Non-contact Infrared Body Thermometer, model:HTD8823US
K200159 · Hetaida Technology Co., Ltd. · FLL · Jan 5, 2021 · General Hospital
Device Facts
Record ID
K200159
Device Name
Non-contact Infrared Body Thermometer, model:HTD8823US
Applicant
Hetaida Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jan 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared Body Thermometer, Model: HTD8823US, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the forehead in people of all ages for home setting use.
Device Story
Hand-held, battery-powered infrared thermometer; uses thermopile sensor to detect infrared radiation from forehead. Two modes: forehead mode (non-contact, 1-5 cm distance) and forehead scan mode (contact, sliding probe across forehead). IR sensor output amplified and processed by microcontroller unit (MCU). MCU calculates temperature; displays result on LCD. Features color-coded temperature indication (green/yellow/orange). Used by patients or caregivers in home settings. Provides rapid temperature readings to assist in clinical decision-making regarding fever management.
Clinical Evidence
Clinical accuracy testing performed per ASTM E1965-98 (2016) on 140 subjects across age groups (0-3 months, 3 months-1 year, 1-5 years, >5 years). Results demonstrated compliance with ASTM E1965-98 accuracy requirements.
Indicated for people of all ages for the measurement of body temperature from the forehead in a home setting.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Braun No Touch + Forehead NTF3000 Thermometer (K163516)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
January 5, 2021
HeTaiDa Technology Co., Ltd. % You Yijie Manager Oimmig Medical Consulting Service Co., Ltd RM.1711, Building K, NO.101 Science Ave International Creative Valley Development Zone Guangzhou, 510663 China
Re: K200159
Trade/Device Name: Infrared Body Thermometer, Model: HTD8823US Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 7, 2020 Received: December 7, 2020
Dear You Yijie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200159
Device Name Infrared Body Thermometer, Model: HTD8823US
Indications for Use (Describe)
The Infrared Body Thermometer, Model: HTD8823US, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the forehead in people of all ages for home setting use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary K200159
## 1. Submitter's Information
## Establishment Reqistration Information
Name: HeTaiDa Technology Co., Ltd Address: 4F, BaiShiDa High-Tech Park, XiangDong Industrial Area, DaLingShan Town, DongGuan City, Guangdong, China. 523820
## Contact Person of applicant
Name: Tom Chen Address: 4F, BaiShiDa High-Tech Park, XiangDong Industrial Area, DaLingShan Town, DongGuan City, Guangdong, China TEL: +86 769-82658050 FAX: +86 769-82658050 Email: tomchen@hetaida.com.cn
## Contact Person of the Submission:
Name: Yijie You Address: RM.1711, Building K, NO.101 Science Ave International Creative Valley Development Zone, Guangzhou China TEL: +86 020-8224 5821 FAX: +86 020-8224 5821 Email: Jet.you@qimmiq-med.com
Date prepared: Dec. 16, 2020
## 2. Device Information
| Device Common Name: | Clinical electronic thermometer |
|---------------------|---------------------------------|
| Trade Name: | Infrared Body Thermometer |
| Model: | HTD8823US |
| Regulation name: | Clinical electronic thermometer |
| Review Panel: | General Hospital |
| Product Code: | FLL |
| Regulation Class: | II |
| Regulation Number: | 880.2910 |
## 3. Predicate Device Information
#### Primary predicate device:
510(k) submitter/holder: KAZ USA, Inc., A Helen of Troy Company 510(K) Number: K163516 Model: Braun No Touch + Forehead NTF3000 Thermometer Trade name: Infrared Thermometer Review Panel: General Hospital Product Code: FLL
{4}------------------------------------------------
Requlation Class: . Requlation Number: 880.2910
#### Secondary predicate device:
510(k) submitter/holder: Exergen Corporation 510(K) Number: K011291 Trade name: TemporalScanner Thermometer, SensorTouch Review Panel: General Hospital Product Code: FLL Regulation Class: -Regulation Number: 880.2910
## 4. Device description
Infrared (IR) Body Thermometer, model: HTD8823US, is a hand-held, battery powered, infrared thermometer that measures human body temperature from forehead. The reference body site is oral.
The device can measure temperature with two modes, forehead mode and forehead scan mode, and both modes measure forehead temperature.
The forehead mode measures temperature from center of the forehead.
