Spacer System |
| Applicant | Life Spine, Inc. |
| Product Code | MAX · Orthopedic |
| Decision Date | Mar 17, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
When used as an interbody fusion device, the PROLIFT® Wedge Expandable Implant System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.
Device Story
PROLIFT® Wedge Expandable Spacer System; intervertebral body fusion device. Implanted during spinal fusion surgery by surgeons. Device expands to desired lordosis (0-15°) to match patient anatomy; features teeth on superior/inferior surfaces to prevent migration/rotation. Allows packing of autograft/allogeneic bone graft to promote fusion. Used with supplemental spinal fixation (e.g., pedicle screws). Benefits patient by restoring disc height and facilitating fusion in DDD cases.
Clinical Evidence
Bench testing only. Static/dynamic compression and shear testing performed per ASTM F2077; subsidence testing performed per ASTM F2267.
Technological Characteristics
Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Expandable spacer design with lordotic adjustment (0-15°). Sterile, single-use. Mechanical interface via surface teeth.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1) with or without Grade 1 spondylolisthesis, who have failed at least six months of non-operative treatment. Includes patients with prior non-fusion spinal surgery at the involved level(s).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- Life Spine ProLift Expandable System (K190488)
- Life Spine ProLift Expandable System (K173182)
- NuVasive TLX Interbody System (K153627/K171633)
Related Devices
- K223430 — ProLift Micro Expandable Spacer System · Life Spine, Inc. · May 4, 2023
- K173182 — ProLift® Expandable System · Life Spine, Inc. · Dec 20, 2017
- K243668 — ProLift Pivot Expandable Spacer System · Life Spine, Inc. · Jan 23, 2025
- K212520 — ProLift Micro Expandable Spacer System · Life Spine, Inc. · Oct 8, 2021
- K190488 — ProLift Expandable System · Life Spine, Inc. · Jul 23, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 17, 2021
Life Spine, Inc. Ms. Angela Batker RA/QA Manager 13951 South Quality Drive Huntley, Illinois 60142
Re: K203361
Trade/Device Name: PROLIFT® Wedge Expandable Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 15, 2021 Received: February 22, 2021
Dear Ms. Batker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K203361
Device Name PROLIFT® Wedge Expandable Cage System
#### Indications for Use (Describe)
When used as an interbody fusion device, the PROLIFT® Wedge Expandable Implant System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary PROLIFT® Wedge Expandable Spacer System
| Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
| Date Prepared: | March 10th, 2021 |
| Trade Name: | PROLIFT® Wedge Expandable Spacer System |
| Common Name: | Intervertebral Body Fusion Device |
| Classification: | MAX, 21 CFR 888.3080, Class II |
| Primary Predicate: | Life Spine ProLift Expandable System (K190488) |
| Additional Predicate: | Life Spine ProLift Expandable System (K173182)<br>NuVasive TLX Interbody System (K153627/<br>K171633) |
#### Device Description:
The PROLIFT® Wedge Expandable Spacer System is available in a range of sizes and footprints and can expand to the desired lordosis (0-15°) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.
All implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the PROLIFT® Wedge Expandable Spacer System components with components from any other system or manufacturer. The PROLIFT® Wedge Expandable Spacer System components should never be reused under any circumstances.
{4}------------------------------------------------
### Indications for Use:
When used as an interbody fusion device, the PROLIFT® Wedge Expandable Implant System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.
### Technological Characteristics:
The PROLIFT® Wedge Expandable Spacer System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.
### Material:
This submission seeks clearance of a device made from titanium allov (Ti-6Al-4V ELI) as described by ASTM F136. This this is the same material used in the predicate devices.
### Performance Data:
Static compression, dynamic compression, static shear and dynamic shear testing according to ASTM F2077 and subsidence testing according to ASTM F2267, was presented to demonstrate the substantial equivalency of the PROLIFT® Expandable System (K190488).
#### Substantial Equivalence:
The PROLIFT® Wedge Expandable Spacer System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.
#### Conclusion:
The information presented demonstrates the substantial equivalency of The PROLIFT® Wedge Expandable Spacer System.