Inspire 6M Hollow Fiber oxygenator, Inspire 7M Hollow Fiber oxygenator, Inspire 8M Hollow Fiber oxygenator
K201916 · Sorin Group Italia S.R.L. · DTZ · Aug 13, 2020 · Cardiovascular
Device Facts
| Record ID | K201916 |
| Device Name | Inspire 6M Hollow Fiber oxygenator, Inspire 7M Hollow Fiber oxygenator, Inspire 8M Hollow Fiber oxygenator |
| Applicant | Sorin Group Italia S.R.L. |
| Product Code | DTZ · Cardiovascular |
| Decision Date | Aug 13, 2020 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.4350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The INSPIRE 6M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 6M is intended to be used for 6 hours or less. The INSPIRE 7M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 7M is intended to be used for 6 hours or less. The INSPIRE 8M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 8M is intended to be used for 6 hours or less.
Device Story
The Inspire 6M, 7M, and 8M are hollow fiber oxygenators with integrated heat exchangers used during cardiopulmonary bypass (CPB). The device receives venous or suction blood; a heat exchanger (polyurethane hollow fiber bundle) controls blood temperature via hypothermia or normothermia maintenance; an oxygenating module (coiled polypropylene microporous hollow fiber bundle) performs gas exchange (oxygenation and CO2 removal). The device is operated by perfusionists or surgeons in an operating room setting. The output is oxygenated, temperature-regulated blood returned to the patient's circulatory system. The device supports patient physiological function during surgery. Modifications include a new supplier for the purge line unidirectional valve and updated IFU warnings.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical in vitro performance testing, including biocompatibility (ISO 10993-1), functional testing per ISO 7199, and evaluation of the modified purge line unidirectional valve on sterile aged devices (accelerated aging equivalent to 3 years).
Technological Characteristics
Integrated oxygenator and heat exchanger. Materials: polyurethane hollow fibers (heat exchanger), polypropylene microporous hollow fibers (oxygenator). Single-use, ethylene oxide sterilized, non-pyrogenic fluid path. Connectivity: None. Operation: Passive gas exchange and heat transfer via fluid circulation.
Indications for Use
Indicated for adult and small adult patients undergoing surgical procedures requiring cardiopulmonary bypass (CPB) for gas exchange support and blood temperature control for durations of 6 hours or less.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
- INSPIRE 6M Hollow Fiber Oxygenator, INSPIRE 8M Hollow Fiber Oxygenator (K180448)
- INSPIRE 7M Hollow Fiber Oxygenator (K190690)
Related Devices
- K190690 — INSPIRE 7 Hollow Fiber Oxygenator with integrated Hardshell/Venous Cardiotomy Reservoir, INSPIRE 7M Hollow Fiber Oxygenator, INSPIRE 7 Dual Hollow Fiber Oxygenator with integrated Hardshell/Venous Cardiotomy Reservoir · Sorin Group Italia S.R.L. · Apr 13, 2019
- K113626 — INSPIRE 6 HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY RESERVOIR · Sorin Group Italia S.R.L. · Apr 5, 2012
- K200683 — INSPIRE 7FM Hollow fiber oxygenator with integrated arterial filter, INSPIRE 7F Hollow fiber oxygenator with integrated arterial filter and hardshell reservoir, INSPIRE 7F DUAL Hollow fiber oxygenator with integrated arterial filter and hardshell reservoir · Sorin Group Italia S.R.L. · Jun 12, 2020
- K130209 — INSPIRE 6 HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDHELL RESERVOIR, INSPIRE 6F HOLLOW FIBER OXYGENATOR WITH INTEGRATED · Sorin Group Italia S.R.L. · Feb 12, 2013
- K120185 — INSPIRE 6F HOLLOW FIBER OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND HARDSHELL VENOUS/CARDIOTOMY RESERVIOR · Sorin Group Italia S.R.L. · May 11, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
August 13, 2020
Sorin Group Italia S.r.l Luigi Vecchi Director, Regulatory Affairs Via Statale 12 Nord, 86 Mirandola, Modena 41037 Italy
Re: K201916
Trade/Device Name: Inspire 6M Hollow Fiber oxygenator, Inspire 7M Hollow Fiber oxygenator, Inspire 8M Hollow Fiber oxygenator Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ, DTR Dated: July 9, 2020 Received: July 10, 2020
Dear Luigi Vecchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201916
Device Name INSPIRE 6M Hollow fiber Oxygenator INSPIRE 7M Hollow fiber Oxygenator INSPIRE 8M Hollow fiber Oxygenator
### Indications for Use (Describe)
The INSPIRE 6M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 6M is intended to be used for 6 hours or less.