Forehead scan mode measures temperature by gently positioning the probe flush (flat) on the center of the forehead, midway between the eyebrow and the hairline, press and hold the On/Scan button. Lightly slide the thermometer across the forehead keeping the sensor flat and in contact with the skin until reaching the right hairline, release the On/Scan button and remove the thermometer from the forehead, then the temperature will display on the screen, the whole process takes 3~10 seconds.
## Principle of operation:
Infrared Body Thermometer, model: HTD8823US, is an electronic thermometer uses IR sensor (thermopile) to detect infrared radiation emitting from forehead. The IR sensor outputs electrical signal which is fed to circuit for amplification and then inputted to the microcontroller unit (MCU). For forehead mode, the MCU captures the temperature measured from center of forehead. The measured temperature will finally appear on LCD display. For forehead scan mode, the MCU captures the highest temperature during scanning sampling process and calculation, finally the measured temperature will appear on LCD display.
## 5. Indications for Use
The Infrared Body Thermometer, Model: HTD8823US, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the forehead in people of all ages for home setting use.
## 6. Summary of technological characteristics of device compared to the predicate devices (K163516 and K011291)
| SE<br>Comparisons | Subject device | Primary predicate device | Secondary predicate<br>device | Discussion<br>of difference |
|-------------------|---------------------------|-------------------------------------------------|-------------------------------|-----------------------------|
| | K200159<br>(Infrared Body | (K163516, Infrared<br>Thermometer, Model: Braun | (K011291, | |
{5}------------------------------------------------
| | Thermometer,<br>model: HTD8823US) | No Touch + Forehead<br>NTF3000 Thermometer)<br>(compared with Forehead<br>mode of HTD8823US) | TemporalScanner<br>Thermometer,<br>SensorTouch)<br>(compared with Forehead<br>scan mode of<br>HTD8823US) | |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Classification | 21CFR 880.2910 | 21CFR 880.2910 | 21CFR 880.2910 | Same |
| Product Code | FLL | FLL | FLL | Same |
| FDA Class | II | II | II | Same |
| Intended Use | The Infrared Body<br>Thermometer, Model:<br>HTD8823US, is an<br>electronic clinical<br>thermometer using an<br>infrared sensor to detect<br>body temperature from the<br>forehead in people of all<br>ages for home setting use. | The Braun No Touch +<br>Forehead NTF3000<br>Thermometer is a non-sterile,<br>reusable clinical thermometer<br>intended for the intermittent<br>determination of human body<br>temperature in a touch and no<br>touch mode on the center of<br>the forehead as the<br>measurement site on people<br>of all ages. | The TemporalScanner<br>Thermometer is an<br>infrared thermometer<br>intended for the<br>intermittent<br>measurement of human<br>body temperature of<br>people of all ages | Similar<br>(Discussion is<br>indicated in<br>D1) |
| Principle of<br>operation | Forehead mode: Measure<br>temperature by reading<br>infrared radiation emitting<br>from the forehead when<br>the thermometer is placed<br>within few centimeters of<br>forehead. | Measure temperature by<br>reading infrared radiation<br>emitting from the forehead<br>when the thermometer is<br>placed within few centimeters<br>of forehead. | NA | Same |
| | Forehead scan mode:<br>Measure temperature by<br>reading infrared radiation<br>emitting from the forehead<br>when the thermometer is<br>placed on the forehead. | NA | Measure temperature by<br>reading infrared radiation<br>emitting from the<br>forehead when the<br>thermometer is placed<br>on the forehead. | Same |
| target<br>population | people of all ages | people of all ages | people of all ages | Same |
| Measurement<br>site | forehead | forehead | forehead | Same |
