K201107 · Gtg Wellness Co., Ltd. · OHS · Jun 22, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201107
Device Name
Opera Lebody (Gold), Opera Lebody (Zafiro)
Applicant
Gtg Wellness Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jun 22, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OPERA LEBODY is an over the counter device that is intended for the use in the treatment of full face wrinkles.
Device Story
OPERA LEBODY is a wearable LED face mask for wrinkle reduction; utilizes red (630nm) and IR (830nm) light to induce photobiological effects. Device components include mask with internal LEDs, controller for power/operation, cradle, pouch, and USB cable. Operated by user in home setting; device features protective eye guides to prevent ocular exposure. Controller manages treatment duration, with automatic shut-off after 10 minutes. Output is non-ionizing LED light; intended to improve skin appearance by reducing wrinkles.
Clinical Evidence
Bench testing only. Evaluated wavelength, output power, and total irradiance. Safety and performance verified against IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 62471 (photobiological safety), ISO 14971 (risk management), and IEC 60601-1-6 (usability). Biocompatibility addressed via use of materials (silicone and ABS plastics) with history of safe use in medical devices.
Technological Characteristics
LED-based light therapy device. Emits red (630nm) and IR (830nm) light. Materials: silicone and ABS plastics. Power: USB-charged battery. Connectivity: None. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62471, ISO 14971, IEC 60601-1-6. Form factor: wearable face mask with controller.
Indications for Use
Indicated for the treatment of full face wrinkles in the general population using an over-the-counter device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 72 (K163329)
{0}------------------------------------------------
June 22, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
GTG Wellness Co., Ltd. % Paweena U-Thainual CEO MDR Solutions Co., Ltd. 1435 Kanchanapisek Rd., Bang Khae Nuea Bangkok, Bang Khae 10160, Thailand
Re: K201107
Trade/Device Name: Opera Lebody (Gold), Opera Lebody (Zafiro) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: April 23, 2020 Received: April 24, 2020
Dear Paweena U-Thainual:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201107
Device Name OPERA LEBODY
Indications for Use (Describe)
OPERA LEBODY is an over the counter device that is intended for the use in the treatment of full face wrinkles.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201107: 510(k) Summary
## 1. General Information
| Applicant/Submitter: | GTG Wellness Co., Ltd. |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 7, Dongwon-ro, Bundang-gu,<br>Seongnam-si, Gyeonggi-do,<br>Rep. of Korea (South Korea)<br>Tel: +82 70-4733-8244 |
| Contact Person: | Paweena U-Thainual, Ph.D.<br>MDR Solutions Co., Ltd. |
| Address: | 1435 Kanchanapisek Rd.,<br>Bang Khae Nuea, Bang Khae,<br>Bangkok 10160, THAILAND<br>Tel: +66 2-804-2101<br>Email: paweena@mdrsolutions.co.th |
| Preparation Date: | April 3rd, 2020 |
## 2. Device Name and Code
| Device Trade Name: | OPERA LEBODY |
|----------------------|------------------------------------------------|
| Common Name: | OPERA LEBODY |
| Classification Name: | Light Based Over The Counter Wrinkle Reduction |
| Product Code: | OHS |
| Regulation Number: | 878.4810 |
| Classification: | Class II |
| Review Panel: | General & Plastic Surgery |
# 3. Predicate Device
OPERA LEBODY is substantially equivalent to the following legally marketed predicate devices
| | Primary Predicate Device | Predicate Device |
|-----------------|--------------------------------------------------------|-------------------|
| 510(K) Number | K163329 | K133896 |
| Manufacturer | Pulsaderm LLC | Trophy Skin, Inc. |
| Device Name | Pulsaderm Wrinkle Mask 28 and<br>Wrinkle Mask 72 | Rejuvalite MD |
| Model Name | Pulsaderm Wrinkle Mask 28<br>Pulsaderm Wrinkle Mask 72 | Rejuvalite MD |
| Clearance Date: | April 14, 2017 | November 13, 2014 |
Table 1 Predicate devices
{4}------------------------------------------------
#### K201107: 510(k) Summary
#### 4. Device Description
The OPERA LEBODY is a device which allows emission of LED light in the red (630nm) and IR (830nm) spectrum onto the face, which induces photobiological effect to the face for reduction of wrinkle. OPERA LEBODY includes the mask that contains LEDs on the inner surface, the controller which turns on and off the device, a cradle that holds the face mask, a pouch that keeps the product, and USB cable that delivers electrical power to the controller for operation or for charging the battery contained in the controller.
