Infinix™ Universal Composite: Restorations in the posterior region (Class I and II) Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Veneering of discolored anterior teeth Splinting of mobile teeth Extended fissure sealing in molars and premolars Repair of composite/ceramic veneers The addition of the QASi particles to the Infinix™ Universal Composite reduces demineralization, which is part of the caries-formation process. Infinix™ Flowable Composite: Class III and V restorations Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations) As base/liner under direct or indirect restorations Repair of small defects in esthetic indirect restorations Pit and fissure sealant Blocking out of undercuts Repair of resin and acrylic temporary materials The addition of the QASi particles to the Infinix™ Flowable Composite reduces demineralization, which is part of the caries-formation process. Infinix™ Bulk Fill Flow Composite: Base under Class I and II direct restorations Liner under direct restorative materials Pit and fissure sealant Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations) Class III and V restorations Blocking out of undercuts Repair of small enamel defects Repair of small defects in esthetic indirect restorations Repair of resin and acrylic temporary materials As a core build-up where at least half the coronal tooth structure is remaining to provide structural support for the crown The addition of the OASi particles to the Infinix™ Bulk Fill Flow Composite reduces demineralization, which is part of the caries-formation process.
Device Story
Infinix Family of Dental Composites (Universal, Flowable, Bulk Fill Flow) are light-cured, resin-based restorative materials. Key feature: silica-based filler particles decorated with quaternary ammonium functional groups (QASi) covalently bound via silane linkers. Mechanism: QASi particles remain entrapped in cured composite; positively charged surface electrostatically attracts negatively charged microbial cell membranes, causing lysis and reducing demineralization. Used by dentists in clinical settings for various cavity preparations and restorations. Output is a cured, antimicrobial dental restoration. Benefits include reduced demineralization and caries-formation process support.
Clinical Evidence
In-situ clinical study (n=25) compared Infinix Universal Composite to a control composite without antibacterial particles. Subjects wore partial dentures with enamel slabs and composite inserts for 4 weeks. Primary endpoint: mineral loss (ΔZ) in enamel slabs adjacent to composite, measured via cross-section microhardness. Results: Mean mineral loss was 233.8 ± 330.9 (Nobio) vs 730.8 ± 558 (control). Difference was statistically significant (p<0.0001).
Technological Characteristics
Light-cured, resin-based dental composites. Contains silica-based filler particles (max 1.2% wt/wt) functionalized with quaternary ammonium groups (QASi) via silane linkers. Form factors: Universal, Flowable, and Bulk Fill Flow. Standalone application.
Indications for Use
Indicated for dental patients requiring restorative procedures including Class I-V restorations, veneering, splinting, fissure sealing, base/liner placement, core build-up, and repair of existing restorations or temporary materials.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 14, 2020
Nobio Ltd. % Shoshan (Shosh) Friedman Senior Regulatory Consultant ProMedoss, Inc. 3521 Hatwynn Rd. Charlotte, North Carolina 28269
Re: K201010
Trade/Device Name: Infinix Universal Composite, Infinix Flowable Composite, Infinix Bulk Fill Flow Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: April 14, 2020 Received: April 17, 2020
Dear Shoshan (Shosh) Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201010
Device Name Infinix™ Universal Composite
Indications for Use (Describe)
- · Restorations in the posterior region (Class I and II)
- · Anterior restorations (Class III, IV)
- · Class V restorations (cervical caries, root erosion, wedge-shaped defects)
- · Veneering of discolored anterior teeth
- · Splinting of mobile teeth
- Extended fissure sealing in molars and premolars
- · Repair of composite/ceramic veneers
The addition of the QASi particles to the Infinix™ Universal Composite reduces demineralization, which is part of the caries-formation process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:inline-block"><input checked="true" type="checkbox"/></div> Registration (Per 49 CFR 390.21 when Change) | <div style="display:inline-block"><input type="checkbox"/></div> One-Time Notification (Per 49 CFR 390.19) |
|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201010
Device Name Infinix™ Flowable Composite
Indications for Use (Describe)
- · Class III and V restorations
- · Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations)
- As base/liner under direct or indirect restorations
- · Repair of small defects in aesthetic indirect restorations
- Pit and fissure sealant
- · Blocking out of undercuts
- · Repair of resin and acrylic temporary materials
The addition of the QASi particles to the Infinix™ Flowable Composite reduces demineralization, which is part of the caries-formation process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201010
Device Name Infinix™ Bulk Fill Flow Composite
Indications for Use (Describe)
- · Base under Class I and II direct restorations
- · Liner under direct restorative materials
- · Pit and fissure sealant
- · Restoration of minimally invasive cavity preparations (including small, non-stress-bearing occlusal restorations)
