Microflex Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Green Pink Colored Tested for
K200671 · Ansell Healthcare Products, LLC · LZA · Jun 19, 2020 · General Hospital
Device Facts
Record ID
K200671
Device Name
Microflex Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Green Pink Colored Tested for
Applicant
Ansell Healthcare Products, LLC
Product Code
LZA · General Hospital
Decision Date
Jun 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, powder-free nitrile butadiene rubber examination gloves; available in blue, green, and pink. Features beaded cuff, straight fingers, and finger-textured surface; coated with polyacrylic polymer for donning and infused with aloe and chamomile. Used by healthcare professionals in clinical settings to prevent cross-contamination. Tested for permeation resistance against various chemotherapy drugs per ASTM D6978-05; pink variant also tested against Fentanyl Citrate. Output is a physical barrier; healthcare providers use gloves during patient examinations and handling of hazardous drugs. Benefits include protection against patient-to-examiner contamination and chemical permeation during drug handling.
Clinical Evidence
Bench testing only. No clinical data required. Performance validated via ASTM D6319-10 (dimensions, physical properties, freedom from holes, powder residual), ASTM D6978-05 (chemotherapy drug permeation), and ISO 10993 series (cytotoxicity, skin irritation, sensitization, acute systemic toxicity).
Indicated for use as a disposable, powder-free patient examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs and (for the pink variant) Fentanyl Citrate. Contraindicated for use with Carmustine.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 19, 2020
Ansell Healthcare Products LLC Donald Cronk Associate Director, Regulatory Affairs - Americas 2301 Robb Drive Reno, Nevada 89523
Re: K200671
Trade/Device Name: Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl CitrateRegulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO (for pink glove only) Dated: May 21, 2020 Received: May 22, 2020
Dear Donald Cronk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200671
#### Device Name
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Tested chemotherapy drugs are as follows:
| Test Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
|----------------------------------------|-----------------------------------------------|
| Carmustine (BCNU) - 3.3 mg/ml | 17.4 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) 2.0 mg/ml | >240 |
| Etoposide (Toposar) - 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) - 6.0 mg/ml | >240 |
| Thiotepa (THT) - 10.0 mg/ml | 67.1 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmusine (BCNU) : 17.4 minutes and Thiotepa: 67.1 minutes. Warning: Do not use with Carmustine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K200671
#### Device Name
Microllex® Nitrile Patient Examination Gloves with Aloe and Chamonile Green Colored Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Tested chemotherapy drugs are as follows: | |
|-------------------------------------------|-----------------------------------------------|
| Test Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
| Carmustine (BCNU) - 3.3 mg/ml | 27.9 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) 2.0 mg/ml | >240 |
| Etoposide (Toposar) - 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) - 6.0 mg/ml | >240 |
| Thiotepa (THT) - 10.0 mg/ml | 48.6 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 27.9 minutes and Thiotepa: 48.6 minutes. Warning: Do not use with Carmustine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200671
#### Device Name
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Druss and Fentanyl Citrate
#### Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Tested chemotherapy drugs are as follows: | |
|-------------------------------------------|-----------------------------------------------|
| Test Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
| Carmustine (BCNU) - 3.3 mg/ml | 23.4 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) 2.0 mg/ml | >240 |
| Etoposide (Toposar) - 20.0 mg/ml | >240 |
| Fentanyl Citrate Injection - 100 mcg/2mL | >240 |
| Fluorouracil (Adrucil) - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) - 6.0 mg/ml | >240 |
| Thiotepa (THT) - 10.0 mg/ml | 64.9 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation (BCNU): 23.4 minutes and Thiotepa: 64.9 minutes. Warning: Do not use with Carmustine.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510K Summary
## 510(k) Number:
K200671
# Submitter:
Ansell Healthcare Products LLC. 2301 Robb Drive Reno, NV 89523
# Contact Person:
Don Cronk Associate Director, Regulatory Affairs - Americas Phone: (775) 470-7106 Email: don.cronk@ansell.com
## Date Prepared:
6/19/2020
## Name of the Device:
| Trade Names: | Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile<br>Blue Colored Tested for Use with Chemotherapy Drugs<br>Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile<br>Green Colored Tested for Use with Chemotherapy Drugs<br>Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile<br>Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl<br>Citrate |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Patient Examination Glove |
| Classification Name: | Patient Examination Glove |
| Classification Regulation: | 21 CFR 880.6250 |
| Device Class: | I |
| Product Code: | LZA, LZC, QDO |
| Classification Panel: | Non-powdered patient examination glove |
## Legally Marketed Predicate Device:
| Company: | Ansell Healthcare Products LLC |
|----------------|------------------------------------------------------------------------------------------------|
| Trade Name: | Micro-Touch NitraTex Sterile Nitrile Powder-Free Blue Examination<br>Gloves (Chemotherapy Use) |
| 510(k) Number: | K082457 |
| Device Class: | Class I |
| Product Code: | LZA (Nitrile) |
| Device Name: | Patient Examination Glove (21 CFR 880.6250) |
## Device Description:
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable, powder free examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy.
