Permanent obturation of the root canal following vital pulp-extirpation Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.
Device Story
Endoseal MTA is a premixed, pre-loaded syringe-based endodontic sealer; based on mineral trioxide aggregate (MTA). Used by dentists/endodontists in clinical settings for permanent root canal obturation. Device is injected directly into root canal system; fills accessory and lateral canals. Hardens via reaction with moisture (calcium oxide hydration); releases calcium silicate oxide and calcium hydroxide. Eugenol-free formulation ensures compatibility with adhesive materials. Provides biocompatible seal; facilitates successful endodontic treatment outcomes.
Clinical Evidence
Bench testing only. Testing included ISO 6876 (flow, setting time, film thickness, solubility, radiopacity), shelf-life testing, and ISO 10993 biocompatibility series (cytotoxicity, delayed hypersensitivity, acute systemic toxicity, implantation, genotoxicity). All results met preset criteria.
Technological Characteristics
Premixed, pre-loaded syringe. Composition: Natural pure cement, zirconium dioxide, bismuth trioxide, bentonite clay, n-Methyl-2-Pyrrolidone, hypromellose. Setting principle: Hydraulic cement hydration (calcium oxide reaction with water). Non-sterile. Standards: ISO 6876.
Indications for Use
Indicated for permanent root canal obturation in patients requiring endodontic treatment following vital pulp-extirpation or removal of infected/necrotic pulp.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{1}------------------------------------------------
February 26, 2018
Maruchi Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc 800 Roosevelt STE 417 Irvine, California 92620
Re: K170175
Trade/Device Name: Endoseal MTA Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: January 17, 2018 Received: January 26, 2018
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K170175
Device Name
ENDOSEAL MTA
Indications for Use (Describe)
· Permanent obturation of the root canal following vital pulp-extirpation
• Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| V D--------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | O -- TL- O- -- L--- II-- II-- IO4 OFF OOL O I---- |
|| | Prescription Use (Part 21 CFR 801 Subpart D)
|__| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
(K170175)
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: ___01/17/2018_____
#### Applicant / Submitter: 1.
MARUCHI 2-208, 42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, 26311, Republic of Korea Tel: +82-33-734-0330 Fax:+82-33-746-2804
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 800 Roosevelt STE 417 Irvine CA 92620 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### 3. Device:
| Proprietary Name: | ENDOSEAL MTA |
|------------------------------|---------------------------|
| Common Name: | Root Filling Material |
| Classification Name: | Root Canal Filling Resin |
| Classification: | Class II, 21 CFR 872.3820 |
| Classification Product Code: | KIF |
#### 4. Predicate Device:
ENDOSEAL (K133054) by MARUCHI MTA Fillapex (K113568) by Angelus Industria de Produtos Odontologicos IROOT SP (K080917) by Innovative BioCeramix Inc.
{5}------------------------------------------------
#### 5. Device Description:
ENDOSEAL MTA is an endodontic sealer based on MTA, providing a biocompatible and effective root canal filling. It is premixed and pre-loaded in a syringe, which allows a complete filling of the entire root canal including accessory and lateral canals. The product is eugenol-free and will not impede adhesion inside the root canal.
#### 6. Indications for Use
- Permanent obturation of the root canal following vital pulp-extirpation
- · Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.
#### 7. Performance Data(Non-Clinical):
The following properties were tested based on the referenced standard. All the test results met the preset test criteria.
- = ISO 6876 Visual, Capacity, Packaging, Flow, Setting time, Film Thickness, Solubility and Radiopacity
- = Shelf Life Test ISO 6876 (Visual, Packaging, Setting time, Solubility)
- ISO 10993-5 Cytotoxicity
- ISO 10993-10 Maximization test for delayed hypersensitivity (LLNA)
- ISO 10993-11 Acute systemic toxicity
- ISO 10993-6 Implantation
- ISO 10993-3 AMES & Mammalian Erythrocyte Micronucleus
{6}------------------------------------------------
#### 8. Substantial Equivalence
| | Proposed Device | Predicate Device | Predicate Device | Predicate Device | Discuss/Justify the |
|----------------|-------------------------|-------------------------|-------------------------|------------------------|---------------------|
| | (K170175) | (K133054) | (K113568) | (K080917) | Differences |
| Trade Name | ENDOSEAL MTA | ENDOSEAL | MTA Fillapex | IROOT SP | |
| Device | ENDOSEAL MTA is | The raw materials of | MTA Fillapex is a | iRoot SP Root Canal | Equivalent |
| Description | an endodontic sealer | the ENDOSEAL are | mineral trioxide | Sealer(IRoot SP) is | |
| | based on MTA, | Natural Pure Cement, | aggregate(MTA) and | premixed ready-to-use | |
| | providing a root canal | Zirconium dioxide, | resin root canal sealer | injectable white | |
