DIXI Medical Microdeep Micro-Macro Depth Electrodes

K202087 · Dixi Medical · GZL · Nov 18, 2021 · Neurology

Device Facts

Record IDK202087
Device NameDIXI Medical Microdeep Micro-Macro Depth Electrodes
ApplicantDixi Medical
Product CodeGZL · Neurology
Decision DateNov 18, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1330
Device ClassClass 2

Indications for Use

The DIXI Medical Microdeep® Micro-Macro Depth Electrodes are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping.

Device Story

Invasive depth electrode for temporary (<30 days) intracranial use; records, monitors, and stimulates brain electrical signals. Input: neural electrical signals; Output: electrical signals transmitted to external recording/monitoring equipment. Used in intraoperative and neurological monitoring settings by physicians trained in intracranial neurophysiology. Device features macro-contacts on the exterior and micro-contacts flush at the surface; includes a microdrive for micro-contact adjustment. Output facilitates identification of epileptogenic foci and brain mapping, aiding clinical decision-making for epilepsy treatment. Benefits include precise subsurface neural signal acquisition.

Clinical Evidence

Bench testing only. Testing included dimensional verification, mechanical functionality (microdrive, water tightness, tensile strength), electrical characteristics (continuity, resistance, stimulation integrity), and biocompatibility per ISO 10993. Comparative testing against the predicate confirmed equivalent tensile strength and mechanical functionality after simulated use.

Technological Characteristics

Single-use, sterile, ethylene oxide-sterilized depth electrode. Macro-contacts: Platinum/Iridium 90/10; Micro-contacts: Tungsten. Anchor bolt: Titanium alloy Ti6Al4V (ASTM F136). One-piece design with microdrive. Dimensions: 0.8 mm macro-contact diameter, 20-micron micro-contact diameter. Connectivity: connects to external recording/stimulation equipment via connection system. No software/algorithm.

Indications for Use

Indicated for patients requiring temporary (<30 days) intracranial recording, monitoring, and stimulation of electrical signals to define epileptogenic foci and perform brain mapping. Used by physicians trained in intracranial neurophysiology.