{6}------------------------------------------------
Characteristic:
- Ambidextrous with beaded cuff and straight fingers
- Finger-textured, ●
- Blue, green or pink colored
- Containing aloe and chamomile.
- Five (5) sizes – extra-small, small, medium, large, and extra-large.
- . Tested against chemotherapy drugs and fentanyl citrate (pink only).
High levels of ozone will degrade rubber material of the glove, therefore the glove should be protected from ozone in particular.
The gloves are designed to meets the specifications of ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.
# Indications for Use Statements:
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
Tested chemotherapy drugs are as follows:
| Test Chemotherapy Drug & Concentration | Average Breakthrough Detection Time (Min) |
|------------------------------------------|-------------------------------------------|
| Carmustine (BCNU) – 3.3 mg/ml | 23.4 |
| Cyclophosphamide – 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) – 2.0 mg/ml | >240 |
| Etoposide (Toposar) – 20.0 mg/ml | >240 |
| Fentanyl Citrate Injection – 100 mcg/2mL | >240 |
| Fluorouracil (Adrucil) – 50.0 mg/ml | >240 |
| Methotrexate – 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) – 6.0 mg/ml | >240 |
| Thiotepa (THT) – 10.0 mg/ml | 64.9 |
| Vincristine Sulfate – 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 23.4 minutes and Thiotepa: 64.9 minutes. Warning: Do not use with Carmustine.
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
{7}------------------------------------------------
Tested chemotherapy drugs are as follows:
| Test Chemotherapy Drug & Concentration | Average Breakthrough Detection Time (Min) |
|------------------------------------------|-------------------------------------------|
| Carmustine (BCNU) – 3.3 mg/ml | 17.4 |
| Cyclophosphamide – 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) – 2.0 mg/ml | >240 |
| Etoposide (Toposar) – 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) – 50.0 mg/ml | >240 |
| Methotrexate – 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) – 6.0 mg/ml | >240 |
| Thiotepa (THT) – 10.0 mg/ml | 67.1 |
| Vincristine Sulfate – 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 17.4 minutes and Thiotepa: 67.1 minutes. Warning: Do not use with Carmustine.
Microflex® Nitrile Patient Examination Glove with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs.
Tested chemotherapy drugs are as follows:
| Test Chemotherapy Drug & Concentration | Average Breakthrough Detection Time (Min) |
|------------------------------------------|-------------------------------------------|
| Carmustine (BCNU) – 3.3 mg/ml | 27.9 |
| Cyclophosphamide – 20.0 mg/ml | >240 |
| Doxorubicin HCl (Adriamycin) – 2.0 mg/ml | >240 |
| Etoposide (Toposar) – 20.0 mg/ml | >240 |
| Fluorouracil (Adrucil) – 50.0 mg/ml | >240 |
| Methotrexate – 25.0 mg/ml | >240 |
| Paclitaxel (Taxol) – 6.0 mg/ml | >240 |
| Thiotepa (THT) – 10.0 mg/ml | 48.6 |
| Vincristine Sulfate – 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 27.9 minutes and Thiotepa: 48.6 minutes. Warning: Do not use with Carmustine.
