K201976 · Kent Imaging · MUD · Nov 10, 2020 · Cardiovascular
Device Facts
Record ID
K201976
Device Name
SnapshotNIR
Applicant
Kent Imaging
Product Code
MUD · Cardiovascular
Decision Date
Nov 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
SnapshotNIR is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of: oxygen saturation (StO2), relative oxyhemogoblin level (HbO2), and relative deoxyhemoglobin (Hb) level in superficial tissue. SnapshotNIR displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions. SnapshotNIR is indicated for use to determine oxygenation levels in superficial tissues.
Device Story
SnapshotNIR is a non-invasive, battery-powered multispectral imaging system used by healthcare professionals to assess superficial tissue oxygenation. The device consists of a camera (with light source, CMOS sensor, and touchscreen PC), a recharger, and a calibration card. It illuminates the tissue surface with LEDs across six wavelengths (600nm–1000nm) and captures returned light via a CMOS sensor. The system performs spectral analysis at each point in the scanned area to calculate approximate oxygen saturation (StO2), oxyhemoglobin (HbO2), and deoxyhemoglobin (Hb) levels. Output is displayed as a two-dimensional, color-coded map of tissue oxygenation and numeric data. The device supports freehand region-of-interest (ROI) selection and area measurements. It is used in clinical settings to provide objective oxygenation data, aiding healthcare providers in clinical decision-making regarding tissue health. Reports can be emailed via Wi-Fi connectivity.
Clinical Evidence
A clinical study on 38 volunteers using a forearm ischemia-reperfusion challenge compared the modified device (KD204) to the predicate (KD203). The study demonstrated a linear relationship between the two devices over a clinically meaningful dynamic range of StO2 (R2 > 0.98). RMSE for relative hemoglobin estimates was 0.000239 for Hb and 0.00208 for HbO2. Bland-Altman analysis confirmed agreement between the algorithms.
Technological Characteristics
Multispectral imaging system; 6 wavelengths (600nm–1000nm); CMOS image sensor with global shutter; LED light source; DC battery-powered; non-contact; DICOM image format; Wi-Fi connectivity for reporting. Compliant with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 62133, and UN38.3.
Indications for Use
Indicated for use by healthcare professionals to determine oxygenation levels (StO2, HbO2, Hb) in superficial tissues.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 10, 2020
Kent Imaging Liz Newell Director, Clinical Research 804B 16th Ave SW Calgary, Alberta t2r 0s9 Canada
Re: K201976
Trade/Device Name: SnapshotNIR Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: July 15, 2020 Received: July 16, 2020
Dear Liz Newell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201976
Device Name SnapshotNIR
Indications for Use (Describe)
SnapshotNIR is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurem tsystem that reports an approximate value of:
- oxygen saturation (StO2),
- relative oxyhemogoblin level (HbO2), and
- relative deoxyhemoglobin (Hb) level
in superficial tissue. SnapshotNIR displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions.
SnapshotNIR is indicated for use to determine oxygenation levels in superficial tissues.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## SnapshotNik (November 2nd, 2020)
#### Submittal Information:
Post-approval contact: Liz Newell Kent Imaging Inc. 804B 16th Ave SW Calgary, AB, Canada T2R 0S9
Phone: 403-455-7610 Fax: 877-664-5450
### Device Name and Classification
| Proprietary Name: | SnapshotNIR |
|-----------------------|---------------------------------------------------------------------|
| Common Name: | Tissue Oximeter |
| Classification Name: | Oximeter, Tissue Saturation (21 CFR 870.2700, Product Code: 74 MUD) |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
#### Device Description
Snapshotwa, Model KD204 (K201976), is a modification to the Kent Camera, Model KD203 (K163070). The changes made to create the modified snapshot include modifications to the software. Both devices have similar hardware.
Snapshotwa is based on multispectral imaging technology and performs spectral analysis at each point in a two-dimensional scanned area producing an image displaying information derived from the analysis. Snapshot;we determines the approximate values of oxygen saturation (S-O2), oxyhemoglobin levels (HbO₂), and deoxyhemoglobin levels (Hb) in superficial tissues and displays a two-dimensional, color-coded image of the tissue oxygenation (S-O2).
The camera consists of:
- Camera: Contains light source, camera and touchscreen PC
- Recharger: Used to recharge the camera ●
- Reference Card: To calibrate the camera
### Intended Use
Snapshotma is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of:
- oxygen saturation (StO2),
- oxyhemoglobin level (HbO₂), and
- deoxyhemoglobin (Hb) level
{4}------------------------------------------------
in superficial tissue. Snapshotwa displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions.
