Medical Surgical Masks-Non Sterile

K202594 · Shandong T&F Nonwoven Co., Ltd. · FXX · Jan 6, 2021 · General, Plastic Surgery

Device Facts

Record IDK202594
Device NameMedical Surgical Masks-Non Sterile
ApplicantShandong T&F Nonwoven Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateJan 6, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The Medical Surgical Masks-Non Sterile is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Medical Surgical Masks-Non Sterile is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile.

Device Story

Medical Surgical Masks-Non Sterile; single-use, disposable, non-sterile face mask. Three-layer construction: inner/outer layers polypropylene non-woven fabric; middle layer polypropylene melt-blown fabric. Features flat pleated design, nose piece (polypropylene/iron wire), and elastic ear loops (polyamide/polyurethane). Worn over mouth and nose to provide physical barrier against microorganisms, body fluids, and particulates. Used in clinical settings for infection control. No electronic components or software.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ASTM F2100-19, ASTM F1862-17 (fluid resistance 120 mmHg), ASTM F2299-03 (PFE ≥99%), ASTM F2101-19 (BFE ≥99%), MIL-M-36945C (Delta P <4.8 mmH2O/cm2), and 16 CFR 1610 (Flammability Class 1).

Technological Characteristics

Three-layer non-woven polypropylene construction. Nose piece: polypropylene/iron wire. Ear loops: 30% polyamide/70% polyurethane. Flat pleated form factor. ASTM F2100 Level 2 compliant. Non-sterile. No software or electronic components.

