K140247 · Angelus Industria DE Productos Odontologicos SA · KIF · Apr 25, 2014 · Dental
Device Facts
Record ID
K140247
Device Name
MTA FILLAPEX
Applicant
Angelus Industria DE Productos Odontologicos SA
Product Code
KIF · Dental
Decision Date
Apr 25, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MTA-Fillapex is a root canal sealer intended for the permanent scaling of root canals and may be used in combination with root canal obturation materials.
Device Story
MTA Fillapex is a mineral trioxide aggregate (MTA) and resin-based root canal sealer. It is supplied as a two-component paste (base and catalyst) in a dual-barrel syringe for consistent dispensing. The device is used by dentists during endodontic procedures to permanently fill the root canal system following debridement and disinfection. It is hydrophilic, allowing for use in non-isolated, moist environments. The sealer is applied in conjunction with gutta-percha or silver points during obturation. The material sets in the presence of moisture, providing a permanent seal. The device benefits patients by facilitating effective root canal obturation without requiring a strictly dry field.
Clinical Evidence
Bench testing only. Testing compared subject device to predicate (K113568) regarding flowability, working time, setting time, dimensional change, solubility, shelf life, and biocompatibility. Results demonstrated substantial equivalence. Biocompatibility testing confirmed the material is non-mutagenic and well-tolerated.
Indicated for permanent sealing of root canals in patients undergoing endodontic treatment; used in combination with root canal obturation materials (e.g., gutta-percha or silver points).
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{1}------------------------------------------------
140247
APR 2 5 2014
Image /page/0/Picture/2 description: The image contains the logo for Angelus Science and Technology. The logo features a stylized letter 'A' inside of a circle on the left. To the right of the circle is the word 'angelus' in lowercase letters, and below that is the phrase 'science and technology' in smaller letters.
### 510(k) Summary
| Proprietary Name | MTA Fillapex |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| K Number | K140247 |
| Date Prepared | April 25, 2014 |
| Submitter | Angelus Industria de Productos Odontologicos<br>Waldir Landgraf, 101<br>Londrina, Brazil 86031-3200 |
| Telephone: | +55 43 2101 3200 |
| Official Contact | Tara Conrad<br>TechLink International Consulting<br>18851 NE 29th Avenue Suite 720<br>Aventura, FL 33180<br>TEL- (305) 377-0077 |
| Common Name: | Resin, Root Canal Filling |
| Trade Name: | MTA Fillapex |
| Classification: | Class II |
| Product Code: | KIF |
| Classification Panel: | Dental |
| Regulation Numbers: | 21 CFR 872.3820 |
| Substantial Equivalence: | K113568 MTA Fillapex by Angelus<br>K010940 Sealapex by Sybron |
### Description of Proposed Device
MTA Fillapex is a mineral trioxide aggregate (MTA) and resin toot canal sealer used during endodontic treatment to permanently fill the canal system following debridement and disinfection. It consists of two component pastes that are combined in a dual barrel syringe for ease of dispensing and consistent dosage. Being hydrophilic in nature, MTA-FILLAPEX is desirable as a root filling material because an isolated dry field is not necessary for use. Moisture does not negatively affect the sealing ability and is required for proper setting. It is used in combination with gutta-percha or silver points during root canal obturation.
### Indications for Use
MTA-Fillapex is a root canal sealer intended for the permanent scaling of root canals and may be used in combination with root canal obturation materials.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Angelus Science and Technology. The logo features a stylized "A" inside of a black circle on the left. To the right of the circle is the word "angelus" in lowercase letters. Below the word "angelus" is the phrase "science and technology" in a smaller font.
# Device Comparison Table
| | Subject Device | Predicate Device | |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MTA Fillapex by Angelus | K113568 MTA Fillapex by | K010940 Sealapex by |
| | K140247 | Angelus | Sybron |
| Indications for use | MTA-Fillapex is a root canal<br>sealer intended for the<br>permanent scaling of root<br>canals and may be used in<br>combination with root canal<br>obturation materials. | MTA-Fillapex is a root canal<br>sealer intended for the<br>permanent scaling of root<br>canals and may be used in<br>combination with root canal<br>obturation materials. | Sealapex 4 is a<br>calcuium hydroxide,<br>polymeric resin, root<br>canal filling material<br>that is used in<br>conjunction with gutta<br>percha or silver<br>endodontic points. |
| Design | Two paste system<br>(base/catalyst) | Two paste system<br>(base/catalyst) | Two paste system<br>(base/catalyst) |
| Flow | 33.38 mm | 27.66mm | 25.15 (±1.73) mm (1) |
| Material components | Dicalcium and<br>tricalcium silicate,<br>tricalcium aluminate,<br>calcium oxide,<br>pentaerythritol<br>rosinate, n-ethyl-o,<br>p-toluenesulfonamide<br>salicylate resin,<br>bismuth oxide, fumed<br>silica, titanium dioxide | Dicalcium and<br>tricalcium silicate,<br>tricalcium aluminate,<br>calcium oxide,<br>pentaerythritol<br>rosinate, n-ethyl-o,<br>p-toluenesulfonamide<br>salicylate resin,<br>bismuth oxide, fumed<br>silica, titanium dioxide | Titanium Dioxide,<br>Fumed Silica, Bismuth<br>Trioxide, Isobutyl<br>salicylate resin, N -<br>ethyl toluene<br>solfanamide resin,<br>Calcium Oxide, Zinc<br>Dioxide |
| Working time | 23 minutes | 35 minutes | Minimum 120 minutes |
| Setting time | Minimum 130 minutes | Minimum 130 minutes | 45 minutes after<br>placement |
| Solubility | 2.47% | 0.1% | >4% (2) |
| Shelf Life | 2 years | 2 years | 2 years |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible |
| Standards | ISO 6876:2012 | ISO 6876: 2001 | ISO 6876: 2001 |
# Substantial Equivalence
The subject device and the predicate devices have intended use and have the same technological characteristics, including the same range of physical and chemical properties. The subject and predicate devices are packaged in similar materials and utilize similar methods of application. Any differences in the technological characteristics do not raise new issues of safety or efficacy.
