White ENDOSEAL MTA
K202015 · Maruchi · KIF · Mar 24, 2021 · Dental
Device Facts
| Record ID | K202015 |
| Device Name | White ENDOSEAL MTA |
| Applicant | Maruchi |
| Product Code | KIF · Dental |
| Decision Date | Mar 24, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3820 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Permanent root canal obturation
Device Story
White ENDOSEAL MTA is a root canal sealer supplied in a pre-loaded syringe. It is designed for use by dental professionals in a clinical setting. The device is a paste that does not require mixing; it sets by absorbing moisture present within the root canal. By filling the root canal space, it acts as a root filling material to achieve permanent obturation. The device is non-sterile and has a shelf life of 2 years.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included ISO 6876 standards (capacity, pH, package, extraneous matter, flow, setting time, film thickness, solubility, radio-opacity) and ISO 10993 biocompatibility testing (cytotoxicity, skin sensitization, acute systemic toxicity, and genotoxicity). All results met preset criteria.
Technological Characteristics
Calcium silicate-based root canal sealer. Composition includes zirconium dioxide, dimethyl sulfoxide, tricalcium silicate, water, bentonite clay, polyvinyl alcohol, polyvinyl pyrrolidone, and lithium carbonate. Form factor: pre-loaded syringe paste. Conforms to ISO 6876. Non-sterile. Shelf life: 2 years.
Indications for Use
Indicated for permanent root canal obturation in patients requiring endodontic treatment.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
Predicate Devices
Related Devices
- K212229 — White Endoseal MTA · Maruchi · Aug 13, 2021
- K080203 — MTA ROOT CANAL SEALER · Dentsply Interntional · Mar 26, 2008
- K170175 — ENDOSEAL MTA · Maruchi · Feb 26, 2018
- K231480 — Bright MTA Sealer Plus · Genoss Co., Ltd. · Jul 28, 2023
- K213757 — Endocem MTA Premixed Regular · Maruchi · Feb 14, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 24, 2021
Maruchi Sung Jang CEO 2-208, Medical Industry Complex Bldg., 42-10, Taejanggongdan-gil Wonju-si, Gangwon-do 26311 REPUBLIC OF KOREA
Re: K202015
Trade/Device Name: White ENDOSEAL MTA Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: January 15, 2021 Received: January 27, 2021
Dear Sung Jang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 4. INDICATION FOR USE STATEMENT
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2023 | | | | | |
|-------------------------------------------------------------------------|---------------------|--|-----------------------------------------------------------------|--|--------------------------|--|--|--|
| | Indications for Use | | | | See PRA Statement below. | | | |
| 510(k) Number (if known) | | | | | | | | |
| K202015 | | | | | | | | |
| Device Name | | | | | | | | |
| White ENDOSEAL MTA | | | | | | | | |
| Indications for Use (Describe) | | | | | | | | |
| Permanent root canal obturation | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
## k202015
## 510(k) Summary
Date: August 20, 2020
# 1. SUBMITTER
MARUCHI
2-208, Medical Industry Complex Bldg., 42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, 26311, Republic of Korea TEL : +82-33-734-0330
FAX : +82-33-746-2804 Contact Name: Sung Wook, Jang Email: ra@endocem.com
## 2. DEVICE
·Trade Name: White ENDOSEAL MTA
·Common Name: Root Filling Material
· Classification Name: Resin, Root canal filling
·Regulation Number 872.3820
·Class: 2
·Classification Product Code: KIF
## 3. PREDICATE DEVICE
K170175, ENDOSEAL MTA, MARUCHI
# 4. DEVICE DESCRIPTION
This product is the root canal sealer conforming to ISO 6876. This is a pre-loaded syringe type that does not require mixing and set by absorbing the moisture around the root canal.
# 5. INDICATIONS FOR USE
Permanent root canal obturation
{4}------------------------------------------------
# 6. PERFORMANCE TESTING (NON-CLINICAL)
The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.
- Testing institution's method Capacity, pH •
- ISO 6876 - Package, Extraneous matter, Flow, Setting time, Film Thickness, Solubility, Radio-opacity
- ISO 10993-5 Cytotoxicity (MTT) •
- . ISO 10993-10 - Guinea Pig Maximization Test for Skin sensitization (GPMT)
- ISO 10993-11 Acute systemic toxicity •
- ISO 10993-3 Genotoxicity (Mammalian chromosome aberration test & . Bacterial Reverse Mutation)
| | Proposed Device | Predicate Device | Discuss/Justify the Differences |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| 510(k) Number | K202015 | K170175 | - |
| Trade Name | White ENDOSEAL MTA | ENDOSEAL MTA | - |
| Manufacturer | MARUCHI | MARUCHI | - |
| Common Name | Root Filling Material | Root Filling Material | Equivalent |
| Classification Name | Resin, Root canal filling | Resin, Root canal filling | Equivalent |
| Device Class | 2 | 2 | Equivalent |
| Product Code | KIF | KIF | Equivalent |
| Device Description | This product is the root canal sealer conforming to ISO 6876. This is a pre-loaded syringe type that does not require mixing and set by absorbing the | ENDOSEAL MTA is an endodontic sealer based on MTA, providing a root canal filling. It is premixed and pre-loaded in a syringe, which allows a complete | Equivalent |
| | moisture around the root canal. | filling of the entire root canal including accessory and lateral canals. | |
| Indications for Use | Permanent root canal obturation | * Permanent obturation of the root canal following vital pulp-extirpation<br>* Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings. | Equivalent |
| Intended user | Dental professional | Dental professional | Equivalent |
| Standards | ISO 6876 | ISO 6876 | Equivalent |
| Physical properties | Setting time: 29 min<br>Flow: 22 mm<br>Radiopacity: 7.7 mm<br>Flim thickness: 14 μm<br>Solubility: 0.3 % | Setting time: 12.31 min<br>Flow: 21 mm<br>Radiopacity: 10.14 mm<br>Flim thickness: 15 μm<br>Solubility: 0.7 % | Differences |
| Chemical Composition | -Zirconium dioxide<br>-Dimethyl sulfoxide<br>-Calcium silicate<br>(Tricalcium silicate)<br>-Water<br>-Bentonite Clay<br>-Polyvinyl alcohol<br>-Polyvinyl pyrrolidone<br>-Lithium Carbonate | -Natural Pure Cement<br>-Zirconium dioxide<br>-Bismuth trioxide<br>-Bentonite Clay<br>-n-Methyl-2-Pyrrolidone<br>-Hypromellose | Differences |
| Liquid Formula | Paste type | Paste type | Equivalent |
| Packaging | Pre-loaded syringe | Pre-loaded syringe | Equivalent |
| Sterile | Non-sterile | Non-sterile | Equivalent |
| Shelf Life | 2 years | 2 years | Equivalent |
# 7. SUBSTANITAL EQUIVALENCE
{5}------------------------------------------------
{6}------------------------------------------------
# 8. SUBSTANTIAL EQUIVALENCE DISCUSSION
White ENDOSEAL MTA has the same Indications for Use and the principle of operations as the predicate devices. It is intended purpose as they are placed into the root canal as a root filling materials which met the requirement according to ISO 6876. It has similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices.
The chemical compositions might slightly different from the predicate devices, both are used calcium silicate as base material, and additional components are used to improve flowability.
The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that White ENDOSEAL MTA is substantially equivalent to the predicate devices. Hence, its equivalent is acceptable.