The Snapshotwa is indicated for use to determine oxygenation levels in superficial tissues.
| Comparative<br>Feature | Kent Imaging Inc.<br>Modified SnapshotNIR-<br>KD204<br>(K201976) | Kent Imaging Inc.<br>Predicate Kent Camera<br>- KD203<br>(K163070) | Significant<br>Differences |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Kent Camera is<br>intended for use by<br>healthcare professionals<br>as a non-invasive tissue<br>oxygenation<br>measurement system<br>that reports an<br>approximate value of<br>oxygen saturation<br>(StO2), oxyhemoglobin<br>level (HbO2), and<br>deoxyhemoglobin (Hb)<br>level in superficial tissue.<br>The Kent Camera<br>displays two-dimensional<br>color-coded images of<br>tissue oxygenation of the<br>scanned surface and<br>reports multispectral<br>tissue oxygenation<br>measurements for<br>selected tissue regions.<br>The Kent Camera is<br>indicated for use to<br>determine oxygenation<br>levels in superficial<br>tissues. | The Kent Camera is<br>intended for use by<br>healthcare professionals<br>as a non-invasive tissue<br>oxygenation<br>measurement system<br>that reports an<br>approximate value of<br>oxygen saturation<br>(StO2), oxyhemoglobin<br>level (HbO2), and<br>deoxyhemoglobin (Hb)<br>level in superficial<br>tissue. The Kent Camera<br>displays two-<br>dimensional color-coded<br>images of tissue<br>oxygenation of the<br>scanned surface and<br>reports multispectral<br>tissue oxygenation<br>measurements for<br>selected tissue regions.<br>The Kent Camera is<br>indicated for use to<br>determine oxygenation<br>levels in superficial<br>tissues. | Same |
| Measurements | Oxygen saturationOxyhemoglobin levelDeoxyhemoglobin<br>level | Oxygen saturationOxyhemoglobin<br>levelDeoxyhemoglobin<br>level | Same |
| Method of<br>Measurement | Non-invasive, non-<br>patient contacting<br>imaging head illuminates<br>the surface and receives | Non-invasive, non-<br>patient contacting<br>imaging head<br>illuminates the surface | Same |
| | returned light | and receives returned light | |
| | 6 wavelengths between<br>600nm and 1000nm | 6 wavelengths between<br>600nm and 1000nm | Same |
| | A CMOS image sensor<br>with global shutter is<br>used as the detector | A CMOS image sensor<br>with global shutter is<br>used as the detector | Same |
| | Spectral analysis at<br>specific wavelengths of<br>light returned from the<br>target tissue | Spectral analysis at<br>specific wavelengths of<br>light returned from the<br>target tissue | Similar -<br>improved signal<br>to noise in<br>modified device |
| Light Source | LEDs | LEDs | Same |
| Lens | Corrected and optimized<br>for use from 400nm to<br>1000nm | Corrected and optimized<br>for use from 400nm to<br>1000nm | Same |
| | ● Two-dimensional<br>color-coded map of<br>estimated oxygen<br>saturation | ● Two-dimensional<br>color-coded map of<br>estimated oxygen<br>saturation | Same |
| Output Display | ● Numeric data | ● Numeric data | |
| Power Source | DC (battery-powered) | DC (battery-powered) | Same |
| Patient Contact | None | None | Same |
| Control Method | Computer controlled | Computer controlled | Same |
| | Preformed at start-up by<br>operator and repeated if<br>operating conditions of<br>camera change to the<br>degree that necessitates<br>recalibration | Preformed at start-up by<br>operator. Performed<br>periodically during<br>extended picture<br>capturing sessions | Similar;<br>Modified device<br>monitors<br>operating<br>conditions and<br>requires<br>recalibration<br>based on<br>changed<br>operating<br>conditions |
| Calibration | | | |
| Sterility | Device is not considered<br>sterile | Device is not considered<br>sterile | Same |
| Software<br>Language | C# | LabVIEW | Similar;<br>The change in<br>programming<br>languages had no<br>effect on<br>performance.<br>This change does<br>not present any<br>additional safety |
| | | | or effectiveness<br>concerns. |
| Selection Tools | Circle (StO2%)<br>Region/Perimeter (StO2%<br>and cm²)<br>Line (cm) | Circle (%) | Similar;<br>Modified device<br>displays<br>approximate 2D<br>measures in<br>addition to %StO2 |
| Region of<br>Interest (ROI)<br>Area | Support freehand<br>drawing ROI selection<br>(cm²) | None | Similar; users can<br>draw a custom<br>region of<br>interest.<br><br>The change to<br>include freehand<br>ROI drawing had<br>no effect on the<br>performance of<br>the device. |
| System<br>Components | Camera, Calibration card<br>and recharger | Camera, Calibration card<br>and recharger | Same |
| Image Format | DICOM | DICOM | Same |
| Connectivity | Wi-Fi available only<br>when emailing a report | Wi-Fi not enabled | Similar; Wi-Fi<br>only available<br>when emailing<br>reports.<br><br>The change to<br>turn on Wi-Fi<br>does not affect<br>the performance<br>of the device and<br>does not present<br>any effectiveness<br>concerns or<br>unacceptable<br>risks. |
## Comparison with the Predicate Device
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
# Similarities and Differences
The changes made to create the modified Snapshotwa are software specific. Both devices have the same hardware (LEDs, image sensor, enclosure) and are the same mechanically. Both cameras use spectral analysis to determine oxygenation levels in near-surface tissues. Both cameras display numeric values of approximate oxygen saturation of the hemoglobin as well as displaying the related approximate oxyhemoglobin and deoxyhemoglobin levels necessary for
{7}------------------------------------------------
the oxygen saturation calculation. Both cameras provide two-dimensional mapping of colorcoded oxygenation levels. Both cameras do not come in contact with the patient. Both cameras have the same indications for use and the same fundamental scientific technology.
