AcuSee AS-P1000 System

K202119 · Weipeng (Suzhou) Medical Devices Co., Ltd. · IYO · Oct 7, 2020 · Radiology

Device Facts

Record IDK202119
Device NameAcuSee AS-P1000 System
ApplicantWeipeng (Suzhou) Medical Devices Co., Ltd.
Product CodeIYO · Radiology
Decision DateOct 7, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1560
Device ClassClass 2

Intended Use

The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization. The AcuSee AS-P1000 system is intended to be used in a clinical setting.

Device Story

AcuSee AS-P1000 is an ultrasound-guided instrument tracking system. It uses stereo-vision optical detection to track markers on mounts attached to an ultrasound probe and a needle-like instrument. The system processes these inputs to calculate the relative position and trajectory of the needle in real-time. It outputs an overlay of the instrument's position and projected path onto the B-scan ultrasound image on a display. Used in clinical settings by physicians during ultrasound-guided procedures. The system is mounted on a movable cart and acts as an accessory to existing ultrasound machines. By providing real-time visual guidance, it assists the clinician in navigating the needle to the target, potentially improving procedural accuracy and patient outcomes.

Clinical Evidence

No clinical data. Bench testing only, including software verification/validation, system verification/validation, and compliance testing for electrical and electromagnetic safety (IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014).

Technological Characteristics

Optical tracking system consisting of a camera, PC, and tracking mounts. Uses stereo-vision optical detection to track markers on ultrasound probes and needles. Real-time processing of spatial geometry. Non-sterile, cart-mounted system. Complies with IEC 60601-1 and IEC 60601-1-2 standards.

