K100235 · 3M Company · EBF · May 12, 2010 · Dental
Device Facts
Record ID
K100235
Device Name
FILTEK SUPREME ULTRA FLOWABLE RESTORATIVE
Applicant
3M Company
Product Code
EBF · Dental
Decision Date
May 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
Class III and V restorations Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations) Base/liner under direct restorations Repair of small defects in esthetic indirect restorations Pit and fissure sealant Undercut blockout Repair of resin and acrylic temporary materials
Device Story
Low-viscosity, visible-light activated, radiopaque, flowable nanocomposite restorative material; used by dentists in clinical settings. Input: light energy (curing light). Transformation: light-induced polymerization of methacrylate-based resins and surface-treated fillers (ytterbium trifluoride, silica, zirconia/silica clusters) to form a hard, bonded composite. Output: cured dental restorative material. Benefit: provides tooth-colored restoration, fluorescence, and polish retention for various dental defects and cavity preparations.
Clinical Evidence
Bench testing only. Performance compared to predicate devices 3M™ LVR System and Revolution Formula 2. Evaluated properties: Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Sorption, Water Solubility, Polymerization Shrinkage, Polish Retention, and Fluorescence. Biocompatibility assessment conducted per ISO 10993 Parts 3, 5, 6, 10, and 11.
Indicated for dental patients requiring Class III and V restorations, minimally invasive cavity preparations, base/liner placement, repair of esthetic indirect restorations, pit and fissure sealing, undercut blockout, and repair of resin/acrylic temporary materials.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
**3M ESPE**
K100235
#### 510(k) Summary vi
3M ESPE Dental Products 3M Center St. Paul, MN 55144-1000 651 733 1110
MAY 1 2 2010
Image /page/0/Picture/6 description: The image shows the logo for 3M ESPE. The logo is in black and white and features the text "3M ESPE" in a bold, sans-serif font. The letters are evenly spaced and the overall design is simple and clean.
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Submitter................ | 3M Company<br>3M ESPE Dental Products<br>3M Center, Bldg. 275-2W-08<br>St. Paul, MN 55144-1000 USA<br>Establishment Registration Number<br>2110898 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person................ | Scott Erickson<br>Regulatory Affairs Specialist<br>Phone: (651) 736-9883<br>Fax: (651) 736-1599<br>sterickson@mmm.com |
| Date Summary was Prepared...... | January 25, 2010 |
| Trade Name................ | FiltekTM Supreme Ultra Flowable<br>Restorative |
| Common Name(s)................. | Tooth shade resin material |
| Recommended Classification...... | Tooth shade resin material<br>(21 CFR 872.3690,<br>Product Code: EBF) |
Predicate Devices: 3M™ LVR System, Revolution Formula 2
{1}------------------------------------------------
# 3M ESPE
#### Indications for Use:
- · Class III and V restorations
- · Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations)
- · Base/liner under direct restorations
- · Repair of small defects in esthetic indirect restorations
- Pit and fissure sealant
- Undercut blockout
- · Repair of resin and acrylic temporary materials
### Description of Device:
Filtek Supreme Ultra Flowable Restorative, is a low viscosity, visible-light activated, radiopaque, flowable nanocomposite. The restorative is packaged in capsules and syringes. It is available in a variety of toothcolored shades. The shades offered with Filtek Supreme Ultra flowable restorative were designed to coordinate with shades offered with Filtek™ Supreme Ultra Universal Restorative. Filtek Supreme Ultra flowable restorative contains bisGMA, TEGDMA and Procrylat resins. The fillers are a combination of ytterbium trifluoride filler with a range of particles sizes from 0.1 to 5.0 microns, a non-agglomerated/non-aggregated surface modified 20nm silica filler, a non-agglomerated/non-aggregated surface modified 75nm silica filler, and a surface modified aggregated zirconia/silica cluster filler (comprised of 20 nm silica and 4 to 11 nm zirconia particles). The aggregate has an average cluster particle size of 0.6 to 10 microns. The inorganic filler loading is approximately 65% by weight (46% by volume).
Filtek™ Supreme Ultra Flowable Restorative is a minor modification of predicate device 3M™ LVR System. The formulation was modified to improve physical properties, such as fluorescence and polish retention and to provide additional tooth colored shades of the restorative material.
