Clinical literature; Post-market surveillance data (US and Europe)
The sponsor utilized clinical literature and post-market surveillance data to support the safety and effectiveness of removing Type A trauma fractures from the device's contraindications, demonstrating a history of safe use since 2008.
Post-market surveillance and clinical literature review; Retrospective analysis of post-market data and literature; Follow-up/Duration: Since 2008; Study Period: 2008-2020
Patients with Type A traumatic vertebral fractures
Not applicable for this study
Safety and effectiveness in Type A traumatic fractures
Indications for Use
The SpineJack® Expansion Kit is indicated for use in the reduction of painful osteoporotic vertebral compression fractures. It is intended to be used in combination with Stryker Vertaplex® and Vertaplex® HV Bone Cement.
Device Story
Implantable fracture reduction system; used to restore vertebral body height in osteoporotic compression fractures. Device consists of expandable metal implant (Ti-6Al-4V) inserted into vertebral body; expanded via plastic deformation of struts to restore height; stabilized by injection of Stryker VertaPlex® or VertaPlex® HV PMMA bone cement at low pressure. Used in clinical setting by surgeons. Output is restored vertebral structure; benefits patient by reducing pain and stabilizing fracture. Modification involves removing Type A traumatic fractures from contraindications list based on clinical literature and post-market surveillance data.
Clinical Evidence
No clinical testing performed for this 510(k). Evidence supporting modification consists of clinical literature review, US and European post-market surveillance data, and ISO 14971 risk analysis.
Technological Characteristics
Material: Ti-6Al-4V; PMMA bone cement (sold separately). Expansion mechanism: continuous plastic deformation of struts. Dimensions: 14-20mm length, 4-6mm width, 4-20mm height (surgeon determined). Implantable device. No software or electronic components.
Indications for Use
Indicated for reduction of painful osteoporotic vertebral compression fractures in patients. Contraindicated for type B or C traumatic vertebral fractures (Magerl classification).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
October 20, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Stryker Corporation Kristi Ashton Senior Staff RA Specialist 1926 Stryker Way Kalamazoo, Michigan 49002
Re: K202393
Trade/Device Name: SpineJack® Expansion Kit Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: NDN Dated: August 18, 2020 Received: August 21, 2020
Dear Kristi Ashton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202393
Device Name
SpineJack® Expansion Kit (Spinejack ®)
Indications for Use (Describe)
The SpineJack® Expansion Kit is indicated for use in the reduction of painful osteoporotic vertebral compression fractures. It is intended to be used in combination with Stryker Vertaplex® HV bone cement.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">×</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. Above the word "stryker" is the alphanumeric code "K202393". The word "stryker" is in all lowercase letters and is the most prominent element in the image. The background is plain white.
## 510(K) SUMMARY
| 510(K) SUMMARY: SPINEJACK® EXPANSION KIT | |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Stryker Instruments<br>1941 Stryker Way<br>Kalamazoo, MI 49002 USA<br>Contact Name: Kristi Ashton<br>Title: Senior Staff RA Specialist<br>Phone: 269-800-1692<br>E-mail: Kristi.Ashton@Stryker.com |
| Date prepared: | October 20, 2020 |
| Subject Device Trade Name: | SpineJack Expansion Kit (SpineJack) |
| Common Name: | Implantable Fracture Reduction System |
| Classification Name: | Cement, Bone, Vertebroplasty |
| Product Code: | NDN |
| Regulatory Class: | II |
| Regulation Name: | Polymethylmethacrylate (PMMA) |
| Regulation Number: | 21 CFR 888.3027 |
| Predicate Device: | SpineJack Expansion Kit (SpineJack), K181262 |
| Device Description: | The SpineJack Expansion Kit (herein referred to as “SpineJack”) is an<br>implanted fracture reduction system, intended to reduce vertebral<br>compression fractures. The SpineJack Expansion Kit is used with the<br>Preparation Kit. The Expansion Kit is available in three sizes, to<br>accommodate different vertebral body sizes, Ø4.2mm, Ø5mm, and<br>Ø5.8mm. After the SpineJack implant is inserted, it is expanded, and a<br>validated Stryker PMMA bone cement, such as Stryker VertaPlex® or<br>VertaPlex® HV Bone Cement, is injected at a low pressure to stabilize<br>the restored vertebral body. The bone cement and its delivery system are<br>intended to be used with the SpineJack but are sold separately. |
| Indications for Use | The SpineJack® Expansion Kit is indicated for use in the reduction of<br>painful osteoporotic vertebral compression fractures. It is intended to be<br>used in combination with Stryker Vertaplex® and Vertaplex® HV Bone<br>Cement. |
| Non-Clinical Testing | Non-clinical testing was deemed not required for this 510(k). The<br>SpineJack implant was previously tested to be non-pyrogenic. |
| Clinical Testing | Clinical testing was deemed not required for this 510(k) |
| Other Evidence: | Clinical literature, post-market surveillance data and a risk analysis per<br>ISO 14971 were used in support of the subject modification. |
| Summary of Technological Characteristics | |
Table 1 below details the comparisons between the subject and predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "stryker" in a bold, sans-serif font. Above the word "stryker" is the text "K202393" in a smaller font size. The word "stryker" is the main focal point of the image.