| Material of<br>Patient<br>contact<br>components | ABS (device housing/ handle<br>and power button);<br>PMMA(LCD screen) | ABS (device housing/ handle and<br>power button); TPR (temperature<br>button and nose / forehead touch<br>bumper). | ABS | Similar<br>(Discussion is<br>indicated in<br>D2) |
| Biocompability<br>testing | Meets ISO 10993- 5<br>ISO 10993-10 | Meets ISO 10993 and<br>FDA Bluebook memo<br>G95-1 | Not public | Similar<br>(Discussion is<br>indicated in<br>D3) |
| Environment | home | home | home | Same |
| Design | Handheld | Handheld | Handheld | Same |
| Measurement<br>method | Infrared radiation<br>detection | Infrared radiation detection | Infrared radiation<br>detection | Same |
| Display Type | LCD | LCD | LCD | Same |
| Measurement | Forehead mode | Forehead mode | NA | Same |
| Mode | Forehead scan mode | NA | Forehead scan mode | Same |
| Key | One button(ON/Scan | Two button(Power button, | One button(SCAN | Similar<br>(Discussion is<br>indicated in<br>D4) |
| | button) | Temperature button) | Button) | indicated in<br>D4) |
| Scale<br>selection | °C/°F | °C/°F | °C/°F | Same |
| Display unit | °C/°F | °C/°F | °C/°F | Same |
| High<br>temperature<br>warning | Yes | Yes | Yes | Same |
| Low battery<br>indicator | Yes | Yes | Yes | Same |
| Case Material | ABS | ABS | ABS | Same |
| Sensor Type | Thermopile | Thermopile | Thermopile | Same |
| Performance<br>Testing | Meets ASTM E1965-98<br>and ISO 80601-2-56 | Meets ASTM E1965-98 and<br>ISO 80601-2-56 | Meets ASTM E1965-98 | Similar<br>(Discussion is<br>indicated in D5) |
| Electrical<br>Safety | ANSI AAMI ES60601-1 | ANSI AAMI ES60601-1 | ANSI AAMI ES60601-1 | Same |
| EMC Meets | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Measuring<br>Range for<br>Forehead<br>mode | 34.0°C ~ 43.0°C<br>(93.2°F ~109.4°F) | 34.4°C ~42.2°C<br>(93.9°F to 108.0°F) | NA | Similar<br>(Discussion is<br>indicated in D6) |
| Measuring<br>Range for<br>Forehead<br>scan mode | 34.0°C ~ 43.0°C<br>(93.2°F ~109.4°F) | NA | Forehead scan:<br>15.5°C to 42°C<br>(60°F to 107.6°F) | Similar<br>(Discussion is<br>indicated in<br>D7) |
| Display<br>resolution | 0.1°F (0.1°C) | 0.1°F (0.1°C) | 0.1°F (0.1°C) | Same |
| Measuring<br>accuracy | ±0.2°C (0.4°F) within<br>35°C ~ 42°C<br>(95°F~107.6°F);<br>±0.3°C (0.5°F) for other<br>range | ±0.2°C (0.4°F) within 35°C<br>~42°C (95°F ~107.6°F);<br>±0.3°C (0.5°F) for other<br>range | ±0.2°C (0.4°F) within<br>35~42°C (95°F ~107.6°F);<br>±0.3°C (0.5°F) for other<br>range | Same |
| Measure time | ≤2S for Forehead mode;<br>3~10s for Forehead scan<br>mode; | ≤2S | Seconds | Similar<br>(Discussion is<br>indicated in D8) |
| Color<br>Indication | Green: ≥ 35.8°C -<br>37.4 °C (96.4°F -99.3°F)<br>Yellow: ≥ 37.5°C- 38.5°C<br>(99.5°F -101.3°F)<br>Orange: ≥ 38.6°C - 43.0°C<br>(≥101.5°F -109.4°F) | Green: > 35.7 °C -37.4 °C<br>(> 96.3°F - 99.3 °F)<br>Yellow: > 37.4 °C - 38.5 °C<br>(> 99.4°F - 101.3 °F)<br>Red: > 38.5 °C - 42.2 °C<br>(> 101.3°F - 108.0 °F) | NA | Similar<br>(Discussion is<br>indicated in D9) |
| Measuring<br>Distance for<br>forehead<br>mode | 1 CM -5CM | 1 CM -5CM | NA | Same |
| Measuring<br>Distance for<br>forehead scan<br>mode | 0 cm | NA | 0 cm | Same |
{6}------------------------------------------------
{7}------------------------------------------------
| Power source | 3 V d.c. (2X AAA batteries) | Two (2) AA batteries | 9 volt Alkaline | Similar<br>(Discussion is<br>indicated in<br>D10) |
|---------------------|------------------------------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------|
| Operating condition | Temperature: 5°C ~40°C<br>(41°F ~104°F)<br>Humidity: ≤ 95%(no<br>condensing) | Temperature: 15°C ~40°C<br>(59°F ~104°F)<br>Humidity: Not public | Temperature:15.5°C<br>~40°C (60°F ~104°F)<br>Humidity: ≤ 95%(no<br>condensing) | Different<br>(Discussion is<br>indicated in<br>D11) |
## The discussion of differences exist between the subject and predicate devices is listed in followina:
- D1: There is only difference in description, all three devices are infrared electronic clinical thermometer used to measure body temperature of all ages from forehead.
