K200158 · GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC · IYN · Apr 17, 2020 · Radiology
Device Facts
Record ID
K200158
Device Name
LOGIQ E10
Applicant
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC
Product Code
IYN · Radiology
Decision Date
Apr 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
LOGIO E10 is intended for use by a qualified physician for ultrasound evaluation of Fetal / Obstetrics; Abdominal (including Renal, Gynecology/Pelvic); Pediatric; Small Organ (Breast, Testes, Thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (Adult and Pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including Prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (Abdominal and Vascular).
Device Story
LOGIQ E10 is a mobile, track 3, general-purpose diagnostic ultrasound system. It acquires, processes, and displays ultrasound signals using various linear, curved, phased, dual, and matrix array transducers. Operated by physicians in clinical settings, the system features a keyboard, specialized controls, touch screen, and widescreen monitor. It performs digital image acquisition, fluid flow analysis, and measurements. New features include UGAP for liver attenuation measurement and migrated software features (SonoNT, SonoIT, SonoAVC, SonoRenderLive). The device aids clinical decision-making by providing diagnostic images and quantitative data for various anatomical evaluations. It benefits patients by enabling non-invasive, real-time diagnostic imaging.
Clinical Evidence
No clinical data was required to support substantial equivalence. The device underwent extensive non-clinical bench testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with recognized standards (e.g., IEC 60601-1, ISO 10993-1).
Technological Characteristics
Mobile console with digital acquisition/processing. Supports linear, curved, phased, dual, and matrix array transducers. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, and ISO 14971. Connectivity includes DICOM (NEMA PS 3.1-3.20). Patient contact materials are biocompatible. Software features include UGAP (attenuation measurement), SonoNT, SonoIT, SonoAVC, and SonoRenderLive.
Indications for Use
Indicated for ultrasound evaluation of fetal/obstetrics, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, urology, transrectal, transvaginal, transesophageal, and intraoperative applications by qualified physicians.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue.
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC % Tracey Ortiz Regulatory Affairs Director 9900 Innovation Drive WAUWATOSA WI 53226
Re: K200158
Trade/Device Name: LOGIO E10 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, ITX, IYO, Dated: January 21, 2020 Received: January 22, 2020
Dear Tracey Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
April 17, 2020
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K200158
P 100000
Device Name LOGIO E10
Indications for Use (Describe)
LOGIQ E10 is intended for use by a qualified physician for ultrasound evaluation of Fetal / Obstetrics; Abdominal (including Renal, GynecologyPelvic); Pediatric; Small Organ (Breast, Testes, Thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (Adult and Pediatric); Peripheral Vasculo-skeletal Conventional and Superficial; Urology (including Prostate); Transvaginal; Transesophageal and Intraoperative (Abdominal and Vascular).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff | |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff@fda.hhs.gov
{3}------------------------------------------------
# GE Healthcare 510(k) Premarket Notification Submission
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The circle is surrounded by a decorative border of swirling lines. The logo is simple and recognizable, and it is associated with a well-known and established company.
## 510(k) Summary
| In accordance with 21 CFR 807.92 the following summary of information is provided: | |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 21, 2020 |
| Submitter: | GE Medical Systems Ultrasound and Primary Care Diagnostics,<br>LLC<br>9900 Innovation Dr<br>Wauwatosa, WI 53226 |
| Manufacturer: | GE Medical Systems Ultrasound and Primary Care Diagnostics,<br>LLC<br>9900 Innovation Dr<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T: (262)676-6120 |
| Alternate Contact Person: | Elizabeth Wentworth<br>Regulatory Affairs Specialist<br>GE Healthcare |
| Device: | |
| Trade Name: | LOGIQ E10 |
| Common/Usual Name: | LOGIQ E10 |
| Classification Names: | Class II |
| Product Code: | IYN (primary), IYO, ITX (secondary)<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550,<br>90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-<br>IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate Device: | K173555 LOGIQ E10 Diagnostic Ultrasound System |
| Reference Device(s): | K170445 LOGIQ S8 Diagnostic Ultrasound System<br>K192159 Voluson E10 Diagnostic Ultrasound System<br>K152309 LOGIQ E9 Diagnostic Ultrasound System<br>K161843 Aplio i900/i800/i700 Diagnostic Ultrasound System<br>V2.0 (ATI)<br>K190442 Koios DS for Breast |
| Device Description: | The LOGIQ E10 is a full featured, track 3, general purpose<br>diagnostic ultrasound system which consists of a mobile console<br>that provides digital acquisition, processing and display<br>capability. The user interface includes a computer keyboard,<br>specialized controls, LCD touch screen and color widescreen |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The logo is colored in a light blue hue, and the border has decorative white swirls around it.
