InRoad® Dental Synthetic Bone Graft is intended for the filling and reconstruction of multi-walled bone defects such as: - Defects after removal of bone cysts - Augmentation of the atrophied alveolar ridge - Sinus lift and sinus floor elevation (subantral augmenation) - Filling of alveolar defects following tooth extraction for alveolar ridge preservation - Filling of extraction defects to create an implant bed - Filling of two- or multi- walled infra-bony pockets. and bi- and trifurcation defects - Support function for a membrane in controlled tissue regeneration (CTR) - Defects after surgical removal of retained teeth or corrective osteotomies - Other multi-walled bone defects of the alveolar processes
Device Story
InRoad® Dental Synthetic Bone Graft is a porous, osteoconductive bioceramic bone graft material. It consists of granules (90% hydroxyapatite, 10% β-tricalcium phosphate) designed to mimic trabecular bone structure with interconnected pores. Used by dental professionals in clinical settings to fill and reconstruct multi-walled bone defects in the alveolar process. The material acts as a scaffold, allowing bone cell migration and new bone growth into the matrix. It is provided sterile in granule form and is partially resorbable. The device is applied directly to the graft site to support bone regeneration and membrane function in controlled tissue regeneration procedures.
Clinical Evidence
No clinical testing performed. Evidence consists of bench testing (biocompatibility per ISO 10993-1, sterilization validation per ISO 11137-2, packaging integrity per ASTM F1140/D3078) and an animal study (one-wall periodontal defect model in dogs). Animal study compared InRoad® to the predicate at 4, 8, and 12 weeks using micro-CT, histopathology, and histomorphometry, demonstrating equivalent performance in new bone growth and residual graft material.
Technological Characteristics
Synthetic bioceramic granules (90% HA, 10% β-TCP). Porosity ~60-80% with interconnecting pore structure. Granule sizes: 0.3-1.0 mm and 0.8-1.8 mm. Sterile (gamma irradiation, SAL 10^-6). Non-pyrogenic. Single-patient use. No electrical or software components.
Indications for Use
Indicated for filling and reconstruction of multi-walled bone defects in the alveolar process, including post-cyst removal, alveolar ridge augmentation, sinus lift, post-extraction ridge preservation, implant bed preparation, infra-bony pockets, bi/trifurcation defects, and corrective osteotomies. For use in human oral/periodontal procedures.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Curasan, Osbone® Dental Synthetic Bone Graft (K102872)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name "U.S. Food & Drug Administration" in a smaller font.
March 04, 2022
Osteogene Tech Corp % James O'connor Principal and Manager JMC Medical, LLC 65 Elm Street Chelmsford, Massachusetts 01824
Re: K202675
Trade/Device Name: InRoad® Dental Synthetic Bone Graft Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: LYC Dated: February 8, 2022 Received: February 9, 2022
Dear James O'connor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202675
Device Name InRoad® Dental Synthetic Bone Graft
#### Indications for Use (Describe)
InRoad® Dental Synthetic Bone Graft is intended for the filling and reconstruction of multi-walled bone defects such as:
- Defects after removal of bone cysts
- Augmentation of the atrophied alveolar ridge
- Sinus lift and sinus floor elevation (subantral augmenation)
- Filling of alveolar defects following tooth extraction for alveolar ridge preservation
- Filling of extraction defects to create an implant bed
- Filling of two- or multi- walled infra-bony pockets. and bi- and trifurcation defects
- Support function for a membrane in controlled tissue regeneration (CTR)
- Defects after surgical removal of retained teeth or corrective osteotomies
- Other multi-walled bone defects of the alveolar processes
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
|----------------------------------------------|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY OSTEOGENE™ TECH CORP InRoad® Dental Synthetic Bone Graft 510(k) K202675
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Substantial Equivalence for the use of OSTEOGENE™ TECH InRoad® Dental Synthetic Bone Graft.