- · Class III and V restorations
- · Blocking out of undercuts
- Repair of small enamel defects
- · Repair of small defects in esthetic indirect restorations
- · Repair of resin and acrylic temporary materials
• As a core build-up where at least half the coronal tooth structure is remaining to provide structural support for the crown The addition of the OASi particles to the Infinix™ Bulk Fill Flow Composite reduces demineralization, which is part of the caries-formation process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(K) SUMMARY [as required by section 807.92(c)] Infinix™ Family of Dental Composites 510(k) Number K201010
#### 5.1 SUBMITTER
### Applicant's Name:
Nobio Ltd. 8 Hamatechet St. POB 50502 Kadima, Israel 6092000 Phone: +972-3-9059966
## Contact Person:
Shoshana (Shosh) Friedman Senior Regulatory Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 704-430-8695 s.friedman@promedoss.com
Date Prepared:
August 7, 2020
#### 5.2 DEVICE
## Trade Name:
Infinix™ Family of Dental Composites which includes:
- Infinix™ Universal Composite
- Infinix™ Flowable Composite ●
- Infinix™ Bulk Fill Flow Composite
Classification: Name: Material, Tooth Shade, Resin Product Code: EBF Regulation No: 872.3690 Class: 2 Review Panel: Dental
#### 5.3 PREDICATE DEVICES
- K182921 InfinixTM Universal Composite (previously called Novidia™ Universal Composite)
{6}------------------------------------------------
- K182580 InfinixTM Flowable Composite (previously called Novidia™ Flowable ● Composite)
- K182714 Infinix™ Bulk Fill Flow Composite (previously called Novidia™ Bulk ● Fill Flow Composite)
#### 5.4 DEVICE DESCRIPTION
Composite restorations are performed with two main components: a resin-based composite for filling the prepared tooth cavity and the bonding system which is applied to the cavity before the placement of the composite filling material.
The composites are resin based and the proportion of the inorganic filler material is the main differentiator among the composite materials.
The Infinix™ Family of Dental Composites include the following light-cured, resinbased composite materials:
- . Universal Composite
- Flowable Composite
- . Bulk Fill Flow Composite
There are no changes to the formulation, manufacturing processes, and packaging of the three composites submitted under K182921, K182714, and K182580.
Some of the silica-based filler particles (max 1.2% wt/wt of the total formulation) contained in all the Infinix™ composites are decorated with quaternary ammonium functional groups that are covalently bound to the filler particles' silica core via a silane linker (herein referred to as "QASi"). Following light induced polymerization, QASi particles (produced by Nobio) remain permanently entrapped within the composite restoration and destroy microorganisms by contact. As the cell membrane of microorganisms is negatively charged and the QASi particles have strong positive charges, microorganisms are electrostatically attracted to the positively charged cured material's surface resulting in disruption of their cell membrane and immediate microbial lysis.
#### 5.5 INDICATIONS FOR USE
The indications for Use of the Infinix™ Family of Dental Composites is substantially similar to these of their predicate devices with the exception of the inclusion of an antimicrobial claim based on data collected from in situ clinical study.
| Subject Devices | Predicate Devices |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Infinix™ Universal Composite | Infinix™ Universal Composite (K182921) |
| • Restorations in the posterior region<br>(Class I and II) | • Restorations in the posterior region (Class I<br>and II) |
| • Anterior restorations (Class III, IV) | • Anterior restorations (Class III, IV) |
| • Class V restorations (cervical caries,<br>root erosion, wedge-shaped defects) | • Class V restorations (cervical caries, root<br>erosion, wedge-shaped defects) |
| • Veneering of discolored anterior teeth | • Veneering of discolored anterior teeth |
| • Splinting of mobile teeth | • Splinting of mobile teeth |
| • Extended fissure sealing in molars and | • Extended fissure sealing in molars and |
| premolars | premolars |
| • Repair of composite/ceramic veneers | • Repair of composite/ceramic veneers |
| The addition of the QASi particles to the | |
| Infinix™ Universal Composite reduces | |
| demineralization, which is part of the | |
| caries-formation process. | |
| Infinix™ Flowable Composite | Infinix™ Flowable Composite (K182580) |
| • Class III and V restorations | • Class III and V restorations |
| • Restoration of minimally invasive | • Restoration of minimally invasive |
| cavity preparations (including small, | cavity preparations (including small, |
| non-stress-bearing occlusal | non-stress-bearing occlusal |
| restorations) | restorations) |
| • As base/liner under direct or indirect | • As base/liner under direct or indirect |
| restorations | restorations |
| • Repair of small defects in esthetic | • Repair of small defects in esthetic |
| indirect restorations | indirect restorations |
| • Pit and fissure sealant | • Pit and fissure sealant |
| • Blocking out of undercuts | • Blocking out of undercuts |
| • Repair of resin and acrylic temporary | • Repair of resin and acrylic temporary |
| materials | materials |
| The addition of the QASi particles to the | |
| Infinix™ Flowable Composite reduces | |
| demineralization, which is part of the | |
| caries-formation process. | |
