The Novidia™ Flowable Composite is indicated for: 1. Class III and V restorations 2. Restoration of minimally invasive cavity preparations (including small, non-stress bearing occlusal restorations) 3. As base/liner under direct or indirect restorations 4. Repair of small defects in esthetic indirect restorations 5. Pit and fissure sealing 6. Blocking out of undercuts 7. Repair of resin and acrylic temporary materials
Device Story
Novidia™ Flowable Composite is a low-viscosity, radiopaque, visible light-activated dental restorative material. Composed of a resin matrix (UDMA, Bis-GMA, TEGDMA) and inorganic fillers (alumino-silicate glass, silica dioxide, pigments). A portion of silica filler is covalently bound to quaternary ammonium residues (QASi) to maintain restoration integrity. Provided in syringes or single-dose capsules in three shades. Used by dentists in clinical settings for restorative procedures. Applied to cavity preparations and cured using a dental light source (430-490 nm). QASi incorporation aims to prevent microbial degradation and preserve surface integrity, though clinical outcomes regarding this benefit are not established. Output is a hardened, tooth-colored restoration.
Clinical Evidence
No clinical or animal data provided. Bench testing only. Performance assessed via ISO 4049 standards for flexural strength, elastic modulus, compression strength, shade stability, polymerization conversion, viscosity, water solubility/sorption, depth of cure, radio-opacity, and Knoop hardness. Biocompatibility confirmed via ISO 10993/7405 testing including cytotoxicity, irritation, systemic toxicity, pyrogenicity, and genotoxicity (Bacterial Reverse Mutation, Mouse Lymphoma Assay). Atomic force microscopy used to demonstrate resistance to microbial degradation.
Indicated for patients requiring Class III and V restorations, minimally invasive cavity preparations, base/liner placement, repair of esthetic indirect restorations, pit and fissure sealing, undercut blockout, and repair of resin/acrylic temporary materials.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2019
Nobio Ltd. % Shoshanna Friedman CEO ProMedoss, Inc. 3521 Hatwynn Rd. Charlotte, North Carolina 28269
Re: K182580
Trade/Device Name: Novidia Flowable Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: June 3, 2019 Received: June 4, 2019
Dear Shoshanna Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K182580
Device Name Novidia™ Flowable Composite
Indications for Use (Describe)
The Novidia™ Flowable Composite is indicated for:
1. Class III and V restorations
2. Restoration of minimally invasive cavity preparations (including small, non-stress bearing occlusal restorations)
3. As base/liner under direct or indirect restorations
- 4. Repair of small defects in esthetic indirect restorations
- 5. Pit and fissure sealing
- 6. Blocking out of undercuts
- 7. Repair of resin and acrylic temporary materials
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY [as required by section 807.92(c)] Novidia™ Flowable Composite 510(k) Number K182580
#### 5.1 SUBMITTER
### Applicant's Name:
Nobio Ltd. 8 Hamatechet St. POB 50502 Kadima, Israel 6092000 Phone: +972-3-9059966
## Contact Person:
Shoshana (Shosh) Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 704-430-8695 s.friedman@promedoss.com
Date Prepared:
June 3, 2019
#### 5.2 DEVICE
Trade Name: Novidia™ Flowable Composite
Classification: Name: Material, Tooth Shade, Resin Product Code: EBF Regulation No: 872.3690 Class: 2 Review Panel: Dental
#### 5.3 PREDICATE DEVICE
Filtek™ Supreme Ultra Flowable Restorative, manufactured by 3M Company, cleared under K100235.
Additionally, we are using the following reference devices as examples of FDAcleared devices that incorporate quaternary ammonium in their formulation:
- Clearfil Protect Bond cleared under K033938
- . Orthodontic Acrylic cleared under K141439 and Orthodontic Acrylic 2 cleared under K163482.
{4}------------------------------------------------
#### DEVICE DESCRIPTION 5.4
Novidia™ Flowable Composite is a low viscosity, visible light activated, radiopaque, flowable composite indicated for minimally invasive cavity preparations as well as various other indications.
The resin matrix of the Novidia™ Flowable Composite contains Urethane dimethacrylate (UDMA), Bisphenol-a glycidyl dimethacrylate (Bis-GMA) and triethylene glycol dimethacrylate (TEGDMA).
The inorganic filler of the Novidia™ Flowable Composite is a mix of particles of alumino-silicate-based glasses, silica dioxide and pigments. A small percentage of the silica-based filler particles are covalently bound to quaternary ammonium residues (OASi), added to maintain the integrity of the restoration.
Note: "Clinical studies demonstrating that the presence of OASi in this device improves clinical outcomes have not been conducted".
The Novidia™ Flowable Composite is provided in syringes (2 g) and in single-dose capsules (0.2 g) and is available in three shades (A1, A2 and A3).