When irradiated by light, the methacrylate functionalities of the resins and surface-treated fillers undergo, in conjunction with the photoinitiator system, a light-induced polymerization to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
#### Technological Characteristics:
Filtek Supreme Ultra flowable restorative contains bisGMA, TEGDMA and Procrylat resins. The fillers are a combination of ytterbium trifluoride filler with a range of particles sizes from 0.1 to 5.0 microns, a nonagglomerated/non-aggregated surface modified 20nm silica filler, a nonagglomerated/non-aggregated surface modified 75nm silica filler, and a surface modified aggregated zirconia/silica cluster filler (comprised of 20 nm silica and 4 to 11 nm zirconia particles). The aggregate has an average
{2}------------------------------------------------
# 3M ESPE
cluster particle size of 0.6 to 10 microns. The inorganic filler loading is approximately 65% by weight (46% by volume).
When irradiated by light, the methacrylate functionalities of the resins and surface-treated fillers undergo, in conjunction with the photoinitiator system, a light-induced polymerization to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
Filtek™ Supreme Ultra Flowable Restorative is a minor modification of predicate device 3M™ LVR System. The formulation was modified to improve physical properties, such as fluorescence and polish retention and to provide additional tooth colored shades of the restorative material. As a result of the reformulation, a biocompatibility assessment was developed for FiltekTM Supreme Ultra Flowable Restorative using standard risk assessment techniques and consideration of FDA & internationally recognized guidelines, including ISO 10993 Parts 3, 5, 6, 10 and 11. The conclusion of the assessment is that Filtek™ Supreme Ultra Flowable Restorative is safe for its intended use.
#### Summary of Physical Tests:
This 510(k) submission includes data from bench testing to evaluate the performance of Filtek™ Supreme Ultra Flowable Restorative compared to predicate devices 3M™ LVR System and Revolution Formula 2. Properties evaluated include Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Sorption, Water Solubility, Polymerization Shrinkage, Polish Retention and Fluorescence.
#### Substantial Equivalence:
Information provided in this 510(k) submission shows that Filtek™ Supreme Ultra Flowable Restorative is substantially equivalent to the predicate device 3M™ LVR System in terms of intended use, indications for use, composition, physical properties and technological characteristics. Filtek™ Supreme Ultra Flowable Restorative is substantially equivalent to the predicate device Revolution Formula 2 in terms of intended use, indications for use, physical properties and technological characteristics. Physical properties evaluated include Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Sorption, Water Solubility, Polymerization Shrinkage, Polish Retention and Fluorescence. A comparison of technological characteristics is provided below:
{3}------------------------------------------------
# ЗМ ESPE
| Technological property | Filtek™<br>Supreme<br>Ultra<br>Flowable | 3M™<br>LVR | Revolution<br>Formula 2 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|------------|-------------------------|
| Camphorquinone/amine photoinitiator system | X | X | - |
| Methacrylate-based resin matrix | X | X | X |
| Silane treated fillers | X | X | - |
| Bonded with a permanent dental adhesive | X | X | X |
| When irradiated by light, the methacrylate<br>functionalities of the resins and surface-treated<br>fillers undergo, in conjunction with the<br>photoinitiator system, a light-induced<br>polymerization to form a hard composite that is<br>bonded to the tooth structure with a permanent<br>dental adhesive. | X | X | X |
:
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a bird or abstract human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
3M Company Mr. Scott Erickson, RAC Regulatory Affairs 3M ESPE Dental Products 3M Center, Building 275-2W-08 St. Paul. Minnesota 55144-1000
MAY 1 2 2010
Re: K100235
Trade/Device Name: Filtek TM Supreme Ultra Flowable Restorative Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: April 23, 2010 Received: April 27, 2010
Dear Mr. Scott Erickson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Scott Erickson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 3M ESPE
#### Indications for Use Statement 4.
#### Indications for Use
510(k) Number (if known): K |00235
Device Name: Filtek™ Supreme Ultra Flowable Restorative
Indications for Use:
- · Class III and V restorations
- · Restoration of minimally invasive cavity preparations (including small, non stressbearing occlusal restorations)
- · Base/liner under direct restorations
- · Repair of small defects in esthetic indirect restorations
- Pit and fissure sealant
- Undercut blockout
- · Repair of resin and acrylic temporary materials
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Kein Nule for HSR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
Page *_ of _*
510(k) Number: K100235
Filtek™ Supreme Ultra Flowable Restorative 510(k)
Page 13 of 111
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.