Indications for Use

Indicated for use by healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material in infection control settings.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 6, 2021 Shandong T&F Nonwoven Co., LTD % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building #15, XiYueHui, No. 5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China Re: K202594 Trade/Device Name: Medical Surgical Masks-Non Sterile Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: December 3, 2020 Received: December 7, 2020 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202594 Device Name Medical Surgical Masks-Non Sterile ### Indications for Use (Describe) The Medical Surgical Masks-Non Sterile is intended to be worn to protect and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Medical Surgical Masks-Non Sterile is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile. Type of Use (Select one or both, as applicable) | | <div></div> | |--|-------------| | | <div></div> | | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ The assigned 510(k) Number: K202594 # 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. - 1. Date of Preparation:2021/01/06 - 2. Sponsor Identification # SHANDONG T&F NONWOVEN CO., LTD. SOUTH OF SHUANGYONG ROAD, CHENGYANG DIST. QINGDAO SHANDONG PROVINCE, 266100, CHINA Contact Person: Minsheng Zhao Position: General Manager Tel: +86-18005326009 Fax: +86-0532-55676260 Email: mins 01@126.com - 3. Designated Submission Correspondent Mr. Ray Wang # Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China Tel: +86-18910677558 Fax: +86-10-56335780 Email: Ray.Wang(@believe-med.com {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Medical Surgical Masks-Non Sterile Common Name: Mask, Surgical Regulatory Information Classification Name: Mask, Surgical Classification: 2 Product Code: FXX Regulation Number: 878.4040 Review Panel: General Hospital Indication for use Statement: The Medical Surgical Masks-Non Sterile is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Medical Surgical Masks-Non Sterile is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile. Device Description The Medical Surgical Masks-Non Sterile is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Medical Surgical Masks-Non Sterile is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile. The proposed device(s) are White color, and Flat Pleated type mask, utilizing Ear Loops' way for wearing, and they all have Nose Piece design for fitting the facemask around the nose. The proposed device(s) are manufactured with three layers, the inner and outer layers are made of polypropylene non-woven fabric, and the middle layer is made of polypropylene melt-blown fabric. The Medical Surgical Masks-Non Sterile is held in place over the user's mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are made with 30% polyamide and 70% polyurethane. The nose piece contained in the proposed device(s) is in the layers of the facemask to allow the user to fit the facemask around their nose, which is made of polypropylene and iron wire. The proposed device(s) are sold non-sterile and are intended to be single-use, disposable devices. - Identification of Predicate Device(s) న్. Predicate Device {5}------------------------------------------------ K153496 Disposable Surgical Face Mask Xiantao Rayxin Medical Products Co., ltd. - Non-Clinical Test Conclusion 6. Applied Standards List: - > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity - > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. - > ASTM F2100-19, Standard Specification For Performance Of Materials Used In Medical Face Masks. - > ASTM F1862-17, Standard Test Method For Resistance Of Medical Face Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity) - > MIL-M-36945C, Method 1 Military Specifications: Surgical Mask disposable; - ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency > (Bfe) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus; - > ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres; - A 16 CFR 1610, Standard for the Flammability of clothing textiles; - > Bench Testing for the performance of Dimensions. | Title of the Test | Purpose of the Test | Acceptance criteria and<br>source of references | Results | |--------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------------------------------------------| | In Vitro<br>Cytotoxicity | Verify the Biological performance<br>of proposed device | No cytotoxicity effect as the<br>test method of ISO<br>10993-10 | Under the conditions of<br>the study, not cytotoxicity<br>effect | | Skin Irritation | Verify the Biological performance<br>of proposed device | No irritation effect as the<br>test method of ISO 10993-5 | Under the conditions of<br>the study, not an irritant | | Skin Sensitization | Verify the Biological performance<br>of proposed device | No sensitization effect as the<br>test method of ISO 10993-5 | Under conditions of the<br>study, not a sensitizer. | | Fluid Resistance | Verify the fluid resistance<br>performance of proposed device | 120 mmHg as the test<br>method of ASTM F1862 | 120 mmHg | | Particulate<br>Filtration Efficiency | Verify the Particulate filtration<br>efficiency performance of proposed<br>device | ≥98% as the test method of<br>ASTM F2299 | ≥99% | | Bacterial Filtration<br>Efficiency | Verify the bacterial filtration<br>efficiency performance of proposed<br> | ≥98% as the test method of<br>ASTM F2101 | ≥99% | ### Test Result Summary: {6}------------------------------------------------ | | device | | | |------------------------------------|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------| | Differential<br>Pressure (Delta P) | Verify the differential pressure<br>performance of proposed device | < 6.0mmH2O/cm2 as the test<br>method of MIL-M-36945C | <4.8 mmH2O/cm2 | | Flammability | Verify the flammability<br>performance of proposed device | Class 1 as the test method of<br>16 CFR 1610 | Class 1 | | Dimensions | Verify the Physical specifications of<br>proposed device | Length: 17.5 cm ±5mm<br>Width: 9.5 cm±5mm<br>Length of loop: 17 cm±5mm<br>Length of Nosepiece: 10<br>cm±5mm | Meet the acceptance<br>criteria | #### 7. Clinical Test Conclusion No clinical study is included in this submission. {7}------------------------------------------------ #### Substantially Equivalent (SE) Comparison 8. | ITEM | | Proposed Device K202594 | Predicate Device K153496 | Remark | |--------------------|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------| | Indication For Use | | The Medical Surgical Masks-Non Sterile<br>is intended to be worn to protect both the<br>patient and healthcare personnel from the<br>transfer of microorganisms, body fluids,<br>and particulate material. The Medical<br>Surgical Masks-Non Sterile is intended for<br>use in infection control practices to reduce<br>the potential exposure to blood and body<br>fluids. This is a single-use, disposable<br>device(s), provided non-sterile. | The Disposable Surgical Face Masks are<br>intended to be worn to protect both the<br>patient and healthcare personnel from<br>transfer of microorganisms, body fluids<br>and particulate material. These face masks<br>are intended for use in infection control<br>practices to reduce the potential exposure<br>to blood and body fluids. This is a single<br>use, disposable device(s), provided<br>non-sterile. | SAME | | Basic Design | | Ear Loops, Flat Pleated, 3 layers | Ear Loops, Tie-On, Flat Pleated, 3 layers | SAME | | Materials | Outer Facing Layer | polypropylene non-woven fabric | Spun-bond polypropylene | Analysis | | | Middle Layer | polypropylene melt-blown fabric | Melt blown polypropylene filter | | | | Inner Facing Layer | polypropylene non-woven fabric | Spun-bond polypropylene | | | | Nose Piece | PP + iron wire | Malleable aluminum wire | | | | Ear Loops | polyamide and polyurethane | Polyester | | | Color | | White | Blue | Analysis | | Dimension (Length) | | 17.5 cm +/- 5mm | 17.5 cm +/- 1cm | Similar | | Dimension (Width) | | 9.5 cm +/- 5mm | 9.5 cm +/- 1cm | | | OTC use | | Yes | Yes | SAME | | Single Use | | Yes | Yes | SAME | | Sterile | | No | No | SAME | | ASTM F2100 Level | | Level 2 | Level 2 | SAME | Table 1 General Comparison Table 2 Performance Characteristic Comparison | ITEM | Proposed Device<br>K202594 | Predicate Device<br>K153496 | ASTM F2100<br>Requirements for<br>Level 2 Classification | Remark | |----------------------------------------------------|----------------------------|-----------------------------|----------------------------------------------------------|--------| | Fluid Resistance<br>Performance ASTM<br>F1862 | 120 mmHg | 120 mmHg | 120 mmHg | SAME | | Particulate Filtration<br>Efficiency ASTM<br>F2299 | ≥99% | 98.46% | ≥ 98% | | | Bacterial Filtration<br>Efficiency ASTM<br>F2101 | ≥99% | 98.7% | ≥ 98% | | | Differential Pressure | < 4.8 mmH2O/cm² | 4.2 mmH2O/cm² | < 6.0 mmH2O/cm² | | {8}------------------------------------------------ | (Delta P)<br>MIL-M-36954C | | | | | |-----------------------------|---------|---------|---------|------| | Flammability<br>16 CFR 1610 | Class 1 | Class 1 | Class 1 | SAME | ## Table 3 Biocompatibility Comparison | ITEM | Proposed Device K202594 | Predicate Device K153496 | Remark | |---------------|----------------------------------------------------|--------------------------|--------| | Cytotoxicity | Under the conditions of the study, not cytotoxic | Comply with ISO 10993-5 | SAME | | Irritation | Under the conditions of the study, not an irritant | Comply with ISO 10993-10 | SAME | | Sensitization | Under conditions of the study, not a sensitizer. | | SAME | Analysis: The Medical Surgical Masks-Non Sterile is substantially equivalent to the Xiantao Rayxin Medical Products Disposable Surgical Face Mask. Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Disposable Surgical Face Mask cleared under K153496. #### 9. Substantially Equivalent (SE) Conclusion The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...