#### ર્સ્ડ INDICATIONS FOR USE
Novidia™ Flowable Composite is indicated for use in:
- 1) Class III and V restorations
- 2) Restoration of minimally invasive cavity preparations (including small, nonstress-bearing occlusal restorations)
- 3) As base/liner under direct or indirect restorations
- 4) Repair of small defects in esthetic indirect restorations
- 5) Pit and fissure sealant
- 6) Blocking out of undercuts
- 7) Repair of resin and acrylic temporary materials
#### 5.6 SUBSTANTIAL EQUIVALENCE
The Novidia™ Flowable Composite has the same indications as the Filtek™ Supreme Ultra Flowable Restorative.
The technological characteristics of the Novidia™ Flowable Composite are substantially equivalent to these of the predicate device and other methacrylate-based products currently on the market. Table 5-1 below shows a comparison of Novidia™ Flowable Composite and the predicate device.
{5}------------------------------------------------
| Feature | Novidia™ Flowable Composite | Filtek Supreme Ultra Flowable<br>Restorative |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K100235 |
| Classification | EBF | EBF |
| Indications | • Class III and V restorations<br>• Restoration of minimally invasive<br>cavity preparations (including<br>small, non-stress-bearing occlusal<br>restorations)<br>• As base/liner under direct or<br>indirect restorations<br>• Repair of small defects in esthetic<br>indirect restorations<br>• Pit and fissure sealant<br>• Blocking out of undercuts<br>• Repair of resin and acrylic<br>temporary materials | • Class III and V restorations<br>• Restoration of minimally invasive<br>cavity preparations (including<br>small, non-stress- bearing occlusal<br>restorations)<br>• Base/liner under direct<br>restorations<br>• Repair of small defects in esthetic<br>indirect restorations<br>• Pit and fissure sealant<br>• Undercut blockout<br>• Repair of resin and acrylic<br>temporary materials |
| Composition | Methacrylate resins, photo-<br>initiators, inorganic fillers | Methacrylate resins, photo-<br>initiators, inorganic fillers |
| Packaging | Syringe and single-dose capsule | Syringe and single-dose capsule |
| Flexural<br>Strength | ISO 4049:2009 | ISO 4049:2009 |
| Intensity for curing | ≥550 mW/cm2 | 400 -1000 mW/cm2 |
| Wavelength for<br>curing | 430-490 nm | 400-500 nm |
| Curing time | 20 sec. | Opaque shades 40 sec;<br>All other shades 20 sec. |
| Radio-opacity | ISO 4049: 2009 | ISO 4049: 2009 |
| Depth of Cure | ISO 4049: 2009 | ISO 4049: 2009 |
| Water Sorption | ISO 4049:2009 | ISO 4049:2009 |
| Water<br>Solubility | ISO 4049:2009 | ISO 4049:2009 |
| Spontaneous<br>Polymerization<br>Sensitivity at<br>Ambient Light | ISO 4049:2009 | ISO 4049:2009 |
# Table 5-1: Comparison of Novidia™ Flowable Composite and Filtek™ Supreme Ultra Flowable Restorative
{6}------------------------------------------------
#### 5.7 PERFORMANCE DATA
## Non-Clinical Performance Testing:
Non-clinical and biological testing was completed to assess the performance and biocompatibility of the Novidia™ Flowable Composite and to support substantial equivalence. The data provided in this 510(k) submission shows that the composite is biocompatible based on the biocompatibility assessment conducted as per ISO 10993 and ISO 7405 and performs as intended based on the bench testing per ISO 4049 and FDA guidance document "Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions". The list of these tests is provided in Table 5-2.
| Biocompatibility |
|---------------------------------------------------------|
| Cytotoxicity |
| Oral Mucosal Irritation Test |
| Acute Systemic Toxicity |
| Material Mediated Pyrogenicity |
| Bacterial Reverse Mutation |
| Mouse Lymphoma Assay |
| Biological Risk Assessment |
| Bench Testing |
| Flexural strength |
| Elastic modulus |
| Compression strength |
| Shade and color stability |
| Polymerization conversion degree |
| Viscosity |
| Spontaneous polymerization sensitivity at ambient light |
| Water solubility |
| Water sorption |
| Depth of cure |
| Radio-opacity |
| Knoop hardness |
Table 5-2: List of Tests Completed on Novidia™ Flowable Composite
An additional test that was performed is "Preservation of Surface Integrity". This study used the atomic force microscopy (AFM) method to test the hypothesis that incorporation of QASi into the Novidia Flowable Composite prevents the increase of surface roughness (a surrogate measure of material degradation) caused by biofilm. The study concluded that the Novidia Flowable Composite with 1.2% (wt/wt) QASi filler is resistant to microbial degradation evidenced by preserved surface roughness in presence of continuous microbial challenge and therefore supports the thesis that 1.2% QASi filler improves the integrity of the dental composite restorations and preserves its functionality over time. Clinical studies demonstrating that the addition of 1.2% wt/wt QASi results in enhanced clinical outcomes have not been conducted.
# Animal and Clinical Performance Testing:
Animal and clinical performance data was not included.
{7}------------------------------------------------
#### 5.8 CONCLUSION
Nobio Ltd. believes that Novidia™ Flowable Composite is substantially equivalent to the Filtek™ Supreme Ultra Flowable Restorative and other legally marketed products. It does not introduce new indications for use, has similar technological characteristics and does not introduce any new safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.