The Pulsaderm Wrinkle Mask 28 and 72 are intended for the use in the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I,II and/or III.
Device Story
Pulsaderm Wrinkle Mask 28 and 72 are over-the-counter light therapy devices for facial wrinkle reduction. Devices consist of wearable masks containing red (620-630 nm) and infrared (850 nm) LED diodes powered by rechargeable Ni-MH batteries. User places mask over face for 15-minute daily treatment sessions. Constant waveform energy output stimulates skin; intended to improve appearance of facial wrinkles. Designed for home use by lay users. Self-selection and usability studies confirm lay users can correctly interpret labeling and operate device without professional assistance. Benefits include non-invasive, at-home wrinkle treatment.
Clinical Evidence
No clinical efficacy trials performed. Bench testing confirmed compliance with IEC 60601-1, IEC 60601-1-2, IEC 62471, and ISO 10993 biocompatibility standards. A lay-user study and self-selection study (IRB approved) demonstrated that intended users could correctly read labeling, self-select for treatment, and operate the device according to the user manual without assistance.
Technological Characteristics
Wearable mask form factor; red (620-630 nm) and infrared (850 nm) LED light source; constant waveform. Powered by rechargeable Ni-MH batteries. Biocompatibility per ISO 10993. Safety testing per IEC 60601-1 (electrical), IEC 60601-1-2 (EMC), and IEC 62471 (photobiological safety). Quality management system per ISO 14971.
Indications for Use
Indicated for treatment of facial wrinkles in individuals with Fitzpatrick Skin Types I, II, and III.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Pulsaderm LLC Gloria Avendano Regulatory Affairs Manager 12801 Commonwealth Dr. Units 2-6 Fort Myers, Florida 33913
Re: K163329
Trade/Device Name: Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: March 10, 2017 Received: March 13, 2017
Dear Gloria Avendano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) Pending
Device Name
Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72
Indications for Use (Describe)
The Pulsaderm Wrinkle Mask 28 and 72 are intended for the use in the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I,II and/or III.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72
#### 1. General Information
Submitter: Pulsaderm LLC 12801 Commonwealth Dr. Units 2-6 Fort Myers, FL 33913
Contact Person: Gloria Avendano Regulatory Affairs Manager Pulsaderm LLC Gloria@pulsaderm.com
Summary Preparation Date: March 10, 2017
#### 2. Device
| Device Name: | Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 7 |
|----------------------------|------------------------------------------------|
| Common/Usual Name: | Wrinkle Light Therapy System |
| Classification Name: | Light Based Over the Counter Wrinkle Reduction |
| Classification Regulation: | 21 CFR 878.4810 |
| Classification Panel: | General and Plastic Surgery |
| Code: | OHS |
## 3. Predicate Devices:
| Device Name | K Number | Manufacturer |
|--------------------------------------------------------------------------------------|----------|-----------------|
| Pro X OTC 5 Light Therapy | K140471 | La Lumiere, LLC |
| Quasar Calypso C100 Wrinkle<br>Reduction Device, Baby Quasar<br>Plus, Quasar MD Plus | K130225 | Silver Bay, LLC |
## 4. Device Description:
The Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 72 are over the counter devices and consist of a collection of red and infrared diodes for the treatment of facial wrinkles. Pulsadern Wrinkle Masks 28 and 72 are powered rechargeable Ni-MH batteries. To receive treatment, the user simply places the mask over the face for a specific period to receive treatment.
#### 5. Indications for Use:
The Pulsaderm Wrinkle Masks 28 and 72 are intended for the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I, II and III.
{4}------------------------------------------------
# 6.Technological Characteristics and substantial equivalence:
Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 72 share the same indications for use, treatment regimen as the predicate devices. Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 72 are as safe and effective as the predicate devices. Pulsaderm LLC is certain that the difference in number of diodes and slightly increase in power output does not affect the safety or efficacy of the device as there are a wide range of number of diodes and power output in predicate K130225 and devices cleared under device code OHS.
Based on technology, similar design, wavelength parameters and treatment regimen the different number of LED Diodes between devices does not adversely affect the safety of the devices.
| Device Name | Pulsaderm Wrinkle Mask 28 | Pulsaderm Wrinkle Mask 72 | Pro X OTC 5 Light Therapy Device | C100, Baby Quasar Plus and Quasar MD Plus |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Indications for Use | The Pulsaderm Wrinkle Mask 28 is intended for the use in the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I, II and/or III. | The Pulsaderm Wrinkle Mask 72 is intended for the use in the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I, II and/or III. | Intended for use in the treatment of facial wrinkles. It is for people with wrinkles on their face and who have Fitzpatrick Skin Types I, II and III. | Intended to emit energy in the RED/IR spectrum, specifically for the treatment of full-face wrinkles. |
| Wavelength | 620-630 nm red and 850 nm Infrared | 620-630 nm Red and 850 nm Infrared | 620-630 nm Red and 850 nm Infrared | 610 nm Red and 850 nm Infrared |
| Waveform | Constant | Constant | Constant | Constant |
| Number of LED's | 28 | 72 | 18 | 70, 40 and 24 |
| Rechargeable | Yes, Ni-MH Batteries | Yes, Ni-MH Batteries | NO | Yes, Ni-MH Batteries |
| Power Density (mW/cm²) | Red: 15.94<br>Infrared: 5.24 | Red: 18.71<br>Infrared:6.61 | Red: 13.20<br>Infrared: 6.60 | unknown |
| Total power (mW/cm²) for entire mask | 21.18 | 25.35 | 19.80 | unknown |
| Total energy per 15-minute treatment session | 0.10 joules | 0.11 joules | 0.09 joules | unknown |
| Overall total Energy for 60 treatments (J) | 5.75 J | 6.87 J | 5.37 J | Unknown |
| Treatment protocol | 15 minutes everyday | 15 minutes everyday | 15 minutes everyday | 3 minute per treatment per area. Daily. |
| Number of treatments device will allow | 60 or more | 60 or more | 30 sessions per device | 60 or more |
The reference table above will confirm the devices share similar output and technology characteristics.
{5}------------------------------------------------
# 7. Performance Data:
Performance testing was conducted and confirm compliance to design specifications; similar wavelength, energy type and safety characteristics.
Pulsaderm Wrinkle Masks 28 and Wrinkle Mask 72 were tested and found in conformance with test standards: IEC 60601-1, IEC 60601-1-2, and IEC 62471. Biocompatibility testing to ISO 10993.
Pulsaderm Wrinkle Masks 28 and Wrinkle Mask 72 were designed and developed under a quality management system conforming to ISO 14971.
## 8. Non-clinical-testing
A lay-user study and self-selection study was conducted with Institutional Review Board (IRB) approval and oversight to determine if lay users could read the product labeling and then self-assess if the Pulsaderm Wrinkle Masks 28 and 72 would be beneficial for them to use. Study data was collected and demonstrated that the intended users of the device could successfully follow the instructions and use the device as intended. The performance data supplied in this 510(k) demonstrated that the majority of lav users were able to properly self-select themselves using the box labeling and the majority of lay users were able to properly use the device by reading instructions in the user manual without any assistance.
In summary, the conclusion was that an average lay user can read and comprehend correctly the user manual and package labeling.
## 9. Conclusion
Based upon the analysis of the performance characteristics, safety characteristics and intended use for the Pulsaderm Wrinkle Masks 28 and 72, Pulsaderm LLC believes that no significant differences exist between this device and the predicates. Therefore, substantial equivalency is requested.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.