GXR-Series Diagnostic X-Ray System, is a stationary X-ray imaging system, for the purpose of acquiring X-ray images of the desired parts of a patient's anatomy. This device is not intended for mammography or bone density applications.
Device Story
Stationary digital radiographic X-ray system; consists of high-frequency inverter generator, tube support, beam limiter, patient table, wall Bucky stand, and X-ray tube. Operates by acquiring X-ray images of patient anatomy; images processed by RADMAX software. Used in clinical settings by trained operators. Operator selects exposure parameters via console (LCD panel/membrane switches); system supports Anatomical Programming (APR) and optional Automatic Exposure Control (AEC) to optimize exposure factors. Software provides image management, processing (enhancement, stitching, annotation), and DICOM connectivity. System includes remote diagnosis capabilities via internet. Output assists clinicians in diagnostic assessment of anatomical structures.
Clinical Evidence
Bench testing only. System verified against IEC 60601-1, 60601-1-3, 60601-2-28, 60601-2-54, 60601-1-2 (EMC), 62366, 62304, 62494-1, and NEMA PS 3.1 standards. Validation confirmed that all predetermined acceptance criteria were met.
Indicated for acquiring X-ray images of patient anatomy. Not for mammography or bone density applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
September 24, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
DRGEM Corporation % Mr. Carl Alletto Consultant OTech Inc. 8317 Belew Drive MCKINNEY TX 75071
Re: K202572
Trade/Device Name: GXR-Series Diagnostic X-Ray System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: August 27, 2020 Received: September 4, 2020
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202572
Device Name GXR-Series Diagnostic X-Ray System
#### Indications for Use (Describe)
GXR-Series Diagnostic X-Ray System, is a stationary X-ray imaging system, for the purpose of acquiring X-ray images of the desired parts of a patient's anatomy. This device is not intended for mammography or bone density applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared: September 18, 2020
### I. SUBMITTER
DRGEM Corporation 7F E-B/D Gwangmyeong Techno-Park 60, Haan-ro Gwangmyeong-si, Gyeonggi-do, 14322 Korea Email: radcheck@drgem.co.kr
Contact Person: Mr. Ki-Nam YANG, Director | QM representative
### II. DEVICE
| Product Name: | GXR-Series Diagnostic X-Ray System |
|--------------------|------------------------------------|
| Common Name: | Digital Diagnostic X-ray System |
| Regulation Name: | Stationary X-Ray System |
| Product Code: | KPR |
| Regulation number: | 892.1680 |
| Regulatory Class: | II |
### III. PREDICATE DEVICE
| Product Name: | GXR-Series Diagnostic X-Ray System |
|--------------------|------------------------------------|
| Submission Number: | K192364 |
| Decision Date: | 09/26/2019 |
| Common Name: | Digital Diagnostic X-ray System |
| Regulation Name: | Stationary X-Ray System |
| Product Code: | KPR |
| Regulation Number: | 892.1680 |
| Regulatory Class: | II |
### IV. DEVICE DESCRIPTION
GXR Series Diagnostic X-ray System is a digital radiographic system. There are 5 power output configurations which are reflected in the model's designation "GXR-XX" The models have 5 different output power ratings:
| DRGEM System<br>Model | GXR-32SD | GXR-40SD | GXR-52SD | GXR-68SD | GXR-82SD |
|----------------------------|----------|----------|----------|----------|----------|
| DRGEM<br>Generator Model | GXR-32 | GXR-40 | GXR-52 | GXR-68 | GXR-82 |
| Generator<br>Output Rating | 32kW | 40kW | 52kW | 68kW | 82kW |
The subject device, GXR Series Diaqnostic X-ray System, is designed to diagnose the human body by providing radiographic x-ray image with anatomical structure.
The subject device has the same x-ray hardware components and image management software as the predicate device. The subject device consists of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, a patient table, wall Bucky stand, and an x-ray tube, that operates on a high-frequency inverter method.