# Performance Data
MTA Fillapex is a modification of K113568. They are designated for the equivalent dental applications and have comparable chemical and physical properties, and performance specifications. An additional predicate device (K010940) includes each specific chemical component that is equivalent to those found in MTA Fillapex, providing evidence that these materials are safe and effective for medical device use. Furthermore, MTA Fillapex and K113568 have equivalent packaging containers and delivery systems.
MTA Fillapex has undergone the following testing:
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Angelus Science and Technology. The logo consists of a black circle with a white "A" inside, followed by the word "angelus" in a sans-serif font. Below the word "angelus" is the phrase "science and technology" in a smaller font.
These tests have shown that MTA Fillapex and K113568 are substantially equivalent to the new MTA Fillapex. Both devices have comparable flowability, working times, setting times, dimensional change following setting, solubility, shelf life and biocompatibility properties.
Since MTA Fillapex's chemical composition is based on the principal chemical components in K113568, the biocompatibility test provides evidence that the product is non-mutenagenic, does not cause an allergenic potential after multiple uses and has a good tolerance. The shelf life, bench and biocompatibility testing, provide evidence that the chemical and physical properties are substantially equivalent.
# Design Control Activities
The risk analysis method used to assess the impact of the modifications was a Failure Modes and Effects Analysis (FMEA) The design verification tests that were performed as a result of the risk analysis assessment are listed in the table below:
| Modification | Test Performed | Acceptance criteria |
|-------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Titanium dioxide was<br>added to the base paste | 1. Working time<br>2. Setting time<br>3. Flow<br>4. Radiopacity<br>5. Solubility | 1. 23 minutes<br>2. 130 minutes<br>3. Minimum of 20 mm<br>4. Minimum of 3 mm<br>5. Maximum 3% |
| | Internal Tests | |
| | 1. Flow/Viscosity | 1. Base Paste 20-22<br>mm / Catalyst Paste<br>27-29 mm<br>Mixed pastes after<br>30 min 21-23 mm |
| | 2. Radiopacity | 2. Greater than 3mm<br>of the aluminium<br>scale |
| | 3. Setting time | 3. 130 minutes |
| | 4. Free lime content | 4. 3-5% |
The test methods are the same as those submitted in the original submission.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Angelus. The logo consists of a circular icon with a stylized "A" inside, followed by the word "angelus" in a bold, sans-serif font. Below the word "angelus" are the words "science and technology" in a smaller, lighter font.
# Conclusion
MTA Fillapex has the equivalent indications for use, comparable chemical compositions, physical properties and performance specifications to K113568. The additional chemical component found in MTA Fillapex were found to be as safe and effective as the predicate.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, M1) 20993-0002
#### April 25, 2014
Angelus Industria de Productos Odontologicos C/O Tara Conrad TechLink International Consulting 18851 NE 29th Avenue Suite 720 Aventura, FL 33180
Re: K140247
:
Trade/Device Name: MTA Fillapex Regulation Number: 21 CFR 872.3820 Regulation Name: Resin, Root Canal Filling Regulatory Class: II Product Code: KIF Dated: March 14, 2014 Received: April 4, 2014
Dear Ms. Conrad:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Ms. Conrad
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Susan Rinno, DDS, MA
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use Statement
Device Name: MTA Fillapex
510(k) Number: K140247
Indications for Use:
MTA-Fillapex is a root canal sealer intended for the permanent scaling of root canals and may be used in combination with root canal obturation materials.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEED
AND/OR
oncurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
> Sheena A. Green -S 2014.04.25 12:35:13 -04'00'
> > Page 12 of 131 Special 510(k) MTA Fillapex
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.