| | filling. It is premixed | and | used during | hydraulic cement | |
| | and pre-loaded in a | Citric acid anhydrous. | endodontic treatment | paste developed for | |
| | syringe, which allows | It is prepared as a | to permanently fill the | permanent root canal | |
| | a complete filling of | mixture of powder | canal system | filling and sealing | |
| | the entire root canal | and water, and it is | following | applications. iRoot SP | |
| | including accessory | used in a putty | debridement and | is an insoluble. | |
| | and lateral canals. | form which gradually | disinfection. It | radiopaque and | |
| | | hardens in the oral | consists of two | aluminum-free | |
| | | environment. | component pastes that | material based on a | |
| | | | are combined in a dual | calcium silicate | |
| | | | barrel syringe for ease | composition, which | |
| | | | of dispensing and | requires the presence | |
| | | | consistent dosage. | of water to set and | |
| | | | | harden. iRoot SP is | |
| | | | | packaged in a pre- | |
| | | | | loaded syringe and is | |
| | | | | supplied with | |
| | | | | disposable Intra Canal | |
| | | | | Tips. | |
| Common Name | Root Filling Material | Root Filling Material | Root Filling Material | Root Filling Material | Equivalent |
| Classification | Root Canal Filling | Root Canal Filling | Root Canal Filling | Root Canal Filling | Equivalent |
| Name | Resin | Resin | Resin | Resin | |
| Class | Class II | Class II | Class II | Class II | Equivalent |
| Product Code | KIF | KIF | KIF | KIF | Equivalent |
| Indications | ▪ Permanent | ■ Repair of perforation | MTA Fillapex is a | ▪ Permanent | Equivalent |
| for Use | obturation of the root | ▪ Root canal filling | root canal sealer | obturation of the root | |
{7}------------------------------------------------
| | canal following vital<br>pulp-extirpation<br>▪ Permanent<br>obturation of the root<br>canal following<br>removal of infected or<br>necrotic pulp and<br>placement of<br>intracanal dressings. | | intended for the<br>permanent sealing of<br>root canals and may<br>be used in<br>combination with root<br>canal obturation<br>materials. | canal following vital<br>pulp-extirpation<br>▪ Permanent<br>obturation of the root<br>canal following<br>removal of infected or<br>necrotic pulp and<br>placement of<br>intracanal dressings.<br>iRoot SP is suitable<br>for use in the single<br>cone and lateral<br>condensation<br>technique. | |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shelf Life | 2 years | 2 years | 2 years | | Equivalent |
| Standards | ISO 6876 | ISO 6876 | ISO 6876 | ISO 6876 | Equivalent |
| Basic Chemical<br>Composition | Natural pure cement<br>Zirconium dioxide<br>Bismuth trioxide<br>Bentonite Clay | Natural pure cement<br>Zirconium dioxide<br>Citric acid anhydrous | Dicalcium and<br>tricalcium silicate,<br>tricalcium aluminate,<br>calcium oxide, | Zirconium oxide<br>Calcium silicates<br>Calcium phosphate<br>monobasic | The main ingredients<br>are similar but some<br>other ingredients are<br>different. However, |
| | n-Methyl-2-<br>Pyrrolidone<br>Hypromellose | | pentaerythritol<br>rosinate, n-ethyl-o,<br>p-toluenesulfonamide<br>salicylate resin,<br>bismuth oxide, fumed<br>silica, titanium<br>dioxide | Calcium hydroxide<br>Filler and thickening<br>agents | the biocompatibility<br>and the performance<br>test results supported<br>that the subject device<br>is substantially<br>equivalent to the<br>predicate devices. |
| Liquid Formula | Premixed type | Mixing with distilled<br>water | Mixing through a dual<br>barrel syringe | Premixed type | Equivalent |
| Packaging | Pre-loaded syringe | Vial | Dual barrel syringe | Pre-loaded syringe | Equivalent |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Equivalent |
{8}------------------------------------------------
ENDOSEAL MTA has the same indications for use and the principle of operations as the predicate devices. It has similar physical and biocompatible properties, and demonstrates comparable performance specifications to the predicate devices. In addition, ENDOSEAL MTA has a comparable delivery system to IROOT SP.
Endoseal MTA and all of the predicate devices are equivalent as they all release calcium silicate oxide and calcium hydroxide by undergoing setting reactions when water reacts with calcium oxide. The reaction results in near identical composition of products after setting, as all of liquid components would have expelled during the setting reaction. Hence, all of the products resulted in similar test results including the cytotoxicity test.
The material compositions might slightly different from the predicate devices, however, despite of this difference, the test results of biocompatibility and performance tests show that the subject device is substantially equivalent to the predicate devices.
The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that ENDOSEAL MTA is substantially equivalent to the predicate devices.
#### 9. Conclusion:
Based on the testing results, MARUCHI concludes that the ENDOSEAL MTA is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.