The operator control console is designed to be user-friendly, and the user can select or change x-ray parameters easily using a large graphic LCD panel display and a soft membrane switch. The GXR Series high frequency X-ray generator (manufactured by
{4}------------------------------------------------
DRGEM) features accuracy, reproducibility and long-term stability with capacitor assisted general line power supply. The APR (Anatomical Programming) and the optional AEC (Automatic Exposure Control) features gives the user control of exposure factors, automatically optimized for the radiological study selected.
The digital flat panel detectors provide spatial resolution, MTF, DQE and stability based on fine pixel pitch. Selection of an anatomical study on the imaging software automatically sets up the x-ray generator's pre-programmed exposure technique setting and post image processing for selected study. The subject device is able to use a total of 10 different digital detectors, (8 new plus 2 cleared in the predicate, which have been previously cleared by the 510(k) process:
- VAREX, 4343R v3 K172951(predicate) ●
- VAREX, 4336W v4- K161459 (predicate) .
- . VAREX, XRpad2 3025 HWC-M- K161942
- VAREX, XRpad2 4336 HWC-M- K161966 .
- VAREX, XRpad2 4343 HWC-M- K181526 ●
- i-Ray, Mano4336W- K201004 ●
- . i-Ray, Mano4343W- K201043
- . Vieworks, VIVIX-S1417N(NAW,NBW)-K163703
- Vieworks, VIVIX-S1717N(NAW,NBW)- K152894 ●
- VAREX, 4343W- K161459 .
The GXR Series Diagnostic X-ray System consists of a combination of an x-ray generator. and associated equipment such as tube stand, patient table, and, digital imaging system. The main power cabinet contains the HT tank and control circuits, the filament drivers, the low speed starter, and interface connections to the room equipment.
- . Tube stand and patient table allows the operator to position the patient.
- Full Featured Imaging Software & Digital Image Processing .
- Control console: .
- The control console allows the operator to select the technique factors, O image receptors, etc., and to initiate an X-ray exposure.
- Real-time monitoring self-diagnosis function and Error code display o
- o Overload & HU protection and error message display
- o Support programmable max. 1,280 APR conditions with APR utility software
- Automatic calibration without measurement equipment. o
- Adaptable calibration keeps up accuracy through long-term usage o
- Smaller, lighter and convenient modular design and user-friendly system O configuration
- O Constant dose output due to kV and mA requlation during exposures.
- Large graphic LCD panel user-friendly controls for APR and technique o display at a glance,
- Time and mA / mAs selections are based on R'10 rule of ISO 497. o
- Remote diagnosis software for system diagnosis via internet. O
- o Easy parameter setting and Firmware upgrade
- System diagnosis, Error log and Statistical data display O
The image manaqement software, RADMAX Digital Imaging Software (K182537) by DRGEM, is used in both the predicate and subject device to serve as a convenient interface to the hardware and images. Anatomical view-based digital image processing automatically optimizes and enhances the quality of the captured images. RADMAX
{5}------------------------------------------------
(K182537) Digital Imaging Software is designed for acquiring images and processing the acquired images. The software can be used together with a digital X-ray detector and or an X-Ray generator. The main features of the RADMAX software are controlling and interfacing the detector, acquiring images after X-ray, storing acquired images, managing data, and image processing. It can also perform system control such as the collimation size, and filter selection. The following are general features.
- . Windows based graphic user interface
- Multi-image display (1x1 ~ 4x4) ●
- Multi-image selection ●
- . Auto display layout changing function
- . X-ray generator control panel
- . Unlimited procedure step
- Quick step-add feature and image maintenance feature by popup menu ●
- ROI changing and creation feature ●
- Marker feature (support the creation of unlimited number of markers by user) ●
- Multi-language support .
- EXCEL sheet for language support (only possible on Microsoft Office automation ● environment)
- . DAP meter (optional)
- . Unlimited PACS code (CPT code)
- Default anatomic program more than 700
- Support DICOM Worklist SCU, DICOM Storage SCU and transfer function
- . Support DICOM Multi-transfer function
- High-performance post-processing feature ●
- Copy & Move Images
- Dose monitoring function ●
- Built-in memory function ●
- Grid line suppression function .