The INSPIRE 7M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 7M is intended to be used for 6 hours or less.
The INSPIRE 8M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 8M is intended to be used for 6 hours or less.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(in accordance with 21 CFR 807.92)
## 510(k) Number: K201916
#### l. Applicant Information
| APPLICANT: | Sorin Group Italia S.r.l.<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) ITALY |
|----------------------------|------------------------------------------------------------------------------------|
| CONTACT PERSON: | Luigi Vecchi<br>Phone: +39 0535 29957<br>e-mail: luigi.vecchi@livanova.com |
| APPLICATION CORRESPONDANT: | Sorin Group Italia S.r.l.<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) ITALY |
| CONTACT PERSON: | Luigi Vecchi<br>Phone: +39 0535 29957<br>e-mail: luigi.vecchi@livanova.com |
| DATE PREPARED: | July 6, 2020 |
#### II. Subject Devices Identification
| PROPRIETARY NAME: | INSPIRE 6M Hollow Fiber Oxygenator<br>INSPIRE 7M Hollow Fiber Oxygenator<br>INSPIRE 8M Hollow Fiber Oxygenator |
|-----------------------|----------------------------------------------------------------------------------------------------------------|
| COMMON/ USUAL NAME: | INSPIRE 6M; INSPIRE 7M; INSPIRE 8M. |
| CLASSIFICATION NAME: | Cardiopulmonary Bypass Oxygenator,<br>Cardiopulmonary Bypass Heat Exchanger. |
| REGULATION NUMBER: | 21 CFR 870.4350 |
| PRODUCT CODE: | DTZ, DTR |
| CLASSIFICATION: | Class II |
| CLASSIFICATION PANEL: | Cardiovascular |
#### III. Predicate Devices
The INSPIRE 6M Hollow Fiber Oxygenator, INSPIRE 7M Hollow Fiber Oxygenator and INSPIRE 8M Hollow Fiber Oxygenator are substantially equivalent to the following cleared predicate devices. All models have the same
{4}------------------------------------------------
fundamental scientific technology and intended use
| 510(k) NUMBER: | K180448 |
|-----------------------|----------------------------------------------------------------------------|
| PROPRIETARY NAME: | INSPIRE 6M Hollow Fiber Oxygenator,<br>INSPIRE 8M Hollow Fiber Oxygenator. |
| COMMON/ USUAL NAME: | INSPIRE 6M; INSPIRE 8 M. |
| CLASSIFICATION NAME: | Cardiopulmonary Bypass Oxygenator |
| REGULATION NUMBER: | 21 CFR 870.4350 |
| CLASSIFICATION: | Class II |
| CLASSIFICATION PANEL: | Cardiovascular |
| 510(k) NUMBER: | K190690 |
|-----------------------|------------------------------------|
| PROPRIETARY NAME: | INSPIRE 7M Hollow Fiber Oxygenator |
| COMMON/ USUAL NAME: | INSPIRE 7M |
| CLASSIFICATION NAME: | Cardiopulmonary Bypass Oxygenator |
| REGULATION NUMBER: | 21 CFR 870.4350 |
| CLASSIFICATION: | Class II |
| CLASSIFICATION PANEL: | Cardiovascular |
#### Device Description IV.
The INSPIRE 6M, INSPIRE 7M and INSPIRE 8M oxygenators (hereinafter identified as INSPIRE) consist of an oxygenator with an integrated heat exchanger.