#### I. SUBMITTER
| Submitted By: | OSTEOGENE™ TECH CORP |
|-----------------|-----------------------------|
| | 75 Oak Street |
| | Norwood, NJ 07648 |
| | Phone: 201-367-9321 |
| Contact Person: | Dr. Daniel S. Oh, CSO |
| | e-mail: dr.oh@osteogene.com |
| Date Prepared: | March 2, 2022 |
II. DEVICE
| Proprietary Name: | InRoad® Dental Synthetic Bone Graft |
|----------------------|-------------------------------------|
| Common Name: | Bone grafting material, synthetic |
| Classification Name: | Bone grafting material |
| Regulation: | 21 CFR 872.3930 |
| Product Code: | LYC |
#### III. PREDICATE DEVICE
Curasan, Osbone® Dental Synthetic Bone Graft, K102872 Predicate Device: This predicate device has not been subject to a design-related recall. Reference Device: GENOSS, OSTEON III Synthetic Bone Graft, K153676 This reference device has not been subject to a design-related recall.
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
InRoad® Dental Synthetic Bone Graft (SBG) is a synthetic, semi-dense porous bioceramic for bone regeneration. It is comprised of 90 ± 5% hydroxyapatite (HA) and 10 ± 5% ß-Tricalcium Phosphate (B-TCP). The granules are provided sterile (gamma irradiated) in a capped vial packaged in a sealed tray with lid.
InRoad® Dental SBG granules are designed to resemble trabecular bone structure. The granule structure consists of interconnected primary open pores, a secondary channel-like structure, and tertiary small holes on the surface resulting in a biocompatible, osteoconductive biomaterial. The multi-porous structure of InRoad® Dental SBG makes it possible for the bone cells to migrate into the matrices and for new bone to grow.
InRoad® Dental Synthetic Bone Graft is available in granule form in sizes of 0.3 - 1.0 mm (Small) and 0.8 - 1.8 mm (Large). Reference numbers, descriptions, and volume of granules in the vials are listed below.
| REFERENCE # | Description | Volume |
|-------------|-----------------------------------------------------------------|--------|
| HA90D05S | InRoad® Small (0.3~1.0 mm) Dental Synthetic Bone Graft granules | 0.5 cc |
| HA90D10S | InRoad® Small (0.3~1.0 mm) Dental Synthetic Bone Graft granules | 1.0 cc |
| HA90D30S | InRoad® Small (0.3~1.0 mm) Dental Synthetic Bone Graft granules | 3.0 cc |
| HA90D05L | InRoad® Large (0.8~1.8 mm) Dental Synthetic Bone Graft granules | 0.5 cc |
| HA90D10L | InRoad® Large (0.8~1.8 mm) Dental Synthetic Bone Graft granules | 1.0 cc |
| HA90D30L | InRoad® Large (0.8~1.8 mm) Dental Synthetic Bone Graft granules | 3.0 cc |
#### V. INDICATIONS FOR USE
InRoad® Dental Synthetic Bone Graft is intended for the filling and reconstruction of multiwalled bone defects such as:
- -Defects after removal of bone cysts
- -Augmentation of the atrophied alveolar ridge
- -Sinus lift and sinus floor elevation (subantral augmentation)
- Filling of alveolar defects following tooth extraction for alveolar ridge preservation -
- Filling of extraction defects to create an implant bed -
- Filling of two- or multi-walled infra-bony pockets and bi- and trifurcation defects -
- Support function for a membrane in controlled tissue regeneration (CTR) -
- -Defects after surgical removal of retained teeth or corrective osteotomies
- Other multi-walled bone defects of the alveolar processes -
{6}------------------------------------------------
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Principle of Use: The InRoad® Dental SBG, the subject device, and Osbone® Dental SBG, the predicate device, are similar in technology and principle of use. For both bone grafts, granule size is selected based on the size of the bony defect to be filled. The bone graft is then applied to the prepared graft bed. The porous structure of the bone graft makes it possible for the bone cells to grow into the matrices. This process is referred to as osteoconductive bone formation.
InRoad® Dental SBG and Curasan Osbone® Dental (K102872) are similar in intended use, area of application (human: oral, periodontal), source (synthetic), form (small and large granules), physical morphology (interconnecting pore structures of varying size), resorption (partially resorbable), packaging (capped vial in a sealed tray and lid in an outer carton), and sterility (gamma irradiation, non-pyrogenic, and intended for one-time single patient use).
Physical form: InRoad® Dental SBG varies slightly from the predicate device Curasan Osbone® Dental in that InRoad® comes in granule sizes between 0.3 mm to 1.8 mm where Osbone has a slightly wider granule size range of 0.25 to 2.0 mm.