Regulatory Classification

Identification

A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ November 18, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Steven Plymale CEO Dixi Neurolab, Inc. 145 Howland Pines Drive Oxford, MI 48371 Re: K202087 Trade/Device Name: DIXI Medical Microdeep Micro-Macro Depth Electrodes Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: October 20, 2021 Received: October 22, 2021 Dear Steven Plymale: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {2}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202087 Device Name DIXI Medical Microdeep® Micro-Macro Depth Electrodes Indications for Use (Describe) The DIXI Medical Microdeep® Micro-Macro Depth Electrodes are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping. Type of Use (Select one or both, as applicable) | <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | |-----------------------------------------------------------------------------------------------| | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue rectangle on the left, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in a smaller, light gray font. The logo is simple and modern, with a focus on the company name. # 510(k) Summary #### I. SUBMITTER 510(k) Owner DIXI Medical 2A route de Pouligney 25640 Chaudefontaine, France Office: 033-3-81889890 Fax: 033-3-81889899 info@diximedical.com Official correspondent: Lucie Petegnief R&D Manager l.petegnief@diximedical.com Submission Correspondent Name: Steve Plymale CEO Dixi Neurolab Inc. 145 Howland Pines Drive Oxford, MI 48371 Email: steven.plymale@dixineurolab.com Phone: 514 882 3258 Date Prepared Oct. 20, 2021 > II. DEVICE | Name of Device: | DIXI Medical Microdeep® Micro-Macro Depth Electrode | |-----------------------|-----------------------------------------------------| | Common or Usual Name: | Depth Electrode | | Classification Name: | 21 CFR §882.1330 Depth Electrode | | Regulatory Class: | II | | Product Code: | GZL | #### III. PREDICATE DEVICE The predicate device is the Ad-Tech Medical Instrument Corporation's Macro Micro Depth Electrodes (K163355). The DIXI Medical Microdeep® Depth Electrode (K170959) was a reference device for this submission. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue rectangle on the left, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. The logo is simple and modern, with a focus on the company name. ### IV. DEVICE DESCRIPTION The Microdeep® Micro-Macro Depth Electrode is a single patient use, sterile and disposable device. The Microdeep® Micro-Macro Depth Electrode is invasive as it is placed in contact with nerve tissue (brain) and must only be used during an SEEG procedure with Anchor Bolts. The Microdeen® Micro-Macro Depth Electrode is intended to connect to the user's recording. monitoring and stimulation equipment using the Connection System. This product is intended to be used only by physicians in the area of biopotential recording, monitoring and stimulation / response studies who are trained in intracranial neurophysiology. The DIXI Medical Microdeep® Micro-Macro Depth Electrode is comprised of the following components: - Macro-contacts - located on the outside of the device. - . Micro-contacts - sit flush at the surface of the device between the macro electrodes. - Microdrive for micro-contacts ● - . Cap - Connectors for micro and macro-contacts ● #### V. INDICATIONS FOR USE The DIXI Medical Microdeep® Micro-Macro Depth Electrodes are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping. {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square on the left, followed by the word "DIXI" in black, bold letters. Below the word "DIXI" is the word "medical" in gray, smaller letters. The logo is simple and modern, with a focus on the company name. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI. The technological characteristics comparing to the predicate device are outlined in the Substantial Equivalent table. | Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) | |--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulation Number | 21 CFR §882.1330 | 21 CFR §882.1330 | 21 CFR §882.1330 | | Regulation Name | Depth electrode | Depth electrode | Depth electrode | | Regulatory Class | Class II | Class II | Class II | | Product Code | GZL | GZL | GZL | | Indications for Use | The DIXI Medical Microdeep®<br>Micro-Macro Depth<br>Electrodes are intended for<br>temporary (<30 days) use<br>with recording, monitoring<br>and stimulation equipment<br>for the recording, monitoring<br>and stimulation of electrical<br>signals at the subsurface level<br>of the brain. The recording of<br>electrical activity supports<br>definition of the location of<br>epileptogenic foci and brain<br>mapping. | The AD-TECH Depth<br>Electrodes (Depth Electrodes,<br>Foramen Ovale Depth<br>Electrodes, Macro Micro<br>Depth Electrodes, Spencer<br>Probe Depth Electrodes,<br>Wyler Sphenoidal Depth<br>Electrodes) are intended for<br>temporary (< 30 days) use<br>with recording, monitoring<br>and stimulation equipment for<br>the recording, monitoring and<br>stimulation of electrical<br>signals at the subsurface level<br>of the brain. The recording of<br>electrical activity supports<br>definition of the location of | The DIXI Medical Microdeep<br>Depth Electrodes are intended<br>for temporary (<30 days) use<br>with recording, monitoring<br>and stimulation equipment for<br>the recording, monitoring and<br>stimulation of electrical signals<br>at the subsurface level of the<br>brain. | | Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) | | Environment of Use | Intraoperative and<br>Neurological monitoring<br>locations | Intraoperative and<br>Neurological monitoring<br>locations<br>epileptogenic foci and brain<br>mapping. | Intraoperative and<br>Neurological monitoring<br>locations | | Provided Sterile | Yes | Yes | Yes | | Method of sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide | | Single Use, disposable | Yes | Yes | Yes | | Duration of Use | < 30 days | < 30 days | < 30 days | | Maximum Stimulation<br>Charge Density | $\leq 30 \mu C/cm^2$ | $\leq 30 \mu C/cm^2$ | $\leq 30 \mu C/cm^2$ | | Electrode Characteristics | | | | | General characteristic | One piece design: depth<br>electrode that has electrical<br>macro-contacts collars placed<br>on the outside and micro-<br>contacts cut flush at the<br>surface, between macro-<br>contacts, which can be<br>released | Two kinds of Macro Micro<br>Depth Electrodes:<br>1. BEHNKE FRIED DEPTH<br>ELECTRODES, two piece<br>design:<br>- Macro portion: depth<br>electrode that has electrical<br>macro-contacts collars placed<br>on the