The modified Snapshotwa has several software specific differences compared to the predicate Kent Camera. Snapshotwa application is written in C# where the predicate device was written in LabVIEW. Snapshotma has a modified algorithm for calculating SQ2. The updated algorithm was implemented to increase the signal to noise ratio and provide better image quality is relies on the same fundamental scientific principles as the predicate algorithm. Lastly, Snaphotwa has an updated user interface which allows for freehand selection of areas and approximate 2D size measurements and can be connected to a wireless network. The user follows the same workflow for capturing images with the modified device as with the predicate.
### Non-Clinical Tests
The modified device went through and passed both internal testing for user and design requirements, as did the predicate device. The modified device compared to the predicate device has no change to the hardware or to the fundamental way it interacts with or configures the hardware therefore although testing wasn't repeated, it is still compliant with the international standards below.
- · Electrical safety and essential performance: ANSI/AAMI ES60601-1:2005/(R)2012
- · Electromagnetic compatibility: IEC 60601-1-2: 2014
Both the predicate and modified devices are battery (DC) powered. The same battery is used in both devices and it is compliant with the following standards:
- Battery safety testing: IEC 62133:2012
- Transportation safety testing of lithium batteries: UN38.3:2009 T
#### Performance Data
A pre-clinical study was conducted comparing tissue oxygen hemoglobin saturation (StO2) measurements taken with the predicate Kent Camera (KD203) and the modified device (KD204). The agreement study used a forearm ischemia protocol to evaluate the performance of the devices both within the expected normal range of S-O2 as well as situations where S-O2 is depressed. The forearm ischemia protocol was intended to test the devices over the clinically meaningful dynamic range of StO2.
The study objectives were as follows:
- . Demonstrate the new StO2 algorithm has superior performance and stability when measuring low reflectivity tissue than the predicate algorithm.
- . Demonstrate the linear relationship between the S-O2 measurements from the two devices over a clinically meaningful dynamic range of StO2.
- Through the use of Bland-Altman plots quantify any scale shift (slope) and bias (difference in mean values) between the devices and estimate the 95% levels of agreement.
{8}------------------------------------------------
# 510(k) Summary
Field data acquired on 38 volunteer subjects undergoing a forearm ischemia - reperfusion challenge confirmed stimulation and lab bench testing that the replacement algorithm provides estimates for relative oxy- [Hbo] and deoxy- [Hb] hemoglobin linearly related to the original (predicate) algorithm R2 > 0.98 with RMSE r[Hb] = 0.000239 and RMSE r[Hbo] = 0.00208. These hemoglobin estimates are used to calculate an estimated tissue hemoglobin oxygenation (StO2). S:O2 = r[Hbo] / (r[Hb] + r[Hbo]). Linear regression in conjunction with Bland – Altman analysis was used to compare StO2 estimates for the predicate algorithm and the proposed replacement algorithm.
The agreement study concluded that StO2 measurements from the modified device (KD204) and the predicate Kent Camera (KD203) show a linear relationship over a wide and clinically meaningful dynamic range of S-O2. These findings support the use of the modified device (KD204) to non-invasively measure superficial tissue hemoglobin oxygen saturation.
### Basis of Substantial Equivalence
The KD204 (K201976) near infrared imaging device has the same indications for use, similar technology characteristics and principles of operation as the predicate device, KD203 (K163070). The modifications between the KD204 and KD203 do not raise new or different types of questions regarding safety and effectiveness. Non-clinical performance data supports the safety and effectiveness of the KD204 device.
Therefore the KD204 (K201976) is considered substantially equivalence to the KD203 (K23432).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.