ARROW REVERSE SHOULDER SYSTEM

K112193 · Fournitures Hospitalieres Industrie · PHX · Jul 12, 2012 · Orthopedic

Device Facts

Record IDK112193
Device NameARROW REVERSE SHOULDER SYSTEM
ApplicantFournitures Hospitalieres Industrie
Product CodePHX · Orthopedic
Decision DateJul 12, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ARROW® reverse shoulder prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cemented of cementless application while the metal-back glenoid baseplate is intended for cementless application with the addition of bone screws for fixation.

Device Story

Reverse shoulder prosthesis; replaces damaged shoulder joint components. System includes humeral stems, humeral inserts, glenospheres, metal-back glenoid bases, and fixation screws. Implanted by orthopedic surgeons in clinical/hospital settings using dedicated instrumentation. Device restores joint function in patients with rotator cuff deficiency by reversing joint anatomy (glenosphere on glenoid side, humeral cup on humeral side). Relies on deltoid muscle for movement. Benefits include pain relief and improved shoulder mobility for patients with severe arthropathy.

Clinical Evidence

Bench testing only. Performance evaluated per ASTM F1378-05 and ASTM F1829-98 for glenoid components. No clinical data presented.

Technological Characteristics

Materials: titanium alloy (humeral stem, glenoid base, screws), cobalt chromium alloy (glenosphere), polyethylene/titanium alloy (humeral insert). Standards: ASTM F1378-05, ASTM F1829-98. Semi-constrained prosthesis. Fixation: cemented or cementless humeral stem; cementless glenoid baseplate with bone screws.

Indications for Use

Indicated for patients with severe shoulder arthropathy and grossly deficient rotator cuff, or failed shoulder joint replacement with grossly deficient rotator cuff. Requires functional deltoid muscle and adequate glenoid bone stock.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 2, 2016 Fournitures Hospitalieres Industrie Mr. Franck Hunt General Manager 6 Rue Nobel. Z.I. de Kernevez 29000 Quimper FRANCE Re: K112193 Trade/Device Name: Arrow® Reverse Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, KWS, KWT Dated: June 27, 2012 Received: July 3, 2012 Dear Mr. Hunt: This letter corrects our substantially equivalent letter of July 12, 2012. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {2}------------------------------------------------ # K112193 (11) ### Indications for Use 510(k) Number (if known) : K112193 Device Name : ARROW® reverse shoulder system Indications for Use : The ARROW® reverse shoulder prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cemented of cementless application while the metal-back glenoid baseplate is intended for cementless application with the addition of bone screws for fixation. X Prescription Use (Part 21-CFR 801 Subpart D) AND/OR Over the counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Eyaluation (ODE) 510(k) Number Page _ / of (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices and Restorative Devices × 112193 - IQ . {3}------------------------------------------------ Page 2 - Mr. Franck Hunt CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ### Lori A. Wiggins -S for - Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ . ( Fournitures Hospitalières industrie ARROW® Reverse Shoulder system Traditional 510(k) Premarket Notification ### JUL 1 2 2012 - 1 - ## 5. 510(k) Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance to the requirements of SDMA 1990 and 21 CFR 807.92. Date prepared: July 9th 2012 The assigned 510(k) number is: K112193 5-1. Applicant: Fournitures Hospitalières industrie 6 Rue Nobel, Z.I. de Kernévez 29000 QUIMPER - FRANCE Tel: (+33) 2.98.55.68.95 Fax: (+33) 2.98.53.42.13 - 5-2. Company Contact: Franck HUNT, General Manager Tel: (+33) 2.98.55.68.95 5-3. | Product : | | |-----------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Trade name: | ARROW® Reverse Shoulder System | | Common name: | Shoulder prosthesis | | Classification: | The ARROW® Reverse Shoulder System components are<br>included in the following classifications : | | | - Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis<br>Product code: KWS<br>Regulation: 21 CFR 888.3660<br>Class: II | | | - Shoulder joint metal/polymer non-constrained<br>cemented prosthesis | Product code: KWT Regulation: 21 CFR 888.3650 Class: II {5}------------------------------------------------ ### 5-4. Predicate/ Legally Marketed Devices : Information on devices to which substantial equivalence is claimed: | Manufacturer: | Zimmer | |--------------------|------------------------------------------------| | Device Trade Name: | Anatomical Shoulder® Inverse/Reverse<br>system | | 510 (K): | K053274 | | Manufacturer: | Exactech, Inc. | | Device Trade Name: | Equinoxe™ Reverse Shoulder system | | 510 (K): | K073688/K063569 | | Manufacturer: | Depuy | | Device Trade Name: | Delta Xtend™ Reverse Shoulder system. | | 510 (K): | K062250 | | Manufacturer: | Fournitures Hospitalières Industrie | | Device Trade Name: | ARROW® Anatomical shoulder system | | 510 (K): | K093599 | #### 5-5. Device Description: The ARROW® reverse shoulder system is a shoulder joint prosthesis, composed of the following elements : - Humeral stems, ・ - STD humeral inserts, • - Glenospheres - - Metal-back glenoid bases - - And fixation screws for the metal back glenoid base (cancellous and cortical bone screws) The ARROW® reverse shoulder system is intended to be implanted using the dedicated instrumentation supplied by the manufacturer. #### 5-6. Indications for Use/ Intended Use: The ARROW® reverse shoulder prosthesis is indicated for patients with severe shoulder arthropathy and a grossly deficient rotator cuff or a previously failed shoulder joint replacement with a grossly deficient rotator cuff. A functional deltoid muscle and adequate glenoid bone stock are necessary to use this device. The humeral stem is intended for cementless application while the metal-back glenoid baseplate is intended for cementless application with the addition of bone screws for fixation. > ﺔ -2- {6}------------------------------------------------ #### 5-7. Comparison of Technological Characteristics: The ARROW® reverse shoulder prosthesis and the above selected predicate devices have the same intended use and substantial similar indications for use and share the following similatirities : - They are made out of the same materials (titanium alloy for the humeral stem, the . metel-back glenoid base and the fixation screws, cobalt chromium alloy for the glenosphere, polyethylene and titanium alloy for humeral insert). - They are available in similar ranges of sizes, • - They bear design features similarities - #### 5-8. Performances: The ARROW® reverse shoulder prosthesis was tested according to ASTM F1378-05, ASTM F1829-98 for the glenoid components. After the testing was completed, it was determined that the ARROW reverse shoulder system performances were substantially equivalent to those of the selected predicated devices. Risks to health have been addressed through the specified materials, processing controls. quality assurance and compliance to the Medical Device Good Manufacturing Practices Regulations. #### 5-9. Substantial Equivalence: The substantial equivalence of our product, when compared to the selected predicate devices, has been established following manufacturers' commercial documents, 510(k) submission's information available on FDA's website as well as conformance to standards in force. The analysis of these technical data allows us to submit the ARROW® reverse shoulder prosthesis as being substantially equivalent to the already cleared predicate devices selected to a draw a comparison. All data on predicate devices which has been used to establish equivalence is available in appendix 8. #### 5-10. Conclusion: Following the examination of all the above mentioned information, we believe that the ARROW® reverse shoulder prosthesis is substantially equivalent to the selected predicate devices in terms of desing, ranges of sizes, materials, intended use, performances, safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...