- . Reject analysis function
### V. INDICATIONS FOR USE
GXR-Series Diagnostic X-Ray System, is a stationary X-ray imaging system, for the purpose of acquiring X-ray images of the desired parts of a patient's anatomy. This device is not intended for mammography or bone density applications.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
### Summary of differences:
The subject device can interface to one of eight previously cleared digital flat panel detectors in addition to the 2 detectors used in the predicate device as listed above.
- . VAREX, 4343R v3 - K172951(predicate)
- VAREX, 4336W v4- K161459 (predicate) .
- VAREX, XRpad2 3025 HWC-M- K161942 .
- . VAREX, XRpad2 4336 HWC-M- K161966
- . VAREX, XRpad2 4343 HWC-M- K181526
- . i-Ray, Mano4336W- K201004
{6}------------------------------------------------
- i-Ray, Mano4343W- K201043 .
- . Vieworks, VIVIX-S1417N(NAW,NBW)-K163703
- Vieworks, VIVIX-S1717N(NAW,NBW)- K152894 ●
- . VAREX, 4343W- K161459
The following information compares the subject device to the predicate.
Any differences between the subject device and the predicated device have no negative impact on safety or efficacy of the subject device and does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
| Specification | Predicate device K192364 | Subject Device | Impact of Differences |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | GXR-Series Diagnostic<br>Imaging System | GXR-Series Diagnostic<br>Imaging System | None |
| Manufacturer | DRGEM Corporation | DRGEM Corporation | None |
| Model Number | GXR-SD/CSD/USD | GXR-SD/CSD/USD | None |
| High Frequency X-<br>ray Generator | | | |
| Output Power | 32KW, 40KW, 52KW, 68KW,<br>82KW | 32KW, 40KW, 52KW,<br>68KW, 82KW | None |
| Generator models<br>(manufactured by<br>DRGEM) | GXR-32, GXR-40, GXR-52,<br>GXR-68, GXR-82<br>GXR-C32, GXR-C40, GXR-<br>C52 | GXR-32, GXR-40, GXR-<br>52, GXR-68, GXR-82<br>GXR-C32, GXR-C40,<br>GXR-C52 | None |
| Line voltage | 220~230VAC<br>380/400/480VAC | 220~230VAC<br>380/400/480VAC | None |
| kV Range | 40~125kV, 1kV step (Optional<br>40~150kV) | 40~125kV, 1kV step<br>(Optional 40~150kV) | None |
| mA Range | GXR-32-10 to 400mA<br>GXR-40-10 to 500mA<br>GXR-52-10 to 640mA<br>GXR-68-10 to 800mA<br>GXR-82-10 to 1,000mA | GXR-32-10 to 400mA<br>GXR-40-10 to 500mA<br>GXR-52-10 to 640mA<br>GXR-68-10 to 800mA<br>GXR-82-10 to 1,000mA | None |
| Image Acquisition | | | |
| Detectors | VAREX, 4343R v3 - K172951<br>VAREX, 4336W v4- K161459 | VAREX, 4343R v3 -<br>K172951<br>VAREX, 4336W v4-<br>K161459<br>VAREX, XRpad2 3025<br>HWC-M- K161942<br>VAREX, XRpad2 4336<br>HWC-M- K161966<br>VAREX, XRpad2 4343<br>HWC-M- K181526<br>i-Ray, Mano4336W-<br>K201004<br>i-Ray, Mano4343W-<br>K201043<br>Vieworks, VIVIX-S1417N<br>(NAW, NBW)-K163703<br>Vieworks, VIVIX-S1717N<br>(NAW, NBW)- K152894<br>VAREX, 4343W- K161459 | Yes, there is a difference.<br>An additional eight digital<br>detectors plus the two from<br>the predicate can be used<br>with the system. All the flat<br>panel detectors have been<br>previously cleared by<br>510(k). The system has<br>been tested and a risk<br>analysis performed. There is<br>"No negative impact on<br>safety or efficacy" and no<br>new potential or increased<br>safety risks concerning were<br>raised because of this<br>difference. |
| Image Management<br>Software | | | |
| Specification | Predicate device K192364 | Subject Device | Impact of Differences |
| Horizontal Flip | Available | Available | None |
| Vertical Flip | Available | Available | None |
| Rotate CW/CCW | Available | Available | None |
| Text Annotation | Available | Available | None |
| Ruler: Distance tool | Available | Available | None |
| Angle measurement<br>tool | Available | Available | None |
| Zoom | Available | Available | None |
| Magnify | Available | Available | None |
| Image panning | Available | Available | None |
| Auto fitting to window<br>size | Available | Available | None |
| Image crop/cut<br>function | Available | Available | None |
| Image Copy | Available | Available | None |
| Recover the original<br>image | Available | Available | None |
| Window level<br>CD Burning | Available | Available | None |
| DICOM Print | Available | Available | None |
| Image Stitching | Available | Available | None |
{7}------------------------------------------------
### VII. PERFORMANCE DATA
Nonclinical Testing:
### Summary:
Based on the performance as documented in the testing, the subject device was found to have a safe and effectiveness profile that is similar to the predicate device.