outside and an inner<br>lumen throughout the entire<br>length<br>- Micro portion: wire bundle | One piece design: depth<br>electrode that has electrical<br>macro-contacts collars placed<br>on the outside, no micro-<br>contacts | | Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) | | | | (micro-contacts) that passes<br>through the inner lumen of the<br>macro-electrode<br><br>2. MACRO-MICRO DEPΤΗ<br>ELECTRODES: one piece<br>design: depth electrode that<br>has electrical macro-contacts<br>collars placed on the outside<br>and micro-contacts cut flush<br>at the surface, between macro-<br>contacts | | | Removable stylet in the<br>electrode | No | Yes | No | | Contacts Material | Platinum/Iridium 90/10<br>(Macro-contact)<br>Tungsten (Micro-contact) | Platinum (Macro-contact)<br>Platinum/Iridium (Micro-<br>contact) | Platinum/Iridium 90/10<br>(Macro-contact) | | Electrode body diameter<br>(brain contact) | 0.8 mm (Macro-contact)<br>20 microns (Micro-contact) | 1.3 mm (Macro-contact)<br>38 to 51 microns (Micro-<br>contact) | 0.8 mm (Macro-contact) | | Number of electrode<br>contacts | From 6 to 9 (Macro-contact)<br>From 8 to 12 (Micro-contact) | From 6 to 12 (Macro-contact)<br>From 8 to 24 (Micro-contact) | From 5 to 18 (Macro-contact) | | Electrode contact length<br>(along body of the<br>electrode) | 2 mm (Macro-contact) | 1.57 mm (Macro-contact) | 2 mm (Macro-contact) | | Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes | DIXI Medical<br>Microdeep Depth Electrode | | | (Subject Device) | (Predicate device K163355) | (Reference Device K170959) | | Overall length | ~ 400 mm | ≤ 660 mm | ≤ 1050 mm | | Device Accessories - Anchor Bolt | | | | | Product Code | GZL | GZL | GZL | | Regulation Number | 21 CFR §882.1330 | 21 CFR §882.1330 | 21 CFR §882.1330 | | Regulatory Class | II | II | II | | Material | Titanium alloy Ti6Al4V (ELI,<br>ASTM F136) | Titanium<br>Silicone (inner lumen gasket)<br>Parylene | Titanium alloy Ti6Al4V (ELI,<br>ASTM F136) | | Length | From 15 mm to 35 mm | From 13 mm to 26 mm | From 15 mm to 35 mm | | Required Drill Hole<br>Diameter | 2.1 mm | 2.4 mm or 2.8 mm | 2.1 mm | | Compatible Depth<br>Electrode Body Diameter | 0.8 mm | 0.86 mm to 1.3 mm | 0.8 mm | | Attachment onto the skull | Threaded into a pre-drilled<br>hole in the skull | Threaded into a pre-drilled<br>hole in the skull | Threaded into a pre-drilled<br>hole in the skull | | Anchor bolt placement | Long Screwdriver | Anchor bolt placement<br>wrench | Long Screwdriver | | Anchor bolt removal | Wrench and Short<br>Screwdriver | Anchor bolt removal wrench | Wrench and Short Screwdriver | | Implantation duration | Up to 30 days | Up to 30 days | Up to 30 days | | Single use | Yes | Yes | Yes | | Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) | | Method of sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide | | Device Accessories | | | | | List of Accessories | Instrument guide, Bone<br>starter, Sliding Ruler,<br>Coagulation Electrode and<br>Stop, Flat Screwdriver (for<br>stop), Drill and Stop, Long<br>Screwdriver, Anchor Bolt,<br>Marking Rod, Cap for Anchor<br>Bolt, Depth Report device,<br>Stylet, Stop, Sheath Holder,<br>Extension cable, Adapter and<br>Connection cable, and Wrench<br>and Short Screwdriver for<br>Anchor Bolt. | Guide, Channeled Ruler, Drill<br>Bit with Stop and Wrench,<br>Anchor placement/removal<br>Wrench, Anchor Bolt,<br>Obturator and Cabrio and<br>Tech-Attach Connection<br>Systems | Instrument guide, Bone<br>starter, Sliding Ruler,<br>Coagulation Electrode and<br>Stop, Flat Screwdriver (for<br>stop), Drill and Stop, Long<br>Screwdriver, Anchor Bolt,<br>Marking Rod, Cap for Anchor<br>Bolt, Depth Report device,<br>Stylet, Stop, Extension cable,<br>Adapter and Connection cable,<br>and Wrench and Short<br>Screwdriver for Anchor Bolt. | | Single-Use Accessories | Coagulation Electrode and<br>Stop, Drill and Stop, Anchor<br>Bolt, Cap for Anchor Bolt,<br>Stylet, Sheath Holder | Drill Bit with Stop and<br>Wrench, Anchor Bolt and<br>Obturator | Coagulation Electrode and<br>Stop, Drill and Stop, Anchor<br>Bolt, Cap for Anchor Bolt,<br>Stylet | | Method of Sterilization<br>for Single Use Accessories | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | | Reusable Accessories | Instrument guide, Bone<br>starter, Sliding Ruler, Flat<br>Screwdriver (for stop), Long<br>Screwdriver, Marking Rod | Guide, Channeled Ruler,<br>Anchor placement/removal<br>Wrench, and Cabrio and Tech-<br>Attach Connection Systems | Instrument guide, Bone<br>starter, Sliding Ruler, Flat<br>Screwdriver (for stop), Long<br>Screwdriver, Marking Rod | | Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) | | | Depth Report Device, Stop,<br>Extension Cable, Adapter,<br>Connection Cable, Wrench<br>and Short Screwdriver for<br>Anchor Bolt | | Depth Report Device, Stop,<br>Extension Cable, Adapter,<br>Connection Cable, Wrench and<br>Short Screwdriver for Anchor<br>Bolt | | Method of Sterilization<br>for Reusable Accessories | Steam Sterilization (with the<br>exception of the Adapter,<br>which does not require<br>sterilization) | Unknown (except for Cabrio<br>and Tech-Attach Connection<br>Systems: Ethylene Oxide or<br>Sterrad) | Steam Sterilization (with the<br>exception of the Adapter,<br>which does not require<br>sterilization) | {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for DIXI medical. The word "DIXI" is in bold, black letters, and the word "medical" is in gray letters below it. To the left of the word "DIXI" is a blue square. {9}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. {10}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. {11}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. {12}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square on the left, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. The logo is simple and modern, with a clear emphasis on the company name. #### VII. NON-CLINICAL TESTING DIXI Medical has performed the following non-clinical laboratory testing to determine substantial equivalence. | Test | Summary of Verifications | Result | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Dimensional<br>Characteristics | Verification of electrodes dimensional<br>characteristics | Pass | | | | | | Mechanical characteristics | - Verification of electrode microdrive<br>functioning | Pass | | | - Verification of electrode water tightness | | | | - Verification of cap maintenance on the<br>electrode sheath | | | | - Verification of the antirotation function of<br>electrode cap | | | | - Verification of tensile strength of electrode<br>macro-contacts | | | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...