The prostheses from FH Industrie are designed for specific indications such as: SIMPLE HUMERAL PROSTHESIS - Humeral head necrosis without injury to the glenoid cavity. - Extensive humeral head cartilage damage without injury to the glenoid cavity - Centred osteoarthritis with a glenoid cavity not allowing implantation of a glenoid implant. - Rheumatoid polyarthritis with thin rotator cuff. - Off-centred osteoarthritis with irreparable cuff, and with maintained active elevation of at least 120°. TOTAL ANATOMICAL PROSTHESIS (CEMENTED GLENOID IMPLANT WITH 4 PEGS) - Centred glenohumeral osteoarthritis with functional rotator cuff - Rheumatoid polyarthritis with functional rotator cuff - Post-traumatic sequela, functional rotator cuff with glenoid injury. TOTAL ANATOMICAL PROSTHESIS (POROUS GLENOID IMPLANT) - Centred glenohumeral osteoarthritis - Rheumatoid polyarthritis - Post-traumatic sequela with glenoid injury - Revision for glenoid loosening - Glenoid bone loss, where bone graft is needed - A functional rotator cuff is necessary to use this device REVERSE PROSTHESIS (METAL-BACK OR POROUS GLENOID IMPLANT) The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. For all types of prosthesis, the glenoid baseplate (metal-back or porous) is intended for cementless use with the addition of bone screws for fixation, the humeral short stem (metaphyseal stem) is intended for cementless use. At least 2/3 of the metaphyseal component must be implanted in the proximal humeral bone to allow for adequate humeral component fixation.
Device Story
The ARROW Short Stem Humeral System is a modular orthopedic implant for shoulder arthroplasty. It consists of a metaphyseal component and a cylindrical diaphyseal stem, designed to replace the humeral head in anatomical or reverse shoulder configurations. The device is implanted by orthopedic surgeons in a clinical/surgical setting. The metaphyseal component features fins for rotational stability and is intended for cementless fixation in the proximal humerus. The system provides surgeons with modular sizing options to better match patient anatomy. By restoring joint structure and function, the device aims to alleviate pain and improve mobility in patients with severe shoulder arthropathy or joint damage. The system is used in conjunction with various glenoid components (cemented or porous) depending on the specific clinical indication.
Clinical Evidence
Bench testing only. Mechanical testing included connection strength (ASTM F2009), fatigue testing (ASTM F2580), range of motion (ASTM F1378), corrosion testing, and torsional resistance. Biocompatibility (ISO 10993-1), sterilization (ISO 11137), and packaging (ISO 11607) validations were performed.
Technological Characteristics
Modular humeral system manufactured from titanium alloy (Ti6Al4V-ELI) per ASTM F-136/ISO 5832-3. Metaphysis features pure titanium plasma spray coating per ASTM 1580. Designed for cementless fixation. No software or electronic components.
Indications for Use
Indicated for patients requiring shoulder arthroplasty due to humeral head necrosis, osteoarthritis, rheumatoid polyarthritis, post-traumatic sequelae, or failed prior replacement. Requires functional rotator cuff for anatomical configurations; requires functional deltoid and adequate glenoid bone stock for reverse configurations.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Fournitures Hospitalieres Industrie: ARROW Glenoid base porous (K162068)
Fournitures Hospitalieres Industrie: ARROW Glenoid base porous (K171789)
Submission Summary (Full Text)
{0}------------------------------------------------
November 30, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
FH Industrie % Christine Scifert Partner MRC Global, LLC 9085 E. Mineral Cir., Suite 110 Centennial, Colorado 80112
Re: K202024
Trade/Device Name: ARROW Short Stem Humeral System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, HSD, KWS Dated: October 27, 2021 Received: October 28, 2021
Dear Christine Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Jiping Chen, Ph.D., M.P.H. Acting Division Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202024
Device Name ARROW Short Stem Humeral System
#### Indications for Use (Describe)
The prostheses from FH Industrie are designed for specific indications such as:
SIMPLE HUMERAL PROSTHESIS
- Humeral head necrosis without injury to the glenoid cavity.
- Extensive humeral head cartilage damage without injury to the glenoid cavity
- Centred osteoarthritis with a glenoid cavity not allowing implantation of a glenoid implant.
- Rheumatoid polyarthritis with thin rotator cuff.
- Off-centred osteoarthritis with irreparable cuff, and with maintained active elevation of at least 120°.
#### TOTAL ANATOMICAL PROSTHESIS (CEMENTED GLENOID IMPLANT WITH 4 PEGS)
- Centred glenohumeral osteoarthritis with functional rotator cuff
- Rheumatoid polyarthritis with functional rotator cuff
- Post-traumatic sequela, functional rotator cuff with glenoid injury.
#### TOTAL ANATOMICAL PROSTHESIS (POROUS GLENOID IMPLANT)
- Centred glenohumeral osteoarthritis
- Rheumatoid polyarthritis
- Post-traumatic sequela with glenoid injury
- Revision for glenoid loosening
- Glenoid bone loss, where bone graft is needed
- A functional rotator cuff is necessary to use this device
## REVERSE PROSTHESIS (METAL-BACK OR POROUS GLENOID IMPLANT)
The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device.
