K240601 · Kent Imaging, Inc. · MUD · Apr 2, 2024 · Cardiovascular
Device Facts
Record ID
K240601
Device Name
SnapshotNIR model KD205
Applicant
Kent Imaging, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Apr 2, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
SnapshotNIR (KD205) is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of: - oxygen saturation (St02), - relative oxyhemogoblin level (Hb02), and - relative deoxyhemoglobin (Hb) level in superficial tissue. SnapshotNIR (KD205) displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions. SnapshotNIR (KD205) is indicated for use to determine oxygenation levels in superficial tissues.
Device Story
SnapshotNIR (KD205) is a non-contact, multispectral imaging device used by healthcare professionals to assess superficial tissue oxygenation. The device illuminates tissue with six near-infrared wavelengths (600nm-1000nm) and white light LEDs, capturing reflected light via a CMOS sensor. It performs pixel-wise spectral analysis to calculate tissue oxygen saturation (StO2), oxyhemoglobin (HbO2), and deoxyhemoglobin (Hb) levels. Results are displayed as two-dimensional, color-coded maps and numeric data on an integrated touchscreen LCD. The device operates from 12 inches away, requiring no patient contact. It improves upon the predicate (KD204) by eliminating the need for manual calibration and utilizing an embedded computer module. Data is stored in an encrypted database on the device, with reporting capabilities via Wi-Fi. Clinical benefit includes non-invasive visualization of tissue perfusion, aiding healthcare providers in clinical decision-making regarding tissue viability.
Clinical Evidence
Bench testing only. Non-clinical engineering-based tests (e.g., IEC 60601-1-2) were performed. A sub-analysis of a pre-clinical dataset compared the KD205 to the KD204 predicate using 2-way ANOVA models across ischemia-reperfusion tests. Results showed no significant differences in StO2 measurements between low and high tissue scattering cohorts for the KD205, representing an improvement over the predicate, which exhibited non-physiological measurements in high scattering conditions due to suboptimal signal-to-noise ratios.
Technological Characteristics
Multispectral imaging system using six near-infrared wavelengths (600nm-1000nm) and white light LEDs. CMOS image sensor with global shutter. Embedded CPU with touchscreen LCD. Non-contact operation. Battery-powered (DC). Connectivity via Wi-Fi for reporting and cloud/USB for software updates. Factory calibrated. Not sterile.
Indications for Use
Indicated for use by healthcare professionals to determine oxygenation levels (StO2, HbO2, Hb) in superficial tissues.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
SnapshotNIR (KD204)
Reference Devices
Kent Camera (KD203)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
April 2, 2024
Kent Imaging Inc % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K240601
Trade/Device Name: SnapshotNIR model KD205 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: March 3, 2024 Received: March 4, 2024
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha L Tanisha L. Hit Hithe -S
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240601
Device Name
SnapshotNIR (KD205)
Indications for Use (Describe)
SnapshotNIR (KD205) is intended for use by healthcare professive tissue oxygenation measurement system that reports an approximate value of:
- oxygen saturation (St02),
- relative oxyhemogoblin level (Hb02), and
- relative deoxyhemoglobin (Hb) level
in superficial tissue. SnapshotNIR (KD205) displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions.
SnapshotNIR (KD205) is indicated for use to determine oxygenation levels in superficial tissues.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> </span> </span> </span> </span> | |
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> </span> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Kent Imaging. The logo features a red square with a white stylized "K" inside. Below the square, the word "KENT" is written in a bold, sans-serif font, with the word "IMAGING" written in a smaller font size directly underneath.