# Technological Characteristics:
Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Blue Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Green Colored Tested for Use with Chemotherapy Drugs, Microflex® Nitrile Patient Examination Gloves with Aloe and Chamomile Pink Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate have the following technological characteristics as compared to ASTM or equivalent standards:
{8}------------------------------------------------
| Characteristics | Standard/Test/<br>FDA Guidance | Result Summary |
|--------------------------------------|--------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Physical Characteristics: | | |
| Dimensions: | ASTM D6319-10 | Meets ASTM D6319-10 requirements for<br>length, width and thickness |
| Length | Minimum 230mm | Minimum 240mm |
| Palm width (mm) | | |
| Size - XS | 70 ± 10 | 75 ± 5 |
| Size - S | 80 ± 10 | 85 ± 5 |
| Size - M | 95 ± 10 | 95 ± 5 |
| Size - L | 110± 10 | 105 ± 5 |
| Size - XL | 120 ± 10 | 115 ± 5 |
| Thickness (mm) - single-wall | | |
| Finger | minimum 0.05 | Finger - min 0.09 |
| Palm | minimum 0.05 | Palm - min 0.06 |
| Cuff | - | Cuff – min 0.05 |
| Physical Properties: | ASTM D6319-10 | Meets ASTM D6319-10 requirements for tensile<br>strength and ultimate elongation before and<br>after accelerated aging: |
| Tensile Strength | | |
| Before Aging | minimum 14 MPa | minimum 16 MPa |
| After Aging | minimum 14 MPa | minimum 14 MPa |
| Ultimate Elongation | | |
| Before Aging | minimum 500% | minimum 500% |
| After Aging | minimum 400% | minimum 400% |
| Freedom from holes | ASTM D6319-10<br>ASTM D5151-06 | Meets or exceeds ASTM D6319-10 and ASTM<br>D5151-06 requirements of AQL 2.5 |
| Powder Residual | ASTM D6319-10 | Meets applicable requirement for powder free;<br>≤ 2 mg per glove |
| | ASTM D6124-06 | |
| Biocompatibility: | | |
| ISO In Vitro Cytotoxicity | ISO 10993-5:2009 | Under the conditions of the study, undiluted, 1:2, 1:4, 1:8, 1:16 dilution was cytotoxic., 1:32 and 1:64 are not cytotoxic |
| ISO Skin Irritation Study | ISO10993-10:2010 | Under the conditions of the study, not an irritant |
| ISO Maximization Sensitization Study | ISO 10993-10:2010 | Under the conditions of the study, not a sensitizer |
| ISO acute systemic toxicity | ISO 10993-11: 2006 | Under the conditions of the study, no evidence of systemic toxicity |
{9}------------------------------------------------
# Substantial Equivalence:
| | Predicate Device | Proposed Subject Device | Substantial<br>Equivalence to<br>Predicate |
|---------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Trade name | Micro-Touch NitraTex Sterile<br>Nitrile Powder-Free Blue<br>Examination Gloves<br>(Chemotherapy Use) | Microflex® Nitrile Patient<br>Examination Gloves with Aloe<br>and Chamomile Blue Colored<br>Tested for Use with<br>Chemotherapy Drugs<br>Microflex® Nitrile Patient<br>Examination Gloves with Aloe<br>and Chamomile Green Colored<br>Tested for Use with<br>Chemotherapy Drugs<br>Microflex® Nitrile Patient<br>Examination Gloves with Aloe<br>and Chamomile Pink Colored<br>Tested for Use with<br>Chemotherapy Drugs and<br>Fentanyl Citrate | Not applicable |
| 510k Number | K082457 | K200671 | Not Applicable |
| Product Owner | Ansell Healthcare | Ansell Healthcare | Ansell Healthcare |
| Product Code | LZA | LZA, LZC, QDO | Similar |
{10}------------------------------------------------
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Yes |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Regulatory Class | I | I | Yes |
| Regulation Name | Patient Examination Glove | Patient Examination Glove | Yes |
| Indications for use | The patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner | A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner | Yes |
| Material<br>Composition | Synthetic nitrile rubber | Synthetic nitrile rubber | Yes |
| Coating | Polyacrylic polymer inner coating to aid donning | Polyacrylic polymer inner coating to aid donning | Yes |
