OSTEON III is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects. - Periodontal/infrabony defects - Ridge augmentation - Extraction Sites(implant preparation/ placement) - Sinus lifts - Cystic cavities
Device Story
OSTEON III is a synthetic, osteoconductive bone graft substitute composed of hydroxyapatite (HA) and beta-tricalcium phosphate (β-TCP). It features an interconnected porous structure mimicking human cancellous bone. Available as irregular particles (0.2–2.0mm), it is used by clinicians in oral and periodontal surgical procedures to fill, augment, or reconstruct bone defects. The device acts as a scaffold to support bone healing and formation. It is supplied in glass vials or syringes and sterilized via gamma radiation. By providing a biocompatible, osteoconductive matrix, the device facilitates natural bone regeneration in the surgical site, benefiting patients requiring dental or maxillofacial reconstruction.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, pyrogenicity, intracutaneous reactivity, systemic toxicity, subchronic toxicity, genotoxicity, implantation, and endotoxin). Physical/chemical characterization included porosity, particle size distribution, Ca/P ratio, crystalline phase composition, compressive strength, and resorption testing.
Technological Characteristics
Synthetic biphasic calcium phosphate (60% HA, 40% β-TCP). Porosity ~80%; pore size 200–400 μm. Particle size 0.2–2.0mm. Ca/P ratio 1.59. Crystallinity >70%. Sterilized by gamma irradiation. Shelf life 1.5 years. Osteoconductive scaffold.
Indications for Use
Indicated for patients requiring bone grafting to fill, augment, or reconstruct periodontal or oral/maxillofacial defects, including periodontal/infrabony defects, ridge augmentation, extraction sites, sinus lifts, and cystic cavities.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 14, 2016
Genoss Co., Ltd. Eunsang Lee Manager 1F Gyeonggi R&DB Center, 105 Gwanggyo-ro, Yeongtong-gu Suwon-si, Gyeonggi-do, 433-270 KR
Re: K153676
Trade/Device Name: OSTEON III Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: LYC Dated: August 8. 2016 Received: August 15, 2016
Dear Eunsang Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Michael J.DRyan -S
for Tina Kiang, PhD Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153676
Device Name
OSTEON III
Indications for Use (Describe)
OSTEON III is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects.
- Periodontal/infrabony defects
- Ridge augmentation
- Extraction Sites(implant preparation/ placement)
- Sinus lifts
- Cystic cavities
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GENOSS. The logo is in green and consists of the word "GENOSS" in a sans-serif font. The letters "GEN" are in a darker green, while the letters "OSS" are in a lighter, lime green color. The logo has a modern and clean design.
# K153676 510(k) Summary
# 1. Company
| | Submitter |
|----------------|--------------------------------------------------------------------------------------------------|
| Name | GENOSS Co., Ltd. |
| Address | 1F Gyeonggi R&DB center, 105 Gwanggyo-ro, Yeongtong-gu, Suwon-si,<br>Gyeonggi-do, 433-270, Korea |
| Phone/Fax | +82-31-888-5100/ +82-31-888-5595 |
| Contact person | Eunsang Lee / RA<br>eslee@genoss.com |
| Summary Date | 14th September, 2016 |
# 2. Device Name
| Proprietary name: | OSTEON III |
|----------------------|-----------------------------------|
| Common name: | Bone grafting material |
| Classification name: | Bone grafting material, Synthetic |
| Product Code: | LYC |
| Regulation: | 21 CFR 872.3930 |
# 3. Predicated Device
| K051885 | MBCP (Primary) |
|---------|--------------------------------|
| K040646 | Straumann Granules (Reference) |
# 4. Description
This product is a synthetic osteoconductive bone graft substitute composed of hydroxyapatite (HA) and beta-tricalcium phosphate (ß-TCP). This product presents an interconnected porous structure, similar to that human cancellous bone. This product is available as irregular shape particle of size 0.2~2.0mm. It is sterilized by gamma-radiation.
#### 510(k) Summary
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "GENOSS" in green font. The first half of the word is a darker green, while the second half is a lighter green. Below the word is the text "5. Indication for use" in a bold, black font.
OSTEON III is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects.