For all types of prosthesis, the glenoid baseplate (metal-back or porous) is intended for cementless use with the addition of bone screws for fixation, the humeral short stem (metaphyseal stem) is intended for cementless use. At least 2/3 of the metaphyseal component must be implanted in the proximal humeral bone to allow for adequate humeral component fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary FH Industrie ARROW Short Stem Humeral System 29 November 2021
| Company: | FH INDUSTRIE<br>ZI DE KERNEVEZ-6 RUE NOBEL<br>QUIMPER Finistere, FRANCE 29000 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Patricia DONNARD -- FH Industrie<br>Naoual RAHIMI- FH Industrie |
| | +33 (0)2 98 55 68 95 |
| Official Correspondent: | Christine Scifert – MRC Global, LLC<br>(901) 831-8053 |
| Trade Name: | ARROW Short Stem Humeral System |
| Common Name: | Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented<br>Shoulder Prosthesis, Reverse Configuration<br>Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3690 (Shoulder joint humeral (hemi-shoulder) metallic<br>uncemented prosthesis)<br>21 CFR 888.3660 (Shoulder Prosthesis, Reverse Configuration) |
| Panel: | Orthopedic |
| Product Code: | PHX, HSD,KWS |
## Device Description:
The ARROW Humeral Short Stem Humeral System is an extension of humeral stem range of the Arrow prosthesis. The short stem is composed of a metaphyseal part and a cylindrical diaphyseal part. Both components are offered in various sizes to accommodate patient anatomy. This modular stem helps for better adaptation to each patient's anatomy. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI) per ASTM F-136/ISO 5832-3. The metaphysis has fins to provide rotational stability; it has a pure titanium plasma spray coating per ASTM 1580.
## Indications for Use:
The prostheses from FH Industrie are designed for specific indications such as: SIMPLE HUMERAL PROSTHESIS
- Humeral head necrosis without injury to the glenoid cavity. -
- -Extensive humeral head cartilage damage without injury to the glenoid cavity
{5}------------------------------------------------
- Centred osteoarthritis with a glenoid cavity not allowing implantation of a glenoid implant.
- Rheumatoid polyarthritis with thin rotator cuff. -
- Off-centred osteoarthritis with irreparable cuff, and with maintained active elevation of at least -
- 120°.
TOTAL ANATOMICAL PROSTHESIS (CEMENTED GLENOID IMPLANT WITH 4 PEGS)
- Centred glenohumeral osteoarthritis with functional rotator cuff -
- Rheumatoid polyarthritis with functional rotator cuff -
- -Post-traumatic sequela, functional rotator cuff with glenoid injury.
TOTAL ANATOMICAL PROSTHESIS (POROUS GLENOID IMPLANT)
- Centred glenohumeral osteoarthritis -
- -Rheumatoid polyarthritis
- -Post-traumatic sequela with glenoid injury
- -Revision for glenoid loosening
- -Glenoid bone loss, where bone graft is needed
A functional rotator cuff is necessary to use this device
REVERSE PROSTHESIS (METAL-BACK OR POROUS GLENOID IMPLANT)
The ARROW Reverse Shoulder Prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device.
For all types of prosthesis, the glenoid baseplate (metal-back or porous) is intended for cementless use with the addition of bone screws for fixation, the humeral short stem (metaphyseal stem and diaphyseal stem) is intended for cementless use.
At least 2/3 of the metaphyseal component must be implanted in the proximal humeral bone to allow for adequate humeral component fixation.
## Substantial Equivalence:
The subject ARROW Humeral Short Stem Humeral System components are substantially equivalent with respect to indications for use, design, dimension, and materials to the following devices, previously cleared by the FDA:
## Primary Predicate:
FH Industrie, Arrow Humeral Stem - K112193
## Additional Predicates:
- FH Industrie, Arrow Humeral Stem - K093599, K150569
- Exactech, Equinoxe Preserve Stem – K162726
- DJO, Altivate Preserve - K190290
- Biomet, Comprehensive Micro - K060692, K080642
- Depuy Delta III – K021478
#### Reference devices /Biocompatiblity Predicates:
Fournitures Hospitalieres Industrie: ARROW Glenoid base porous (K162068, K171789)
{6}------------------------------------------------
The Indications for Use, Materials, and Geometry for predicate devices are all similar to those of the subject device. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.
## Performance Testing:
Sterilization (ISO 11137) and packaging (ISO 11607) validations , and biocompatibility (ISO 10993-1) assessment and testing were conducted and provided to demonstrate substantial equivalence. Bacterial endotoxin levels were evaluated using LAL pyrogen testing.
Mechanical testing, including the following was performed on the subject devices:
- connection between stem and head per ASTM F 2009
- connection between stem and humeral insert per ASTM F 2009 ●
- static test on the connection between metaphysis and diaphysis
- fatigue test on short stem - below the connection
- fatigue test on short stem - above the connection
- range of motion for anatomical prosthesis per ASTM F1378
- range of motion for reverse prosthesis per ASTM F1378
- fatigue testing per ASTM F2580
- corrosion testing
- connection dynamic torsional resistance
- connection static torsion
The results of all mechanical tests have shown them to be substantially equivalent to the predicate device.
## Conclusion
Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.