Clear Guide SCENERGY

K171677 · Clear Guide Medical · JAK · Dec 8, 2017 · Radiology

Device Facts

Record IDK171677
Device NameClear Guide SCENERGY
ApplicantClear Guide Medical
Product CodeJAK · Radiology
Decision DateDec 8, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The Clear Guide SCENERGY is software that provides fusion of images from Computed Tomography (CT), Magnetic Resonance (MR), and Ultrasound (US) modalities. US images can be fused with either CT or MR. The Clear Guide SCENERGY utilizes the Clear Guide CORE and Clear Guide SuperPROBE platform to display images of the target regions and the projected path of the interventional instrument, while taking into account patient movement and deformation. Instrumentation used with the Clear Guide SCENERGY might include an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle, or an ablation needle. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting.

Device Story

Clear Guide SCENERGY is an accessory to ultrasound systems for image-guided interventions. It uses optical detection (Clear Guide CORE and SuperPROBE) to track interventional instruments (needles) in real-time. The system inputs include pre-operative CT or MR image volumes and real-time ultrasound data. Proprietary software performs automated segmentation and registration to fuse modalities, accounting for patient movement and deformation. The system displays the fused image and the projected instrument path to the clinician. Used in clinical settings by physicians to guide needle placement; the output assists in targeting regions of interest. Benefits include improved visualization and guidance during procedures, potentially increasing accuracy and reducing procedural risk.

Clinical Evidence

Bench testing only. No clinical or animal data provided. Performance evaluated using phantom datasets to validate automated segmentation of fiducial markers, fusion quality (Tissue Registration Error), and systematic tip-to-tip tracking error (comparing device guidance to ground truth CT/MR).

Technological Characteristics

Optical detection system consisting of Clear Guide CORE and Optical Head. Software-based fusion of US with CT or MR. Automated segmentation and registration algorithms. Non-sterile, reusable, clinical-use accessory. Connectivity via integration with existing ultrasound systems.

Indications for Use

Indicated for patients undergoing interventional or diagnostic procedures requiring image fusion of Ultrasound with CT or MR, excluding brain procedures. Used with rigid interventional instruments (e.g., biopsy, aspiration, or ablation needles).