- D2: The proposed device has been validated for cytotoxicity per ISO 10993- 5 and Irritation as well as Sensitization per ISO 10993-10. The difference will not affect the safety and effectiveness.
- D3: The proposed device has been validated for the lasted version of ISO 10993-5: 2009 and ISO 10993-10:2010. the difference will not affect the safety and effectiveness.
- D4: The different button number will not affect the safety and effectiveness.
- D5: The proposed device conducted performance testing in accordance with ASTM E1965-98. ISO 80601-2-56 to demonstrate the difference does not affect the safety and effectiveness.
- D6: The proposed device conducted performance testing in accordance with ASTM E1965-98, ISO 80601-2-56 to demonstrate the difference does not affect the safety and effectiveness.
- D7: The proposed device conducted performance testing in accordance with ASTM E1965-98, ISO 80601-2-56 to demonstrate the difference does not affect the safety and effectiveness.
- D8: The Measure time of proposed device was validated through performance testing in accordance with ASTM E1965-98, ISO 80601-2-56 to demonstrate the difference does not affect the safety and effectiveness.
- D9: The Color Indication of proposed device in upper temperature range ≥ 38.6℃ 43.0℃ is orange and the predicate is red in upper temperature range > 38.6℃ - 42.2℃, it's only different in visual effect of color, this difference does not affect the safety and effectiveness.
- D10: The proposed device was electromagnetic compatility and electrical safety in accordance with IEC 60601-1-2 and ANSI AAMI ES60601-1. The difference does not affect the safety and effectiveness.
- D11: The operating condition of subject device has passed the safety test, and the subject device complies the standard IEC 60601-1-11, so the difference between the operating conditions of subject device and predicate device will not affect the safety and effectiveness.
## 7. Discussion of Non-Clinical Tests Performed for Safety and effectiveness are as follows:
The recognized consensus standards for safety of medical equipment: ANSI AAMI ES60601-1, IEC 60601-1-11 for safety, IEC 60601-1-2 for electromagnetic compatibility, ASTM E 1965-98 and ISO 80601-2-56 for performance and IEC 62304 are complied, and see below table for details.
| Standards | Standards Name |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI<br>ES60601-1:2005/(R)2012 and<br>A1:2012 | Medical Electrical Equipment - Part 1: General Requirements For<br>Basic Safety And Essential Performance |
| IEC 60601-1-2: 2014 | Medical Electrical Equipment -- Part 1-2: General Requirements For<br>Basic Safety And Essential Performance -- Collateral Standard:<br>Electromagnetic Disturbances -- Requirements And Tests |
{8}------------------------------------------------
| IEC 60601-1-11: 2015 | Medical Electrical Equipment - Part 1-2: General Requirements For<br>Basic Safety And Essential Performance - Collateral Standard:<br>Electromagnetic Disturbances - Requirements And Tests |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 80601-2-56: 2017 | Medical Electrical Equipment - Part 2-56: Particular Requirements<br>For Basic Safety And Essential Performance Of Clinical<br>Thermometers For Body Temperature Measurement. |
| ASTM E1965-98:2016 | Standard Specification For Infrared Thermometers For Intermittent<br>Determination Of Patient Temperature |
| IEC 62304:2006+A1:2015 | Medical device software - Software life cycle processes |
Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff, May 2005.
## 8. Discussion of Clinical Accuracy Testing Performed:
The clinical accuracy test report and data analysis followed the requirements of the ASTM E 1965-98 (2016).
The clinical accuracy testing evaluated 140 of subjects. Each model was evaluated in 0 up to 3 months, 3 months up to one year, older than 1 year and younger than 5 years, and older than 5 years age groups. The test data showed the clinical accuracy of the subject device complied with the requirements of ASTM E1965-98 (2016).
## 9. Conclusions
Based on performance testing, comparison and analysis, the subject device Infrared Body Thermometer, model HTD8823US is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.