Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)
K201931 · Dixi Medical · GYC · Oct 2, 2021 · Neurology
Device Facts
Record ID
K201931
Device Name
Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)
Applicant
Dixi Medical
Product Code
GYC · Neurology
Decision Date
Oct 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1310
Device Class
Class 2
Indications for Use
The DIXI Medical Intraoperative Subdural Electrodes (Strips and Grids) are intended for intraoperative use for less than or equal to 24 hours with recording and stimulation equipment for the recording and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports brain mapping.
Device Story
Intra-cranial electrodes used intraoperatively on brain surface; designed for EEG recording and brief stimulation for brain mapping. Device consists of circular stainless steel contacts sandwiched between two layers of silicone substrate; insulated wires extend from contacts through flexible tube to connectors for equipment interface. Used in OR by neurosurgeons/clinicians. Output is electrical signal data transmitted to external recording/stimulation equipment; clinicians interpret signals to map brain function. Benefits include localized brain activity monitoring during surgery.
Clinical Evidence
Bench testing only. Included dimensional verification, electrical characteristics (stimulation signal transmission, resistance, dielectric strength, corrosion stability), and mechanical testing (torsion, bending, traction). Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, intracutaneous reactivity, pyrogenicity, hemolysis, and acute systemic toxicity. All tests passed.
Indicated for intraoperative use (≤24 hours) in patients requiring recording and stimulation of electrical signals on the brain surface to support brain mapping.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
Predicate Devices
Ad-Tech Medical Instrument Corporation's Subdural Cortical Electrodes (K191186)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2021
DIXI Medical % Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough. California 94010
Re: K201931
Trade/Device Name: DIXI Medical Intraoperative Subdural Electrodes (Strips and Grids) Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: Class II Product Code: GYC Dated: September 1, 2021 Received: September 2, 2021
## Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201931
### Device Name
DIXI Medical Intraoperative Subdural Electrodes (Strips and Grids)
Indications for Use (Describe)
The DIXI Medical Intraoperative Subdural Electrodes (Strips and Grids) are intended for intraoperative use for less than or equal to 24 hours with recording and stimulation equipment for the recording and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports brain mapping.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☑</span> Prescription Use (Part 21 CFR 201 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;"> ☐ Over-The-Counter Use (21 CFR 201) </span> </div> |
| X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### I. SUBMITTER
510(k) Owner DIXI Medical 2A route de Poulignev 25640 Chaudefontaine, France Office: 033-3-81889890 Fax: 033-3-81889899 info@diximedical.com
Official Correspondent: Lucie Petegnief R&D Manager l.petegnief@diximedical.com
Submission Correspondent Cindy Domecus, R.A.C. (US & EU) Domecus Consulting Services LLC Phone: (650) 343-4813 Fax: (650) 343-7822 Email: Cindy@DomecusConsulting.com
Date Prepared September 24, 2021
II. DEVICE
Name of Device: DIXI Medical Intraoperative Subdural Electrodes (Strips and Grids) Common or Usual Name: Intraoperative Subdural Electrode Classification Name: 21 CFR §882.1310 Cortical Electrode Regulatory Class: II Product Code: GYC
III. PREDICATE DEVICE
The predicate device is the Ad-Tech Medical Instrument Corporation's Subdural Cortical Electrodes (K191186). There have been no recalls for the predicate device.
DEVICE DESCRIPTION IV.
The DIXI Medical Intraoperative Subdural Electrode (Strips and Grids) is an intra-cranial electrode used intraoperatively on the surface of the brain. The device is designed for
{4}------------------------------------------------
electroencephalography (EEG) recording and brief stimulation for brain mapping purposes.
