K201020 · Aidoc Medical , Ltd. · QAS · Aug 26, 2020 · Radiology
Device Facts
Record ID
K201020
Device Name
BriefCase
Applicant
Aidoc Medical , Ltd.
Product Code
QAS · Radiology
Decision Date
Aug 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K201020 · Aug 26, 2020
BriefCase
Aidoc Medical , Ltd.
Retrospective clinical chest CT images; Hospital clinical workflow data (time-to-exam-open)
Retrospective analysis of 268 clinical cases to validate algorithm performance (sensitivity/specificity) and assess potential clinical benefit by comparing device notification time against standard-of-care exam opening times.
Retrospective, blinded, multicenter study: 268 cases (74 positive, 194 negative) from 2 clinical study sites in the US.
—
Indications for Use
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced chest CTs (but not dedicated CTPA protocol). The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspect cases of incidental Pulmonary Embolism (iPE) pathologies. For the iPE pathology, the software is only intended to be used on single-energy exams. The device is intended to work with GE and Siemens scanners only. BriefCase uses an artificial intelligence algorithm to analyze images and flag suspect cases on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Device Story
Software-based radiological triage system; processes contrast-enhanced chest CT DICOM images; utilizes AI algorithm to detect suspected incidental Pulmonary Embolism (iPE). Operates in parallel to standard-of-care; does not alter original images; does not remove cases from reading queues. System components: Aidoc Hospital Server (image acquisition), Aidoc Cloud Server (processing), Aidoc Worklist Application (desktop notification). Radiologists receive pop-up notifications with patient details and low-quality, unannotated preview images. Preview images are for informational triage purposes only; not for diagnosis. Enables preemptive prioritization of suspected positive cases; reduces time-to-exam-open compared to standard FIFO workflows. Intended for hospital networks; used by trained radiologists.
Clinical Evidence
Retrospective, blinded, multicenter study (N=268; 74 positive, 194 negative). Sensitivity: 90.5% (95% CI: 81.4%-96.2%); Specificity: 88.7% (95% CI: 83.3%-92.8%); NPV: 99.7%; PPV: 17.6%. Time-to-notification (4.7 min) compared to standard-of-care time-to-exam-open (223.3 min) for 63 true positive cases, showing a statistically significant mean reduction of 220.9 minutes.
Technological Characteristics
Software-only device; runs on Windows Server 2012 64-bit. Components: PostgreSQL, DICOM module, Image Processing Application. Connectivity: Networked (Hospital Server to Cloud Server to Worklist Application). Processing: AI-based algorithm for image analysis. Compatibility: GE and Siemens CT scanners (single-energy). Input: Contrast-enhanced chest CT (0.5-2.0mm axial slice thickness). Output: Desktop notification and compressed preview image.
Indications for Use
Indicated for adults/transitional adults (≥18 years) undergoing contrast-enhanced chest CT (excluding dedicated CTPA) on GE or Siemens scanners to triage incidental Pulmonary Embolism (iPE).
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
August 26, 2020
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Aidoc Medical, Ltd. % John J. Smith, M.D., J.D. Regulatory Counsel Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004
Re: K201020
Trade/Device Name: BriefCase for iPE Triage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer-assisted triage and notification software Regulatory Class: Class II Product Code: QAS Dated: July 24, 2020 Received: July 24, 2020
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K201020
#### Device Name
#### BriefCase for iPE Triage
Indications for Use (Describe)
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of contrast enhanced chest CT images (but not dedicated CTPA protocol) CT. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive cases of Incidental Pulmonary Embolism (iPE) pathologies. For the iPE pathology, the software is only intended to be used on singleenergy exams. The device is intended to work with GE and Siemens scanners only.
BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summarv Aidoc Medical, Ltd.'s BriefCase (K201020)
### Submitter:
| Aidoc Medical, Ltd.<br>3 Aminadav St.<br>Tel-Aviv, Israel<br>Phone: | +972-73-7946870 |
|---------------------------------------------------------------------|---------------------------------------------------------------------------|
| Contact Person: | N. Epstein, Ph.D. |
| Date Prepared: | July 24, 2020 |
| Name of Device: | BriefCase for iPE Triage |
| Classification Name: | Radiological computer-assisted triage and notification software<br>device |
| Regulatory Class: | Class II |
| Product Code: | QAS (21 C.F.R. 892.2080) |
| Predicate Device: | BriefCase (K190072, for PE) |
#### Device Description
BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and is comprised of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS) for image acquisition; (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Worklist Application for workflow integration, installed on the radiologist' desktop and provides the user interface in which notifications from the BriefCase software are received.
DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the Worklist desktop application, thereby prompting preemptive triage and prioritization by the attending radiologist. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.g. a chest radiologist may choose to filter out notifications on LVO cases, and a neuro-radiologist would opt to divert PE notifications. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out notification by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc worklist application.
The Worklist Application displays the pop-up text notifications of new suspected studies when they come in. Notifications are in the form of a small pop-up containing patient name, accession number and the relevant pathology (e.g., iPE). A list of all incoming suspect cases is also displayed. Hovering over a notification or a case in the worklist pops up a compressed, lowquality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does
{4}------------------------------------------------
not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.
Presenting the radiologist with notification facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
#### Intended Use / Indications for Use
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced chest CTs (but not dedicated CTPA protocol). The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspect cases of incidental Pulmonary Embolism (iPE) pathologies. For the iPE pathology, the software is only intended to be used on single-energy exams. The device is intended to work with GE and Siemens scanners only.
BriefCase uses an artificial intelligence algorithm to analyze images and flaq suspect cases on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
#### Comparison of Technological Characteristics
The subject BriefCase for iPE triage and predicate device BriefCase for PE triage (K190072) are identical in all aspects and defer only with respect to the training of the algorithm on iPE and PE images, respectively. .
Both devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence algorithms incorporated software packages for use with DICOM 3.0 compliant CT scanners, PACS, and radiology workstations. Both devices are intended to aid in triage and prioritization of radiological images. The predicate device processes CTPAs and is indicated for Pulmonary Embolism triage, while the subject device processes contrast-enhanced chest images and is indicated for incidental Pulmonary Embolism (iPE) triage. Both devices are intended to provide radiologists with notifications and unannotated preview images of suspect studies for the purpose of preemptive triage.
Both software devices notify the attending radiologist of the availability of time sensitive radiological images for review based on computer aided image analysis. Both devices send notifications and low-quality compressed previews to the radiology workstations' desktop. Both devices feature a notification filter in the user interface. Notifications are for informational purpose only and are meant to prompt the radiologist to start preemptive triage of a flagged case, upon which he may decide after observing the unannotated, low quality preview on his desktop, to turn to the local PACS to perform evaluation of the original series earlier than would have been the case in the standard of care alone.
Thus, the subject and predicate BriefCase raise the same types of safety and effectiveness
{5}------------------------------------------------
questions, namely, accurate detection of findings within the processed study. It is important to note that, like the predicate, the subject device does not remove cases from the standard of care reading queue and does not modify them. Both devices operate in parallel with the standard of care, which remains the default option for all cases.
A table comparing the key features of the subject and predicate devices is provided below.
| Table 1. Key feature comparison | | |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device<br>Aidoc Briefcase for PE triage<br>(K190072) | Subject Device<br>Aidoc Briefcase for iPE triage<br>(K201020) |
| Intended Use /<br>Indications for Use | BriefCase is a radiological<br>computer aided triage and<br>notification software indicated for<br>use in the analysis of CTPA<br>images. The device is intended<br>to assist hospital networks and<br>trained radiologists in workflow<br>triage by flagging and<br>communication of suspected<br>positive findings of Pulmonary<br>Embolism (PE) pathology. The<br>software is only intended to be<br>used on single-energy exams.<br><br>BriefCase uses an artificial<br>intelligence algorithm to analyze<br>images and highlight cases with<br>detected findings on a<br>standalone desktop application in<br>parallel to the ongoing<br>standard of care image<br>interpretation. The user is<br>presented with notifications for<br>cases with suspected findings.<br>Notifications include compressed<br>preview images that are<br>meant for informational purposes<br>only and not intended for<br>diagnostic use<br>beyond notification. The device<br>does not alter the original<br>medical image and is not<br>intended to be used as a<br>diagnostic device.