- Periodontal/infrabony defects
- Ridge augmentation
- Extraction Sites(implant preparation/ placement)
- Sinus lifts
- Cystic cavities
# 6. Technological Characteristics
OSTEON III has the similar technological characteristics as the predicated device; main material, intended use, device design, composition of Materials, physical properties is as following.
| Device name | OSTEON III | MBCP | Straumann Granules |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | GENOSS | Biomatlante | Straumann |
| 510(k) Number | New Device | K051885 | K040646 |
| Intended Use | | | |
| Indication for Use | -Periodontal/<br>Infrabony defects<br>- Ridge augmentation<br>-Extraction sites<br>(implant preparation /<br>placement)<br>- Sinus lifts | - Osteogenic scaffold<br>for bone healing<br>- Promotes<br>osteoblastic<br>proliferation and<br>differenciation.<br>- Stimulates the<br>osteoblastic response. | - Sinus lift<br>- Bony defects of the<br>alveolar ridge<br>- Intraosseous defects<br>- Peri-implant defects<br>- Extraction sockets |
| Applicable | Human Oral, periodontal | | |
| Function | Osteoconductive | | |
| Device Design | | | |
| Physical morphology | Interconnecting<br>macro and micro pores | | |
| Performance | Bone formation | | |
| Surgical site | Oral, Periodontal | | |
| Package | Glass vial,<br>Glass syringe | Glass, Plastic vial<br>Glass syringe | Glass, Plastic vial |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GENO" are in a dark green color, while the letters "SS" are in a lighter, lime green color. The font appears to be sans-serif and slightly rounded, giving the word a modern and clean look.
| Composition of Materials | | | |
|----------------------------------|------------------------------------------------|---------------------|---------------------|
| Composition | Biphasic Calcium Phosphate (HA 60%+ β-TCP 40%) | | |
| Physical Properties | | | |
| Bone material | Synthetic bone | | |
| Phase | Biphasic (HA+ β-TCP) | | |
| Porosity | ~80% | ~70% | ~90% |
| Pore size | 200~400 μm | 300~600 μm | 100~500 μm |
| Particle of Size | 0.2~2.0mm | 0.5~2.0mm | 0.4~1.0mm |
| Ca/P | 1.59 | 1.6 | 1.61 |
| Crystalline phase<br>composition | HA 60%: | HA 60%: | HA 60%: |
| Crystallinity | > 70% | > 70% | > 70% |
| Elution-pH | pH difference ≤ 1.5 | pH difference ≤ 1.5 | pH difference ≤ 1.5 |
| Others | | | |
| Biocompatible | Biocompatible | | |
| Sterilization | Gamma Irradiation | | |
| Shelf life | 1.5 years | | |
#### 7. Performance Data
Biocompatibility testing on the proposed OSTEON III has been completed. Requirements for biological evaluation of the proposed device were based on the Blue Book memorandum G95-1 issued on May 1, 1995, Use of International Standard ISO10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological bone and tissues with its intended use. The following biocompatibility tests were completed:
- -ISO Cytotoxicity -ISO Sensitization -ISO Pyrogencity -ISO Intracutaneous reactivity -ISO Systemic toxicity -ISO Subchronic toxicity -ISO Genotoxicity-Ames -ISO Genotoxicity-Micro nucleus -ISO Implantation -ISO Genotoxicity-Chromosomal aberration -ISO Endotoxin (LAL)
#### 510(k) Summary
{7}------------------------------------------------
The proposed OSTEON III was evaluated using the following performance bench testing to confirm the performance characteristics:
| - Visual Test | - Size |
|----------------------------------------|-----------------------------------|
| - Capacity | - Density |
| - Trace Elements | - Ca/P ratio |
| - Crystalline phase Composition | - Crystallinity |
| - Morphology & Surface Characteristics | - Water Solubility |
| - Elution Test (pH, Pb) | - Potassium permanganate consumed |
| - Ultraviolet rays absorption spectrum | - Resorption |
| - Particle size distribution | - Compressive strength |
| - Shear and Elastic modulus | |
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the OSTEON III met the established specifications necessary for consistent performance according to its intended use.
### 8. Substantial Equivalence
While the Indications for Use statements for the proposed and predicate devices differ, their meaning is equivalent as they share the same intended use, namely to serve as a bone grafting material to fill, augment or reconstruct various types of periodontal or oral/maxillofacial defects. The primary difference between OSTEON III and its primary predicate device MBCP (K051885) is related to slight mechanical and physical characteristics, such as small differences in pore and particle sizes. The proposed device features a pore size slightly lower than that available in the primary predicate device. However, a reference predicate is provided which features a range of pore sizes capturing those of the proposed device. Similarly, the proposed device features particle size which is slightly lower than that of the primary predicate. However, additional nonclinical testing results demonstrate that the any differences noted between OSTEON III and its predicate devices do not raise new concerns.
### 9. Conclusion
Based on the information provided in this premarket notification of GENOSS Co., Ltd. conclude that OSTEON III is substantially equivalent to its predicate devices.
#### 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.