The user wears a mask and operates OPERA LEBODY using the controller. The device will automatically turn off after 10 minutes. Emitted light from LEDs are not intended for ocular or ophthalmic treatment. To prevent irradiation of LED lights to eyes during the treatment, OPERA LEBODY has protective eye guide which blocks light energy from LEDs.
#### 5. Indications / Intended Use
The OPERA LEBODY is an over the counter device that is intended for the use in the treatment of full face wrinkles.
## 6. Technical Characteristics in Comparison to Predicate Devices
| | Primary Predicate Device | Predicate Device | Proposed Device |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| 510(K)<br>Number | K163329 | K133896 | N/A |
| Manufacturer | Pulsaderm LLC | Trophy Skin, Inc. | GTG Wellness Co., Ltd. |
| Device Name | Pulsaderm Wrinkle Mask 28<br>and Wrinkle Mask 72 | Rejuvalite MD | OPERA LEBODY |
| Model Name | Pulsaderm Wrinkle Mask 28<br>Pulsaderm Wrinkle Mask 72 | Rejuvalite MD | OLG-200, OLZ-200 |
| Clearance<br>Date: | April 14, 2017 | November 13, 2014 | N/A |
| Product Code | OHS | OHS | OHS |
| Classification<br>/ Regulation | Class II | Class II | Class II |
| Intended Use<br>/ Indications<br>for Use: | The Pulsaderm Wrinkle<br>Masks 28 and 72 are<br>intended for the use in the<br>treatment of facial wrinkles<br>and for people with<br>Fitzpatrick Skin Types I, II<br>and III | The Rejuvalite MD is an<br>over-the-counter device<br>intended for the use in the<br>treatment of full-face<br>wrinkles | OPERA LEBODY is an over<br>the counter device that is<br>intended for the use in the<br>treatment of full face<br>wrinkles. |
| Type of Use | Over-The-Counter use | Over-The-Counter use | Over-The-Counter use |
| Light Source | LEDs | LEDs | LEDs |
| LED<br>emission | 2 types:<br>- Red (620~630nm)<br>- IR (850nm) | 2 types:<br>- Red (600, 622,<br>660nm)<br>- IR (860nm) | 2 types:<br>- Red (630nm)<br>- IR (830nm) |
| Number of<br>LEDs | Pulsaderm Wrinkle Mask<br>28: Total 28 LEDs | Total 120 LEDs | Total 78 LEDs |
The OPERA LEBODY is substantially equivalent to the following legally marketed predicate devices
{5}------------------------------------------------
#### OPERA LEBODY
#### K201107: 510(k) Summary
| | Pulsaderm Wrinkle Mask<br>72: Total 72 LEDs | | |
|----------------------------------|---------------------------------------------|-------------------------------------------------|--------------------------------------------------|
| Power<br>Density<br>(Irradiance) | 21.18 - 25.35 mW/cm² | 62 mW/cm² | 50 mW/cm² |
| Treatment<br>time | 15 minutes every day, 60<br>sessions | 3 minutes daily, 5 days per<br>week for 8 weeks | 10 minutes daily, 3 days per<br>week for 8 weeks |
| Treatment<br>area | Full Face | Full Face | Full Face |
#### 7. Performance Data
Non-clinical tests: Measurement of wavelength, average output power, and total irradiance of treatment LEDs were performed. Other performance, such as safety of lamps and lamp systems, electromagnetic compatibility and electrical safety, were tested using following consensus standards:
- -Basic safety and essential performance of the OPERA LEBODY is tested and evaluated according to the IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007+A1:2012.
- Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2 Edition 4.0 2014-02.
- Photobiological safety of lamps and lamp systems is evaluated in accordance with -IEC 62471: 2006 (First Edition).
- Risk management was recorded by referring to ISO 14971: 2012. -
- Usability was documented by referring to IEC 60601-1-6: 2010, AMD1: 2013. -
The portion of the device that touches patient body is made of silicone and ABS plastics, which have been used for other medical devices without any biocompatibility risk.
#### 8. Substantial Equivalence
The proposed device uses similar or identical technology as the predicate devices and has same intended uses. Based upon the predicted overall performance characteristics for the OPERA LEBODY, GTG Wellness Co., Ltd. believes that no significant differences in usage of its underlying technological principles between OPERA LEBODY and the predicate devices.
#### 9. Conclusions
On the basis of the information provided in this Summary, GTG Wellness Co., Ltd. Believes that OPERA LEBODY is substantially equivalent to legally commercialized predicate devices for the purposes of this 510 (k) submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.