<br><br>The results of BriefCase are<br>intended to be used in<br>conjunction with other<br>patient information and based on<br>professional judgment, to assist<br>with triage/prioritization of<br>medical images. Notified<br>clinicians are responsible for<br>viewing full images per the<br>standard of care. | BriefCase is a radiological<br>computer aided triage and<br>notification software indicated for<br>use in the analysis of contrast-<br>enhanced chest CTs (but not<br>dedicated CTPA protocol).<br>The device is intended to<br>assist hospital networks and<br>trained radiologists in workflow<br>triage by flagging and<br>communication of suspected<br>positive cases of incidental<br>Pulmonary Embolism (iPE)<br>pathologies. For the iPE<br>pathology, the software is only<br>intended to be used on single-<br>energy exams. The device is<br>intended to work with GE and<br>Siemens scanners only.<br><br>BriefCase uses an artificial<br>intelligence algorithm to analyze<br>images and flag suspect cases on<br>a standalone desktop application<br>in parallel to the ongoing<br>standard of care image<br>interpretation. The user is<br>presented with notifications for<br>suspect cases. Notifications<br>include compressed preview<br>images that are<br>meant for informational purposes<br>only and not intended for<br>diagnostic use beyond notification.<br>The device does not alter the<br>original medical image and is not<br>intended to be used as a<br>diagnostic device.<br><br>The results of BriefCase are<br>intended to be used in conjunction<br>with other patient information and<br>based on their professional<br>judgment, to assist with<br>triage/prioritization of medical<br>images. Notified clinicians are |
| | Predicate Device | Subject Device |
| | Aidoc Briefcase for PE triage<br>(K190072) | Aidoc Briefcase for iPE triage<br>(K201020) |
| | | responsible for viewing full images<br>per the standard of care. |
| User population | Radiologist | Radiologist |
| Anatomical region of<br>interest | Chest | Chest |
| Inclusion/<br>Exclusion criteria | Inclusion Criteria<br>- CTPA protocols.<br>- Single energy exams.<br>- Scans performed with 64-<br>slice scanner or greater<br>number of detectors.<br>- Scans performed on<br>adults/transitional adults ≥ 18<br>years of age.<br>- Slice thickness 0.5 - 3.0 mm<br>axial.<br>Exclusion Criteria<br>- All studies that are technically<br>inadequate, including studies<br>with motion artifacts, severe<br>metal artifacts, sub-optimal<br>bolus or inadequate field of<br>view. | Inclusion criteria<br>- Contrast-enhanced chest CTs<br>(but not dedicated CTPA<br>protocol.<br>- Single energy exams.<br>- Scans performed with a 64<br>slice or greater number of<br>detectors.<br>- Scans performed on<br>adults/transitional adults ≥ 18<br>years of age.<br>- Slice thickness: 0.5mm -<br>2.0mm axial.<br>Exclusion Criteria<br>- All studies that are technically<br>inadequate, including studies<br>with motion artifacts, severe<br>metal artifacts, or inadequate<br>field of view. |
| Data acquisition<br>protocol | CTPA protocol | Contrast-enhanced chest CTs (but<br>not dedicated CTPA protocol) |
| View DICOM data | DICOM Information about the<br>patient, study and current image | DICOM Information about the<br>patient, study and current image |
| Segmentation of region<br>of interest | No; device does not mark,<br>annotate, or direct users'<br>attention to a specific location in<br>the original image | No; device does not mark,<br>annotate, or direct users' attention<br>to a specific location in the original<br>image |
| Algorithm | Artificial intelligence algorithm<br>with database of images | Artificial intelligence algorithm with<br>database of images |
| Notification/Prioritization | Yes | Yes |
| Preview images | Presentation of a low-quality,<br>compressed, grayscale preview<br>image that is captioned "Not for<br>diagnostic use". | Presentation of a low-quality,<br>compressed, grayscale preview<br>image that is captioned "Not for<br>diagnostic use". |
| Alteration of original<br>image | No | No |
| Removal of cases from<br>worklist queue | No. The device operates in<br>parallel with the standard of care,<br>which remains the default option<br>for all cases. Unflagged cases<br>are not de-prioritized. | No. The device operates in<br>parallel with the standard of care,<br>which remains the default option<br>for all cases. Unflagged cases are<br>not de-prioritized. |
| | Predicate Device<br>Aidoc Briefcase for PE triage<br>(K190072) | Subject Device<br>Aidoc Briefcase for iPE triage<br>(K201020) |
| Structure | - AHS module (image<br>acquisition).<br>- ACS module (image<br>processing).<br>- Aidoc Worklist application for<br>workflow integration (worklist<br>and non-diagnostic basic<br>Image Viewer). | - AHS module (image<br>acquisition).<br>- ACS module (image<br>processing).<br>- Aidoc Worklist application for<br>workflow integration (worklist<br>and non-diagnostic basic<br>Image Viewer). |
# Table 1. Kev feature comparison
{6}------------------------------------------------
{7}------------------------------------------------
## Performance Data
## Pivotal Study Summary
Aidoc conducted a retrospective, blinded, multicenter study with the BriefCase software with the primary endpoint to evaluate the software's performance in identifying Contrast-enhanced chest CTs (but not dedicated CTPA protocol), containing Incidental Pulmonary Embolism in 268 cases from 2 clinical study sites in the US. There were 74 positive cases and 194 negative cases (images with iPE versus without iPE) included in the analysis.