The porosity of InRoad® Dental SBG (~ 60 - 80%) is similar to that of Osbone® Dental (~ 80 ±5%). The porosity of the Osteon III, the reference device, is >70%. All three devices have interconnecting pore structure morphology and are synthetic bone grafting material.
Composition: The nominal composition of InRoad® Dental SBG is 90 ± 5% HA and 10 ± 5% B-TCP. InRoad® composition is closest to the predicate Osbone® Dental (100% HA), the difference being that InRoad® Dental SBG has a small amount of β-TCP (<15%). GENOSS Osteon III, the reference device, by comparison has a larger amount of β-TCP than the subject device as it is composed of 40% B-TCP and 60% HA. The subject device, the predicate device, and the reference device have similar HA crystalline composition.
| | Subject Device | Predicate Device | Reference<br>Device |
|------------------------|-----------------------------------|--------------------|---------------------|
| Device Name | InRoad® Dental SBG | Osbone® Dental SBG | OSTEON III SBG |
| Manufacturer | OSTEOGENE™ TECH | Curasan | GENOSS |
| 510(k) Number | K202675 | K102872 | K153676 |
| Product Code | LYC | | |
| Classification<br>Name | Bone grafting material, synthetic | | |
| | | | Table. Comparison of InRoad® with Predicate Device and Reference Device | | | |
|--|--|--|-------------------------------------------------------------------------|--|--|--|
|--|--|--|-------------------------------------------------------------------------|--|--|--|
{7}------------------------------------------------
| | Subject Device | Predicate Device | Reference<br>Device |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Filling and<br>reconstruction of multi-<br>walled bone defects such<br>as:<br>-Defects after removal of<br>bone cysts<br>-Augmentation of the<br>atrophied alveolar ridge<br>-Sinus lift and sinus floor<br>elevation (subantral<br>augmentation)<br>-Filling of alveolar<br>defects following tooth<br>extraction for alveolar<br>ridge preservation<br>-Filling of extraction<br>defects to create an<br>implant bed<br>-Filling of two- or multi-<br>walled infra-bony<br>pockets. and bi- and<br>trifurcation defects<br>-Support function for a<br>membrane in controlled<br>tissue regeneration<br>(CTR)<br>-Defects after surgical<br>removal of retained teeth<br>or corrective osteotomies<br>-Other multi-walled bone<br>defects of the alveolar<br>processes | Filling and reconstruction<br>of multi-walled bone<br>defects, e.g.:<br>-Defects after removal of<br>bone cysts<br>-Augmentation of the<br>atrophied alveolar ridge<br>-Sinus lift and sinus floor<br>elevation (subantral<br>augmentation)<br>-Filling of alveolar<br>defects following tooth<br>extraction for alveolar<br>ridge preservation<br>-Filling of extraction<br>defects to create an<br>implant bed<br>-Filling of two- or multi-<br>walled infra-bony<br>pockets. and bi- and<br>trifurcation defects<br>-Support function for a<br>membrane in controlled<br>tissue regeneration<br>(CTR)<br>-Defects after surgical<br>removal of retained teeth<br>or corrective osteotomies<br>-Other multi-walled bone<br>defects of the alveolar<br>processes | -Periodontal/<br>Infrabony defects<br>- Ridge<br>augmentation<br>-Extraction sites<br>(implant preparation<br>/<br>placement)<br>- Sinus lifts |
| Application | Human: Oral, Periodontal | | |
| Performance<br>function | Osteoconductive bone formation | | |
| Source | | Synthetic | |
| Composition | $90 \pm 5\%$ HA | 100% HA | 60% HA |
| | Subject Device | Predicate Device | Reference<br>Device |
| | $10 \pm 5\% \beta$ -TCP | | 40% $\beta$ -TCP |
| Form | Granules, Small and<br>Large | Granules, Small and<br>Large | Granules, Small,<br>Medium, and Large |
| Particle size | 0.3~1.0 mm<br>0.8~1.8 mm | 0.25-1.0 mm<br>1.0-2.0 mm | 0.2~0.5 mm<br>0.5~1.0 mm<br>1.0~2.0 mm |
| Porosity<br>(Nominal) | ~60 - 80% | ~ 80 ± 5%1 | ~80%2 |
| Physical<br>morphology | Interconnecting pore structures | | |
| Crystallinity<br>(%) | Small granules: 71 – 91<br>Large granules: 98 – 993 | 71 – 973<br>$\ge 95^4$ | >702 |
| Resorption | Partially resorbable | Partially resorbable | Partially resorbable5 |
| Packaging | Vial in sealed tray with<br>outer carton | Vial in sealed tray with<br>outer carton | Vial in sealed tray,<br>syringe in sealed<br>tray with outer<br>cartons |
| Sterility | Sterile (Gamma Irradiation); Non-pyrogenic; One-time single patient use | | |
{8}------------------------------------------------
1Nominal from K102872, 2Nominal from K153676, 3From testing, 4From IFU, 5Based on composition
# VII.PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility Testing: InRoad® Dental SBG falls under the category of implant device in contact with tissue and/or bone with permanent contact duration (>30 days) (FDA 2016). Based on this categorization, InRoad® Dental Synthetic Bone Graft has been evaluated on a risk basis for biocompatibility endpoints according to ISO 10993-1 and FDA's modified biocompatibility matrix Table A.1 as listed in Attachment A of the FDA guidance, Use of International Standard ISO 10993- 1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
The battery of biocompatibility testing for InRoad® Dental SBG included the following.