Indications for Use

Indicated for augmenting ultrasonic images of interventional needles or rigid needle-like devices (e.g., biopsy, aspiration, ablation needles) and predicting their future path on a display alongside B-scan ultrasound images. Intended for use in clinical settings where ultrasound visualization is required.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 7, 2020 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Weipeng (Suzhou) Medical Devices Co., Ltd. % Geetha Rao, Ph.D. Regulatory and Quality Representative Springborne Life Sciences 750 Menlo Avenue, Suite 200 MENLO PARK CA 94025 ## Re: K202119 Trade/Device Name: AcuSee AS-P1000 System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO Dated: July 24, 2020 Received: July 30, 2020 Dear Dr. Rao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202119 Device Name AcuSee AS-P1000 System #### Indications for Use (Describe) The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization. The AcuSee AS-P1000 system is intended to be used in a clinical setting. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|----------------------------------------------------------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"><b>X</b></div> | | Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"></div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY #### Submitter | Name | Weipeng (Suzhou) Medical Devices Co., Ltd.<br>Unit 207, Building B2 | |----------------|----------------------------------------------------------------------| | Address | 218 Xinghu St., Suzhou Industrial Park<br>Suzhou, 215123, P.R. China | | Phone Number | +86 0512-67311267 | | Fax Number | NA | | Contact Person | Bin Yang | | Email | binyang@vibronixinc.com | | Phone Number | +86 0512-67311267 | | Fax Number | NA | | Date Prepared | July 15, 2020 | ### Device Information | Trade Name | AcuSee AS-P1000 System | |--------------|--------------------------------------------------------| | Common Name | Tracking or Guidance System | | Product Code | IYO, Ultrasonic pulsed echo imaging system | | Regulation | 21 CFR 892.1560, Ultrasonic pulsed echo imaging system | | Device Class | II | #### Predicate Information | Device Name | Clear Guide ONE | |---------------|-----------------| | 510(k) Number | K141806 | #### Device Overview The AcuSee AS-P1000 system is a medical device that tracks instrument positioning during ultrasoundguided procedures to provide instrument guidance to the end user. With accessories attached to the transducer probe and needle or needle-like rigid device, the AcuSee AS-P1000 system identifies and tracks the probe and the device within the field of view through optical detection technology. By coupling positioning information with the ultrasound image, the projected instrument pathway is displayed to the user for guiding instrumentation. The AcuSee AS-P1000 system is an accessory to any compatible ultrasound machine and transducer probe. #### Indications for Use The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle, and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization. {4}------------------------------------------------ # VibroniX The AcuSee AS-P1000 system is intended to be used in a clinical setting. ### Technological Characteristics The AcuSee AS-P1000 System applies optical tracking technology to augment the ultrasonic image of an interventional needle or needle-like rigid device and for predicting its future path on a display and also show the image of a B-scan (or similar display) of a medical ultrasound imaging system. The relative position and trajectory of the needle is calculated using stereo-vision optical detection of markers on tracking mounts attached to the ultrasound probe and needle device. | Feature | Subject Device<br>AcuSee AS-P1000 System | Predicate Device<br>Clear Guide One | SE Comparison/<br>Comment | |--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------| | 510(k) Number | NA | K141806 | NA | | Product Code | IYO | IYO | Same | | Regulation No. | 892.1560 | 892.1560 | Same | | Risk Class | II | II | Same | | Regulatory<br>Pathway | 510(k) | 510(k) | Same | | Indications for<br>Use | The AcuSee AS-P1000 System<br>is indicated for augmenting the<br>ultrasonic image of an<br>interventional needle or<br>needle-like rigid device, such as<br>a biopsy needle, an aspiration<br>needle, or ablation needle, and<br>for predicting its future path<br>on a display, which also shows<br>the image of a B-scan (or<br>similar display) of a medical<br>ultrasound imaging system.<br>The device is intended to be<br>used in procedures where<br>ultrasound is currently used for<br>visualization. | The Clear Guide ONE is<br>indicated for augmenting the<br>ultrasonic image of an<br>interventional needle or<br>needle-like rigid device, such as<br>a biopsy needle, an aspiration<br>needle, or ablation needle, and<br>for predicting its future path<br>on a display, which also shows<br>the image of a B-scan (or<br>similar display) of a medical<br>ultrasound imaging system.