The device is used for the endoscopic removal of biliary stones and foreign bodies.
Device Story
Disposable stone extraction basket for endoscopic removal of biliary stones and foreign bodies. Device consists of stainless steel or nitinol basket at distal end of PTFE sheath with handle. Operator advances basket out of sheath and retracts it using handle actuator to capture stones/foreign bodies. Used in clinical settings by physicians during endoscopic procedures. Output is physical capture and retrieval of biliary objects. Benefits include minimally invasive removal of stones/foreign bodies. Device is single-use and EO sterilized.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing (tensile strength, flexibility, rotation) and biocompatibility testing (ISO 10993-5 cytotoxicity, ISO 10993-10 sensitization and intracutaneous reactivity).
Technological Characteristics
Materials: Stainless steel (SUS 303, SUS 304), Nitinol, PTFE. Sensing/Actuation: Mechanical manual actuation via handle. Dimensions: Sheath OD 1.8-3.2mm; basket open width 10-30mm; working length 700-1950mm. Connectivity: None. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for the endoscopic removal of biliary stones and foreign bodies in patients requiring biliary intervention.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
December 16, 2020
Hangzhou AGS MedTech Co., Ltd. Chunqi Han R&D Director Building 5, Building 6, No.597 Kangxin Road Yuhang District Hangzhou. Zhejiang 311106 China
Re: K201509
Trade/Device Name: Disposable Stone Extraction basket Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: Class II Product Code: LQR Dated: November 18, 2020 Received: November 20, 2020
#### Dear Chunqi Han:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K201509
Device Name
Disposable Stone Extraction Basket
Indications for Use (Describe)
The device is used for the endoscopic removal of biliary stones and foreign bodies.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue, surrounding a green triangle in the center. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters. The text is in a dark color, contrasting with the white background.
# K201509 510(k) Summary
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes forming a triangle in the center, with a green triangle in the middle. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters.
# 510(k) Summary
We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.
# 5.1 Submitter
| Submitted by/Owner: | Hangzhou AGS MedTech Co., Ltd.<br>Building 5, Building 6, NO.597 Kangxin Road Yuhang<br>District, Hangzhou, Zhejiang 311106 China |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 3010288205 |
| Registration Status: | Active |
| Contact Person: | Yanping Fu<br>Phone: 0086-15958493282<br>Fax: 0086-0571-87671225<br>Email: fuyp@bioags.com |
| Date Prepared: | July 18, 2019 |
## 5.2 Proposed Device
| Trade Name: | / |
|-------------------------|------------------------------------|
| Device Name: | Disposable Stone Extraction Basket |
| Common Name: | Disposable Stone Extraction Basket |
| Regulation class: | Class II |
| Regulation Number: | 876.5010 |
| Regulation Description: | Biliary Catheter and Accessories |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | LQR |
| Product Code Name: | Dislodger, Stone, Biliary |
## 5.3 Predicate Device
| Trade Name: | MEMORY BASKET 5F SOFT WIRE |
|-------------------------|-----------------------------------|
| Device Name: | MEMORY BASKET 5F SOFT WIRE |
| Common Name: | Non-Lithotripsy Extraction Basket |
| 510(k) Number: | K171969 |
| Regulation class: | Class II |
| Regulation Number: | 876.5010 |
| Regulation Description: | Biliary Catheter and Accessories |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | LQR |
| Product Code Name: | Dislodger, Stone, Biliary |
## 5.4 Device Description
The subject Disposable Stone Extraction Basket is comprised of a stainless steel/nitinol
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with a circular design on the left and text on the right. The circular design is composed of three curved shapes, two in blue and one in green, arranged in a triangular pattern. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by Chinese characters below it. The text is in a dark color, contrasting with the white background.
K201509
basket at the distal end of a sheath with handle. Basket is advanced out of and retracted into the single lumen polytetrafluoroethylene catheter sheath using the handle component's actuator. The device is not compatible with any mechanical lithotripter. EO Sterilization and use for single use only.
## 5.5 Indication for use statement
The device is used for the endoscopic removal of biliary stones and foreign bodies.