Sensitivity and specificity exceeded the 80% performance goal. Specifically, sensitivity was 90.5% (95% Cl: 81.4%, 96.2%) and specificity was 88.7% (95% Cl: 83.3%, 92.8%).
#### Secondary Endpoint
Briefcase's potential clinical benefit of worklist prioritization for true positive iPE cases was evaluated by comparing the standard-of-care metric of time-to-exam-open to the software's timeto-notification metric for iPE, in the study sites where the time-to-exam-open information was available.
- -The BriefCase time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the worklist application.
- -The standard of care time-to-open-exam consists of the from scan acquisition to when the radiologist first opened the exam for review.
The standard of care metric was compared to the BriefCase time-to-notification in the two study sites (both in the US) for 63 True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase device), and the results are reported in the Table 2 below.
The BriefCase time-to-notification for iPE was 4.7 minutes (95% Cl: 4.4-5.1; Median: 5.0, IQR: 2.3). In contrast, standard of care time-to-exam-open was much longer, 223.3 minutes (95% Cl: 125.8-320.7; Median: 70.4, IQR: 217.6). The mean difference of 220.9 minutes (95% Cl: 122.0-319.9; Median: 63.2, IQR: 219.8) for these two metrics is statistically significant and assuming the radiologist receives a notification on a true positive iPE case and acts on it immediately, it can on average save more than three hours compared to the time-to-exam-open in a first in first out (FIFO) reading queue. The value of 220.9 is based on the study of 63 cases from 2 study sites and may vary in practice.
{8}------------------------------------------------
| Parameter | N | Mean<br>estimate | Lower<br>Confidence<br>Limit | Upper<br>Confidence<br>Limit | Median | IQR |
|-------------------------------------------------|----|------------------|------------------------------|------------------------------|--------|--------|
| Time-to-exam-open<br>in the standard of<br>care | 63 | 223.3 | 125.8 | 320.7 | 70.4 | 217.65 |
| Time-to-notification<br>BriefCase iPE | 63 | 4.7 | 4.4 | 5.1 | 5.0 | 2.3 |
| Difference | 63 | 220.9 | 122.0 | 319.9 | 63.2 | 219.8 |
Table 2. Time saving data
NPV was 99.7% (95% Cl: 99.5%-99.9%) and PPV was 17.6% (95% Cl: 9.5%-22.1%).
Thus, the reported time savings data demonstrates that radiologists may have the opportunity to be involved in the clinical workflow substantially earlier due to the notifications from the BriefCase device. Performance validation data suggest that when using the subject BriefCase for iPE triage, the radiologists may have the same benefit in time saving as with using the BriefCase for PE triage.
# Conclusions
The subject BriefCase for iPE triage and the predicate BriefCase for PE triage devices are both intended to aid in prioritization and triage of radiological images for the indications of incidental Pulmonary Embolism and Pulmonary Embolism respectively. Both devices are software packages with similar technological characteristics and principles of operation, both incorporating deep learning AI algorithms that process images, and software to send notifications and unannotated compressed preview images to the radiologists' workstation. In both devices, the labeling states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.
Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, and do not remove images from the standard of care FIFO queue, thus not disturbing standard interpretation of the images by the attending radiologists. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.
The BriefCase device for iPE triage is thus substantially equivalent to the BriefCase for PE triage.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.