# GE Healthcare 510(k) Premarket Notification Submission
monitor. The system utilizes a variety of linear, curved, phased, dual, and matrix array transducers to support the broad imaging capabilities.
## Intended Use:
LOGIO E10 is intended for use by a qualified physician for ultrasound evaluation of Fetal / Obstetrics; Abdominal (including Renal, Gynecology/Pelvic); Pediatric; Small Organ (Breast, Testes, Thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (Adult and Pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including Prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (Abdominal and Vascular).
- Technology: The LOGIQ E10 employs the same fundamental scientific technology as its predicate device(s).
#### Determination of Substantial Equivalence:
#### Comparison to Predicates
The proposed LOGIO E10 is substantially equivalent to the predicate devices. The following is an overview of the differences between the proposed LOGIQ E10 and the predicate LOGIQ E10 (K173555). The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- · The LOGIO E10 and predicate LOGIO E10 systems have the same clinical intended use, except Intraoperative (Abdominal) that is available on predicate LOGIQ E9 K152309).
- · The LOGIQ E10 and predicate LOGIQ E10 systems have the same imaging modes.
- · The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- · The systems have acoustic power levels which are below the applicable FDA limits.
- · The LOGIQ E10 and predicate LOGIQ E10 systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- · The LOGIQ E10 and predicate LOGIQ E10 systems have been designed in compliance with approved electrical and physical safety standards.
- · The transducers supported in proposed LOGIO E10 and predicate LOGIQ E10 are identical except:
- the following transducers migrated from LOGIQ E9 -(K152309): L3-9i-D, 6S-D, C2-6b-D, P6D;
- the following transducers migrated from LOGIQ S8 -(K170455): L3-12-D, BE9CS-D
- adding a new transducer, L6-24-D, which is similar to the -
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by decorative swirls or flourishes, giving it a classic and recognizable appearance.
# GE Healthcare
510(k) Premarket Notification Submission
previously cleared L8-18i-D
- adding B-Flow (hybrid) as a new mode to: C1-6-D, C1-6VN-D, C2-9-D, C2-9VN-D, L2-9-D, L2-9VN-D, L3-12-D. ML6-15-D. L8-18i-D:
- · The software features supported in proposed LOGIQ E10 and predicate LOGIQ E10 are identical except: - the following software features have been migrated from Voluson E10 (K192195): SonoNT, SonoIT, SonoAVC for renal cyst, SonoRenderLive.
- · UGAP is a new feature similar to ATI on Aplio i900/i800/i700 Diagnostic Ultrasound System V2.0 (K161843) that measures the attenuation value in the liver.
- · Other minor software feature modifications are: improved the Vnav image based registration, enabled Device Management, add a type 2 to SRI-HD
## Summary of Non-Clinical Tests:
been evaluated for acoustic output, The device has biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIO E10 and its applications comply with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment -● Part 1: General Requirements for Safety, 2005/(R)2012 And A1:2012
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: ● General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbance - Requirements and Tests, 2014
- IEC 60601-2-37, Medical Electrical Equipment Part 2-● 37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015
- ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within A Risk Management Process, 2009
- ISO 14971, Application of risk management to medical ● devices, 2007
- NEMA PS 3.1 3.20, ● -Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology) 2016
FDA Guidance's used:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a series of curved lines that resemble water droplets or waves. The logo is simple and recognizable, and it is often used to represent the company's brand.
# GE Healthcare
510(k) Premarket Notification Submission
- o Off-The-Shelf Software Use in Medical Devices -Guidance for Industry and Food and Drug Administration Staff - Issued on September 27, 2019
- Marketing Clearance of Diagnostic Ultrasound Systems ● and Transducer- Guidance for Industry and Food and Drug Administration Staff Document - Issued on June 27, 2019
- o Guidance for Industry and Food and Drug Administration Staff. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices - Issues on October 2, 2014
- Guidance for the Content of Premarket Submissions for ● Software Contained in Medical Devices - Issued on May 11,2005
- Guidance for Industry - Cyber Security for Networked Medical Devices Containing Off-The-Shelf (OTS) Software issued January 14, 2005
The following quality assurance measures are applied to the development of the system:
- Risk Analysis ●
- Requirements Reviews ●
- Design Reviews
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Performance testing (Verification) ●
- Safety testing (Verification) .
Transducer materials and other patient contact materials are biocompatible.
# Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ E10, did not require clinical studies to support substantial equivalence.
- GE Healthcare considers the LOGIQ E10 to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.