The DIXI Medical Intraoperative Subdural Electrode consists of circular contacts sandwiched between two layers of silicone substrate. The brain contacting side of the silicone substrate body has material removed to expose an amount of contact surface area. Insulated wires extend from each contact through a flexible tube which terminates in connectors for direct connection to user's equipment.
| Product<br>Reference | Configuration | Schematic drawing |
|----------------------|-----------------------------------------------------------------|----------------------------------------------------------------------|
| C10-08AIOM | 1 strip x 8<br>contacts<br>Contacts<br>numbered from 1<br>to 8 | Image: Schematic drawing of a connector with 1 strip of 8 contacts. |
| C10-16AIOM | 2 strip x 8<br>contacts<br>Contacts<br>numbered from 1<br>to 16 | Image: Schematic drawing of a connector with 2 strips of 8 contacts. |
| C10-06BIOM | 1 strip x 6<br>contacts<br>Contacts<br>numbered from 1<br>to 6 | Image: Schematic drawing of a connector with 1 strip of 6 contacts. |
| C10-12BIOM | 2 strip x 6<br>contacts<br>Contacts<br>numbered from 1<br>to 12 | Image: Schematic drawing of a connector with 2 strips of 6 contacts. |
| C10-04CIOM | 1 strip x 4<br>contacts<br>Contacts<br>numbered from 1<br>to 4 | Image: Schematic drawing of a connector with 1 strip of 4 contacts. |
A summary of the subject device configurations is provided below.
{5}------------------------------------------------
| Product<br>Reference | Configuration | Schematic drawing |
|----------------------|-----------------------------------------------------------------|----------------------------------------|
| C10-08CIOM | 2 strip x 4<br>contacts<br>Contacts<br>numbered from 1<br>to 8 | Image: Schematic drawing of C10-08CIOM |
| C10-16CIOM | 4 strip x 4<br>contacts<br>Contacts<br>numbered from 1<br>to 16 | Image: Schematic drawing of C10-16CIOM |
#### INDICATIONS FOR USE V.
The Indications for Use for the subject and predicate devices are noted below.
| DIXI Medical Intraoperative Subdural<br>Electrodes<br>SUBJECT DEVICE | AD-TECH Subdural Cortical Electrodes<br>PREDICATE DEVICE, K191186 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The DIXI Medical Intraoperative Subdural<br>Electrodes (Strips and Grids) are<br>intended for intraoperative use for less<br>than or equal to 24 hours with recording<br>and stimulation equipment for the<br>recording and stimulation of electrical<br>signals on the surface level of the brain.<br>The recording of electrical activity<br>supports brain mapping. | The Ad-Tech Subdural Electrodes<br>(Strip/Intraoperative Strip,<br>Grid/Intraoperative Grid, Dual-Sided<br>Interhemispheric, Multi-Strip and Split Grid,<br>Intraoperative Mapping Grid) are intended<br>for temporary (< 30 days) use with<br>recording, monitoring, and stimulation<br>equipment for the recording, monitoring,<br>and stimulation of electrical signals on the<br>surface of the brain. The recording of<br>electrical activity supports definition of the<br>location of epileptogenic foci and brain<br>mapping. |
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The predicate device clearance is for a family of electrodes, whereas the subject device is for a single electrode type. As such, the below comparisons are between the subject device and the corresponding electrode type within the predicate device family of electrodes (i.e., the Intraoperative Strip and Intraoperative Grid).
{6}------------------------------------------------
Both the subject and predicate devices are intended for use with recording and stimulation equipment for the recording and stimulation of electrical signals at the surface level of the brain. At a high level, the subject and predicate devices are based on the following same technological elements:
- Both the subject and predicate device are subdural cortical electrodes for the same . duration of use in the same environment of use.
- . Both the subject and predicate device are provided sterile and for single-use only and use the same method of sterilization.
- . Both the subject and predicate device have a maximum stimulation charge density of < 30 µC/cm².
- . The subject and predicate device have the same contact spacing, maximum number of contacts per electrode cable and diameter of electrode cable.
The following technological differences exist between the subject and the Intraoperative Strip and Intraoperative Grid predicate devices:
- The number of contacts per electrode is similar. The number of contacts for strips is . the same. The maximum number of contacts for predicate Intraoperative Grid is unknown but, in any case, the number of contacts for the subject device is within the range of the number of contacts for the predicate device.
- . The contact materials are similar but not identical; the subject device is stainless steel only, whereas the predicate device includes 90:10 Platimum:Iridium not present in the subject device.
