Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0

K161843 · Toshibamedical Systems Corporation · IYN · Sep 21, 2016 · Radiology

Device Facts

Record IDK161843
Device NameAplio i900/i800/i700 Diagnostic Ultrasound System, V2.0
ApplicantToshibamedical Systems Corporation
Product CodeIYN · Radiology
Decision DateSep 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K161843 · Sep 21, 2016Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0Toshibamedical Systems CorporationVolunteer clinical imagingClinical evaluations were conducted on volunteer liver images to confirm the performance and display capabilities of the Attenuation Imaging and Normalized Local Variance (NLV) features.Clinical evaluation; Volunteer imaging; Liver imaging

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical evaluation of Attenuation Imaging and NLV; Clinical evaluation using volunteer subjectsVolunteersNot applicable for this studyConfirmation of color map display and numerical results for attenuation coefficient and NLV values

Indications for Use

The Diagnostic Ultrasound Systems Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body usinq ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small orqans, trans-vaqinal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic.

Device Story

Mobile diagnostic ultrasound system; uses wide array of probes (flat linear, convex linear, sector) with 2-20 MHz frequency range. Inputs: ultrasound signals from patient anatomy. Transforms inputs via beamforming and signal processing to generate 2D, M-mode, Doppler, and 3D/4D images. Used in clinical settings by trained healthcare professionals. Features include Sensor 3D, Shadow Glass, Attenuation Imaging, Shear Wave Dispersion, Tissue Intensity Analysis (NLV), and Contrast Harmonic Imaging (CHI). Output displayed on system monitor for diagnostic interpretation. Assists clinicians in visualizing structures/dynamic processes; supports clinical decision-making through quantitative measurements (e.g., attenuation coefficients, volume measurements). Benefits include improved diagnostic accuracy and workflow efficiency.

Clinical Evidence

Clinical evaluations conducted using representative in-vivo images of volunteer livers. Confirmed Attenuation Imaging displays color maps of attenuation coefficients and numerical results. Confirmed Tissue Intensity Analysis (NLV) displays acceptable color map images and mean NLV values. Bench testing included phantom studies for all new features (Sensor 3D, Attenuation Imaging, NLV, Fusion Auto Track, Auto Volume Measurement, CHI, Shear Wave Dispersion, Shadow Glass) to verify performance and measurement accuracy.

Technological Characteristics

Mobile ultrasound system; frequency range 2-20 MHz. Transducers: flat linear, convex linear, sector arrays. Connectivity: supports fusion with CT/MR data. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-2-37, IEC 62304, ISO 10993-1, AIUM RTD2-2004. Features: Beamforming, Doppler, Elastography, 3D/4D imaging, Fusion imaging.

