K202701 · Covidien · GDW · Dec 7, 2020 · General, Plastic Surgery
Device Facts
Record ID
K202701
Device Name
GIA Stapler with Tri-Staple Technology
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Dec 7, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GIA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis.
Device Story
Manual, single-use surgical stapler; used in abdominal and thoracic procedures for tissue resection, transection, and anastomosis. Device places two triple-staggered rows of titanium staples while simultaneously cutting tissue between rows. Operated by surgeon via single-handle squeeze mechanism; firing knob slides forward to actuate staple formation and knife blade. Available in 60mm staple line length with two cartridge types (medium/thick or extra thick) accommodating varying tissue thicknesses. Provides audible feedback during operation. Device is sterile (ethylene oxide), disposable, and supports up to 8 firings per instrument via reloading. Facilitates tissue closure and hemostasis, reducing surgical time and potential complications associated with manual suturing.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including visual inspection, IFU walkthrough, staple formation, firing force, multi-fire evaluation, burst pressure, pneumostasis, hemostasis, and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Manual, single-use linear stapler. Materials: Titanium (ASTM F67 Grade I) staples, stainless steel knife and anvil. Staple configuration: 2 triple-staggered rows. Staple heights: 3.0-4.0mm (medium/thick) or 4.0-5.0mm (extra thick). Sterilization: Ethylene oxide. Shelf life: 5 years. No software or electronic components.
Indications for Use
Indicated for patients undergoing abdominal or thoracic surgical procedures requiring resection, transection, or creation of anastomosis.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
GIA™ Stapler with Tri-Staple™ Technology (K192720)
Reference Devices
GIA™ Stapler with DST Series™ Technology (K111825)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
December 7, 2020
Covidien Katherine Choi Principal Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K202701
Trade/Device Name: GIA Stapler with Tri-Staple Technology Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: November 6, 2020 Received: November 9, 2020
Dear Katherine Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202701
Device Name GIATM Stapler with Tri-Staple™ Technology
Indications for Use (Describe)
The GLA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis.
| Type of Use (Select one or both, as applicable) | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <table><tr><td><label><input checked="" type="checkbox"/> Residential Use (18 CCR 91.1 Subpart B)</label></td></tr><tr><td><label><input type="checkbox"/> On-Site Construction (18 CCR 91.1 Subpart C)</label></td></tr></table> | <label><input checked="" type="checkbox"/> Residential Use (18 CCR 91.1 Subpart B)</label> | <label><input type="checkbox"/> On-Site Construction (18 CCR 91.1 Subpart C)</label> |
| <label><input checked="" type="checkbox"/> Residential Use (18 CCR 91.1 Subpart B)</label> | | |
| <label><input type="checkbox"/> On-Site Construction (18 CCR 91.1 Subpart C)</label> | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
### Date Prepared:
Sep 10, 2020
### Submitter:
Leo Chen Covidien Rooms 501, 502, 601, 602, No.3 building No.2388 Chen Hang Road Min Hang District, Shanghai, 201114, China
#### Contact:
Katherine Y. Choi (U.S. Agent) on behalf of Leo Chen Covidien 60 Middletown Avenue North Haven, CT 06473, USA Principal Regulatory Affairs Specialist Telephone: (917) 841-6315 Fax: (203) 492-5029 katherine.y.choi@medtronic.com Email:
#### Name of Device:
Proprietary/Trade Name: Model Numbers: Classification Name: Regulations Number: Product Codes: FDA Panel Number: Device Class: Review Panel: Common Name:
#### Predicate Device:
Proprietary/Trade Name: 510(k) Number: Classification Name: Regulations Number: Product Codes: FDA Panel Number: Device Class: Review Panel: Common Name:
GIA™ Stapler with Tri-Staple™ Technology K192720 Staple, Implantable 21 CFR 878.4750 GDW 79 Class II General and Plastic Surgery Surgical Stapler
GIA™ Stapler with Tri-Staple™ Technology
Staple, Implantable
General and Plastic Surgery
21 CFR 878.4750
Surgical Stapler
GDW
Class II
79
GIA60MTS, GIA60XTS, GIA60MTC, GIA60XTC
#### Reference Device:
Proprietary/Trade Name: 510(k) Number: Classification Name: Regulations Number:
GIA™ Stapler with DST Series™ Technology K111825 Staple, Implantable 21 CFR 878.4750
{5}------------------------------------------------
| Product Codes: | GDW |
|-------------------|-----------------------------|
| FDA Panel Number: | 79 |
| Device Class: | Class II |
| Review Panel: | General and Plastic Surgery |
| Common Name: | Surgical Stapler |
#### Device Description:
The GIA™ stapler with Tri-Staple™ technology places two triple staggered rows of titanium staples and simultaneously cuts and divides tissue between these two triple rows. The subject GIA™ stapler and cartridges with Tri-Staple™ technology are available in 60 mm staple line length and two staple sizes to accommodate tissue thicknesses: medium/thick and extra thick. Staplers for medium/thick tissue (purple cartridge) deploy three height-progressive rows of 3.0 mm. 3.5 mm and 4.0 mm titanium staples on either side of the tissue cut line. Staplers for extra thick tissue (black cartridge) deploy three height-progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples on either side of the cut line. Each GIA™ stapler with Tri-Staple™ technology may be reloaded with a GIA™ Cartridge with Tri-Staple™ Technology up to 7 times for a total of 8 firings per instrument.
The GIA™ Stapler with Tri-Staple™ Technology is manufactured with essentially the same patient contact materials that are utilized within the predicate device (K192720).
