The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs(thyroid, breast and testicle), trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatic), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and Thoracic/Pleural. This system provides high-quality ultrasound images in the following modes B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
Device Story
Mobile, compact diagnostic ultrasound system; uses various probes (2-30 MHz) to acquire ultrasound signals. Transforms signals into B-mode, M-mode, Doppler, speckle-tracking, elastography, and attenuation maps. Operated by physicians/qualified personnel in clinical settings. Features include Smart Fetal Heart (automated view acquisition), LI-RADS Checklist (liver characterization), Fluctuational imaging, Workflow Navigator (ASE-based reporting), Smart Optimization (image quality), and Prostate Biopsy Report (3D modeling). Output displayed on system monitor for diagnostic visualization. Assists clinicians in anatomical assessment and procedural guidance; improves workflow efficiency and diagnostic standardization.
Clinical Evidence
Bench testing and volunteer studies conducted to validate new and improved features. No clinical trials on patient outcomes reported. Performance testing confirmed that improvements met design specifications and functional requirements.
Technological Characteristics
Mobile ultrasound system; probes (flat/convex/sector) 2-30 MHz. Modes: B, M, CW/PW/Color/Combination Doppler, Speckle-tracking, Tissue Harmonic, Shear wave, Elastography, Attenuation mapping. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, ISO 10993-1. Software: Moderate Level of Concern. Cybersecurity controls per FDA guidance.
Indications for Use
Indicated for visualization of structures and dynamic processes in fetal, abdominal, intra-operative, pediatric, small organs (thyroid, breast, testicle), trans-rectal, neonatal/adult cephalic, cardiac, peripheral vascular, transesophageal, musculo-skeletal, laparoscopic, and thoracic/pleural applications. For use by physicians or qualified persons in hospital/clinical settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
October 8, 2020
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Canon Medical Systems Corporation % Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive TUSTIN CA 92780
# Re: K201972
Trade/Device Name: Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V5.1 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 13, 2020 Received: July 15, 2020
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K201972
## Device Name
Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V5.1
Indications for Use (Describe)
The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs(thyroid, breast and testicle), trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatic), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and Thoracic/Pleural. This system provides high-quality ultrasound images in the following modes B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# AL SYSTEMS USA, INC.
# 510(k) SUMMARY
K201972
- 1. SUBMITTER'S NAME Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi-ken, Japan 324-8550
- OFFICIAL CORRESPONDENT 2. Fumiaki Teshima
- 3. ESTABLISHMENT REGISTRATION 9614698
#### 4. CONTACT PERSON
Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
#### 5. Date Prepared
July 13, 2020
#### 6. DEVICE NAME
Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V5.1
#### 7. TRADE NAME(S)
Diagnostic Ultrasound System Aplio i900 Model TUS-AI900 Software Version V5.1 Aplio i800 Model TUS-AI800 Software Version V5.1 Aplio i700 Model TUS-AI700 Software Version V5.1
#### 8. COMMON NAME
System, Diagnostic Ultrasound
#### 9. DEVICE CLASSIFICATION
Class II
Ultrasonic Pulsed Doppler Imaging System – Product Code: 90-IYN [per 21 CFR 892.1550] Ultrasonic Pulsed Echo Imaging System – Product Code: 90-IYO [per 21 CFR 892.1560] Diagnostic Ultrasonic Transducer – Product Code: 90-ITX [per 21 CFR 892.1570]
{4}------------------------------------------------
# 10. PREDICATE DEVICE
| Product | Marketed by | 510(k) Number | Clearance Date |
|------------------------------------------------------------------------------------------|------------------------------|---------------|--------------------|
| Aplio i900/i800/i700 Diagnostic<br>Ultrasound System, V4.0<br>(Primary Predicate Device) | Canon Medical<br>Systems USA | K191467 | July 19, 2019 |
| Versana Premier<br>(Reference Predicate Device) | GE Healthcare | K182277 | November 21, 2018 |
| Voluson E10<br>(Reference Predicate Device) | GE Healthcare | K172342 | September 29, 2017 |
| AIXPLORER® MACH Ultrasound<br>Diagnostic Systems<br>(Reference Predicate Device) | Super Imagine | K180572 | May 29, 2018 |
| Epiq 7<br>(Reference Predicate Device) | Philips Medical<br>Systems | K181485 | July 27, 2018 |
# 11. DEVICE DESCRIPTION
The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800, and Aplio i700 Model TUS-Al700, V5.1 are mobile, compact diagnostic ultrasound systems which implement the latest technologies. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 30 MHz.