| Aloe and<br>Chamomile | N/A | Aloe and Chamomile coated on the donning surface | As noted |
| Design | Non-sterile | Non-sterile | Yes |
| | Single use | Single use | Yes |
| | Powder-free | Powder-free | Yes |
| | Ambidextrous | Ambidextrous | Yes |
| | Beaded cuff | Beaded cuff | Yes |
| Color | Blue | Green, Blue, and Pink | As noted |
| Performance | Meets ASTM D6319-10<br>requirements | Meets ASTM D6319-10<br>requirements | Yes |
| a. Dimensions | | | |
| b. Physical<br>Properties | Meets ASTM D6319-10<br>requirements | Meets ASTM D6319-10<br>requirements | Yes |
| c. Freedom from<br>holes | Meets ASTM D6319-10<br>requirements of GI, AQL 2.5 | Meets ASTM D6319-10<br>requirements of GI, AQL 2.5 | Yes |
| d. Powder Residual | Meets ASTM D6319-10<br>requirements; Not more than<br>2.0mg/glove | Meets ASTM D6319-10<br>requirements; Not more than<br>2.0mg/glove | Yes |
{11}------------------------------------------------
| e. Sterility | Non-sterile | Non-sterile | Yes | | |
|-----------------------|----------------------------------------------------------------------|-------------|-------------------------------------------------------------------------------------------------------------|------|--|
| Biocompatibility | Passes Primary Skin Irritation Test<br>and Dermal Sensitization Test | | Passes Primary Skin Irritation<br>Test and Dermal Sensitization<br>Test and Acute Systemic Toxicity<br>Test | Yes | |
| Chemotherapy<br>Claim | Predicate Device | Pink Device | As Noted (Fentanyl<br>Citrate) | | |
| | Carmustine (BCNU) | 32.1 | Carmustine (BCNU) | 23.4 | |
| | Cyclophosphamide | >240 | Cyclophosphamide | >240 | |
| | Doxorubicin HCl<br>(Adriamycin) | >240 | Doxorubicin HCl<br>(Adriamycin) | >240 | |
| | Etoposide (Toposar) | >240 | Etoposide (Toposar) | >240 | |
| | Fluorouracil (Adrucil) | >240 | Fluorouracil (Adrucil) | >240 | |
| | Methotrexate | >240 | Methotrexate | >240 | |
| | Paclitaxel (Taxol) | >240 | Paclitaxel (Taxol) | >240 | |
| | Thiotepa (THT) | 140.7 | Thiotepa (THT) | 64.9 | |
| | Vincristine Sulfate | >240 | Vincristine Sulfate | >240 | |
| | | | Fentanyl Citrate<br>Injection | >240 | |
| Chemotherapy<br>Claim | Predicate Device | Blue Device | Yes | | |
| | Carmustine (BCNU) | 32.1 | Carmustine (BCNU) | 17.4 | |
| | Cyclophosphamide | >240 | Cyclophosphamide | >240 | |
| | Doxorubicin HCl<br>(Adriamycin) | >240 | Doxorubicin HCl<br>(Adriamycin) | >240 | |
| | Etoposide (Toposar) | >240 | Etoposide (Toposar) | >240 | |
| | Fluorouracil (Adrucil) | >240 | Fluorouracil (Adrucil) | >240 | |
| | Methotrexate | >240 | Methotrexate | >240 | |
| | Paclitaxel (Taxol) | >240 | Paclitaxel (Taxol) | >240 | |
| | Thiotepa (THT) | 140.7 | Thiotepa (THT) | 67.1 | |
| | Vincristine Sulfate | >240 | Vincristine Sulfate | >240 | |
{12}------------------------------------------------
| Chemotherapy<br>Claim | Predicate Device | Green Device | Yes |
|-----------------------|--------------------------------------|--------------------------------------|-----|
| | Carmustine (BCNU) 32.1 | Carmustine (BCNU) 27.9 | |
| | Cyclophosphamide >240 | Cyclophosphamide >240 | |
| | Doxorubicin HCl<br>(Adriamycin) >240 | Doxorubicin HCl<br>(Adriamycin) >240 | |
| | Etoposide (Toposar) >240 | Etoposide (Toposar) >240 | |
| | Fluorouracil (Adrucil) >240 | Fluorouracil (Adrucil) >240 | |
| | Methotrexate >240 | Methotrexate >240 | |
| | Paclitaxel (Taxol) >240 | Paclitaxel (Taxol) >240 | |
| | Thiotepa (THT) 140.7 | Thiotepa (THT) 48.6 | |
| | Vincristine Sulfate >240 | Vincristine Sulfate >240 | |
The subject device meets the applicable requirements for patient examination gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D6319, ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization, irritation and acute systemic toxicity, in accord with the ISO 10993 series of standards.
A clinical study was not required for the subject or predicate devices.
# Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject devices are as safe, as effective, and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.