<br>The device is intended to be<br>used in procedures where<br>ultrasound is currently used for<br>visualization. | Identical | | Intended Users | Physician | Physician | Same | | Use<br>Environment | Clinical Setting | Clinical Setting | Same | | Sterility | Non-Sterile | Non-Sterile | Same | | Movability | Main system mounted on<br>movable cart | Main system mounted on<br>movable cart | Same | | Feature | Subject Device<br>AcuSee AS-P1000 System | Predicate Device<br>Clear Guide One | SE Comparison/<br>Comment | | Fundamental<br>Technology | Optical Detection | Optical Detection | Same | | Device Form<br>Factor | Main system contains camera,<br>and PC with software on<br>movable cart, with power cord<br>for AC power,<br>Separate tracking accessories<br>consist of tracking mounts | Main system (Clear Guide<br>Core) contains hardware and<br>software with power adapter<br>and cord,<br>Separate tracking accessories<br>include camera (Clear Guide<br>SuperProbe) and custom<br>instruments | Similar primary<br>concept with different<br>physical configuration | | Detection<br>Mechanism | Camera located at a distance<br>tracks custom tracking mounts<br>on ultrasound probe and<br>needle | Camera is directly mounted on<br>ultrasound probe and tracks<br>custom needle | Similar concept, with<br>a variant spatial<br>geometry | | Tracking<br>Algorithm | Automatic | Automatic | Similar | | Timing for<br>tracking | Real-time | Real-Time | Same | | Information<br>Displayed | Overlay of the detected<br>instrument positioning and<br>projected path onto the<br>ultrasound image | Overlay of the detected<br>instrument positioning and<br>projected path onto the<br>ultrasound image | Same | | Reusability | Reusable main system with<br>consumable tracking mounts | Reusable main system and<br>probe with consumable<br>tracking needle | Similar | | Instruments<br>tracked | Any compatible needle-like<br>instrument | Requires custom needle (not<br>known what kind of needle) | Similar | | Tracking of<br>instrument tip | Any compatible instrument | Only with custom instrument<br>with CLEAR GUIDE PercepTIP | Similar | | Compatibility<br>with probe | Tracking mount fits any<br>compatible probe | Optical head with camera fits<br>any compatible probe | Same functionality | | Tracking<br>instruments in<br>and out-of-plane | Within +/- 90 deg | Within +/- 45 deg | Subject device has<br>wider performance<br>range | | Feature | Subject Device<br>AcuSee AS-P1000 System | Predicate Device<br>Clear Guide One | SE Comparison/<br>Comment | | Electrical Safety<br>and<br>Electromagnetic<br>Compatibility | IEC 60601-1:2005+A1:2012 -<br>Medical electrical equipment –<br>Part 1: General requirements<br>for basic safety and essential<br>performance<br>IEC 60601-1-2:2014 - Medical<br>electrical equipment - Part 1-2:<br>General requirements for basic<br>safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests. | IEC 60601-1 3rd Edition:<br>Medical Electrical Equipment;<br>Part 1 General Requirements<br>for Safety<br>IEC 60601-1-2 :2007 – Medical<br>electrical equipment - Part 1-2:<br>General requirements for<br>safety – Collateral standard:<br>Electromagnetic compatibility | Similar - Subject<br>device meets updated<br>standard | | Performance<br>Testing | Bench testing | Bench testing | Similar | | Animal Testing | None | None | Same | | Clinical Testing | None | None | Same | ### Device Comparison Table {5}------------------------------------------------ # VibroniX {6}------------------------------------------------ ### Substantial Equivalence Overview The AcuSee AS-P1000 System has the same intended use and indications for use, as well as similar technological characteristics and principles of operation compared to the previously legally cleared predicate device: Clear Guide ONE (K141806, cleared in 2014). The AcuSee AS-P1000 System and the predicate device both apply the same optical detection technology in a similar device form factor. The AcuSee AS-P1000 system overlays the detected instrument positioning data onto an existing ultrasound image through proprietary software algorithms, In the same manner as the predicate device. Both devices track targeted instrumentation and provide real-time visual guidance during a clinical procedure. Both devices have reusable main components and some consumable accessories. The AcuSee AS-P1000 System was tested to demonstrate compliance to the same applicable safety standards as the predicate device, but for updated versions of the standards. The AcuSee AS-P1000 System uses the same primary concept for the detection mechanism but uses a different physical configuration for the camera and tracked targets, resulting in a different spatial geometry. This variation does not alter the fundamental mechanism and raises no new questions of safety. Performance data was collected to demonstrate that the AcuSee AS-P1000 system achieves its intended function in a manner that is as effective as the predicate device. The AcuSee AS-P1000 system uses tracking mounts that are designed to fit ultrasound probes and instruments that are available on the market. Although specific models of probes (and the associated ultrasound machines with which they are used) and instruments are individually validated to be compatible, the mounts themselves are standardized and may be used with any compatible probe or instrument. Hence unlike the predicate device, the system does not require special patterns to be marked or printed to the shaft of the needle or other tracked instrument. {7}------------------------------------------------ # VibroniX Although there are minor differences in technological characteristics of the AcuSee AS-P1000 System, they do not raise new issues of safety or effectiveness. ## Summary of Performance Testing The performance testing for the AcuSee AS-P1000 System includes software verification/validation testing, system verification/validation testing and testing to compliance standards for electrical and electromagnetic safety. Traceability has been documented between the system specification to verification/validation protocols. The results of performance testing show that the AcuSee AS-P1000 system is as safe and as effective as the predicate device. The AcuSee AS-P1000 system complies with the following recognized consensus standards: - 1. IEC 60601-1:2005+A1:2012 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance). - 2. IEC 60601-1-2:2014 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests). # Conclusion The performance of the AcuSee AS-P1000 System is substantially equivalent to that of the Clear Guide ONE system and raises no safety or effectiveness issues and performs as well or better than the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%