The complete system has been assessed and tested at the factory and by Standards testing facilities. GXR-Series Diagnostic X-Ray System, has passed all predetermined testing criteria. Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
The following International Standards were used to develop and verify the system. GXR-Series Diagnostic X-Ray System, device has met all the requirements listed in the Standards except for inapplicable requirements (which are listed in the various test reports):
| Std # | Safety/EMC Standards Description | FDA<br>Consensus<br>Standard # |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| IEC 60601-1 | Medical electrical equipment, Part 1: General requirements for<br>basic safety and essential performance | 19-4 |
| IEC 60601-1-3 | Medical electrical equipment<br>Part 1-3: General Requirements for Radiation Protection in<br>Diagnostic X-Ray Equipment | 12-269 |
| IEC 62366 | This part of IEC 62366-1:2015 specifies a process for a<br>manufacturer to analyze, specify, develop and evaluate the<br>usability of a medical device as it relates to safety | 5-129 |
{8}------------------------------------------------
| Std # | Safety/EMC Standards Description | FDA<br>Consensus<br>Standard # |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| IEC 60601-2-28 | IEC 60601-2-28<br>Medical electrical equipment Part 2: Particular requirements for the<br>safety of X-ray source assemblies and X-ray tube assemblies for<br>medical diagnosis. | 12-204 |
| 60601-2-54 | IEC 60601-2-54 Medical electrical equipment Part 2: Particular<br>requirements for the basic safety and essential performance of X-<br>ray equipment for radiography and radioscopy | 12-296 |
| IEC 60601-1-2<br>(EMC) | IEC 60601-1-2 Edition 4.0 2014-02. Medical electrical equipment -<br>Part 1-2: General requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic disturbances<br>Requirements and tests. | 19-8 |
| IEC 62304 | ANSI AAMI IEC 62304:2015<br>Medical device software - Software life cycle processes | 13-79 |
| NEMA PS 3.1 | NEMA PS 3.1 - 3.20 (2016). Digital Imaging and Communications<br>in Medicine (DICOM) Set DICOM Standard. | 12-300 |
| IEC/ISO10918-1 | JPEG Standard IEC/ISO10918-1 First edition 1994-02-15,<br>Information technology - Digital compression and coding of<br>continuous-tone still images: Requirements and guidelines<br>[Including: Technical Corrigendum 1 | 12-261 |
| IEC 62494-1 | IEC 62494-1 Edition 1.0 (2008-08), Medical electrical equipment -<br>Exposure index of digital X-ray imaging systems - Part 1:<br>Definitions and requirements for general radiography. | 12-215 |
| ISO 14971: | ISO 14971:2007/(R)2010 (Corrected 4 October 2007), Medical<br>devices - Applications of risk management to medical devices. | 5-40 |
| ISO 15223-1 | ISO 15223-1 Third Edition 2016-11-01, Medical devices - Symbols<br>to be used with medical device labels, labelling, and information to<br>be supplied - Part 1: General requirements. | 5-117 |
### Conclusion:
The 510(k) Pre-Market Notification for the GXR-Series Diagnostic X-Ray System, contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device will be manufactured in accordance with the voluntary standards listed in the voluntary standard survey. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs comparably to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.