# 510(k) Summary K240601
April 2, 2024
Device Name: SnapshotNIR (KD205)
### Submittal Information:
Post-approval contact: Hitalo Arume
Kent Imaging Inc. Suite 300, 1210 8 Street SW Calgary, AB CANADA T2R 1L3
Phone: 403-455-7610 Fax: 877-664-5450
#### Device and Classification Name
| | SnapshotNIR (KD205) |
|--------------------------|-----------------------------|
| Classification Name | Oximeter, Tissue Saturation |
| Classified Under | 21CFR 870.2700 |
| Classification of Device | Class II |
| Classification Panel | Cardiovascular |
| Product Code | MUD |
Predicate Device: SnapshotNIR (KD204)
Reference Device: Kent Camera (KD203)
### Indications for Use
SnapshotNIR (KD205) is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of:
- oxygen saturation (St02),
- relative oxyhemogoblin level (Hb02), and
- relative deoxyhemoglobin (Hb) level
in superficial tissue. SnapshotNIR (KD205) displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions.
SnapshotNIR (KD205) is indicated for use to determine oxygenation levels in superficial tissues.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Kent Imaging. The logo features a red square with a white stylized "K" inside. Below the square, the word "KENT" is written in a bold, sans-serif font, with the letters spaced slightly apart. Underneath "KENT", the word "IMAGING" is written in a smaller, less bold font.
## Device Description
SnapshotNIR (KD205) is a medical device that operates like a camera. The SnapshotNIR device is completely non-contact, capturing images from 12 inches away from the patient or other imaging field-ofviews. The device uses six near-infrared wavelengths of light and white light emitting diodes (LEDs) to produce a resultant tissue oxygenation image and a colour-based "RGB" or clinical image, respectfully, that can be viewed on the touchscreen display.
SnapshotNIR is based on multispectral imaging technology and performs spectral analysis at each pixel to determine the approximate values of tissue oxygen saturation (StO2), oxyhemoglobin levels (HbO2), and deoxyhemoglobin levels (Hb) in superficial tissues and displays a two-dimensional, color-coded image of the tissue oxygen saturation (StO2).
SnapshotNIR consists of:
- · SnapshotNIR
- · Recharger
- · User Guide
- · Quick Start Guide
- · Sterile Drape (Optional)
The intended use and user interaction remained the same in the new model (KD205). The user will notice that the modified device no longer requires the user to take a calibration image prior to image capture. SnapshotNIR now uses an embedded computer module and a touchscreen LCD. Additional minor form factor changes were made on the new model (KD205) to accommodate the hardware change. With respect to the software used by SnapshotNIR, the predicate and new model (KD205) operate similarly.
## Fundamental Principles
The SnapshotNIR (KD205) has identical fundamental working principles to the predicate camera. SnapshotNIR builds on the standard principles used in other oximeters and tissue oxygenation measurement systems. Tissue oxygenation devices expose tissue with light of known wavelengths and capture reflected light that is not absorbed by the tissue. Wavelengths are chosen to provide spectral information about the hemoglobin oxygen content. Tissue oxygen saturation (StO2) is calculated as:
$$StO_2 = \frac{[HbO_2]}{[Hb] + [HbO_2]}$$
For viable (blood-perfused) tissue, hemoglobin is a dominant absorber of light within the near infrared (near-IR) region of the electromagnetic spectrum, specifically between 600nm to 1000nm. The well documented and established absorption spectra exhibited by oxygenated hemoglobin (HbO2) and deoxygenated hemoglobin (Hb) enables near-IR spectroscopy to determine the proportions of HbO2 and Hb within the tissue. SnapshotNIR visualizes the tissue oxygen saturation levels at each image pixel reflected from the superficial capillary bed of the tissue being imaged based on the reflectance characteristics of HbO2 and Hb.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Kent Imaging. The logo features a red square with a white stylized "K" inside, positioned above the word "KENT" in a bold, sans-serif font. Below "KENT" is the word "IMAGING" in a smaller, less bold font. The overall design is clean and professional, suggesting a company involved in imaging technology or services.
## Nonclinical data
Nonclinical engineering-based tests (e.g., IEC 60601-1-2, etc) among other performance bench tests support conclusions that the KD205 device is safe for use with a low risk of adverse events occurring during the intended use scenarios. These tests and findings are analogous to those conducted on the predicate KD204 SnapshotNIR device and demonstrate that the KD205 device is similarly safe, and effective compared to a current legally marketed device.