- Cytotoxicity ●
- . Sensitization
{9}------------------------------------------------
- Irritation .
- Acute systemic toxicity .
- Material mediated pyrogenicity
- Subacute / subchronic toxicity ●
- Genotoxicity mouse lymphoma assay ●
- Genotoxicity bacterial reverse mutation study ●
- Implantation ●
- Chronic toxicity
- Carcinogenicity .
InRoad® Dental SBG has been found to be biocompatible for all biological effects evaluated as an implant that stays in the body for more than 30 days.
Animal Testing: Animal testing was conducted to evaluate the substantial equivalence of small and large InRoad® Dental Synthetic Bone Graft in a one-wall periodontal defect model in dogs in conjunction with a US marketed membrane. Results were compared to defects treated with the US marketed predicate device with the same US marketed membrane, and untreated [sham] defects at intervals of 4, 8, and 12 weeks.
The results were analyzed through animal health observation, radiographic assessment, micro-CT evaluation with statistical analysis, histopathological evaluation and histomorphometry.
These studies procedures, implantations, observations, and assessments allowed for the successful evaluation of substantial equivalence for two test article bone grafting materials while conforming to ISO 10993-6 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation.
The results indicate that the two test article bone grafts, small granule InRoad® Dental SBC and large granule InRoad® Dental SBG were biocompatible, and demonstrated equivalent performance for endpoints assessed including morphometric analysis of new bone growth and residual synthetic bone graft in a one-wall periodontal defect model in dogs when compared to sites treated with the US marketed predicate device, Osbone® Dental synthetic bone graft material.
Sterilization and Shelf-life Testing: The VDmax23 gamma sterilization validation method, ISO 11137-2 was used to show InRoad® Dental SBG as packaged results in a sterility assurance level (SAL) 1x106 for all devices. Packaging integrity testing per ASTM F1140 and ASTM D3078 showed that shipping, sterilization, and aging would not compromise the ability of the packaging to maintain sufficient strength and seal characteristics when products are sealed in trays to specification and packaged in shelf cartons and shipping cartons.
Product performance verification was conducted again after sterilization, transportation simulation, and accelerated aging to equivalent to 1-year real time aging confirm that InRoad®
{10}------------------------------------------------
Dental SBG characterization was not significantly changed by these processes. The evaluation showed that the material composition and granule size were not affected by sterilization, transportation simulation, and aging.
The successful transit testing and package integrity testing with accelerated aging equivalent to 1-year real time aging showed that the packaging can last for at least 1 year and can therefore be labeled with a 1-year shelf-life.
InRoad® Dental SBG does not incorporate electrical components or software.
No clinical testing was performed with InRoad® Dental SBG.
## VIII. CONCLUSIONS
InRoad® Dental SBG, the subject device, and Osbone® Dental SBG, the predicate, have been shown to be equivalent in technology, function, and use. They are equivalent in form, structure, and composition. InRoad® Dental SBG and Osbone® Dental SBG performed similarly in an animal study at 12 weeks.
InRoad® Dental SBG has been shown to be biocompatible and through an animal study has been shown to have equivalent performance as the predicate device at 12 weeks. Its packaging has been shown to have sufficient integrity after sterilization, transportation simulation, and accelerated aging to be labeled with a maximum 1-year shelf-life.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.