# 12. INDICATIONS FOR USE
The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs(thyroid, breast and testicle), trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and Thoracic/Pleural. This system provides high-quality ultrasound images in the following modes B mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler, Power Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
{5}------------------------------------------------
# 13. SUBSTANTIAL EQUIVALENCE
The purpose of this submission is the modification of the cleared Aplio i900/i800/i700, Diagnostic Ultrasound System, V4.0, K191467, marketed by Canon Medical Systems USA, to which the subject device is substantially equivalent.
The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800, and Aplio i700 Model TUS-Al700, V5.1 employ the same fundamental scientific technology as the predicate devices, and functions in a manner similar to and is intended for the same use as the predicate devices.
The subject device includes modifications to the cleared device which improves upon existing features. This submission also includes details regarding new features determined to be substantially equivalent to features cleared under the predicate devices referenced within this submission.
- . The Aplio i900/i800/i700 and predicate Aplio i900/i800/i700 have the same clinical intended use and the same imaging modes, except that Thoracic/Pleural is now available on several transducers on the subject Aplio i900/i800/i700, and Shear Wave Elastography mode was expanded to include musculo-skeletal clinical use.
- . The transducers supported in the subject Aplio i900/i800/i700 and predicate Aplio i900/i800/i700 are identical except the addition of new transducer PLI-605BX which shares the same fundamental technological and biocompatibility characteristics as existing transducers supported by the predicate Aplio i900/i800/i700.
- . The software features supported in subject Aplio i900/i800/i700 and predicate Aplio i900/i800/i700 are identical except the following main, new features:
- O Smart Fetal Heart, a new feature which improves upon existing manual workflow to acquire fetal heart views, available on predicate Aplio i900/i800/i700.
- о LI-RADS Checklist, a new feature predicated upon existing features available on predicate Aplio i900/i800/i700, which supports the standardization of liver characterization based on American College of Radiology, (ACR) guidelines.
- O Fluctuational imaging, a new feature similar to functionality available with Attenuation Imaging, a feature available on predicate Aplio i900/i800/i700, which displays the degree of fluctuation in tissue.
- O Workflow Navigator, a new feature predicated on existing workflow functionality available on predicate Aplio i900/i800/i700 which provides workflow and reporting support based on American Society of Echocardiography, (ASE) guidelines for specific medical conditions.
- O Smart Optimization, a new feature similar to "Whizz" on the Versana Premier by GE Healthcare (K182277) which is intended to improve image quality targeting the abdominal region.
- o Prostate Biopsy Report, a new feature which adds functionality existing in predicate Aplio i900/i800/i700, to provide a workflow panel for biopsy procedures and 3D modeling for visualization during these procedures.
- . In addition to these new features, several existing features available with predicate Aplio i900/i800/i700 (K191467) have been modified for improved functionality in the subject device:
- Smart Body Mark, an improvement to manual breast scan workflow by O incorporation of previously cleared functionality
- MSK Protocol movie, to support the display of musculo-skeletal (MSK) instructional O movies with existing functionality
- O Wide View support, an expansion of the scan range available with convex transducers
- O iBeam+, an improvement to conventional iBeam by increased processing speed
{6}------------------------------------------------
- о SMI G4, an improvement to the previous generation of SMI by enhanced clutter suppression, supported by the availability of a higher detection range
- Sensor 3D, modified to support improved resolution o
- O Obstetric Image quality, modified to support improved visualization
- Cardiac 4D Luminance, an improvement to conventional Luminance by migration of o a previously cleared feature
## 14. SAFETY
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the AAMI/ANSI ES60601-1:2012, IEC 60601-1-2 (2014), IEC 60601-2-37 (2015), IEC 62304 (2015), IEC 62359 (2017) and ISO 10993-1(2009) standards.
## 15. TESTING
Risk Analysis, Verification/Validation testing conducted through bench testing and clinical evaluation which are included in this submission demonstrates that the requirements for the new and improved features have been met. Additional performance testing, using test data and volunteer studies, were conducted to assess improvements to existing features. Results of all these studies demonstrated that the improvements met specifications are performed as intended.
FDA guidance document "Marketing Clearance of Diagnostic Ultrasound Systems and Transducer", issued June 27, 2019 was referenced for this submission, along with Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005.
Cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 18, 2018, was also referenced for this submission.
Additionally, testing of this device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and UL systems.
### 16. CONCLUSION
The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800, and Aplio i700 Model TUS-Al700, V5.1 is substantially equivalent to the predicate devices. The subject devices function in a manner similar to and is intended for the same use as the predicate devices, as described in the labeling. Based upon the bench testing, clinical evaluation, successful completion of software validation, application of risk management and design controls, it is concluded that this device is safe and effective for its intended use and performance is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.