K243668 · Life Spine, Inc. · MAX · Jan 23, 2025 · Orthopedic
Device Facts
Record ID
K243668
Device Name
ProLift Pivot Expandable Spacer System
Applicant
Life Spine, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jan 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProLift Pivot Expandable System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the level involved. It is to be used in patients who have had at least six months of nonoperative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.
Device Story
ProLift Pivot Expandable Spacer System is an intervertebral body fusion device used in spinal fusion procedures. The device is implanted by a surgeon in an operating room setting to treat degenerative disc disease. It features an expandable design (8mm to 15mm height) to accommodate patient anatomy and pathology. The spacer is packed with autograft or allogeneic bone graft to promote fusion. Superior and inferior surfaces feature teeth to interface with vertebral bodies, preventing rotation and migration. The device is used in conjunction with a supplemental internal spinal fixation system (e.g., pedicle screws or plating). It provides structural support to the disc space, facilitating bone fusion and potentially alleviating back pain associated with discogenic origin.
Clinical Evidence
Bench testing only. Mechanical performance was demonstrated via static compression shear, static axial compression, dynamic compression shear, and dynamic axial compression testing in accordance with ASTM F2077.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI, ASTM F136) and Cobalt Chrome (CO-26Cr-6MO, ASTM F1537). Expandable spacer design with heights from 8mm to 15mm. Features teeth on superior/inferior surfaces for fixation. Provided sterile, single-use.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including patients with up to Grade 1 spondylolisthesis, who have failed at least six months of non-operative treatment. Includes patients with previous non-fusion spinal surgery at involved levels.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Life Spine ProLift Expandable Spacer System (K190488)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
January 23, 2025
Life Spine, Inc. Angela Batker RA/QA Manager 13951 Quality Drive Huntley, Illinois 60142
Re: K243668
Trade/Device Name: ProLift Pivot Expandable Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: November 21, 2024 Received: November 27, 2024
Dear Ms. Batker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K243668
Device Name
ProLift Pivot Expandable Spacer System
Indications for Use (Describe)
The ProLift Pivot Expandable System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiquous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K243668 - 510(k) Summary ProLift Pivot Expandable Spacer System
| Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
| Date Prepared: | November 26th, 2024 |
| Trade Name: | ProLift Pivot Expandable Spacer System |
| Common Name: | Intervertebral Body Fusion Device |
| Classification: | MAX, CFR 888.3080, Class II |
| Primary Predicate: | Life Spine ProLift Expandable Spacer System (K190488) |
| Additional Predicate: | Spinal Elements Lucent Xp Curved (K202298)<br>Globus Altera Spacer (K140411)<br>Life Spine ProLift Micro (K212520, K223430)<br>Life Spine Steerable Plateau Ti (K193521) |
#### Device Description:
The ProLift® Pivot Expandable System is available in a range of sizes and footprints and can expand to the desired height (8mm to 15mm) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6AI-4V ELI) and Cobalt Chrome (CO- 26Cr-6MO) as described by ASTM F136and ASTM F1537. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.
The ProLift® Pivot Expandable Spacer System is available in a range of sizes and footprints and can expand to the desired height (8mm to 15mm) to suit the individual pathology and anatomical conditions of the patient.
- The ProLift implants are fabricated and manufactured from titanium alloy (Ti-6Al-. 4V ELI) as described by ASTM F136.
{5}------------------------------------------------
. The ProLift Pivot-U implants are fabricated and manufactured from titanium alloy (Ti-6Al-4V ELI) and Cobalt Chrome (CO-26Cr-6MO) as described by ASTM F136 & ASTM F1537.
The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.
All ProLift Pivot implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the ProLift® Pivot Expandable Spacer System components with components from any other system or manufacturer.
## Intended Use of the Device:
The ProLift Pivot Expandable System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the level involved. It is to be used in patients who have had at least six months of nonoperative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.
## Technological Characteristics:
The ProLift Pivot Expandable System is substantially equivalent to the predicate systems in terms of design, materials, and indications for use and sizing.
## Material:
This submission seeks clearance of a device made from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136) and Cobalt Chrome (CO- 26Cr-6MO) as described by ASTM F1537. This is the same material used in the predicate devices.
## Performance Data:
Static compression shear, Static Axial Compression, Dynamic Compression Shear and Dynamic Axial Compression according to ASTM F2077 was presented to demonstrate the substantial equivalency of the Life Spine ProLift Expandable Spacer System (K190488).
## Substantial Equivalence:
The ProLift Pivot Expandable System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.
## Conclusion:
The information presented demonstrates the substantial equivalency of The ProLift Pivot Expandable System.