- . The contact external diameter is similar but not identical; the subject device is 6.0mm-diameter only. The contact exposure diameter is similar but not identical. The subject device exposure diameter is within the range of exposure diameter for the predicate device.
- . Both the subject and predicate device cabling terminate in a 1.5mm female socket, but there is the option with the predicate device to connect to a patient cable.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Nonclinical Testing
The nonclinical testing included in the table below was conducted on subject devices, on baseline and aged devices. Aged devices underwent three sterilization cycles, accelerated aging equivalent to real-time ageing of 5 years (ASTM F1980), extreme climatic conditions (ASTM D4332) and tests of transport (ASTM D4169).
| Test<br>Performed | Tested<br>Device<br>Reference | Objective | Conclusion |
|--------------------------------|-------------------------------|-----------------------------------------------------------|------------|
| Dimensional<br>Characteristics | C10-04CIOM<br>C10-08AIOM | Verification of electrodes<br>dimensional characteristics | Pass |
{7}------------------------------------------------
| Test<br>Performed | Tested<br>Device<br>Reference | Objective | Conclusion |
|-------------------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------|------------|
| Electrical<br>Characteristics | C10-04CIOM<br>C10-08AIOM | Verification of the electrodes ability<br>to transmit a stimulation signal | Pass |
| | | Verification of the electrodes<br>resistance to a stimulation signal<br>under worst case charge density<br>condition | Pass |
| | | Verification of the electrodes<br>resistance to a stimulation signal<br>under worst case current condition | Pass |
| | | Verification of the electrodes<br>dielectric strength | Pass |
| | | Verification of electrodes stability<br>under conditions of use (absence of<br>corrosion) | Pass |
| Mechanical<br>Characteristics | C10-04CIOM<br>C10-08AIOM | Verification of the resistance to<br>torsion of the active part and silicon<br>sheath | Pass |
| | | Verification of electrodes resistance<br>to bending | Pass |
| | | Verification of electrodes resistance<br>to traction | Pass |
# Biocompatibility Testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA's September 2020 guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff. The biocompatibility evaluation included the following tests and results:
{8}------------------------------------------------
| Test | Tested<br>Device<br>Reference | Result | Conclusion |
|------------------------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Cytotoxicity | C10-08AIOM | Test 1: Percent viability of test article<br>was 93.9% of reagent control.<br>Test 1: Percent viability of test article<br>was 98.6% of reagent control.<br>Test 3: Percent viability of test article<br>was 88.5% of reagent control. | Non-cytotoxic |
| Sensitization | C10-16CIOM | Topical application of the 0.9%<br>sodium chloride extract and the<br>sesame oil extract did not induce<br>delayed sensitization in the guinea<br>pig. | Non-sensitizer |
| Intracutaneous<br>reactivity | C10-16CIOM | Difference between each test extract<br>overall mean score and<br>corresponding control blank overall<br>mean score was lower than 1.0 (0.0<br>for the 0.9% sodium chloride extract<br>and sesame oil test extracts). | Non-irritant |
| Pyrogenicity | C10-16CIOM | No rabbit showed an individual<br>temperature rise higher or equal to<br>0.5℃ above its initial temperature. | Non-<br>pyrogenic |
| Indirect<br>(extract)<br>hemolysis | C10-16CIOM | Mean hemolytic index for test article<br>extract was of 0.0%. | Non-<br>hemolytic |
| Acute Systemic<br>Toxicity | C10-16AIOM | No evidence of significant systemic<br>toxicity or mortality after test article<br>extracts injection. | Non-toxic |
The contact classification for the subject device is an external communicating device with tissue/bone contact and cerebrospinal fluid (CSF) contact for a limited (≤24 hours) duration.
# Sterilization and Shelf-Life Testing
The subject device is sterilized using Ethylene Oxide to a SAL = 10-6. A shelf-life of 5 years has been established based on accelerated and real-time aging.
#### CONCLUSIONS VIII.
The Indications for Use for the subject device is a subset of the Indications for Use for the predicate device, and the differences in technological characteristics do not raise new or different questions of safety and effectiveness, with the non-clinical and biocompatibility testing performed demonstrating the continued effectiveness of the subject device as compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.