The INSPIRE devices are intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control.
The INSPIRE consist of the following main components
- a heat exchanger consisting of a bundle of polyurethane hollow fibers ● that are wound on a cylindrical core. it controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery.
- an oxygenating module element made of a coiled bundle of polypropylene microporous hollow fibers rolled on the heat exchanger sub assembly. The hollow fiber membrane provides oxygenation and carbon dioxide removal from venous blood or suction blood.
The modified devices are modified versions of the currently marketed INSPIRE 6M, INSPIRE 7M and INSPIRE 8M products.
{5}------------------------------------------------
#### V. Indications for use
The subject devices (i.e., the INSPIRE 6M, INSPIRE 7M and INSPIRE 8M) are intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. They all provide gas exchange support and blood temperature control. All the devices are intended to be used for 6 hours or less. Note: The intended use of the subject devices is identical to the intended use of the respective predicates
#### VI. Summary of technical characteristics
The INSPIRE subject devices have the same principles of operation and control mechanisms as the INSPIRE unmodified device. The INSPIRE subject devices and the INSPIRE unmodified device share the same fundamental technological characteristics except for some modifications that do not affect the basic device function. These differences are summarized below and do not raise any new issues of safety and effectiveness.
- 1. The purge line unidirectional valve present in the recirculation line of the INSPIRE devices was changed with another valve from a different supplier. made with different materials and having different performances;
- 2. A warning in the Instructions for Use was revised to reflect the evidence found during V&V testing.
No change to the intended use has been made as a result of these modifications. Also, there are no differences in packaging type and material between INSPIRE and INSPIRE unmodified device. Both modified and unmodified device are for single use only, ethylene oxide sterilized and has a non-pyrogenic fluid path
#### VII. Substantial equivalence discussion
Based on equivalent intended use and technological characteristics, as well as on equivalent performance testing, the INSPIRE can be deemed to be substantially equivalent to its predicate device, the Unmodified INSPIRE The INSPIRE as designed and manufactured, does not raise new questions reqarding safety and effectiveness as compared to its predicate device and is determined to be substantially equivalent to its predicate device, the Unmodified INSPIRE
#### VIII. Non clinical performance data
The subject devices were tested to ensure that they can provide all the capabilities necessary to operate safely and effectively. Applicable tests were carried out in accordance with the requirements of ISO 10993-1, the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility
{6}------------------------------------------------
testing of materials, and the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions: Final Guidance for Industry and FDA Staff' issued on November 13, 2000, "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions; Final Guidance for Industry and FDA" issued on November 29, 2000; and ISO 7199 "Cardiovascular implants and artificial organs - Blood-gas exchangers (oxyqenators)".
In vitro testing was performed to compare devices with the modified and unmodified purqe line unidirectional valve.
This performance testing was conducted on sterile aged devices; accelerated aging for a period of time equivalent to at least 3 years as per device labeling The oxygenating module with the modified purge line unidirectional valve.
successfully met all acceptance criteria for the recirculation line handling.
The results of in vitro studies demonstrate that the subject INSPIRE performs in a manner substantially equivalent to the Unmodified INSPIRE predicate device with respect to the relevant functional parameters.
#### IX. Clinical performance data
No clinical testing was conducted in support of the INSPIRE, as the indications for use are equivalent to those of their respective predicates, which have been on the market for many years. The non-clinical testing summarized in this submission supports the substantial equivalence of these devices with their respective predicates in relation to the changes subject of this submission.
#### X. Statement of Substantial Equivalence
As designed and manufactured and based on the intended use, technological characteristics, and performance testing, the modified INSPIRE do not raise new questions reqarding their safety and effectiveness as compared to their predicates devices and are determined to be substantially equivalent to the predicate devices.