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 8, 2017 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and full name on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. Clear Guide Medical, Inc. % Dorothee Heisenberg, Ph.D. Executive Vice President 3600 Clipper Mill Road. Suite 400 BALTIMORE MD 21211 Re: K171677 Trade/Device Name: Clear Guide SCENERGY Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: November 3, 2017 Received: November 6, 2017 Dear Dr. Heisenberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171677 Device Name Clear Guide SCENERGY #### Indications for Use (Describe) The Clear Guide SCENERGY is software that provides fusion of images from Computed Tomography (CT), Magnetic Resonance (MR), and Ultrasound (US) modalities. US images can be fused with either CT or MR. The Clear Guide SCENERGY utilizes the Clear Guide CORE and Clear Guide SuperPROBE play images of the target regions and the projected path of the interventional instrument, while taking into account patient movement and deformation. Instrumentation used with the Clear Guide SCENERGY might include an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Clear Guide Medical. The logo consists of a stylized letter "L" formed by a thick black line and two thin gray lines, followed by the words "CLEAR GUIDE" stacked on top of each other in bold black letters. Below "GUIDE" is the word "MEDICAL" in gray letters. ## 510(k) Summary (per 21 CFR § 807.92) ## Submitter's Information | Name | Clear Guide Medical, Inc. | |----------------|---------------------------------------------------------| | Address | 3600 Clipper Mill Rd., Suite 400<br>Baltimore, MD 21211 | | Phone Number | (410) 504-6540 | | Contact Person | Dorothee Heisenberg, Executive Vice President | | Date Prepared | November 1, 2017 | ## Device Information | Trade Name | Clear Guide SCENERGY | |----------------|--------------------------------------------------------------------------| | Common Name | System, Imaging Software | | Classification | Computed Tomography X-Ray System<br>21 CFR § 892.1750 (Product Code JAK) | ## Predicate Device Information | Device Name | Clear Guide SCENERGY | |---------------|----------------------| | 510(k) Number | K153004 | ## Device Description The Clear Guide SCENERGY guidance system is intended to be an accessory to existing ultrasound imaging systems, to provide image fusion, instrument tracking, and image/instrument guidance functionality to operators during image-guided medical interventions that utilize data from multiple modalities (e.g., ultrasound and computed-tomography (CT), ultrasound and magnetic resonance (MR)). The Clear Guide SCENERGY uses optical detection technology to identify and track objects in the field of view. By pairing this information with the aforementioned imaging data, the Clear Guide SCENERGY executes proprietary software algorithms to display fused images in real-time to the clinician. These segmentation and registration algorithms are automated. Segmentation results are deterministic, meaning that new inputs (e.g., new CT or MR) would be required to change the segmentation output. Registration can be reset by the user at any time during use. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Clear Guide Medical. The logo consists of a stylized letter "L" on the left and the words "CLEAR GUIDE MEDICAL" on the right. The words "CLEAR GUIDE" are in bold black letters, and the word "MEDICAL" is in gray letters. ## Intended Use The device is a stereotaxic accessory intended to provide fusion of images from certain imaging modalities. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting. #### Indications for Use The Clear Guide SCENERGY is software that provides fusion of images from Computed Tomography (CT), Magnetic Resonance (MR), and Ultrasound (US) modalities. US images can be fused with either CT or MR. The Clear Guide SCENERGY utilizes the Clear Guide CORE and Clear Guide SuperPROBE platform to display images of the target regions and the projected path of the interventional instrument, while taking into account patient movement and deformation. Instrumentation used with the Clear Guide SCENERGY might include an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle, or an ablation needle. The device is intended to be used in any interventional or diagnostic procedure where the combination of these modalities is used for visualization, except for procedures on the brain. The device is intended for use in a clinical setting. - Comparison to Predicate: The indications for use statement for the Clear Guide SCENERGY differs from the predicate device to account for feature-level differences and to provide additional clarity. These changes do not represent a difference in the overall intended use of the product compared to the predicate device. #### Technological Characteristics The Clear Guide SCENERGY operates using optical detection technology. The modified Clear Guide SCENERGY enables fusion with two difference image volumes: CT and MRI. Fusion with ultrasound is available with both types, versus just CT for the predicate device. Optical detection technology does not require specialized instruments or calibration at the point of use. As with the predicate device, the Clear Guide SCENERGY overlays instrument positioning data onto an existing ultrasound image through proprietary software algorithms. For certain software algorithms (specifically segmentation and registration), these functions are automated. A comparison of Clear Guide SCENERGY's technological characteristics to its predicate device is provided below. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Clear Guide Medical. The logo consists of a stylized "L" shape on the left, with two horizontal lines extending from it. To the right of the "L" shape, the words "CLEAR GUIDE" are stacked on top of each other in bold, black letters, and below that, the word "MEDICAL" is written in gray. | Category | Modified Clear Guide SCENERGY | Original Clear Guide SCENERGY | |-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name (Full) | Clear Guide SCENERGY | Clear Guide SCENERGY | | 510(k) Number | [not assigned] | K153004 | | Product Code(s) | JAK | JAK | | Classification Regulation | 21 CFR 892.1750 | 21 CFR 892.1750 | | Regulatory Class | II | II | | Intended Use | Image Fusion<br>Instrument Tracking and Guidance | Image Fusion<br>Instrument Tracking and Guidance | | Indications for Use | The Clear Guide SCENERGY is software<br>that provides fusion of images from<br>Computed Tomography (CT), Magnetic<br>Resonance (MR), and Ultrasound (US)<br>modalities. US images can be fused with<br>either CT or MR.