The GIA™ stapler with Tri-Staple™ technology has the same principle operation as the predicate device. The GIA™ Stapler with Tri-Staple™ Technology is a manual single-use device. It is provided sterile (ethylene oxide) with a 5-year shelf life, and intended for multiple use during a single procedure, which is the same as the predicate device (K192720).
#### Indications for Use:
The GIA™ stapler with Tri-Staple™ technology has applications in abdominal and thoracic surgical procedures for resection, transection and creation of anastomosis.
#### Technological and Performance Characteristics:
The subject device GIA™ stapler with Tri-Staple™ technology (staple line length 60mm) is substantially equivalent to the predicate device K192720 (staple line length 80mm) regarding the fundamental stapling technologies employed, intended use and indications for use. Both are single-use manual linear staplers.
The Tri-staple™ technology used in the subject device is exactly the same as the predicate device K192720. The subject disposable manual linear stapler is available in 60mm staple line length, while, the predicate device is available in 80mm staple line length. The 60mm staple line length, however, is not new; the reference device K111825 offers the same length.
#### Substantial Equivalence:
The subject new product models have the same intended use and indications for use as the predicate device.
They also have the same fundamental scientific technology in that they are all sterile, single used, hand-held, manual surgical instruments equipped with titanium staples intended to have applications in abdominal and thoracic surgical procedures for resection and creation of anastomosis. The subject and predicate device are the same in design and are sterilized via ethylene oxide, but different in staple line length.
{6}------------------------------------------------
The below table further summarizes the similarities and differences between the subject and predicate devices.
| Features | Subject Device | Predicate Device (K192720) |
|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | GIA™ Stapler with Tri-Staple™ Technology | |
| Manufacturer | Same as predicate device. | Covidien |
| Constructional<br>(example) | Image: Subject Device | Image: Covidien |
| Indications for Use | Same as predicate device. | The GIA™ stapler with Tri-Staple™ technology<br>has applications in abdominal and thoracic<br>surgical procedures for resection, transection<br>and creation of anastomosis. |
| Operation Method | Same as predicate device. | Manual |
| Anatomical Site | Same as predicate device. | Alimentary tract and Thoracic |
| Surgical Approach | Same as predicate device. | Open surgery |
| Method of<br>Operation | Same as predicate device. | The instruments are activated by sliding the<br>firing knob forward to a complete stop and<br>Immediately after staple formation, the knife<br>blade resects the excess tissue, creating a<br>linear anastomosis. |
| Product Codes | Stapler with Cartridge:<br>GIA60MTS, GIA60XTS<br>Cartridge:<br>GIA60MTC, GIA60XTC | Stapler with Cartridge:<br>GIA80MTS, GIA80XTS<br>Cartridge:<br>GIA80MTC, GIA80XTC |
| Staple Rows | Same as predicate device. | 3 staggered rows of staples on either side of<br>the tissue cut line with different staple height in<br>each staple row |
| Instrument Handle<br>Type | Same as predicate device. | Single-handle squeeze |
| Staple Cartridge<br>Configuration | Same as predicate device. | 2 triple rows staples, step-faced cartridge with<br>different staple size in each staple row |
| Cartridge Color | Same as predicate device. | Purple, Black |
| Staple Size<br>(open leg height) | Same as predicate device. | Purple cartridge: 3.0mm, 3.5mm, 4.0mm<br>Black cartridge: 4.0mm, 4.5mm, 5.0mm |
| Anvil | Same as predicate device. | 2 triple staggered rows of anvil pocket design |
| Staple Line Length | 60mm<br>Same as reference device (60mm), which<br>is selected to be the control device in the<br>performance testing. | 80mm |
| Staple Material | Same as predicate device. | Titanium per ASTM F67 Grade l |
| Features | Subject Device | Predicate Device (K192720) |
| | GIA™ Stapler with Tri-Staple™ Technology | |
| Identification of<br>Materials of Implant<br>(staple) and tissue<br>cutting component<br>(knife) | Same as predicate device. | Staple: Titanium per ASTM F67 Grade I<br>Knife: Stainless Steel<br>Anvil: Stainless Steel |
| Biocompatibility | Same as predicate device. | Evaluated per ISO 10993-1 series and FDA<br>biocompatibility guidance |
| Audible Feedback | Same as predicate device. | Yes |
| Knife | Same as predicate device. | Yes. |
| Single Use | Same as predicate device. | Yes |
| Disposable | Same as predicate device. | Yes |
| Sterile | Same as predicate device. | Ethylene oxide |
| Shelf Life | Same as predicate device. | 5 years |
{7}------------------------------------------------
Tests performed to evaluate and compare technological and performance characteristics: Non-clinical performance data - the following testing has been performed to demonstrate substantial equivalence to the predicate device.
- 1. Performance Test (Bench)
- Visual inspection ●
- . IFU walkthrough
- . Staple formation on test media
- . Firing force test
- Multi-fire evaluation
- 2. Performance Test (Ex-Vivo)
- Burst pressure .
- . Pneumostasis
- . Across staple line evaluation
- 3. Performance Test (In-Vivo)
- Hemostasis test .
- . Staple formation on tissues
- 4. Biocompatibility Tests per ISO 10993-1 and FDA guidance "Use of international Standard ISO 10993-1" issued on September 4, 2020
- Cytotoxicity test .
- Sensitization .
- . Intracutaneous irritation
- Acute system toxicity
- . Pyrogenicity
Clinical performance data - No clinical study is deemed necessary since the substantial equivalence has been sufficiently demonstrated by non-clinical studies.
{8}------------------------------------------------
## Conclusion:
Based upon the supporting data summarized above, we concluded that the subject device GIA™ Stapler with Tri-Staple™ Technology is substantially equivalent to the legally-marketed device K192720 and does not raise different questions or additional risks of safety and effectiveness than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.