The Lucero Medical Cervical Spacer is intended for anterior intervertebral body fusion in skeletally mature patients who have had at least six weeks of non-operative treatment. The Lucero Medical Cervical Spacer is indicated to treat cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The Lucero Medical Cervical Spacer is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone to facilitate fusion and to be used with supplemental fixation cleared for use in the cervical spine.
Device Story
The Lucero Medical Cervical Spacer is an additively manufactured interbody fusion device. It features a keystone cross-sectional shape and an open structural architecture. Available in various heights, lengths, widths, and angulations to match patient anatomy. Used by surgeons in anterior cervical interbody fusion procedures to facilitate spinal fusion. The device is packed with autogenous or allogeneic bone graft and requires supplemental fixation cleared for cervical spine use. It provides structural support to the intervertebral space to alleviate symptoms of disc degeneration or instability. The device is provided sterile.
Clinical Evidence
Bench testing only. Mechanical testing included static and dynamic compression, static torsion (ASTM F2077), subsidence (ASTM F2267), and expulsion testing on worst-case configurations.
Technological Characteristics
Additively manufactured interbody spacer; material: Ti-6Al-4V ELI titanium alloy (ASTM F3001); open structural architecture; keystone cross-sectional shape; provided sterile.
Indications for Use
Indicated for skeletally mature patients with cervical disc degeneration and/or cervical spinal instability (confirmed by imaging) resulting in radiculopathy, myelopathy, or pain at C2-T1 levels, following at least six weeks of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K252737 — DeGen Medical Latitude-C AM Cervical Interbody Fusion System · Degen Medical · Oct 10, 2025
K193359 — SureMAX Family of Cervical Spacers · Additive Implants, Inc. · Jan 3, 2020
K230607 — SPICCA Cervical Fusion Cages · Southern Medical (Pty) , Ltd. · Aug 14, 2023
K213359 — SeaSpine WaveForm, C Interbody System · SeaSpine Orthopedics Corporation · Dec 9, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 16, 2020
Lucero Medical, LLC % Karen E. Warden, PhD President BackRoads Consulting, Inc 12520 Heath Road Chesterland, Ohio 44026
Re: K200549
Trade/Device Name: Lucero Medical Cervical Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: December 9, 2020 Received: December 10, 2020
Dear Dr. Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K200549
Device Name Lucero Medical Cervical Spacer System
## Indications for Use (Describe)
The Lucero Medical Cervical Spacer is intended for anterior intervertebral body fusion in skeletally mature patients who have had at least six weeks of non-operative treatment. The Lucero Medical Cervical Spacer is indicated to treat cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2 - T1. The Lucero Medical Cervical Spacer is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone to facilitate fusion and to be used with supplemental fixation cleared for use in the cervical spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> <line style="stroke:#000000;stroke-width:2" x1="2" x2="13" y1="2" y2="13"></line> <line style="stroke:#000000;stroke-width:2" x1="13" x2="2" y1="2" y2="13"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of*
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Date: | 2 March 2020 |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Lucero Medical, LLC<br>6815 West Streetsboro<br>Richfield, OH 44286<br>Office: 330.659.0855 |
| Sponsor Contact: | Lee Strnad, CEO |
| 510(k) Contact: | Karen E. Warden, PhD<br>BackRoads Consulting Inc.<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457 |
| Proposed Trade Name: | Lucero Medical Cervical Spacer System |
| Common Name: | Cervical interbody fusion device |
| Device Classification: | Class II |
| Regulation Names,<br>Regulation Numbers,<br>Product Codes: | Intervertebral fusion device with bone graft, 888.3080, cervical, ODP |
| Device Description: | The Lucero Medical Cervical Spacer is an additively manufactured<br>interbody device. These cervical implants have basic keystone cross-<br>sectional shape and an open structural architecture. A variety of height,<br>length, width and anteroposterior angulation combinations are available to<br>accommodate the anatomic requirements of individual patients. The devices<br>are provided sterile. |
| Indications for Use: | The Lucero Medical Cervical Spacer is intended for anterior intervertebral<br>body fusion in skeletally mature patients who have had at least six weeks of<br>non-operative treatment. The Lucero Medical Cervical Spacer is indicated to<br>treat cervical disc degeneration and/or cervical spinal instability, as<br>confirmed by imaging studies (radiographs, CT, MRI), that results in<br>radiculopathy, myelopathy, and/or pain at multiple contiguous levels from<br>C2 - T1. The Lucero Medical Cervical Spacer is designed for use with<br>autogenous and/or allogeneic bone graft comprised of cancellous and/or<br>corticocancellous bone to facilitate fusion and to be used with supplemental<br>fixation cleared for use in the cervical spine. |
| Materials: | The Lucero Medical Cervical Spacer implants are manufactured from Ti-<br>6Al-4V ELI titanium alloy (ASTM F3001). |
| Primary Predicate: | Cascadia Interbody System (K2M Inc. - K160125) |
| Additional Predicates: | Tranquil™ Interbody System System (Nexus Spine, LLC. - K181483), MC+<br>(LDR Holding - K091088) |
| Performance Data: | Mechanical testing of the worst case Lucero Medical Cervical Spacer<br>included static and dynamic compression and static torsion according to<br>ASTM F2077. In addition, subsidence according to ASTM F2267 and<br>expulsion tests were performed.<br>The mechanical test results demonstrate that the Lucero Medical Cervical<br>Spacer System performance is substantially equivalent to the predicate<br>devices. |
| Technological<br>Characteristics: | The Lucero Medical Cervical Spacer System possesses the same<br>technological characteristics as one or more of the predicate devices. These<br>include:<br>• intended use (as described above)<br>• basic design (additively manufactured structure),<br>• material (titanium alloy) and<br>• sizes (dimensions are comparable to those offered by the predicate<br>systems)<br>The Lucero Medical Cervical Spacer is the same as previously cleared<br>devices. |
| Conclusion: | The Lucero Medical Cervical Spacer System possesses the same intended<br>use and technological characteristics as the predicate devices. Therefore<br>the Lucero Medical Cervical Spacer System is substantially equivalent for<br>its intended use. |
{4}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.