| Infinix™ Bulk Fill Flow Composite | Infinix™ Bulk Fill Flow Composite (K182714) |
| • Base under Class I and II direct | • Base under Class I and II direct |
| restorations | restorations |
| • Liner under direct restorative materials | • Liner under direct restorative materials |
| • Pit and fissure sealant | • Pit and fissure sealant |
| • Restoration of minimally invasive | • Restoration of minimally invasive |
| cavity preparations (including small, | cavity preparations (including small, |
| non-stress-bearing occlusal | non-stress-bearing occlusal |
| restorations) | restorations) |
| • Class III and V restorations | • Class III and V restorations |
| • Blocking out of undercuts | • Blocking out of undercuts |
| • Repair of small enamel defects | • Repair of small enamel defects |
| • Repair of small defects in esthetic | • Repair of small defects in esthetic |
| indirect restorations | indirect restorations |
| • Repair of resin and acrylic temporary | • Repair of resin and acrylic temporary |
| materials | materials |
| • As a core build-up where at least half | • As a core build-up where at least half |
| | the coronal tooth structure is remaining |
| to provide structural support for the<br>crown<br>The addition of the QASi particles to the<br>InfinixTM Bulk Fill Flow Composite<br>reduces demineralization, which is part of<br>the caries-formation process. | to provide structural support for the<br>crown |
{7}------------------------------------------------
{8}------------------------------------------------
#### 5.6 SUBSTANTIAL EQUIVALENCE
The Infinix™ Universal Composite, Infinix™ Flowable Composite, and InfinixTM Bulk Fill Flow Composite have the exact same formulation, manufacturing processes, and packaging as the three composites submitted under K182921, K182714, and K182580. The only differences implemented in this submission are:
- 1) Removal from the Instructions for Use the statement "Clinical studies have not been conducted to demonstrate that the presence of QASi in this device results in improved clinical outcomes" that was requested by the Agency in the previous 510(k) submissions; and
- 2) Addition of the following statement to the intended use of each composite: "The addition of the QASi particles to the { composite name } reduces demineralization, which is part of the caries-formation process."
#### 5.7 PERFORMANCE DATA
## Non-Clinical Performance Testing:
All non-clinical and biological testing submitted in K182921, K182714, and K182580 apply to this submission as no changes have been implemented in the formulation, manufacturing processes, and packaging.
## Clinical Performance Testing:
An in-situ clinical study comparing Infinix™ Universal Composite to a widely used, similar composite without antibacterial particles was conducted at the School of Dentistry, UCSF (San Francisco, CA). Subjects wearing lower partial dentures with acrylic flanges on both sides of the mouth who met the eligibility criteria were recruited. On each side of the denture an enamel slab was placed next to a composite, separated by a tiny gap. On one side of the denture the composite was Nobio's Infinix Universal Composite ("Nobio group"), on the other side a widely used, similar composite without antibacterial particles served as control material ("control group"). Enamel slab and composite were recessed into the flange, allowing microbial plaque to accumulate on top of it and especially in the gap. After wearing the dentures for four weeks with the slabs, decalcification (mineral loss) in the enamel slabs adjacent to the gap were determined by cross-section microhardness testing in the laboratory. Average mineral loss was calculated for the Nobio group and the control group, and difference tested for statistical significance.
A total of 25 subjects, 17 males and 8 females, were recruited having an average age of 63.7 ± 14.1 years (range 33.66 to 80.98 years) and their ethnicity included 12 subjects
{9}------------------------------------------------
(48%) white and 13 subjects (52%) non-white (7 Black, 3 Hispanic, 2 Asian, and 1 Native American). The subjects wore the gap models in average for 27.9 ± 1.2 days (range 26 to 31 days), thus meeting the targeted 4-week in situ wearing period. The average mineral loss ΔZ (vol% x um) in enamel slabs facing the control composite was 730.8 ± 558 while the average mineral loss ΔZ in enamel slabs facing the Nobio composite with antibacterial particles was 233.8 ± 330.9. The two-tailed, paired t-test determined the significance level for a difference as P<0.0001, which demonstrates that the difference in average mineral loss between control and Nobio sides is statistically significant. Analysis of the individual ΔZ mineral losses for the control side and the Nobio side for all evaluated subjects showed that in all cases the ΔZ mineral loss for the Nobio side was lower than the control side (-268 to 862 in Nobio side vs. - 7 to 2,081 in the control side).
### 5.8 CONCLUSION
Nobio Ltd. believes that the Infinix™ Family of Dental Composites, which includes the Infinix™ Universal Composite, Infinix™ Flowable Composite, and Infinix™ Bulk Fill Flow Composite, are substantially equivalent to their predicate devices without introducing any new safety or effectiveness concerns and, as shown in the in-situ clinical study, the addition of the OASi particles to these composites reduces demineralization, which is part of the carries-formation process.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.