Indications for Use

Indicated for visualization of structures and dynamic processes in fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (adult/pediatric), peripheral vascular, transesophageal, musculoskeletal (conventional/superficial), and laparoscopic applications.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 21, 2016 Toshiba Medical Systems Corporation % Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780 Re: K161843 Trade/Device Name: Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 14, 2016 Received: September 15, 2016 Dear Mr. Tadeo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use # 510(k) Number (if known) K161843 #### Device Name Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0 #### Indications for Use (Describe) The Diaqnostic Ultrasound Systems Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body usinq ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small orqans, trans-vaqinal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic. Type of Use (Select one or both, as applicable) ك Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Transducer: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | Clinica | | |--|--|---------|--| |--|--|---------|--| | Clinical Application | | | | Mode of Operation | | | | | | | | | | | | | | | | | | | Other [Note] | |---------------------------------|---|---|-----|-------------------|-----------------------------------------|---------------|-----------|------------|------|-------|-----|--------------|----------|------------|----|----------------------------|------|------------|----------|--------|------------------|--------|--------------| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Precision Imaging<br>Combined (Specify) | Color Doppler | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI(ADF) | Shear wave | 4D | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 6,8 | | Abdominal | N | N | N | N | N | 2,3 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 4,5,6,7,8 | | Intra-operative (Abdominal) | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | | | Pediatric | N | N | N | N | N | 2,3 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 6,8 | | Small Organ (Note 1) | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 6 | | Neonatal Cephalic | N | N | N | N | N | 3 | N | N | N | N | N | | | | | | | | | | | | | | Adult Cephalic | N | N | N | N | N | 3 | | | | | N | | | | | | | | | | | | | | Trans-rectal | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 8 | | Trans-vaginal | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 8 | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 6 | | Musculo-skeletal (Superficial) | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 6 | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | N | N | N | N | N | 3 | N | | | | N | N | | | | | | | | | | | 7 | | Cardiac Pediatric | N | N | N | N | N | 3 | N | | | | N | N | | | | | | | | | | | 7 | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | N | N | N | N | N | 3 | N | | | | N | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | N | N | N | N | N | 2 | N | N | N | N | N | N | N | N | N | N | N | N | N | N | N | | 6 | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {4}------------------------------------------------ Transducer: PSI-30BX Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | Other [Note] | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|-----------------------------------------|-----------|------------|------|-------|-----|--------------|----------|------------|----------------------|----------------------------|------|------------|----------|--------|------------------|--------|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI(ADF) | Shear wave | 4D<br>(Volume color) | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | N | N | N | N | N | 3 | N | | | N | | N | | | | | | | | N | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | N | N | N | N | N | 3 | N | | | N | | N | | | | | | | | N | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | N | N | N | N | N | 3 | N | | | N | | | | | | | | | | | | | | | Adult Cephalic | N | N | N | N | N | 3 | | | | N | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | N | N | N | N | N | 3 | N | | | N | N | | | | | | | | | | | 7 | | | Cardiac Pediatric | N | N | N | N | N | 3 | N | | | N | N | | | | | | | | | | | 7 | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {5}------------------------------------------------ Transducer: PSI-70BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | Other [Note] | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|-----------------------------------------|-----------|------------|------|-------|-----|--------------|------------------------|----------------------|----------------------------|------|------------|----------|--------|------------------|--------|---| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | Shear wave<br>SMI(ADF) | 4D<br>(Volume color) | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | N | N | N | N | N | 3 N | N | | | N | | | N | | | | | | | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | N | N | N | N | N | 3 N | N | | | N | | | N | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | N | N | N | N | N | 3 N | N | | | N | | | | | | | | | | | | | | Adult Cephalic | N | N | N | N | N | 3 | | | | N | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | N | N | N | N | N | 3 N | N | | | N | N | | | | | | | | | | | 7 | | Cardiac Pediatric | N | N | N | N | N | 3 N | N | | | N | N | | | | | | | | | | | 7 | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {6}------------------------------------------------ Transducer:_ PST-25BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | Other [Note] | | | | | | | | |--------------------------------|-------------------|---|-----|-----|---------------|-----------------------------------------|------------|-----------|------|-------|-----|----------|------------|----------------------|--------------|------|------------|----------|--------|------------------|--------|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) | Micro Pure | Apli Pure | BEAM | Power | TDI | SMI(ADF) | Shear wave | 4D<br>(Volume color) | 3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | P | P | P | P | P | 3 | N | | | P | | | N | | | | | | | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | 3 | N | | | P | | | N | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | P | P | 3 | N | | | P | | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal(Conventional) | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | P | P | P | P | P | 3 | N | | | P | P | | | | | | | | | | | | | Cardiac Pediatric | P | P | P | P | P | 3 | N | | | P | P | | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {7}------------------------------------------------ Transducer: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | Other | [Note] | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|--------------------|-------------------|------------|------|-------|-----|--------------|------------|----------|----------------------|----------|------|------------|----------|--------|------------------|--------|------|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Precision Imaging | Micro Pure | BEAM | Power | TDI | Elastography | Shear wave | SMI(ADF) | 4D<br>(Volume color) | 3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | P | P | P | P | P | 3 | N | | | P | | | N | | | | | | | | | | 7(N) | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | 3 | N | | | P | | | N | | | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | P | P | P | P | P | 3 | | | | P | P | | | | | | | | | | | | 7 | | | Cardiac Pediatric | P | P | P | P | P | 3 | | | | P | P | | | | | | | | | | | | 7 | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {8}------------------------------------------------ Transducer: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation | B | M | PWD | CWD | Color Doppler<br>Combined (Specify) | Precision Imaging | Apli Pure | Micro Pure | BEAM | Power | Elastography | TDI | SMI(ADF) | Shear wave | 4D | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other | [Note] | |------------------------------------------------|-------------------|---|---|-----|-----|-------------------------------------|-------------------|-----------|------------|------|-------|--------------|-----|----------|------------|----|----------------------------|------|------------|----------|--------|------------------|--------|-------|--------| | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | | P | P | P | P | P | 3 | N | | | P | | | | | | | | | | | | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | 3 | N | | | P | | | | | | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | Adult Cephalic | | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | | P | P | P | P | P | 3 | | | | P | P | | | | | | | | | | | | | | | Cardiac Pediatric | | P | P | P | P | P | 3 | | | | P | P | | | | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {9}------------------------------------------------ Transducer: _ PVI-475BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | Other | [Note] | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|--------------------|-------------------|-----------|------------|------|-------|-----|--------------|----------|------------|----|----------------------------|-------|------------|----------|--------|------------------|--------|-----|--|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Precision Imaging | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI(ADF) | Shear wave | 4D | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | | | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | N | N | N | | N | 2 | N | N | | | N | | | N | | | | | | N | | | | | | | | Abdominal | N | N | N | | N | 2 | N | N | | | N | | N | N | | | | | | N | N | N | | 6,7 | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | N | N | N | | N | 2 | N | N | | | N | | | N | | | | | | N | N | N | | 6 | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 ATI Note 5 Tissue Intensity Analysis Note 6 Sensor3D Note 7 CHI (Per FDA approved contrast agent prescribing information) {10}------------------------------------------------ Transducer: _ PVI-475BX Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI(ADF) | Shear wave | 4D<br>(Volume color) | 3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other | [Note] | |------------------------------------------------|-------------------|---|---|-----|-----|---------------|-----------------------------------------|-----------|------------|------|-------|-----|--------------|----------|------------|----------------------|----------|------|------------|----------|--------|------------------|--------|---------|--------| | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | N | N | N | N | N | 2 | N | N | | N | | N | | | | | | N | | | | | | | | Abdominal | | N | N | N | N | N | 2 | N | N | | N | N | N | N | N | | | | | N | N | N | | 4,5,6,7 |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...