Finally, the KD205 SnapshotNIR device has made improvements over the predicate KD204 device. In a sub-analysis of the pre-clinical dataset, the KD205 device demonstrated no significant differences (derived via 2-way ANOVA models) in STO2 across the ischemia-reperfusion test when comparing cohorts of low versus high tissue scattering effects. This outcome is considered an improvement comparative to the KD204 device whereby non-physiological measurements were derived from the high tissue scattering cohort due to suboptimal signal-to-noise ratios of near-infrared light which impairs the calculation of STO2.
### Substantial Equivalence
The modified device (SnapshotNIR, model KD205) has the same regulatory information as the predicate device (Snapshotsin, model KD204) and the same regulatory information as the reference device (Kent Camera, model KD203).
| Comparative Feature | SnapshotNIR - KD205 | Predicate SnapshotNIR – KD204 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Same | SnapshotNIR is intended for use by healthcare professionals as a non-invasive tissue oxygenation measurement system that reports an approximate value of oxygen saturation (StO2), oxyhemoglobin level (HbO2), and deoxyhemoglobin (Hb) level in superficial tissue. SnapshotNIR displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports multispectral tissue oxygenation measurements for selected tissue regions. SnapshotNIR is indicated for use to determine oxygenation levels in superficial tissues. |
| Measurements | Same | Oxygen saturation Oxyhemoglobin level Deoxyhemoglobin level |
| Method of Measurement | Same | Non-invasive, non-patient contacting imaging head illuminates the surface and receives returned light. |
| | Same | 6 wavelengths between 600nm and 1000nm. |
| Comparative Feature | SnapshotNIR -<br>KD205 | Predicate SnapshotNIR – KD204 |
| | Same | A CMOS image sensor with global shutter is<br>used as the detector. |
| | Same | Spectral analysis at specific wavelengths of light<br>returned from the target tissue. |
| Light Source | Same | LEDs |
| Lens | Same | Corrected and optimized for use from 400nm to<br>1000nm. |
| Output Display | Same | Two-dimensional color-coded map of<br>• estimated oxygen saturation<br>• Numeric data |
| Power Source | Same | DC (battery-powered) |
| Patient Contact | Same | None |
| Control Method | Same | Computer controlled. |
| Calibration | Device is factory<br>calibrated to<br>account for range<br>of operating<br>conditions. | Preformed at start-up by operator and repeated if<br>operating conditions of camera change to the<br>degree that necessitates recalibration. |
| Sterility | Same | Device is not considered sterile. |
| Software Language | Same | C# |
| Selection Tools | Same | Circle<br>Region/Perimeter<br>Line |
| Region of Interest (ROI) Area | Same | Support freehand drawing ROI selection (cm2). |
| User Interface | Embedded CPU,<br>touchscreen LCD<br>and battery<br>incorporated into<br>camera enclosure. | Off-the-shelf touchscreen PC incorporated in<br>enclosure with camera and battery. |
| System Components | Camera and<br>recharger. | Camera, calibration card and recharger. |
| Data storage | Same | Patient images are stored in a proprietary secure<br>format on the device and patient information is<br>stored in an encrypted database on the device.<br>Patient images and database are hidden on<br>device. |
| Comparative Feature | SnapshotNIR - KD205 | Predicate SnapshotNIR – KD204 |
| Connectivity and Reporting | Same | User can create reports and email through Wi-Fi connectivity. |
| | Software upgrades performed via USB or through cloud.<br>Log files can also be uploaded to cloud for support. | Software upgrades performed via USB. |
| Expiration Date | Same | No expiration date. |
Detailed Comparative Features between Devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Kent Imaging. The logo features a red square with a white "K" inside it. Below the square is the word "KENT" in a bold, blue font. Underneath "KENT" is the word "IMAGING" in a smaller, less bold, blue font.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Kent Imaging. The logo features a red square with a stylized white "K" inside. Below the square, the word "KENT" is written in a bold, sans-serif font. Underneath "KENT", the word "IMAGING" is written in a smaller, sans-serif font.
## Conclusion
The modified SnapshotNIR (KD205) is substantially equivalent to the predicate Snapshotsm (KD204)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.