## 5.6 Comparison of Technology Characteristics
Our proposed device Disposable Stone Extraction Basket is substantially equivalent to the predicate devices. The differences between the Disposable Stone Extraction Basket and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below:
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a blue and green circular design on the left and the text "AGS MEDTECH" and "安杰思医学" on the right. The circular design consists of three blue curved shapes that form a triangle, with a green triangle in the center. The text "AGS MEDTECH" is in a smaller font size than the Chinese characters "安杰思医学".
| Item | | Proposed device | Predicate device | Comparison |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name | | Disposable Stone Extraction Basket | Non-Lithotripsy Extraction Basket | / |
| Trade name | | / | MEMORY BASKET 5F SOFT WIRE | / |
| Model number | | 5061 series, 5062 series, 50631 series, 5064 series,<br>5065 series, 50610 series, 50611 series, 50613 series,<br>50614 series. | MSB5-2X4 | / |
| 510(k) submitter | | Hangzhou AGS MedTech Co., Ltd. | Wilson-Cook Medical, Inc. | / |
| 510(k) number | | / | K171969 | / |
| Technical | Sheath<br>Max outer<br>diameter | 5061/5062/50610/50611:1.8/2.4/3.2mm;<br>5063:2.4mm;<br>5064/5065/50613/50614:1.8mm; | 5Fr ( $\approx$ 1.67mm ) | Similar.<br>For different size of endoscope channel, we design<br>different sheath max. outer diameter, operator can<br>choose different sheath max. outer diameter according<br>to different clinical use situation. The difference does<br>not raise different questions regarding its safety and<br>effectiveness. |
| | Minimum<br>Accessory<br>Channel | 2.0mm/2.8mm/3.7mm | 2mm | Similar.<br>Difference size of sheath max outer diameter suit for<br>different endoscope channel. The difference does not<br>raise different questions regarding its safety and<br>effectiveness. |
| | Open<br>width | 10,15,20,25,30mm | 20mm | Similar.<br>We design different open width of basket for different<br>size stone or foreign bodies. The difference does not |
| Item | | Proposed device | Predicate device | Comparison |
| | | | | raise different questions regarding its safety and<br>effectiveness. |
| | Working<br>length | 5061/2/3/10/11: 1950mm;<br>5064/5/13/14:700mm; | 2000mm | Similar. |
| | Handle<br>type | Two handle type:<br>One handle type: | Rotatable pin vise, two handle type. | Similar.<br>We design one handle type for simple operation.<br>The difference does not raise different questions<br>regarding its safety and effectiveness. |
| | Performan<br>ce | Tensile strength;<br>Operation smoothly;<br>Rotate smoothly (for rotatable models) ; | Tensile Testing;<br>Flexibility Testing;<br>Rotate smoothly; | Different.<br>We conduct bench performance for our proposed<br>device and the predicate device, please refer to<br>Section 18 of this submission, the test results show<br>that our proposed device is substantial equivalence<br>with the predicate device. |
| Biological | Materials<br>or<br>substances<br>in contact<br>with the<br>same<br>human<br>tissue or<br>body<br>fluids | 5063: SUS 303, SUS 304, PTFE;<br>5061-5062, 5064-5065, 50610, 50611, 50613, 50614:<br>SUS 303, Nitinol, PTFE; | Basket: metal wire;<br>Catheter sheath: polytetrafluoroethylene<br>(PTFE); | Different.<br>Biocompatibility tests have been done for the<br>difference. Biological risks are acceptable, please<br>refer to Section 15 of this submission. |
| Item | Proposed device | Predicate device | Comparison | |
| Biocompatibility | In Vitro Cytotoxicity Test<br>ISO 10993.5-2009;<br>Skin sensitization Test,<br>Intracutaneous Reactivity Test<br>ISO 10993.10-2010; | Unknown | | |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by the same text in Chinese characters. The logo appears to be for a medical technology company.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by Chinese characters below it. The overall design is clean and professional, suggesting a company in the medical technology field.
# 5.7 Applicable Guidance Document
NA
# 5.8 Performance Data
The Disposable Stone Extraction Basket meets all device standards for biocompatibility (ISO 1099) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device.
# 5.9 Clinical Test
No Clinical test is included in this submission.
# 5.10 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS Medtech Co., Ltd has demonstrated that proposed device Disposable is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.