| Table 1 Comparison of Regulatory and Technical Characteristics | | | | |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Element of<br>Comparison | SpineJack<br>Subject Device | SpineJack<br>Predicate Device | Comparison | |
| Indications for Use | The SpineJack®<br>Expansion Kit is<br>indicated for use in the<br>reduction of painful<br>osteoporotic vertebral<br>compression fractures. It<br>is intended to be used in<br>combination with<br>Stryker Vertaplex® and<br>Vertaplex® HV bone<br>cements. | The SpineJack®<br>Expansion Kit is<br>indicated for use in the<br>reduction of painful<br>osteoporotic vertebral<br>compression fractures.<br>It is intended to be used<br>in combination with<br>Stryker Vertaplex® and<br>Vertaplex ® HV bone<br>cements. | Identical | |
| Contraindications | Patients presenting with<br>type B or C traumatic<br>vertebral fractures<br>according to the Magerl<br>classification. | Patient presenting with<br>traumatic fractures<br>(Reference IFU for full<br>list of contraindications) | Different-It is proposed<br>that Type A trauma<br>fractures are removed<br>from the list of<br>contraindications for<br>SpineJack. Clinical<br>literature, post-market data<br>and a risk assessment per<br>ISO 14971 were provided<br>to support the safety of<br>this modification. This<br>change does not raise new<br>types of safety and<br>effectiveness questions<br>regarding the use of the<br>subject device. | |
| Contact | Implantable | Implantable | Identical | |
| Material | Ti-6Al-4V with PMMA<br>(cement injected) | Ti-6Al-4V with PMMA<br>(cement injected) | Identical | |
| Length (mm) | 14 - 20 (plate length) | 14 - 20 (plate length) | Identical | |
| Width (mm) | 4 - 6 | 4 - 6 | Identical | |
| Height pre and post<br>expansion (mm) | 4 - 20 (Surgeon<br>determined) | 4 - 20 (Surgeon<br>determined) | Identical | |
| Expansion Mechanism | Plastic deformation of<br>struts, continuous | Plastic deformation of<br>struts, continuous | Identical | |
| Number of Implants<br>Typically Used | 1-2 per patient | 1-2 per patient | Identical | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. Above the word "stryker" is the alphanumeric code "K202393". The text is black and the background is white. The image appears to be a logo or product identifier.
SpineJack has the same intended use, fundamental scientific technology, principle of operation, and mode of action as the predicate device. An evaluation of clinical literature, post-market data in US and Europe and risk per ISO 14971 provide the evidence to support the removal of Type A trauma fractures from the contraindications. The use of post-market data from Europe demonstrates the history of safe and effective use of the Conclusion product in Type A traumatic fractures since 2008. The evaluation of the post-market data, use history and a clinical literature review provide established evidence which confirm that the subject modification does not raise new types of safety and effectiveness questions regarding the use of the subject device. Therefore, it is proposed that the subject SpineJack is equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.