<br><br>The Clear Guide SCENERGY utilizes the<br>Clear Guide CORE and Clear Guide<br>SuperPROBE platform to display images<br>of the target regions and the projected<br>path of the interventional instrument,<br>while taking into account patient<br>movement and deformation.<br>Instrumentation used with the Clear<br>Guide SCENERGY might include an<br>interventional needle or needle-like rigid<br>device, such as a biopsy needle, an<br>aspiration needle, or an ablation needle.<br>The device is intended to be used in any<br>interventional or diagnostic procedure<br>where the combination of these<br>modalities is used for visualization,<br>except for procedures on the brain. The<br>device is intended for use in a clinical<br>setting. | The Clear Guide SCENERGY is a<br>stereotaxic accessory for the fusion of<br>images from multiple modalities -<br>including Computed Tomography (CT)<br>and Ultrasound (US).<br><br>The Clear Guide SCENERGY utilizes the<br>Clear Guide CORE and Clear Guide<br>SuperPROBE platform to display images<br>of the target regions and the projected<br>path of the interventional instrument,<br>while taking into account patient<br>movement and deformation. Imaging<br>modalities may be fused in various<br>combinations. Instrumentation used<br>with the Clear Guide SCENERGY might<br>include an interventional needle or<br>needle-like rigid device, such as a biopsy<br>needle, an aspiration needle, or an<br>ablation needle. The device is intended<br>to be used in any interventional or<br>diagnostic procedure where the<br>combination of these modalities is used<br>for visualization. The device is intended<br>for use in a clinical setting. | | Fundamental Technology<br>(i.e., Mechanism of Action) | Optical Detection & Software | Optical Detection & Software | | Hardware Components | Clear Guide CORE and Optical Head | Clear Guide CORE and Optical Head | | Operating Principle | Optical Detection | Optical Detection | | Conditions of Use | Optical Detection | Optical Detection | | Fusion Capabilities | US-CT & US-MR | US-CT | | Signal Receiver | Connected/Anchored to Transducer | Connected/Anchored to Transducer | | Segmentation Process | Automatic | Automatic | | Registration Process | Automatic | Automatic | | Category | Modified Clear Guide SCENERGY | Original Clear Guide SCENERGY | | Fusion Algorithms | Automatic | Automatic | | Tracking/Guidance Algorithms | Automatic | Automatic | | Intended User | Physician | Physician | | Environment of Use | Clinical Setting | Clinical Setting | | Duration of Use | ≤ 24 Hours | ≤ 24 Hours | | Number of Uses | Reusable | Reusable | | Sterility | Non-Sterile | Non-Sterile | | Bench Testing | Yes (V&V on modification) | Yes | | Animal Testing | N/A | Yes | | Clinical Testing | N/A | Yes | | Standards | N/A (for software) | N/A (for software) | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for Clear Guide Medical. On the left side of the logo is a stylized letter "L" with two horizontal lines running through it. To the right of the "L" is the text "CLEAR GUIDE MEDICAL" in a stacked format, with "CLEAR GUIDE" in black and "MEDICAL" in gray. Performance data was collected to demonstrate that the Clear Guide SCENERGY achieves its intended function in a manner that is as safe and as effective as the predicate device. ## Performance Data Performance testing of the Clear Guide SCENERGY device demonstrates that the product accurately achieves its intended use. while also showing that differences in technological features from the predicate device did not affect device performance. Results of performance testing show that the subject device is as safe and as effective as the predicate device. ## Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for the Clear Guide SCENERGY was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator, or indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Clear Guide Medical. The logo consists of a stylized letter "L" on the left and the words "CLEAR GUIDE" stacked on top of each other on the right. The word "MEDICAL" is located below the words "CLEAR GUIDE" and is a different color than the other words. ## Segmentation Testing The automatic segmentation of fiducial markers by the Clear Guide SCENERGY was evaluated by comparing software outputs to manual selection in phantom datasets. ## Fusion Testing The Clear Guide SCENERGY's ability to provide fused images from ultrasound and CT/MR modalities was evaluated using fusion testing. Fusion quality was assessed by taking distance measurements between identifiable landmarks or surfaces seen on ultrasound and CT, or ultrasound and MR. This metric is known as Tissue Registration Error (TRE). ## Systematic Error (Tip-to-Tip) Testing Systematic error in the Clear Guide SCENERGY is defined as the cumulative error that would be observed within the entire system, which includes segmentation, registration, fusion, and guidance errors. This performance metric is a "tip-to-tip" distance from the needle point seen by ground truth CT (or MR) to the same needle point seen by Clear Guide SCENERGY's displayed guidance. This metric has been called "tracking error" in literature for the predicate device. Phantom datasets were utilized for this evaluation. ## Conclusions The Clear Guide SCENERGY has the same intended use as the predicate device. Changes to the specific indications for use statement were made to account for technological differences, without impacting intended use. Differences in the product's feature offerings do not raise different questions of safety or effectiveness. Additionally, performance tests confirm that the Clear Guide SCENERGY is as safe and as effective as the predicate device. Therefore, the Clear Guide SCENERGY is substantially equivalent to its predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%